0x230141C0-Failed To Leave Initial Position (Sample Syringe)
0x230141C0-Failed To Leave Initial Position (Sample Syringe)
Use Background
When analyzer gives a false alarm - Failed to leave initial position (Sample syringe).
Mechanism: The motor initial position photocoupler always detects that the photocoupler is blocked.
Declaration
Reference Standard
DH36X,
DF50/DF51/DF52/D
F50VET,
DF50CRP
DH26,DH31/DH33/DH36/DH36VET/DH36X,DF50/DF51/DF52/DF50VET/DF50CRP↑ DH56/UN73/DH76/D5CRP/D7CRP↓
If you have any problem with this file, please contact our engineer ASAP, thank you!
DH31/DH33/DH36/DH36VET,
DF50/DF51/DF52/DF50VET
DH26/DH31/DH33/DH36/DH
36VET,
DH56/UN73/DH76/D5CRP/D7
CRP
DF50/DF51/DF52/DF50VET,
O-ring for 250ul
DH56/UN73/DH76, /
sampling syringe
D5CRP/D7CRP
DH36X,
Sensor(Vertical) 23.99.0007C