Control System
The central control system shall provide an intelligent human machine
interface incorporating on board flash memory and real-time clock
for recording operational parameters in the in built event log. The
central control system shall operate at low voltage and include BMS
connection for common fault. Visualisation of plant inputs, outputs
and status through a web browser, using a simple Ethernet connection
shall be available. The central control unit shall incorporate a user
friendly 5.7” high-definition colour display with clear pictograms
and LED indicators, providing easy access to system operational
information.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 2
Typical Layout
HTM02-01 50 Hz
Model L (ength) W (idth) H (eight) LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 250 T - 50 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
mVAC 330 T - 50 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
mVAC 500 T - 50 Hz 2400 980 1650 3700 1980 500 800 54 54 n/a
mVAC 300 T - 60 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
mVAC 400 T - 60 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 660 Q - 50 Hz 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel
mVAC 1000 Q - 50 Hz 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel
mVAC 2560 Q - 50 Hz 2600 1200 2400 1060 3900 4360 500 800 76 54 76 outl. unit - 54 inl. vessel
mVAC 3300 Q - 50 Hz 2600 1200 2500 1230 3900 4530 500 800 76 54 76 outl. unit - 76 inl. vessel
mVAC 3840 P - 50 Hz 4100 1250 2500 1230 5990 4580 500 800 76 54 76 outl. unit - 76 inl. vessel
mVAC 620 T - 60 Hz 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. Vessel
mVAC 800 Q - 60 Hz 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. Vessel
mVAC 1200 Q - 60 Hz 1910 980 2500 860 3210 3940 500 800 54 54 54 outl. unit - 54 inl. Vessel
mVAC 1500 T - 60 Hz 2200 1100 2500 860 3500 4060 500 800 54 54 54 outl. unit - 54 inl. Vessel
mVAC 3000 Q - 60 Hz 2600 1200 2500 1230 3900 4530 500 800 86 54 76 outl. unit - 76 inl. Vessel
mVAC 3900 Q - 60 Hz 2600 1200 2500 1230 3900 4530 500 800 86 54 76 outl. unit - 76 inl. Vessel
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 3
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 1280 T - 50 Hz 2200 1100 2100 660 4280 3860 500 800 76 54 54 outl. unit - 54 inl. Vessel
mVAC 4950 P - 50 Hz 4100 1250 2500 1230 5990 4580 500 800 108 54 108 outl. unit - 76 inl. Vessel
mVAC 6000 H - 50 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. Vessel
mVAC 4500 P - 60 Hz 4100 1250 2400 1060 5480 4410 500 800 86 108 108 outl. unit - 76 inl. Vessel
mVAC 5850 P - 60 Hz 4100 1250 2500 1230 5990 4580 500 800 86 108 108 outl. unit - 76 inl. Vessel
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 6600 H - 50 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. Vessel
mVAC 7800 H - 60 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. Vessel
Note!
Drawing shows other models not used in this range and are for
illustration purposes only.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 4
HTM2022 50 Hz
Model L (ength) W (idth) H (eight) LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 250 D - 50 Hz 2040 980 1650 3340 1980 500 800 54 54 n/a
mVAC 300 D - 60 Hz 2040 980 1650 3340 1980 500 800 54 54 n/a
Model L (ength) W (idth) H (eight) LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 500 T - 50 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
mVAC 660 T - 50 Hz 2400 980 1900 3700 1980 500 800 54 54 n/a
mVAC 1000 T - 50 Hz 2400 980 1900 3700 1980 500 800 54 54 n/a
mVAC 500 T - 60 Hz 2300 980 1650 3600 1980 500 800 54 54 n/a
mVAC 800 T - 60 Hz 2400 980 1900 3700 1980 500 800 54 54 n/a
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 1500 Q - 50 Hz 1830 980 2400 1060 3130 3840 500 800 54 54 54 outl. unit - 54 inl. vessel
mVAC 1200 T - 60 Hz 1910 980 2400 1060 3210 3840 500 800 54 54 54 outl. unit - 54 inl. vessel
mVAC 1860 Q - 60 Hz 2200 1200 2500 1230 3500 4230 500 800 76 54 76 outl. unit - 76 inl. vessel
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 5
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 2560 T - 50 Hz 2600 1200 2400 1060 3900 4360 500 800 76 54 76 outl. unit - 54 inl. vessel
mVAC 3840 Q - 50 Hz 2600 1200 2500 1230 3900 4530 500 800 76 54 76 outl. unit - 76 inl. vessel
mVAC 3000 T - 60 Hz 2600 1200 2500 1230 3900 4530 500 800 76 54 76 outl. unit - 76 inl. vessel
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 4950 Q - 50 Hz 3400 1250 2500 1230 5990 4580 500 800 76 108 108 outl. unit - 76 inl. vessel
mVAC 6000 P - 50 Hz 4100 1250 2500 1230 5990 4580 500 800 76 108 108 outl. unit - 76 inl. vessel
mVAC 4500 Q - 60 Hz 3400 1250 2400 1060 5480 4410 500 800 76 108 108 outl. unit - 76 inl. vessel
mVAC 5850 Q - 60 Hz 3400 1250 2500 1230 5990 4580 500 800 76 108 108 outl. unit - 76 inl. vessel
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 6
Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø
mVAC 6600 P - 50 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. vessel
mVAC 8000 H - 50 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. vessel
mVAC 7800 P - 60 Hz 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl. unit - 76 inl. vessel
mVAC 9200 H - 60 Hz 4100 1250 2500 1230 7720 4580 500 800 108 54 108 outl. unit - 76 inl. vessel
Note!
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 7
Data Tables
HTM02-01 50 Hz
mVAC-1000-Q
mVAC-3300-Q
mVAC-2560-Q
mVAC-6000-H
mVAC-6600-H
mVAC-500-TH
mVAC-250-TH
mVAC-330-TH
mVAC-3840-P
mVAC-4950-P
mVAC-1280-T
mVAC-660-Q
Model Number
Plant Output (litres/minute) 250 330 500 660 1000 1280 2560 3300 3840 4950 6000 6600
40 52 79 105 159 203 406 524 609 786 952 1047
System Flow (m3/h - l/m)
665 865 1315 1750 2650 3380 6765 8730 10150 13100 15865 17450
Number of pumps 3 3 3 4 4 3 4 4 5 5 6 6
Number of receivers 1(H) 1(H) 1(H) 2 2 3 2 2 2 3 3 4
Total receiver volume (litres) 500 500 500 1000 1000 1500 3000 4000 4000 6000 6000 8000
Receiver connection(s) (mm) n/a n/a n/a 54 54 54 54 76 76 76 76 76
Inlet/service connection (mm) 54 54 54 54 54 54 76 76 76 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 69 70 69 70 72 72 75 72 75 75 75
Weight (kg) 650 690 750 960 1040 1475 2215 2335 2660 3190 3390 3820
Motor rating (kW) 1.1 1.5 2.2 1.5 2.2 5.5 5.5 7.5 5.5 7.5 7.5 7.5
1.5 1.5 1.5 1.5 1.5 1.5 1.5 2.5 1.5 2.5 2.5 2.5
Motor cable size (mm2/Amps)
(13) (13) (13) (13) (13) (13) (13) (17) (13) (17) (17) (17)
Motor rated supply per pump (A) 10 10 10 10 10 20 20 20 20 20 20 20
FLC per pump (A) 2.8 3.6 5.1 3.6 5.1 12 12 14.9 12 14.9 14.9 14.9
Starting current (A) 14 18 26 18 26 60 60 75 60 75 75 75
Central control supply - single phase 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
(mm2/Amps) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.2 0.1 0.2 0.3 0.3 0.4 0.3 0.4 0.4 0.4
8152 1306 04
9820 6365 00
8152 1306 00
9820 6365 00
81521 306 02
9820 6365 00
9820 6368 00
8152 1306 06
9820 6368 00
8152 1306 08
9820 6369 00
8152 1306 10
9820 6368 00
8152 1306 12
9820 6368 00
8152 1306 14
9820 6368 00
8152 1306 16
9820 6369 00
8152 1306 18
9820 6369 00
8152 1306 20
9820 6370 00
8152 1306 22
Installation Proposal
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with two pumps on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with two pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 8
HTM02-01 50 Hz - No Vessels
mVAC-1000-Q
mVAC-3300-Q
mVAC-2560-Q
mVAC-6000-H
mVAC-6600-H
mVAC-3840-P
mVAC-4950-P
mVAC-1280-T
mVAC-660-Q
mVAC-500-T
mVAC-250-T
mVAC-330-T
NV
NV
NV
NV
NV
NV
NV
NV
NV
NV
NV
NV
Model Number
Plant Output (litres/minute) 250 330 500 660 1000 1280 2560 3300 3840 4950 6000 6600
40 52 79 105 159 203 406 524 609 786 952 1047
System Flow (m3/h - l/m)
665 865 1315 1750 2650 3380 6765 8730 10150 13100 15865 17450
Number of pumps 3 3 3 4 4 3 4 4 5 5 6 6
Number of receivers n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Total receiver volume (litres) n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Receiver connection(s) (mm) n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Inlet/service connection (mm) 54 54 54 54 54 54 76 76 76 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 69 70 69 70 72 72 75 72 75 75 75
Motor rating (kW) 1.1 1.5 2.2 1.5 2.2 5.5 5.5 7.5 5.5 7.5 7.5 7.5
1.5 1.5 1.5 1.5 1.5 1.5 1.5 2.5 1.5 2.5 2.5 2.5
Motor cable size (mm2/Amps)
(13) (13) (13) (13) (13) (13) (13) (17) (13) (17) (17) (17)
Motor rated supply per pump (A) 10 10 10 10 10 20 20 20 20 20 20 20
FLC per pump (A) 2.8 3.6 5.1 3.6 5.1 12 12 14.9 12 14.9 14.9 14.9
Starting current (A) 14 18 26 18 26 60 60 75 60 75 75 75
Central control supply - single phase (mm2/ 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
Amps) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.2 0.1 0.2 0.3 0.3 0.4 0.3 0.4 0.4 0.4
8152 1306 48
81521 306 49
8152 1306 50
8152 1306 51
8152 1306 52
8152 1306 53
8152 1306 54
8152 1306 55
8152 1306 56
8152 1306 57
8152 1306 58
8152 1306 59
Installation Proposal
-
-
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with two pumps on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with two pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 9
HTM2022 50 Hz
mVAC-1000-TH
mVAC--6000-P
mVAC-250-DH
mVAC-1500-Q
mVAC-3840-Q
mVAC-4950-Q
mVAC-8000-H
mVAC-500-TH
mVAC-660-TH
mVAC-6600-P
mVAC-2560-T
Model Number
Plant Output (litres/minute) 250 500 660 1000 1500 2560 3840 4950 6000 6600 8000
40 79 105 159 238 406 609 786 952 1047 1270
System Flow (m3/h - l/m)
665 1315 1750 2650 3965 6765 10150 13100 15865 17450 21165
Number of pumps 2 3 3 3 4 3 4 4 5 5 6
Number of receivers 1(H) 1(H) 1(H) 1(H) 1 2 2 3 3 4 4
Total receiver volume (litres) 500 500 1000 1000 1500 3000 4000 6000 6000 8000 8000
Receiver connection(s) (mm) n/a n/a n/a n/a 54 54 76 76 76 76 76
Inlet/service connection (mm) 54 54 54 54 54 54 76 108 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 67 69 70 70 72 72 75 75 75 75
Weight (kg) 540 650 800 860 1120 2105 2560 3090 3340 3770 4080
Motor rating (kW) 1.1 1.1 1.5 2.2 2.2 5.5 5.5 7.5 7.5 7.5 7.5
1.5 1.5 1.5 1.5 1.5 1.5 1.5 2.5 2.5 2.5 2.5
Motor cable size (mm2/Amps)
(13) (13) (13) (13) (13) (13) (13) (17) (17) (17) (17)
Motor rated supply per pump (A) 10 10 10 10 10 20 20 20 20 20 20
FLC per pump (A) 2.8 3.6 5.1 3.6 5.1 12 12 14.9 14.9 14.9 14.9
Starting current (A) 14 18 26 18 26 60 60 75 75 75 75
1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
Central control supply - single phase (mm2/Amps)
(5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.1 0.2 0.2 0.3 0.3 0.4 0.4 0.4 0.4
9820 6364 00
8152 1306 82
9820 6365 00
8152 1306 84
9820 6365 00
8152 1306 86
9820 6365 00
8152 1306 88
9820 6367 00
8152 1306 90
9820 6368 00
8152 1306 92
9820 6368 00
8152 1306 94
9820 6369 00
8152 1306 96
9820 6369 00
8152 1306 98
9820 6370 00
9820 6370 00
8152 1307 00
8152 1307 02
Installation Proposal
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with one pump on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with one pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 10
HTM2022 50 Hz - No Vessels
mVAC--6000-P
mVAC-1500-Q
mVAC-3840-Q
mVAC-4950-Q
mVAC-8000-H
mVAC-6600-P
mVAC-1000-T
mVAC-2560-T
mVAC-250-D
mVAC-500-T
mVAC-660-T
NV
NV
NV
NV
NV
NV
NV
NV
NV
NV
NV
Model Number
Plant Output (litres/minute) 250 500 660 1000 1500 2560 3840 4950 6000 6600 8000
40 79 105 159 238 406 609 786 952 1047 1270
System Flow (m3/h - l/m)
665 1315 1750 2650 3965 6765 10150 13100 15865 17450 21165
Number of pumps 2 3 3 3 4 3 4 4 5 5 6
Number of receivers n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Total receiver volume (litres) n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Receiver connection(s) (mm) n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a n/a
Inlet/service connection (mm) 54 54 54 54 54 54 76 108 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 69 69 70 70 72 72 75 75 75 75
Motor rating (kW) 1.1 1.5 1.5 2.2 2.2 5.5 5.5 7.5 7.5 7.5 7.5
1.5 1.5 1.5 1.5 1.5 1.5 1.5 2.5 2.5 2.5 2.5
Motor cable size (mm2/Amps)
(13) (13) (13) (13) (13) (13) (13) (17) (17) (17) (17)
Motor rated supply per pump (A) 10 10 10 10 10 20 20 20 20 20 20
FLC per pump (A) 2.8 2.8 5.1 3.6 5.1 12 12 14.9 14.9 14.9 14.9
Starting current (A) 14 14 26 18 26 60 60 75 75 75 75
1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
Central control supply - single phase (mm2/Amps)
(5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.1 0.2 0.2 0.3 0.3 0.4 0.4 0.4 0.4
8152 1307 26
8152 1307 27
8152 1307 28
8152 1307 29
8152 1307 30
8152 1307 31
8152 1307 32
8152 1307 33
8152 1307 34
8152 1307 35
8152 1307 36
Installation Proposal
-
-
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with one pump on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with one pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
2005161
16/05/14
Issue 7
Page 11
HTM02-01 60 Hz
mVAC-3000-Q
mVAC-1200-Q
mVAC-3900-Q
mVAC-7800-H
mVAC-300-TH
mVAC-400-TH
mVAC-4500-P
mVAC-5850-P
mVAC-1500-T
mVAC-800-Q
mVAC-620-T
Model Number
Plant Output (litres/minute) 300 400 620 800 1200 1500 3000 3900 4500 5850 7800
48 63 98 127 190 238 476 619 714 928 1238
System Flow (m3/h - l/m)
800 1050 1630 2115 3165 3965 7930 10315 11900 15465 20630
Number of pumps 3 3 3 4 4 3 4 4 5 5 6
Number of receivers 1(H) 1(H) 2 2 2 2 2 2 3 3 4
Total receiver volume (litres) 500 500 1000 1000 2000 2000 3000 4000 4500 6000 8000
Receiver connection(s) (mm) n/a n/a 54 54 54 54 54 76 54 76 76
Inlet/service connection (mm) 54 54 54 54 54 54 76 76 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 75 72 75 72 74 74 77 74 77 77
Weight (kg) 650 690 1050 960 1260 1545 2380 2560 3185 3340 4080
Motor rating (kW) 1.7 2.4 3.4 2.4 3.4 9.2 9.2 10 9.2 10 10
Motor cable size 1.5 1.5 1.5 1.5 1.5 4 4 4 4 4 4
(mm2/Amps) (13) (13) (13) (13) (13) (23) (23) (23) (23) (23) (23)
Motor rated supply per pump (A) 10 10 25 10 25 25 25 25 25 25 25
FLC per pump (A) 4 5.7 7.4 5.7 7.4 18.5 18.5 20.8 18.5 20.8 20.8
Starting current (A) 20 29 37 29 37 93 93 104 93 104 104
Central control supply - single phase 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
(mm2/Amps) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.2 0.1 0.2 0.3 0.3 0.4 0.3 0.4 0.4
9820 6365 00
8152 1306 60
9820 6365 00
8152 1306 62
9820 6368 00
8152 1306 64
9820 6368 00
8152 1306 66
9820 6368 00
8152 1306 68
9820 6368 00
8152 1306 70
9820 6368 00
8152 1306 72
9820 6368 00
8152 1306 74
9820 6369 00
8152 1306 76
9820 6369 00
8152 1306 78
9820 6370 00
81521 306 80
Installation Proposal
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with two pumps on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with two pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.
BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005161
the right is reserved to change the specification of the 16/05/14
items described herein at any time. Please contact us Issue 7
for further information and up to date specifications.
Page 12
HTM2022 60 Hz
mVAC-300-DH
mVAC-4500-Q
mVAC-1860-Q
mVAC-5850-Q
mVAC-9200-H
mVAC-500-TH
mVAC-800-TH
mVAC-7800-P
mVAC-3000-T
mVAC-1200-T
Model Number
Plant Output (litres/minute) 300 500 800 1200 1860 3000 4500 5850 7800 9200
48 79 127 190 295 476 714 928 1238 1460
System Flow (m3/h - l/m)
800 1315 2115 3165 4915 7930 11900 15465 20630 24330
Number of pumps 2 3 3 3 4 3 4 4 5 6
Number of receivers 1(H) 1(H) 1(H) 1 1 2 3 3 4 5
Total receiver volume (litres) 500 500 1000 1500 2000 3000 4500 6000 8000 10000
Receiver connection(s) (mm) n/a n/a n/a 54 76 54 54 76 76 76
Inlet/service connection (mm) 54 54 54 54 54 76 76 108 108 108
Exhaust connection (mm, per pump) 54 54 54 54 54 54 54 54 54 54
Maximum exhaust back pressure (mbar) 60 60 60 60 60 60 60 60 60 60
Noise level/pump (dB[A]) 67 67 75 72 72 74 74 77 77 77
Weight (kg) 540 650 800 980 1480 2565 2935 3090 3880 4510
Motor rating (kW) 1.7 1.7 2.4 3.4 3.4 9.2 9.2 10 10 10
1.5 1.5 1.5 1.5 1.5 4 4 4 4 4
Motor cable size (mm2/Amps)
(13) (13) (13) (13) (13) (23) (23) (23) (23) (23)
Motor rated supply per pump (A) 10 10 10 25 25 25 25 25 25 25
FLC per pump (A) 4 4 5.7 7.4 7.4 18.5 18.5 20.8 20.8 20.8
Starting current (A) 20 20 29 37 37 93 93 104 104 114
1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
Central control supply - single phase (mm2/Amps)
(5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
Maximum Inlet Temperature (0C) 40 40 40 40 40 40 40 40 40 40
Cooling air flow per pump (m3/s) 0.1 0.1 0.2 0.2 0.3 0.3 0.3 0.4 0.4 0.4
9820 6364 00
9820 6365 00
9820 6365 00
9820 6367 00
9820 6367 00
9820 6368 00
9820 6369 00
9820 6369 00
9820 6370 00
9820 6370 00
8152 1307 37
8152 1307 39
8152 1307 41
8152 1307 43
8152 1307 45
8152 1307 47
8152 1307 49
8152 1307 51
8152 1307 53
8152 1307 55
Installation Proposal
Part Number
• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with one pump on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with one pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.