EchoPod ® DL34 Ultrasonic Level Transmitter, Switch & Controller

Application

The general purpose ultrasonic level transmitter


provides continuous level measurement up to 18.0’
(5.5m) with a 4-20 mA signal output, and is configured
via our free Webcal software. The level sensor has 4
programmable relays with selectable hysteresis and
fail-safe logic. The embedded level controller can
lower cost by replacing external control hardware.
This non-contact liquid level sensor is ideally suited
for corrosive, sticky or dirty liquids, and is widely
selected for day tank, skid, IBC, sump and process
tank level applications.

Features Success

• Level detection, switch and control functions up to About 1.3M poly-


18.0’ (5.5m) ethylene tanks are
sold in North Amer-
• Configuration is fast and easy via WebCal software ica each year rang-
and USB adapter ing in size from 1 to
15,000 gallons. Most
• Narrow 3” beam width and short 8” dead band of these are less
optimized for small tanks than 6’ in height with
capacities under 500
• Four programmable relays for switch, pump or valve gallons. The largest
control and fail-safety markets are industrial
manufacturing, mu-
• 1 pump or valve with 3 alarms nicipal, agriculture and residential in chemical, water
• 2 pumps (lead-lag) with 2 alarms & wastewater applications. Over the past 10 years,
• 2 pumps (duplexing) with 2 alarms customers have moved toward the use of smaller
• 4 independent switch point alarms storage tanks with increased automation to lower
cost. Today, mini-bulk tanks are a major product cat-
• PVDF transducer and 6P polycarbonate enclosure for egor. Flowline has developed a new multi-function
corrosive liquids sensor to meets the needs of this growing market.
The DL34 EchoPod puts non-contact level measure-
• Automatic temperature compensation for accurate ment, relay switch and advanced control functions in
measurement the palm of your hand.

Compatible Products

DataLoop™ DataView™ Commander™ DataPoint™


Level Indicator Level Controller Multi-Tank Level Controller Level Controller

LI25 LI55 LI90 LC52

© 2010 Flowline Inc. • 10500 Humbolt Street Los Alamitos, CA 90720 • 562.598.3015 • www.flowline.com • DS204220 REV A
®
EchoPod ® DL34 Ultrasonic Level Transmitter, Switch & Controller

Specifications Dimensions Configuration


Red - Power

Range: 18.0’ (5.5m) Blk


Wht
-
-
Return
TX (Out)
Gry - RX (In)

Accuracy: ± 0.2% of range Blu


Org
-
-
Rly 1
Rly 2

Resolution: 0.079” (2mm)


Yel - Rly 3

(1.2 m)
48”
Pur - Rly 4
Brn - Rly Com

Dead band: 8” (20cm)

(27 mm)
Beam width: 3” (7.62cm)

1.1”
1/2” NPT
Configuration: WebCal® PC LI99-1001
Windows® USB 2.0 1.07”
(27 mm)
The level sensor is configurable via
Memory: Non-volatile

2.7” (68 mm)


our free WebCal PC software and
Supply voltage: 24 VDC (loop)
Fob USB adapter. The sensors are
Consumption: 0.5W
offered with and without Fobs. Fobs
Loop resist.: 400Ω max
are universal and can be used to
Signal output: 4-20 mA, two-wire
3.06” configure any WebCal compatible
Signal invert: 4-20 mA or 20-4 mA (78 mm)
product.

3.4" (88 mm)


Signal fail-safe: 4 mA, 20 mA, 21 mA,

2.7” (68 mm)


2" NPT
22 mA or hold last (2" G)
Download your free copy of WebCal
Contact type: (4) SPST relays in English or Chinese at
Contact rating: 60 VA, 1A maximum www.flowline.com/webcal.php.
Contact fail-safe: Power loss: Hold last
Echo loss: Open, close
or hold last Fittings
Hysteresis: Selectable
Process temp.: F: 20° to 140°
For optimum performance, install the level sensor using the below
C: -7° to 60°
recommended or direct equivalent fittings.
Temp. comp.: Automatic
Ambient temp.: F: -31° to 140° P/N Description
C: -35° to 60°
LM52-2400 3” NPT x 2” NPT (Sch. 40)
Pressure: MWP = 30 PSI (2 bar)
Enclosure rating: Type 6P, encapsulated, LM52-2800 3” NPT x 2” NPT (Sch. 80)
corrosion resistant & LM52-3800 4” NPT x 2” NPT (Sch. 80)
submersible
Encl. material: Polycarbonate LM52-2410 3” Socket x 2” NPT (Sch. 40)
Trans. material: PVDF LM52-3410 4” Socket x 2” NPT (Sch. 40)
Cable jacket mat: Polyurethane
LM52-2810 3” Socket x 2” NPT (Sch. 80)
Cable type: 9-conductor, shielded
LM52-3810 4” Socket x 2” NPT (Sch. 80)
Cable length: 48” (1.2m)
Process mount: 2” NPT (2” G)
Mount gasket: Viton® LM52-2890 2” NPT Bulkhead Fitting, PVC
Classification: General purpose
Compliance: CE, RoHS
Approvals: cFMus
LM50-1001 2” NPT Bracket, Polypropylene

Ordering
EchoPod
Notes

DL34- Install Ordering


installation
1)
the level Notes

fittings or
1) Install
sensor using Flowline
theequivalents.
level sensor using Flowline
Process mount (1) The level sensor 2)
is configured
installation fittings via our
orWebCal
equivalents.
0 NPT (US) software and one LI99-1001 Fob USB adapter.
2) The level sensor is configured via our
1 G (Metric)
The level sensor is offered with and without
a Fob. Fobs WebCal are software
universal and
and onecan LI99-1001
be
Fob USB adapter (2) Fob USB adapter. The level sensor is

used to configure any WebCal compatible
0 Without Fob offered with and without a Fob.
product. WebCal is a free download from our Fobs
1 With Fob are universal and can be used to configure

website at www.flowline.com/webcal.php.
any WebCal compatible product. WebCal
is a free download from our website.
© 2010 Flowline Inc. • 10500 Humbolt Street Los Alamitos, CA 90720 • 562.598.3015 • www.flowline.com • DS204220 REV A