Saudi Aramco Test Report: Calibration Test Report - Pressure Gauge SATR-A-2002 24-Mar-16 Mech
Saudi Aramco Test Report: Calibration Test Report - Pressure Gauge SATR-A-2002 24-Mar-16 Mech
Saudi Aramco Test Report: Calibration Test Report - Pressure Gauge SATR-A-2002 24-Mar-16 Mech
01 6-Sep-15
TASK COMPLETED DATE RFI No. REQUESTED INSPECTION DATE QUANTITY MANHOURS TRAVEL TIME
SPENT
Pressure Gauge #: Service: NOTE: Indicate use. (Pressure Testing, etc) Construction or
Inspection Tool-specify,
CALIBRATION (Note 1)
Applied Load, psi (Standard
Gauge Reading, psi Difference, psi Error, % of full scale Reading
Pressure)
0%
30%
50%
80%
100%
80%
50%
30%
0%
Notes: 1) Calibrated at Ambient Temperature of 23 C (+/- 2 deg C) . 2) Maximum interval is 1 month. Apply legible sticker to gage. Pressure gages
are delicate instruments. As calibrated, they exhibit no evidence of mechanical damage or mishandling. Safety Item: Do not attempt to calibrate a
damaged gage. Repair or replace damaged gages and report them to Quality for entry into equipment log and investigation for root cause of damage.
Test crews handling instrumentation are properly trained in the care and use of the instrument.
Remarks:
REFERENCE DOCUMENTS:
1. SAES-A-004 - General Requirements of Pressure Testing (30 Jan 2011)
2. ASME/ANSI PTC 19.2 - Pressure Measurement Instruments & Apparatus ( 2010 Edition)