Download as pdf or txt
Download as pdf or txt
You are on page 1of 110

Voluson™ E6 BT15

Voluson™ E8 BT15
Advanced Acoustic Output References
English (English)

Part number: 5504378

Revision 3

Software: EC300

© 2015 General Electric


Revision History

Revision Date

Revision 1 January 2014

Revision 2 May 2014

Revision 3 July 2015

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


i-ii 5504378 Revision 3
Table of Contents
Chapter 1 – Legal Notes
Chapter 2 – Acoustic Output & Index Determination Tables
Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Probe 11L-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-3
Probe 3Sp-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8
Probe 4C-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-15
Probe 9L-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-21
Probe C1-5-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-27
Probe C4-8-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-34
Probe IC5-9-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-41
Probe ML6-15-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-47
Probe P2D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-52
Probe P6D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-53
Probe RAB2-5-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-54
Probe RAB6-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-61
Probe RIC5-9-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-68
Probe RNA5-9-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-74
Probe RSP6-16-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-81
Probe S4-10-D - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-86
Chapter 3 – Accuracies
Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Geometric measurements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3
Flow speed measurements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
Other measurements - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-5
Accuracies of acoustic parameters - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6
Chapter 4 – Probe Settings
Probe Settings Voluson™ E6 BT15 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-2
Probe Settings Voluson™ E8 BT15 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-3

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 i-iii
This page was intentionally left blank.

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


i-iv 5504378 Revision 3
Chapter 1

Legal Notes

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 1-1
Legal Notes

Legal Notes
The contents of this publication may not be copied or duplicated in any form, in whole or in
part, without prior written permission of GE Healthcare Austria GmbH & Co OG .
GE Healthcare Austria GmbH & Co OG may revise this publication from time to time without
written notice.

TRADEMARKS
All products and their name brands are trademarks of their respective holders.

COPYRIGHTS
© 2015 General Electric

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


1-2 5504378 Revision 3
Chapter 2

Acoustic Output & Index


Determination Tables

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-1
Acoustic Output & Index Determination Tables

2.1 Overview
This chapter provides Acoustic Output Reporting Tables for IEC60601-2-37 and a Summary of
measured Quantities for Index Determination for all applicable Transducers at different
Operating Modes.

IEC FDA Meaning - IEC / FDA

α a Acoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-


MHz)

Aaprt Aaprt -12db Output Beam Area / Active aperture area

deq deq Equivalent Beam Diameter

fawf fc Acoustic Working Frequency / Center frequency

Ipa Ipa Pulse-Average Intensity

Ipa,α Ipa,3 Attenuated Pulse-Average Intensity

Ipi PII Pulse-Intensity Integral

Ipi,α PII.3 Attenuated Pulse-Intensity Integral

Ita(z) ITA Temporal-Average Intensity

Ita,α(z) ITA,3(z) Attenuated Temporal-Average Intensity at depth z

MI MI Mechanical Index

P Wo Output Power / Time average acoustic power at the source

Ita,α(Z) ITA,3(Z) Attenuated Output Power / derated acoustic power at depth z

pr pr Peak-Rarefactional Acoustic Pressure

prα Pr.3 Attenuated Peak-Rarefactional Acoustic Pressure

prr PRF Pulse Repetition Rate / Pulse repetition frequency

TI TI Thermal Index

TIB TIB Bone Thermal Index

TIC TIC Cranial-Bone Thermal Index

TIS TIS Soft-Tissue Thermal Index

td PD Pulse Duration

X,Y X,Y -12dB Output Beam Dimensions

Z Z Distance from the source to a specified point

Zb Zsp Depth for max TIB

Zbp Zbp Break-Point Depth

ZS Z1 Depth for max TIS

Table 2-1 Table Key for Acoustic Output Reporting Table

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-2 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.2 Probe 11L-D

2.2.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 1.3

Index component value - 0.21* 0.21* 0.21* 0.21* -


0.21 0.21 0.21 0.21
0.21 0.21 0.21 0.21
0.21 0.21 0.21 0.21

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.36 - - - - -

P W0 mW - 14.0 14.0 56.4

P1x1 W1x1 mW - 6.2 6.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.9 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 7.24 7.22 7.22 7.02

Other Information

prr PRF Hz 2310 - - - - -

srr - Hz 55 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 752.6 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 34 - - - - -

Ispta at zpii ITA at zpii mW/cm² 63 - - - - -

pr at zpii pr at zpii MPa 4.25 - - - - -

tMode Components - B B* B* B
u B B
e
B B
v
B B
Table 2-2 11L-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 11.0 11.0 11.0 11.0

B-Image sector angle Degree - - - -

B-Image sector width cm 3.7 3.7 3.7 3.7

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - norm - - penet

S/PRI - - 1 1 -

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 8.2 8.2 8.2

B-Zoom Box end cm 1.1 9.3 9.3 9.3

B-Zoom Box angle Degree - - - -

B-Zoom Box width cm 0.8 2.3 2.3 2.3

Mode Type - B BF BF B

Focal Depth cm 0.9 6.0 6.0 6.0


Table 2-3 Operating Control Conditions : 11L-D, Brightness Mode (2D)

2.2.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 1.6

Index component value - 0.02 0.02 0.02 0.02 -


0.02 0.02 0.02 0.02
0.73* 0.73* 0.73* 0.73*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.07 - - - - -

P W0 mW - 26.5 26.5 74.0

P1x1 W1x1 mW - 21.4 21.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.1 - - - - -

zpii,α zpii,3 cm 1.2 - - - - -

fawf fc MHz 6.05 7.13 7.13 5.35

Other Information

prr PRF Hz 588 - - - - -

srr - Hz 5 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 393.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 20 - - - - -

Ispta at zpii ITA at zpii mW/cm² 33 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-4 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 2.72 - - - - -

tMode Components - B B B B
u B* B B B
e
CFM CFM* CFM* CFM*
v
Table 2-4 11L-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 11.0 11.0 11.0 11.0

B-Image sector angle Degree - - - -

B-Image sector width cm 3.7 3.7 3.7 3.7

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

CFM-Box start cm 0.0 6.6 6.6 0.0

CFM-Box end cm 1.6 8.2 8.2 1.6

CFM-Box width cm 3.7 0.6 0.6 3.7

CFM-Box angle Degree - - - -

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 12

Line Density - 1 10 10 9

Flow Res - low low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm low

CFM-Frequency - mid high high low


Table 2-5 Operating Control Conditions : 11L-D, Color Flow Doppler Mode (CFM)

2.2.3 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.0 1.3

Index component value - 0.68* 0.68* 0.68* 0.68* -


0.18 0.14 0.18 0.30

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.36 - - - - -

P W0 mW - 46.1 46.1 51.9

P1x1 W1x1 mW - 20.3 20.3 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-5
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.9 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 7.24 7.02 7.02 7.02

Other Information

prr PRF Hz 4400 - - - - -

srr - Hz 105 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 752.6 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 65 - - - - -

Ispta at zpii ITA at zpii mW/cm² 121 - - - - -

pr at zpii pr at zpii MPa 4.25 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-6 11L-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 4.9 4.9 2.9

B-Image sector end cm 1.1 6.0 6.0 4.0

B-Image sector angle Degree - - - -

B-Image sector width cm 0.8 2.3 2.3 2.3

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - norm penet penet penet

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 0.9 6.0 6.0 6.0


Table 2-7 Operating Control Conditions : 11L-D, Motion Mode (M)

2.2.4 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.1 0.6 1.3 0.8

Index component value - 0.59 0.41 0.58 1.27 -

Associated Acoustic Parameters

IEC FDA Unit -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-6 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr,α at zMI pr.3 at zsp MPa 2.58 - - - - -

P W0 mW - 24.1 24.2 24.2

P1x1 W1x1 mW - 23.2 23.2 -

zs z1 cm - - 1.1 - - -

zb zsp cm - - - - 1.1 -

zMI zsp cm 1.0 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 5.35 5.35 5.26 5.26

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 297.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 250 - - - - -

Ispta at zpii ITA at zpii mW/cm² 382 - - - - -

pr at zpii pr at zpii MPa 3.10 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-8 11L-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 0.07 1.50 1.50

Gate depth cm 3.30 3.30 7.37 7.37

PRF KHz 0.9 7.0 14.0 14.0

Power % 100 100 100 100


Table 2-9 Operating Control Conditions : 11L-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-7
Acoustic Output & Index Determination Tables

2.3 Probe 3Sp-D

2.3.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.5 1.3

Index component value - 0.49 0.49 0.49 0.49 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.07 - - - - -

P W0 mW - 43.3 43.3 10.4

P1x1 W1x1 mW - 30.7 30.7 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 5.7 - - - - -

zpii,α zpii,3 cm 5.7 - - - - -

fawf fc MHz 2.73 3.33 3.33 2.26

Other Information

prr PRF Hz 1684 - - - - -

srr - Hz 120 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 113.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 34 - - - - -

Ispta at zpii ITA at zpii mW/cm² 77 - - - - -

pr at zpii pr at zpii MPa 3.60 - - - - -

tMode Components - B B B B*
u B
e
B
v
B
Table 2-10 3Sp-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 23.7 23.7 23.7 23.7

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet resol resol CPI

S/PRI - - - - 1

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-8 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 15.0 15.0 8.0

B-Zoom Box end cm 2.5 17.5 17.5 10.4

B-Zoom Box angle Degree 15.0 20.0 20.0 15.0

B-Zoom Box width cm - - - -

Mode Type - B B B Contr

Focal Depth cm 11.0 9.0 9.0 3.0


Table 2-11 Operating Control Conditions : 3Sp-D, Brightness Mode (2D)

2.3.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 1.6

Index component value - 0.01 0.01 0.01 0.01 -


0.65* 0.65* 0.65* 0.65*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 1.77 - - - - -

P W0 mW - 76.7 76.7 60.0

P1x1 W1x1 mW - 54.4 54.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 5.6 - - - - -

zpii,α zpii,3 cm 5.6 - - - - -

fawf fc MHz 2.01 2.51 2.51 2.51

Other Information

prr PRF Hz 255 - - - - -

srr - Hz 6 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 100.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 7 - - - - -

Ispta at zpii ITA at zpii mW/cm² 8 - - - - -

pr at zpii pr at zpii MPa 2.66 - - - - -

tMode Components - B* B B B
u CFM CFM* CFM* CFM*
e
v
Table 2-12 3Sp-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-9
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 23.7 23.7 23.7 23.7

B-Image sector angle Degree 90.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

CFM-Box start cm 0.0 3.4 3.4 0.0

CFM-Box end cm 3.5 6.9 6.9 3.5

CFM-Box width cm - - - -

CFM-Box angle Degree 90.0 12.0 12.0 12.0

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 10

Flow Res - low high high high

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - mid high high high


Table 2-13 Operating Control Conditions : 3Sp-D, Color Flow Doppler Mode (CFM)

2.3.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 2.1 1.6

Index component value - 0.03 0.03 0.03 0.03 -


0.01 0.01 0.01 0.02
0.51* 0.68* 0.48* 2.09*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.19 - - - - -

P W0 mW - 86.7 56.5 56.5

P1x1 W1x1 mW - 43.0 40.3 -

zs z1 cm - - 2.5 - - -

zb zsp cm - - - - 1.4 -

zMI zsp cm 5.9 - - - - -

zpii,α zpii,3 cm 5.9 - - - - -

fawf fc MHz 3.07 2.51 2.51 2.51

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-10 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 1950 - - - - -

srr - Hz 163 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 123.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 34 - - - - -

Ispta at zpii ITA at zpii mW/cm² 74 - - - - -

pr at zpii pr at zpii MPa 4.22 - - - - -

tMode Components - B* B B B
u M M M M
e
cM cM* cM* cM*
v
Table 2-14 3Sp-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 23.7 23.7 4.7 4.7

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 16

Flow Res - high high high high

PRF KHz 0.1 0.9 1.8 1.8

Speed - 6 6 6 6

cM-Frequency - Low High High High


Table 2-15 Operating Control Conditions : 3Sp-D, Color Motion Doppler Mode (cM)

2.3.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.3 1.2 0.9

Index component value - 0.28 0.31 0.28 1.24 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.07 - - - - -

P W0 mW - 42.1 42.1 42.1

P1x1 W1x1 mW - 30.1 30.1 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-11
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 1.7 - - -

zb zsp cm - - - - 2.1 -

zMI zsp cm 4.2 - - - - -

zpii,α zpii,3 cm 4.2 - - - - -

fawf fc MHz 1.94 1.94 1.94 1.94

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 130 - - - - -

Ispta at zpii ITA at zpii mW/cm² 233 - - - - -

pr at zpii pr at zpii MPa 0.09 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-16 3Sp-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-17 Operating Control Conditions : 3Sp-D, Continuous Wave Doppler Mode (CW)

2.3.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.8 1.1

Index component value - 0.46* 0.46* 0.15 0.15 -


0.03 0.03 0.15* 0.65*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.07 - - - - -

P W0 mW - 40.5 21.8 30.5

P1x1 W1x1 mW - 28.7 15.6 -

zs z1 cm - - - - - -

zb zsp cm - - - - 3.1 -

zMI zsp cm 5.7 - - - - -

zpii,α zpii,3 cm 5.7 - - - - -

fawf fc MHz 2.73 3.33 2.01 3.07

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-12 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 1950 - - - - -

srr - Hz 163 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 113.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 39 - - - - -

Ispta at zpii ITA at zpii mW/cm² 89 - - - - -

pr at zpii pr at zpii MPa 3.60 - - - - -

tMode Components - B* B* B B*
u M M M* M
e
v
Table 2-18 3Sp-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 17.7 0.0

B-Image sector end cm 2.5 2.5 20.2 2.5

B-Image sector angle Degree 15.0 90.0 15.0 90.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet resol norm norm

M-Power % 100 100 100 100

Mode Type - M M HI M

Focal Depth cm 11.0 9.0 7.0 3.0


Table 2-19 Operating Control Conditions : 3Sp-D, Motion Mode (M)

2.3.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 2.3 1.6

Index component value - 0.64 0.73 0.61 2.27 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 1.99 - - - - -

P W0 mW - 89.9 82.3 74.0

P1x1 W1x1 mW - 53.5 58.8 -

zs z1 cm - - 2.2 - - -

zb zsp cm - - - - 4.4 -

zMI zsp cm 4.8 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-13
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zpii,α zpii,3 cm 4.8 - - - - -

fawf fc MHz 2.54 2.49 2.17 2.17

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 122.6 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 117 - - - - -

Ispta at zpii ITA at zpii mW/cm² 279 - - - - -

pr at zpii pr at zpii MPa 3.08 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-20 3Sp-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.10 1.00 1.50 1.50

Gate depth cm 0.05 7.31 5.19 2.97

PRF KHz 0.9 2.2 0.9 2.2

Power % 100 100 100 100


Table 2-21 Operating Control Conditions : 3Sp-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-14 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.4 Probe 4C-D

2.4.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 1.5

Index component value - 0.18* 0.18* 0.18* 0.18* -


0.18 0.18 0.18 0.18
0.18 0.18 0.18 0.18
0.18 0.18 0.18 0.18

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.03 - - - - -

P W0 mW - 24.5 24.5 24.5

P1x1 W1x1 mW - 12.0 12.0 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.0 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 2.65 3.13 3.13 3.13

Other Information

prr PRF Hz 1870 - - - - -

srr - Hz 55 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 308.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 30 - - - - -

Ispta at zpii ITA at zpii mW/cm² 65 - - - - -

pr at zpii pr at zpii MPa 2.95 - - - - -

tMode Components - B B* B* B*
u B B B
e
B B B
v
B B B
Table 2-22 4C-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 36.0 36.0 36.0 36.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - high low low low

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-15
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - penet - - -

S/PRI - - 1 1 1

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 17.6 17.6 17.6

B-Zoom Box end cm 3.6 21.1 21.1 21.1

B-Zoom Box angle Degree 15.0 15.0 15.0 15.0

B-Zoom Box width cm - - - -

Mode Type - B BF BF BF

Focal Depth cm 5.0 1.5 1.5 1.5


Table 2-23 Operating Control Conditions : 4C-D, Brightness Mode (2D)

2.4.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 0.9 1.8

Index component value - 0.02 0.02 0.02 0.02 -


0.02 0.02 0.02 0.02
0.88* 0.88* 0.88* 0.88*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 1.80 - - - - -

P W0 mW - 86.3 86.3 83.2

P1x1 W1x1 mW - 58.0 58.0 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 3.7 - - - - -

zpii,α zpii,3 cm 3.8 - - - - -

fawf fc MHz 2.08 3.20 3.20 2.85

Other Information

prr PRF Hz 1023 - - - - -

srr - Hz 102 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 145.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 22 - - - - -

Ispta at zpii ITA at zpii mW/cm² 38 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-16 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 2.36 - - - - -

tMode Components - B B B B
u B B B B
e
CFM* CFM* CFM* CFM*
v
Table 2-24 4C-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 36.0 36.0 36.0 36.0

B-Image sector angle Degree 52.6 25.0 25.0 25.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

CFM-Box start cm 0.0 0.0 0.0 0.0

CFM-Box end cm 5.4 5.4 5.4 5.4

CFM-Box width cm - - - -

CFM-Box angle Degree 52.6 7.3 7.3 7.3

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 10

Flow Res - low low low low

PRF KHz 0.1 0.1 0.1 11.0

CFM-Quality - low norm norm norm

CFM-Frequency - low high high mid


Table 2-25 Operating Control Conditions : 4C-D, Color Flow Doppler Mode (CFM)

2.4.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.1 2.3 1.7

Index component value - 0.03 0.03 0.08 0.08 -


0.01 0.01 0.03 0.10
0.48* 1.02* 0.41* 2.07*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.23 - - - - -

P W0 mW - 150.0 50.5 150.0

P1x1 W1x1 mW - 35.6 38.9 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-17
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 3.5 - - -

zb zsp cm - - - - 1.9 -

zMI zsp cm 4.8 - - - - -

zpii,α zpii,3 cm 7.5 - - - - -

fawf fc MHz 3.18 2.82 2.19 2.82

Other Information

prr PRF Hz 150 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 330.5 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 41 - - - - -

Ispta at zpii ITA at zpii mW/cm² 217 - - - - -

pr at zpii pr at zpii MPa 4.54 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-26 4C-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 36.0 36.0 5.0 36.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 16

Flow Res - high high high high

PRF KHz 0.1 0.9 9.0 0.9

Speed - 6 6 6 6

cM-Frequency - High Mid Low Mid


Table 2-27 Operating Control Conditions : 4C-D, Color Motion Doppler Mode (cM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-18 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.4.4 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.3 1.5

Index component value - 0.54* 0.54* 0.17 0.17 -


0.14 0.12 0.20* 1.09*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.03 - - - - -

P W0 mW - 72.0 47.5 83.1

P1x1 W1x1 mW - 35.2 19.1 -

zs z1 cm - - - - - -

zb zsp cm - - - - 4.1 -

zMI zsp cm 4.0 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 2.65 3.23 2.19 2.65

Other Information

prr PRF Hz 3938 - - - - -

srr - Hz 131 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 308.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 47 - - - - -

Ispta at zpii ITA at zpii mW/cm² 102 - - - - -

pr at zpii pr at zpii MPa 2.95 - - - - -

tMode Components - B* B* B B*
u M M M* M
e
v
Table 2-28 4C-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 4.0 4.0 24.3 0.0

B-Image sector end cm 7.7 7.7 27.9 3.6

B-Image sector angle Degree 20.0 15.0 15.0 15.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet resol penet penet

M-Power % 100 100 100 100

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-19
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

Mode Type - M M HI M

Focal Depth cm 5.0 5.0 10.0 1.5


Table 2-29 Operating Control Conditions : 4C-D, Motion Mode (M)

2.4.5 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.4 2.9 2.3

Index component value - 0.78 1.38 0.83 2.90 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.10 - - - - -

P W0 mW - 184.5 158.8 184.5

P1x1 W1x1 mW - 57.2 61.0 -

zs z1 cm - - 3.1 - - -

zb zsp cm - - - - 4.3 -

zMI zsp cm 4.4 - - - - -

zpii,α zpii,3 cm 4.5 - - - - -

fawf fc MHz 2.82 2.86 2.86 2.86

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 242.4 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 201 - - - - -

Ispta at zpii ITA at zpii mW/cm² 570 - - - - -

pr at zpii pr at zpii MPa 2.62 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-30 4C-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 1.50 1.50

Gate depth cm 10.81 14.55 7.65 11.10

PRF KHz 0.9 7.0 2.2 0.9

Power % 100 100 100 100


Table 2-31 Operating Control Conditions : 4C-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-20 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.5 Probe 9L-D

2.5.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.1 1.1 2.4

Index component value - 0.27* 0.27* 0.27* 0.27* -


0.27 0.27 0.27 0.27
0.27 0.27 0.27 0.27
0.27 0.27 0.27 0.27

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.18 - - - - -

P W0 mW - 11.4 11.4 34.0

P1x1 W1x1 mW - 10.4 10.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.1 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 6.46 5.50 5.50 3.94

Other Information

prr PRF Hz 1295 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 422.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 21 - - - - -

Ispta at zpii ITA at zpii mW/cm² 60 - - - - -

pr at zpii pr at zpii MPa 3.94 - - - - -

tMode Components - B B* B* B*
u B B B
e
B B B
v
B B B
Table 2-32 9L-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 13.9 13.9 13.9 13.9

B-Image sector angle Degree - - - -

B-Image sector width cm 4.3 4.3 4.3 4.3

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-21
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - norm - - CPI

S/PRI - - 1 1 1

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 2.6 2.6 3.1

B-Zoom Box end cm 1.4 4.0 4.0 4.5

B-Zoom Box angle Degree - - - -

B-Zoom Box width cm 0.8 0.8 0.8 2.5

Mode Type - B BF BF Contr

Focal Depth cm 1.2 3.5 3.5 7.0


Table 2-33 Operating Control Conditions : 9L-D, Brightness Mode (2D)

2.5.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.2 1.2 2.5

Index component value - 0.01 0.01 0.01 0.01 -


0.01 0.01 0.01 0.01
1.20* 1.20* 1.20* 1.20*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.44 - - - - -

P W0 mW - 74.1 74.1 175.7

P1x1 W1x1 mW - 47.4 47.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.7 - - - - -

zpii,α zpii,3 cm 1.9 - - - - -

fawf fc MHz 3.81 5.30 5.30 3.81

Other Information

prr PRF Hz 1792 - - - - -

srr - Hz 119 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 276.6 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 156 - - - - -

Ispta at zpii ITA at zpii mW/cm² 253 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-22 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 1.95 - - - - -

tMode Components - B B B B
u B B B B
e
CFM* CFM* CFM* CFM*
v
Table 2-34 9L-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 13.9 13.9 13.9 13.9

B-Image sector angle Degree - - - -

B-Image sector width cm 4.3 4.3 4.3 4.3

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 5.9 5.9 5.9 2.0

CFM-Box end cm 8.0 8.0 8.0 4.1

CFM-Box width cm 4.3 0.7 0.7 4.3

CFM-Box angle Degree - - - -

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 9

Flow Res - low low low low

PRF KHz 0.1 0.1 0.1 13.0

CFM-Quality - low norm norm norm

CFM-Frequency - low high high low


Table 2-35 Operating Control Conditions : 9L-D, Color Flow Doppler Mode (CFM)

2.5.3 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.2 0.9 0.4

Index component value - 0.24 0.17 0.24 0.90 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.10 - - - - -

P W0 mW - 11.8 11.8 11.8

P1x1 W1x1 mW - 11.8 11.8 -

zs z1 cm - - 1.2 - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-23
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zb zsp cm - - - - 2.1 -

zMI zsp cm 2.1 - - - - -

zpii,α zpii,3 cm 2.1 - - - - -

fawf fc MHz 4.24 4.24 4.24 4.24

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 320 - - - - -

Ispta at zpii ITA at zpii mW/cm² 613 - - - - -

pr at zpii pr at zpii MPa 0.14 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-36 9L-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-37 Operating Control Conditions : 9L-D, Continuous Wave Doppler Mode (CW)

2.5.4 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.1 1.3 1.8

Index component value - 0.84* 0.84* 0.83* 0.83* -


0.24 0.23 0.25 0.46

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.18 - - - - -

P W0 mW - 61.9 61.2 91.4

P1x1 W1x1 mW - 33.6 33.3 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.1 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 6.46 5.25 5.25 5.25

Other Information

prr PRF Hz 2700 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-24 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

srr - Hz 104 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 422.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 45 - - - - -

Ispta at zpii ITA at zpii mW/cm² 124 - - - - -

pr at zpii pr at zpii MPa 3.94 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-38 9L-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 7.3 5.8 4.7

B-Image sector end cm 1.4 8.7 7.2 6.1

B-Image sector angle Degree - - - -

B-Image sector width cm 0.8 0.8 0.8 2.5

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - norm penet penet penet

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 1.2 9.5 9.5 9.5


Table 2-39 Operating Control Conditions : 9L-D, Motion Mode (M)

2.5.5 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.0 2.6 1.8

Index component value - 0.95 0.95 0.90 2.56 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.47 - - - - -

P W0 mW - 80.1 80.1 80.1

P1x1 W1x1 mW - 51.2 51.2 -

zs z1 cm - - 1.7 - - -

zb zsp cm - - - - 1.7 -

zMI zsp cm 1.8 - - - - -

zpii,α zpii,3 cm 2.0 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-25
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

fawf fc MHz 3.89 3.89 3.70 3.89

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 300.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 220 - - - - -

Ispta at zpii ITA at zpii mW/cm² 677 - - - - -

pr at zpii pr at zpii MPa 2.65 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-40 9L-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 0.07 1.50 0.07

Gate depth cm 4.20 5.59 6.98 5.59

PRF KHz 0.9 3.3 16.0 22.2

Power % 100 100 100 100


Table 2-41 Operating Control Conditions : 9L-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-26 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.6 Probe C1-5-D

2.6.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.1 1.1 2.8

Index component value - 0.27* 0.27* 0.27* 0.27* -


0.27 0.27 0.27 0.27
0.27 0.27 0.27 0.27
0.27 0.27 0.27 0.27

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.16 - - - - -

P W0 mW - 40.2 40.2 40.2

P1x1 W1x1 mW - 24.3 24.3 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.5 - - - - -

zpii,α zpii,3 cm 4.5 - - - - -

fawf fc MHz 2.98 2.36 2.36 2.36

Other Information

prr PRF Hz 2614 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 213.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 36 - - - - -

Ispta at zpii ITA at zpii mW/cm² 85 - - - - -

pr at zpii pr at zpii MPa 3.51 - - - - -

tMode Components - B B* B* B*
u B B B
e
B B B
v
B B B
Table 2-42 C1-5-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 5.0 5.0 5.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - high low low low

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-27
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - penet CPI CPI CPI

S/PRI - - 1 1 1

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.6 0.6 0.6

B-Zoom Box end cm 3.0 3.1 3.1 3.1

B-Zoom Box angle Degree 15.0 15.0 15.0 15.0

B-Zoom Box width cm - - - -

Mode Type - B Contr Contr Contr

Focal Depth cm 6.5 1.5 1.5 1.5


Table 2-43 Operating Control Conditions : C1-5-D, Brightness Mode (2D)

2.6.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.1 1.1 2.3

Index component value - 0.03 0.03 0.03 0.03 -


1.02* 1.02* 1.02* 1.02*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.24 - - - - -

P W0 mW - 86.5 86.5 162.8

P1x1 W1x1 mW - 75.2 75.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.0 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 3.21 2.85 2.85 2.11

Other Information

prr PRF Hz 2062 - - - - -

srr - Hz 137 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 206.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 33 - - - - -

Ispta at zpii ITA at zpii mW/cm² 83 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-28 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 3.55 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-44 C1-5-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 30.0 30.0 30.0

B-Image sector angle Degree 69.2 45.0 45.0 45.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 0.0 0.0 0.0 19.1

CFM-Box end cm 4.5 4.5 4.5 23.6

CFM-Box width cm - - - -

CFM-Box angle Degree 69.2 7.2 7.2 7.2

CFM-Power % 100 100 100 100

Ensemble - 7 31 31 31

Line Density - 1 10 10 10

Flow Res - high low low low

PRF KHz 11.0 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - high mid mid low


Table 2-45 Operating Control Conditions : C1-5-D, Color Flow Doppler Mode (CFM)

2.6.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.2 2.5 2.1

Index component value - 0.09 0.09 0.06 0.06 -


0.01 0.02 0.02 0.07
0.67* 1.05* 0.51* 2.39*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.24 - - - - -

P W0 mW - 134.8 116.2 31.4

P1x1 W1x1 mW - 49.5 50.0 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-29
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 2.8 - - -

zb zsp cm - - - - 3.6 -

zMI zsp cm 4.7 - - - - -

zpii,α zpii,3 cm 4.7 - - - - -

fawf fc MHz 3.21 2.85 2.15 2.15

Other Information

prr PRF Hz 150 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 206.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 36 - - - - -

Ispta at zpii ITA at zpii mW/cm² 105 - - - - -

pr at zpii pr at zpii MPa 3.83 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-46 C1-5-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 30.0 30.0 5.0

B-Image sector angle Degree 20.0 20.0 69.2 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 8

Flow Res - high high high high

PRF KHz 0.1 0.9 2.4 11.0

Speed - 6 6 6 6

cM-Frequency - High Mid Low Low


Table 2-47 Operating Control Conditions : C1-5-D, Color Motion Doppler Mode (cM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-30 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.6.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.5 2.1 1.0

Index component value - 0.33 0.45 0.33 2.08 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.09 - - - - -

P W0 mW - 65.5 65.5 65.5

P1x1 W1x1 mW - 34.1 34.1 -

zs z1 cm - - 2.4 - - -

zb zsp cm - - - - 4.7 -

zMI zsp cm 4.9 - - - - -

zpii,α zpii,3 cm 4.9 - - - - -

fawf fc MHz 2.01 2.01 2.01 2.01

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 320 - - - - -

Ispta at zpii ITA at zpii mW/cm² 643 - - - - -

pr at zpii pr at zpii MPa 0.12 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-48 C1-5-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-49 Operating Control Conditions : C1-5-D, Continuous Wave Doppler Mode (CW)

2.6.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.4 2.0

Index component value - 0.77* 0.77* 0.73* 0.73* -


0.17 0.14 0.19 0.65

Associated Acoustic Parameters

IEC FDA Unit -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-31
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr,α at zMI pr.3 at zsp MPa 2.16 - - - - -

P W0 mW - 89.3 84.8 144.5

P1x1 W1x1 mW - 53.9 51.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.5 - - - - -

zpii,α zpii,3 cm 4.5 - - - - -

fawf fc MHz 2.98 2.98 2.98 2.98

Other Information

prr PRF Hz 8176 - - - - -

srr - Hz 80 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 213.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 38 - - - - -

Ispta at zpii ITA at zpii mW/cm² 88 - - - - -

pr at zpii pr at zpii MPa 3.51 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-50 C1-5-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 5.6 0.0

B-Image sector end cm 3.0 2.5 8.6 3.0

B-Image sector angle Degree 45.0 15.0 15.0 30.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet penet penet penet

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 6.5 3.0 3.0 1.5


Table 2-51 Operating Control Conditions : C1-5-D, Motion Mode (M)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-32 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.6.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.2 3.0 2.3

Index component value - 0.83 1.19 0.68 3.00 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.24 - - - - -

P W0 mW - 147.3 141.8 158.6

P1x1 W1x1 mW - 61.3 68.1 -

zs z1 cm - - 2.7 - - -

zb zsp cm - - - - 3.6 -

zMI zsp cm 4.0 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 3.20 2.86 2.11 2.07

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 213.6 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 156 - - - - -

Ispta at zpii ITA at zpii mW/cm² 396 - - - - -

pr at zpii pr at zpii MPa 3.56 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-52 C1-5-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 0.20 0.10 1.00

Gate depth cm 3.03 15.00 9.02 29.50

PRF KHz 0.9 12.0 4.4 16.0

Power % 100 100 100 100


Table 2-53 Operating Control Conditions : C1-5-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-33
Acoustic Output & Index Determination Tables

2.7 Probe C4-8-D

2.7.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 1.0 1.0 1.3

Index component value - 0.96 0.96 0.96 0.96 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.39 - - - - -

P W0 mW - 49.1 49.1 74.6

P1x1 W1x1 mW - 43.7 43.7 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 3.7 - - - - -

zpii,α zpii,3 cm 3.9 - - - - -

fawf fc MHz 3.66 4.60 4.60 4.68

Other Information

prr PRF Hz 1954 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 343.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 28 - - - - -

Ispta at zpii ITA at zpii mW/cm² 72 - - - - -

pr at zpii pr at zpii MPa 3.73 - - - - -

tMode Components - B B B B*
u B
e
v
Table 2-54 C4-8-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 3.0 3.0 26.0

B-Image sector angle Degree 20.0 20.0 20.0 75.0

B-Image sector width cm - - - -

B-Line Dens. - high low low high

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet resol resol norm

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-34 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.0 0.0 14.6

B-Zoom Box end cm 2.6 2.0 2.0 17.2

B-Zoom Box angle Degree 15.0 15.0 15.0 75.0

B-Zoom Box width cm - - - -

Mode Type - B B B CE

Focal Depth cm 4.5 4.5 4.5 11.0


Table 2-55 Operating Control Conditions : C4-8-D, Brightness Mode (2D)

2.7.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 1.4

Index component value - 0.05 0.05 0.05 0.05 -


0.62* 0.62* 0.62* 0.62*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.50 - - - - -

P W0 mW - 36.5 36.5 140.0

P1x1 W1x1 mW - 32.5 32.5 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 2.8 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 4.01 4.01 4.01 3.11

Other Information

prr PRF Hz 1690 - - - - -

srr - Hz 113 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 531.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 40 - - - - -

Ispta at zpii ITA at zpii mW/cm² 123 - - - - -

pr at zpii pr at zpii MPa 3.87 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-56 C4-8-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-35
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 26.0 26.0 26.0

B-Image sector angle Degree 75.0 20.0 20.0 75.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 0.0 1.8 1.8 19.3

CFM-Box end cm 3.9 5.7 5.7 23.2

CFM-Box width cm - - - -

CFM-Box angle Degree 75.0 7.4 7.4 75.0

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 9

Flow Res - high high high low

PRF KHz 0.1 0.1 0.1 2.4

CFM-Quality - low norm norm norm

CFM-Frequency - high high high low


Table 2-57 Operating Control Conditions : C4-8-D, Color Flow Doppler Mode (CFM)

2.7.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.7 1.4

Index component value - 0.04 0.04 0.02 0.02 -


0.01 0.02 0.01 0.05
0.61* 0.42* 0.50* 1.61*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.50 - - - - -

P W0 mW - 36.3 36.6 32.2

P1x1 W1x1 mW - 32.2 33.3 -

zs z1 cm - - 1.8 - - -

zb zsp cm - - - - 1.5 -

zMI zsp cm 0.7 - - - - -

zpii,α zpii,3 cm 0.7 - - - - -

fawf fc MHz 4.01 4.01 3.15 3.15

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-36 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 150 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 245.5 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 24 - - - - -

Ispta at zpii ITA at zpii mW/cm² 30 - - - - -

pr at zpii pr at zpii MPa 2.75 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-58 C4-8-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 26.0 3.0 3.0

B-Image sector angle Degree 20.0 75.0 75.0 75.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 12

Flow Res - high high high high

PRF KHz 0.1 1.3 2.4 4.0

Speed - 6 6 6 6

cM-Frequency - High High Low Low


Table 2-59 Operating Control Conditions : C4-8-D, Color Motion Doppler Mode (cM)

2.7.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.0 0.4 1.1 0.8

Index component value - 0.36 0.40 0.36 1.10 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.07 - - - - -

P W0 mW - 46.4 46.4 46.4

P1x1 W1x1 mW - 26.1 26.1 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-37
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 2.3 - - -

zb zsp cm - - - - 3.8 -

zMI zsp cm 6.0 - - - - -

zpii,α zpii,3 cm 6.0 - - - - -

fawf fc MHz 2.87 2.87 2.87 2.87

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 154 - - - - -

Ispta at zpii ITA at zpii mW/cm² 521 - - - - -

pr at zpii pr at zpii MPa 0.12 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-60 C4-8-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-61 Operating Control Conditions : C4-8-D, Continuous Wave Doppler Mode (CW)

2.7.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.0 1.3

Index component value - 0.82* 0.82* 0.82* 0.82* -


0.11 0.07 0.11 0.18

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.39 - - - - -

P W0 mW - 42.1 42.1 124.9

P1x1 W1x1 mW - 37.5 37.5 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 3.7 - - - - -

zpii,α zpii,3 cm 3.9 - - - - -

fawf fc MHz 3.66 4.60 4.60 4.60

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-38 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 2025 - - - - -

srr - Hz 56 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 343.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 29 - - - - -

Ispta at zpii ITA at zpii mW/cm² 75 - - - - -

pr at zpii pr at zpii MPa 3.73 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-62 C4-8-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 13.7 4.9 4.9 0.0

B-Image sector end cm 16.3 7.5 7.5 2.6

B-Image sector angle Degree 15.0 15.0 15.0 75.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet resol resol resol

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 4.5 4.5 4.5 12.0


Table 2-63 Operating Control Conditions : C4-8-D, Motion Mode (M)

2.7.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.7 1.2

Index component value - 0.73 0.53 0.59 1.67 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.22 - - - - -

P W0 mW - 42.3 45.3 38.1

P1x1 W1x1 mW - 38.4 41.1 -

zs z1 cm - - 1.5 - - -

zb zsp cm - - - - 3.4 -

zMI zsp cm 3.5 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-39
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zpii,α zpii,3 cm 3.7 - - - - -

fawf fc MHz 3.15 3.99 3.03 4.04

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 293.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 260 - - - - -

Ispta at zpii ITA at zpii mW/cm² 590 - - - - -

pr at zpii pr at zpii MPa 3.20 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-64 C4-8-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 0.80 0.07

Gate depth cm 2.63 3.20 2.92 0.04

PRF KHz 0.9 0.9 5.5 7.0

Power % 100 100 100 100


Table 2-65 Operating Control Conditions : C4-8-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-40 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.8 Probe IC5-9-D

2.8.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 1.0

Index component value - 0.84 0.84 0.84 0.84 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 39.2 39.2 19.8

P1x1 W1x1 mW - 28.0 28.0 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.7 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.50 6.32 6.32 6.32

Other Information

prr PRF Hz 1952 - - - - -

srr - Hz 98 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 328.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 43 - - - - -

Ispta at zpii ITA at zpii mW/cm² 74 - - - - -

pr at zpii pr at zpii MPa 3.33 - - - - -

tMode Components - B B B B
u
e
v
Table 2-66 IC5-9-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 3.0 3.0 3.0

B-Image sector angle Degree 20.0 146.3 146.3 20.0

B-Image sector width cm - - - -

B-Line Dens. - high high high low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet norm norm norm

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-41
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 1.7 1.7 0.3

B-Zoom Box end cm 1.6 2.7 2.7 1.3

B-Zoom Box angle Degree 15.0 80.7 80.7 20.0

B-Zoom Box width cm - - - -

Mode Type - B B B B

Focal Depth cm 1.0 2.0 2.0 1.0


Table 2-67 Operating Control Conditions : IC5-9-D, Brightness Mode (2D)

2.8.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 1.0

Index component value - 0.05 0.05 0.05 0.05 -


0.79* 0.79* 0.79* 0.79*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.95 - - - - -

P W0 mW - 30.2 30.2 27.4

P1x1 W1x1 mW - 29.9 29.9 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.7 - - - - -

zpii,α zpii,3 cm 2.0 - - - - -

fawf fc MHz 5.58 5.58 5.58 5.58

Other Information

prr PRF Hz 1831 - - - - -

srr - Hz 131 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 668.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 101 - - - - -

Ispta at zpii ITA at zpii mW/cm² 224 - - - - -

pr at zpii pr at zpii MPa 4.07 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-68 IC5-9-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-42 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 16.0

B-Image sector angle Degree 134.4 30.0 30.0 30.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 0.0 1.1 1.1 0.0

CFM-Box end cm 2.4 3.5 3.5 2.4

CFM-Box width cm - - - -

CFM-Box angle Degree 134.4 12.0 12.0 12.0

CFM-Power % 100 100 100 100

Ensemble - 31 31 31 31

Line Density - 1 1 1 10

Flow Res - low low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - norm norm norm norm

CFM-Frequency - mid mid mid mid


Table 2-69 Operating Control Conditions : IC5-9-D, Color Flow Doppler Mode (CFM)

2.8.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.1 0.9

Index component value - 0.22 0.22 0.36 0.36 -


0.04 0.03 0.07 0.09
0.48* 0.27* 0.17* 0.69*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.98 - - - - -

P W0 mW - 16.4 6.7 33.0

P1x1 W1x1 mW - 16.2 6.7 -

zs z1 cm - - 1.4 - - -

zb zsp cm - - - - 1.0 -

zMI zsp cm 0.9 - - - - -

zpii,α zpii,3 cm 1.0 - - - - -

fawf fc MHz 5.67 6.20 5.23 5.50

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-43
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 822 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 354.7 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 154 - - - - -

Ispta at zpii ITA at zpii mW/cm² 225 - - - - -

pr at zpii pr at zpii MPa 3.51 - - - - -

tMode Components - B B B B*
u M M M M
e
cM* cM* cM* cM
v
Table 2-70 IC5-9-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 3.0

B-Image sector angle Degree 20.0 20.0 20.0 146.3

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 12 12 16

Flow Res - high high high high

PRF KHz 0.9 1.8 4.0 0.9

Speed - 6 6 6 6

cM-Frequency - Mid High Low Mid


Table 2-71 Operating Control Conditions : IC5-9-D, Color Motion Doppler Mode (cM)

2.8.4 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.1 1.1

Index component value - 0.86* 0.86* 0.85* 0.85* -


0.08 0.06 0.09 0.27

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 39.8 39.6 55.8

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-44 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

P1x1 W1x1 mW - 28.5 28.3 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.7 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.50 6.32 6.32 6.32

Other Information

prr PRF Hz 2100 - - - - -

srr - Hz 105 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 328.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 46 - - - - -

Ispta at zpii ITA at zpii mW/cm² 79 - - - - -

pr at zpii pr at zpii MPa 3.33 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-72 IC5-9-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 2.4 0.0 0.8

B-Image sector end cm 1.6 4.0 1.0 1.8

B-Image sector angle Degree 15.0 80.7 80.7 146.3

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet norm norm norm

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 1.0 1.0 1.0 1.0


Table 2-73 Operating Control Conditions : IC5-9-D, Motion Mode (M)

2.8.5 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.3 1.0

Index component value - 0.94 0.56 0.69 1.31 -

Associated Acoustic Parameters

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-45
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.04 - - - - -

P W0 mW - 35.6 29.2 35.6

P1x1 W1x1 mW - 35.6 29.2 -

zs z1 cm - - 1.4 - - -

zb zsp cm - - - - 2.1 -

zMI zsp cm 1.9 - - - - -

zpii,α zpii,3 cm 2.1 - - - - -

fawf fc MHz 5.91 5.55 4.99 5.55

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 860.7 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 218 - - - - -

Ispta at zpii ITA at zpii mW/cm² 544 - - - - -

pr at zpii pr at zpii MPa 4.26 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-74 IC5-9-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 0.70 1.50

Gate depth cm 0.04 3.65 3.41 3.65

PRF KHz 0.9 4.4 10.0 4.4

Power % 100 100 100 100


Table 2-75 Operating Control Conditions : IC5-9-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-46 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.9 Probe ML6-15-D

2.9.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.5 0.5

Index component value - 0.14* 0.14* 0.14* 0.14* -


0.14 0.14 0.14 0.14
0.14 0.14 0.14 0.14
0.14 0.14 0.14 0.14

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.41 - - - - -

P W0 mW - 3.5 3.5 14.7

P1x1 W1x1 mW - 3.4 3.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.4 - - - - -

zpii,α zpii,3 cm 1.4 - - - - -

fawf fc MHz 7.43 8.39 8.39 6.43

Other Information

prr PRF Hz 1650 - - - - -

srr - Hz 55 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 578.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 16 - - - - -

Ispta at zpii ITA at zpii mW/cm² 30 - - - - -

pr at zpii pr at zpii MPa 4.86 - - - - -

tMode Components - B* B* B* B*
u B B B B
e
B B
v
B B
Table 2-76 ML6-15-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 16.0

B-Image sector angle Degree - - - -

B-Image sector width cm 5.0 5.0 5.0 5.0

B-Line Dens. - high high high low

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-47
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - High - - Low

S/PRI - - 1 1 -

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 3.0 3.0 0.0

B-Zoom Box end cm 1.6 4.6 4.6 1.6

B-Zoom Box angle Degree - - - -

B-Zoom Box width cm 0.8 0.8 0.8 2.9

Mode Type - HI BF BF HI

Focal Depth cm 1.5 3.6 3.6 4.7


Table 2-77 Operating Control Conditions : ML6-15-D, Brightness Mode (2D)

2.9.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 0.7

Index component value - 0.01 0.01 0.01 0.01 -


0.77* 0.77* 0.77* 0.77*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.14 - - - - -

P W0 mW - 21.2 21.2 24.5

P1x1 W1x1 mW - 20.8 20.8 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 2.4 - - - - -

zpii,α zpii,3 cm 2.4 - - - - -

fawf fc MHz 6.33 7.73 7.73 6.30

Other Information

prr PRF Hz 853 - - - - -

srr - Hz 171 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 613.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 83 - - - - -

Ispta at zpii ITA at zpii mW/cm² 246 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-48 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 5.32 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-78 ML6-15-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 16.0

B-Image sector angle Degree - - - -

B-Image sector width cm 5.0 5.0 5.0 5.0

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

CFM-Box start cm 0.6 11.9 11.9 11.9

CFM-Box end cm 3.0 14.3 14.3 14.3

CFM-Box width cm 0.7 0.7 0.7 0.7

CFM-Box angle Degree - - - -

CFM-Power % 100 100 100 100

Ensemble - 22 25 25 25

Line Density - 1 10 10 10

Flow Res - high low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - low mid mid low


Table 2-79 Operating Control Conditions : ML6-15-D, Color Flow Doppler Mode (CFM)

2.9.3 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.6 0.5

Index component value - 0.24 0.24 0.18* 0.18* -


0.25* 0.25* 0.17 0.17
0.02 0.01 0.06 0.11
0.02 0.01 0.06 0.11

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.41 - - - - -

P W0 mW - 6.8 5.8 8.5

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-49
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

P1x1 W1x1 mW - 6.8 5.8 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.4 - - - - -

zpii,α zpii,3 cm 1.4 - - - - -

fawf fc MHz 7.43 7.59 6.43 6.43

Other Information

prr PRF Hz 6132 - - - - -

srr - Hz 39 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 578.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 10 - - - - -

Ispta at zpii ITA at zpii mW/cm² 18 - - - - -

pr at zpii pr at zpii MPa 4.86 - - - - -

tMode Components - B* B B* B*
u B B* B B
e
M M M M
v
M M M M
Table 2-80 ML6-15-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 2.4 0.6 13.2 9.0

B-Image sector end cm 4.0 2.2 14.8 10.6

B-Image sector angle Degree - - - -

B-Image sector width cm 5.0 0.8 0.8 0.8

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - resol resol penet penet

M-Power % 100 100 100 100

Mode Type - HI HI HI HI

Focal Depth cm 1.5 3.3 3.3 4.7


Table 2-81 Operating Control Conditions : ML6-15-D, Motion Mode (M)

2.9.4 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.7 0.8 0.8 0.8

Index component value - 0.79 0.40 0.77 0.83 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-50 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 1.82 - - - - -

P W0 mW - 21.7 25.8 25.8

P1x1 W1x1 mW - 21.7 25.8 -

zs z1 cm - - 1.3 - - -

zb zsp cm - - - - 2.3 -

zMI zsp cm 2.5 - - - - -

zpii,α zpii,3 cm 2.6 - - - - -

fawf fc MHz 6.31 7.69 6.25 6.25

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 194.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 169 - - - - -

Ispta at zpii ITA at zpii mW/cm² 955 - - - - -

pr at zpii pr at zpii MPa 3.89 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-82 ML6-15-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 1.50 1.50

Gate depth cm 4.80 3.64 3.64 3.64

PRF KHz 0.9 5.5 0.9 5.5

Power % 100 100 100 100


Table 2-83 Operating Control Conditions : ML6-15-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-51
Acoustic Output & Index Determination Tables

2.10 Probe P2D

2.10.1 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.5 2.3 1.4

Index component value - 0.51 0.44 0.51 2.30 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.11 - - - - -

P W0 mW - 54.9 54.9 54.9

P1x1 W1x1 mW - 51.4 51.4 -

zs z1 cm - - 1.5 - - -

zb zsp cm - - - - 1.6 -

zMI zsp cm 2.9 - - - - -

zpii,α zpii,3 cm 2.9 - - - - -

fawf fc MHz 2.09 2.09 2.09 2.09

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 320 - - - - -

Ispta at zpii ITA at zpii mW/cm² 488 - - - - -

pr at zpii pr at zpii MPa 0.13 - - - - -

Mode Components - CW CW CW CW
Table 2-84 P2D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-85 Operating Control Conditions : P2D, Continuous Wave Doppler Mode (CW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-52 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.11 Probe P6D

2.11.1 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.0 0.2 0.7 0.5

Index component value - 0.20 0.16 0.20 0.74 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.10 - - - - -

P W0 mW - 6.8 6.8 6.8

P1x1 W1x1 mW - 6.8 6.8 -

zs z1 cm - - 0.6 - - -

zb zsp cm - - - - 1.1 -

zMI zsp cm 1.1 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 6.21 6.21 6.21 6.21

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 320 - - - - -

Ispta at zpii ITA at zpii mW/cm² 520 - - - - -

pr at zpii pr at zpii MPa 0.13 - - - - -

Mode Components - CW CW CW CW
Table 2-86 P6D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-87 Operating Control Conditions : P6D, Continuous Wave Doppler Mode (CW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-53
Acoustic Output & Index Determination Tables

2.12 Probe RAB2-5-D

2.12.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.5 1.4

Index component value - 0.13* 0.13* 0.13* 0.13* -


0.13 0.13 0.13 0.13
0.13 0.13 0.13 0.13
0.13 0.13 0.13 0.13

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.21 - - - - -

P W0 mW - 18.6 18.6 33.0

P1x1 W1x1 mW - 13.1 13.1 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.2 - - - - -

zpii,α zpii,3 cm 4.3 - - - - -

fawf fc MHz 3.13 2.12 2.12 2.12

Other Information

prr PRF Hz 2090 - - - - -

srr - Hz 55 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 254.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 29 - - - - -

Ispta at zpii ITA at zpii mW/cm² 77 - - - - -

pr at zpii pr at zpii MPa 3.51 - - - - -

tMode Components - B B* B* B*
u B B B
e
B B B
v
B B B
Table 2-88 RAB2-5-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 30.0 30.0 30.0

B-Image sector angle Degree 20.0 20.0 20.0 80.0

B-Image sector width cm - - - -

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-54 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Foc. Zones - 1 1 1 1

Frequency - penet CPI CPI CPI

S/PRI - - 1 1 1

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 16.9 16.9 22.5

B-Zoom Box end cm 3.0 19.9 19.9 25.5

B-Zoom Box angle Degree 15.0 15.0 15.0 15.0

B-Zoom Box width cm - - - -

Mode Type - B Contr Contr Contr

Focal Depth cm 5.0 5.0 5.0 15.0


Table 2-89 Operating Control Conditions : RAB2-5-D, Brightness Mode (2D)

2.12.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.5 1.2

Index component value - 0.02 0.02 0.02 0.02 -


0.49* 0.49* 0.49* 0.49*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.24 - - - - -

P W0 mW - 62.3 62.3 55.0

P1x1 W1x1 mW - 43.6 43.6 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 4.2 - - - - -

zpii,α zpii,3 cm 4.3 - - - - -

fawf fc MHz 3.20 2.37 2.37 2.37

Other Information

prr PRF Hz 819 - - - - -

srr - Hz 410 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 327.7 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 111 - - - - -

Ispta at zpii ITA at zpii mW/cm² 292 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-55
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr at zpii pr at zpii MPa 3.59 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-90 RAB2-5-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 30.0 30.0 30.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

CFM-Box start cm 0.0 2.1 2.1 0.0

CFM-Box end cm 4.5 6.6 6.6 4.5

CFM-Box width cm - - - -

CFM-Box angle Degree 7.6 7.6 7.6 7.6

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 10

Flow Res - high low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - high low low low


Table 2-91 Operating Control Conditions : RAB2-5-D, Color Flow Doppler Mode (CFM)

2.12.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 1.9 1.4

Index component value - 0.03 0.03 0.03 0.03 -


0.01 0.01 0.01 0.02
0.31* 0.54* 0.44* 1.83*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.24 - - - - -

P W0 mW - 76.2 49.6 38.1

P1x1 W1x1 mW - 26.7 38.2 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-56 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 2.9 - - -

zb zsp cm - - - - 3.7 -

zMI zsp cm 5.5 - - - - -

zpii,α zpii,3 cm 5.6 - - - - -

fawf fc MHz 3.20 2.41 2.41 2.41

Other Information

prr PRF Hz 538 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 345.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 154 - - - - -

Ispta at zpii ITA at zpii mW/cm² 552 - - - - -

pr at zpii pr at zpii MPa 4.16 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-92 RAB2-5-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 30.0 30.0 5.0 5.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 16

Flow Res - high high high high

PRF KHz 0.6 0.9 2.4 3.2

Speed - 6 6 6 6

cM-Frequency - High Low Low Low


Table 2-93 Operating Control Conditions : RAB2-5-D, Color Motion Doppler Mode (cM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-57
Acoustic Output & Index Determination Tables

2.12.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.6 2.7 2.0

Index component value - 0.52 0.56 0.52 2.69 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.08 - - - - -

P W0 mW - 71.6 71.6 71.6

P1x1 W1x1 mW - 55.1 55.1 -

zs z1 cm - - 1.4 - - -

zb zsp cm - - - - 1.4 -

zMI zsp cm 1.1 - - - - -

zpii,α zpii,3 cm 1.1 - - - - -

fawf fc MHz 1.96 1.96 1.96 1.96

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 315 - - - - -

Ispta at zpii ITA at zpii mW/cm² 419 - - - - -

pr at zpii pr at zpii MPa 0.09 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-94 RAB2-5-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-95 Operating Control Conditions : RAB2-5-D, Continuous Wave Doppler Mode (CW)

2.12.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.5 0.7 0.8

Index component value - 0.32* 0.32* 0.32 0.32 -


0.14 0.12 0.14* 0.43*

Associated Acoustic Parameters

IEC FDA Unit -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-58 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

pr,α at zMI pr.3 at zsp MPa 2.21 - - - - -

P W0 mW - 28.4 11.9 28.4

P1x1 W1x1 mW - 21.3 9.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - 4.0 -

zMI zsp cm 4.2 - - - - -

zpii,α zpii,3 cm 4.3 - - - - -

fawf fc MHz 3.13 3.13 3.13 3.13

Other Information

prr PRF Hz 2138 - - - - -

srr - Hz 56 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 254.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 23 - - - - -

Ispta at zpii ITA at zpii mW/cm² 59 - - - - -

pr at zpii pr at zpii MPa 3.51 - - - - -

tMode Components - B* B* B B*
u M M M* M
e
v
Table 2-96 RAB2-5-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 13.5 13.5 13.5 13.5

B-Image sector end cm 16.5 16.5 16.5 16.5

B-Image sector angle Degree 20.0 15.0 15.0 15.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet penet penet penet

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 5.0 5.0 5.0 5.0


Table 2-97 Operating Control Conditions : RAB2-5-D, Motion Mode (M)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-59
Acoustic Output & Index Determination Tables

2.12.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 2.1 1.3

Index component value - 0.44 0.61 0.50 2.06 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 1.94 - - - - -

P W0 mW - 81.3 71.8 57.6

P1x1 W1x1 mW - 39.4 45.3 -

zs z1 cm - - 2.5 - - -

zb zsp cm - - - - 4.3 -

zMI zsp cm 4.8 - - - - -

zpii,α zpii,3 cm 4.8 - - - - -

fawf fc MHz 2.41 2.33 2.33 2.32

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 246.9 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 180 - - - - -

Ispta at zpii ITA at zpii mW/cm² 413 - - - - -

pr at zpii pr at zpii MPa 2.92 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-98 RAB2-5-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 0.20 0.20 0.90

Gate depth cm 6.02 9.04 6.06 3.36

PRF KHz 0.9 0.9 3.3 4.4

Power % 100 100 100 100


Table 2-99 Operating Control Conditions : RAB2-5-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-60 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.13 Probe RAB6-D

2.13.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 1.2

Index component value - 0.69 0.69 0.69 0.69 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.30 - - - - -

P W0 mW - 58.2 58.2 64.0

P1x1 W1x1 mW - 42.4 42.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 2.2 - - - - -

zpii,α zpii,3 cm 2.4 - - - - -

fawf fc MHz 3.40 3.40 3.40 2.89

Other Information

prr PRF Hz 2064 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 186.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 25 - - - - -

Ispta at zpii ITA at zpii mW/cm² 60 - - - - -

pr at zpii pr at zpii MPa 3.06 - - - - -

tMode Components - B B B B
u
e
v
Table 2-100 RAB6-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 3.0 3.0 3.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - high low low high

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet penet penet Mid

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-61
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.0 0.0 0.0

B-Zoom Box end cm 2.6 2.0 2.0 2.0

B-Zoom Box angle Degree 15.0 15.0 15.0 15.0

B-Zoom Box width cm - - - -

Mode Type - B B B HI

Focal Depth cm 3.0 6.0 6.0 7.0


Table 2-101 Operating Control Conditions : RAB6-D, Brightness Mode (2D)

2.13.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 0.8 1.4

Index component value - 0.06 0.06 0.06 0.06 -


0.76* 0.76* 0.76* 0.76*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.19 - - - - -

P W0 mW - 59.4 59.4 53.4

P1x1 W1x1 mW - 52.4 52.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 3.9 - - - - -

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 3.07 3.05 3.05 3.05

Other Information

prr PRF Hz 1751 - - - - -

srr - Hz 117 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 225.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 24 - - - - -

Ispta at zpii ITA at zpii mW/cm² 59 - - - - -

pr at zpii pr at zpii MPa 3.34 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-102 RAB6-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-62 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 26.0 26.0 26.0

B-Image sector angle Degree 63.4 25.1 25.1 25.1

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 1.8 1.8 1.8 0.0

CFM-Box end cm 5.7 5.7 5.7 3.9

CFM-Box width cm - - - -

CFM-Box angle Degree 63.4 7.2 7.2 7.2

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 10

Flow Res - high low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - low low low low


Table 2-103 Operating Control Conditions : RAB6-D, Color Flow Doppler Mode (CFM)

2.13.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 2.1 1.5

Index component value - 0.12 0.12 0.02 0.02 -


0.02 0.02 0.02 0.06
0.62* 0.42* 0.62* 2.04*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.19 - - - - -

P W0 mW - 36.7 46.8 30.8

P1x1 W1x1 mW - 32.4 42.5 -

zs z1 cm - - 1.9 - - -

zb zsp cm - - - - 1.4 -

zMI zsp cm 0.6 - - - - -

zpii,α zpii,3 cm 0.6 - - - - -

fawf fc MHz 3.07 4.02 3.07 3.07

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-63
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 545 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 172.5 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 86 - - - - -

Ispta at zpii ITA at zpii mW/cm² 98 - - - - -

pr at zpii pr at zpii MPa 2.33 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-104 RAB6-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 26.0 26.0 3.0 3.0

B-Image sector angle Degree 20.0 20.0 63.4 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 12

Flow Res - high high high high

PRF KHz 0.6 1.3 2.4 6.0

Speed - 6 6 6 6

cM-Frequency - Low Mid Low Low


Table 2-105 Operating Control Conditions : RAB6-D, Color Motion Doppler Mode (cM)

2.13.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.1 0.6 2.2 1.7

Index component value - 0.63 0.55 0.63 2.17 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.10 - - - - -

P W0 mW - 51.9 51.9 51.9

P1x1 W1x1 mW - 47.1 47.1 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-64 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 1.2 - - -

zb zsp cm - - - - 1.2 -

zMI zsp cm 1.2 - - - - -

zpii,α zpii,3 cm 1.2 - - - - -

fawf fc MHz 2.82 2.82 2.82 2.82

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 287 - - - - -

Ispta at zpii ITA at zpii mW/cm² 418 - - - - -

pr at zpii pr at zpii MPa 0.12 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-106 RAB6-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-107 Operating Control Conditions : RAB6-D, Continuous Wave Doppler Mode (CW)

2.13.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.1 1.2

Index component value - 0.53* 0.53* 0.43 0.43 -


0.17 0.13 0.20* 0.69*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.30 - - - - -

P W0 mW - 44.1 15.9 42.4

P1x1 W1x1 mW - 32.6 14.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - 1.8 -

zMI zsp cm 2.2 - - - - -

zpii,α zpii,3 cm 2.4 - - - - -

fawf fc MHz 3.40 3.40 2.89 2.89

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-65
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 3375 - - - - -

srr - Hz 89 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 186.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 31 - - - - -

Ispta at zpii ITA at zpii mW/cm² 74 - - - - -

pr at zpii pr at zpii MPa 3.06 - - - - -

tMode Components - B* B* B B*
u M M M* M
e
v
Table 2-108 RAB6-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 5.8 5.8 13.7 13.7

B-Image sector end cm 8.4 8.4 16.3 16.3

B-Image sector angle Degree 20.1 15.0 15.0 15.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet penet norm norm

M-Power % 100 100 100 100

Mode Type - M M HI HI

Focal Depth cm 3.0 4.5 3.0 3.0


Table 2-109 Operating Control Conditions : RAB6-D, Motion Mode (M)

2.13.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.9 1.5

Index component value - 0.86 0.75 0.86 1.90 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.19 - - - - -

P W0 mW - 78.2 78.2 78.2

P1x1 W1x1 mW - 59.6 59.6 -

zs z1 cm - - 2.0 - - -

zb zsp cm - - - - 4.0 -

zMI zsp cm 3.9 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-66 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zpii,α zpii,3 cm 4.0 - - - - -

fawf fc MHz 3.07 3.03 3.03 3.03

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 230.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 189 - - - - -

Ispta at zpii ITA at zpii mW/cm² 466 - - - - -

pr at zpii pr at zpii MPa 3.33 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-110 RAB6-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 1.50 1.50

Gate depth cm 5.22 13.00 13.00 13.00

PRF KHz 0.9 14.0 14.0 14.0

Power % 100 100 100 100


Table 2-111 Operating Control Conditions : RAB6-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-67
Acoustic Output & Index Determination Tables

2.14 Probe RIC5-9-D

2.14.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.5 0.4 0.4 0.6

Index component value - 0.42 0.42 0.42 0.42 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.03 - - - - -

P W0 mW - 14.1 14.1 13.1

P1x1 W1x1 mW - 14.1 14.1 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.6 - - - - -

zpii,α zpii,3 cm 1.7 - - - - -

fawf fc MHz 4.18 6.30 6.30 5.44

Other Information

prr PRF Hz 965 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 286.4 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 32 - - - - -

Ispta at zpii ITA at zpii mW/cm² 51 - - - - -

pr at zpii pr at zpii MPa 3.77 - - - - -

tMode Components - B* B B B
u B
e
v
Table 2-112 RIC5-9-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 3.0 3.0 3.0

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - Low norm norm penet

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-68 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 2.0 2.0 0.0

B-Zoom Box end cm 1.6 3.0 3.0 1.0

B-Zoom Box angle Degree 15.0 20.0 20.0 20.0

B-Zoom Box width cm - - - -

Mode Type - HI B B B

Focal Depth cm 2.0 8.0 8.0 1.0


Table 2-113 Operating Control Conditions : RIC5-9-D, Brightness Mode (2D)

2.14.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.5 0.5 0.5 0.7

Index component value - 0.02 0.02 0.02 0.02 -


0.46* 0.46* 0.46* 0.46*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.37 - - - - -

P W0 mW - 15.6 15.6 34.4

P1x1 W1x1 mW - 15.4 15.4 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.8 - - - - -

zpii,α zpii,3 cm 1.9 - - - - -

fawf fc MHz 5.20 6.23 6.23 5.05

Other Information

prr PRF Hz 1739 - - - - -

srr - Hz 124 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 706.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 52 - - - - -

Ispta at zpii ITA at zpii mW/cm² 102 - - - - -

pr at zpii pr at zpii MPa 4.54 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-114 RIC5-9-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-69
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 16.0

B-Image sector angle Degree 134.4 30.0 30.0 134.4

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

CFM-Box start cm 0.0 1.1 1.1 13.6

CFM-Box end cm 2.4 3.5 3.5 16.0

CFM-Box width cm - - - -

CFM-Box angle Degree 134.4 8.0 8.0 134.4

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 22

Line Density - 1 1 1 9

Flow Res - high low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm low

CFM-Frequency - low high high low


Table 2-115 Operating Control Conditions : RIC5-9-D, Color Flow Doppler Mode (CFM)

2.14.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.5 0.5 1.0 0.6

Index component value - 0.05 0.05 0.12 0.12 -


0.01 0.01 0.04 0.07
0.40* 0.23* 0.18* 0.78*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.37 - - - - -

P W0 mW - 13.6 7.2 7.3

P1x1 W1x1 mW - 13.4 7.2 -

zs z1 cm - - 1.4 - - -

zb zsp cm - - - - 1.0 -

zMI zsp cm 0.9 - - - - -

zpii,α zpii,3 cm 1.0 - - - - -

fawf fc MHz 5.20 6.20 5.20 5.44

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-70 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 822 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 505.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 245 - - - - -

Ispta at zpii ITA at zpii mW/cm² 348 - - - - -

pr at zpii pr at zpii MPa 3.94 - - - - -

tMode Components - B B B B*
u M M M M
e
cM* cM* cM* cM
v
Table 2-116 RIC5-9-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 16.0 16.0 16.0 3.0

B-Image sector angle Degree 20.0 20.0 146.3 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - penet penet penet penet

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 8 8

Flow Res - high high high high

PRF KHz 0.9 0.9 3.2 5.0

Speed - 6 6 6 6

cM-Frequency - Low High Low Low


Table 2-117 Operating Control Conditions : RIC5-9-D, Color Motion Doppler Mode (cM)

2.14.4 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.5 0.4 0.6 0.6

Index component value - 0.36* 0.36* 0.25 0.25 -


0.07 0.04 0.10* 0.36*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.03 - - - - -

P W0 mW - 12.2 4.0 10.5

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-71
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

P1x1 W1x1 mW - 12.0 4.0 -

zs z1 cm - - - - - -

zb zsp cm - - - - 1.8 -

zMI zsp cm 1.6 - - - - -

zpii,α zpii,3 cm 1.7 - - - - -

fawf fc MHz 4.18 6.30 5.44 5.44

Other Information

prr PRF Hz 1565 - - - - -

srr - Hz 6 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 286.4 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 5 - - - - -

Ispta at zpii ITA at zpii mW/cm² 9 - - - - -

pr at zpii pr at zpii MPa 3.77 - - - - -

tMode Components - B* B* B B*
u B M M* M
e
M
v
M
Table 2-118 RIC5-9-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 12.6 8.4 12.0 6.0

B-Image sector end cm 14.2 10.0 13.6 7.6

B-Image sector angle Degree 146.3 30.0 30.0 22.5

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet norm penet penet

M-Power % 100 100 100 100

Mode Type - HI M M M

Focal Depth cm 2.0 6.5 2.0 1.0


Table 2-119 Operating Control Conditions : RIC5-9-D, Motion Mode (M)

2.14.5 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.2 0.5 0.9 0.6

Index component value - 0.53 0.29 0.45 0.94 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-72 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.88 - - - - -

P W0 mW - 17.7 18.7 18.7

P1x1 W1x1 mW - 17.7 18.7 -

zs z1 cm - - 1.4 - - -

zb zsp cm - - - - 1.9 -

zMI zsp cm 1.9 - - - - -

zpii,α zpii,3 cm 2.0 - - - - -

fawf fc MHz 5.64 6.25 4.99 4.99

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 544.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 143 - - - - -

Ispta at zpii ITA at zpii mW/cm² 315 - - - - -

pr at zpii pr at zpii MPa 4.00 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-120 RIC5-9-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 0.70 0.70

Gate depth cm 0.04 3.65 3.41 3.41

PRF KHz 0.9 2.2 12.0 12.0

Power % 100 100 100 100


Table 2-121 Operating Control Conditions : RIC5-9-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-73
Acoustic Output & Index Determination Tables

2.15 Probe RNA5-9-D

2.15.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 0.6 0.9

Index component value - 0.61 0.61 0.61 0.61 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 24.6 24.6 49.5

P1x1 W1x1 mW - 23.2 23.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.8 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.51 5.51 5.51 5.51

Other Information

prr PRF Hz 3279 - - - - -

srr - Hz 182 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 260.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 62 - - - - -

Ispta at zpii ITA at zpii mW/cm² 100 - - - - -

pr at zpii pr at zpii MPa 3.41 - - - - -

Mode Components - B B B B
Table 2-122 RNA5-9-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 18.0 18.0 18.0 3.0

B-Image sector angle Degree 10.0 10.0 10.0 115.7

B-Image sector width cm - - - -

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - norm norm norm norm

S/PRI - - - - -

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.0 0.0 0.0

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-74 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Zoom Box end cm 1.8 1.8 1.8 1.0

B-Zoom Box angle Degree 10.0 10.0 10.0 115.7

B-Zoom Box width cm - - - -

Mode Type - B B B B

Focal Depth cm 1.0 8.0 8.0 1.0


Table 2-123 Operating Control Conditions : RNA5-9-D, Brightness Mode (2D)

2.15.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 1.0

Index component value - 0.01 0.01 0.01 0.01 -


0.74* 0.74* 0.74* 0.74*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 32.1 32.1 53.5

P1x1 W1x1 mW - 27.9 27.9 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.8 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.51 5.55 5.55 5.55

Other Information

prr PRF Hz 609 - - - - -

srr - Hz 6 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 260.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 1 - - - - -

Ispta at zpii ITA at zpii mW/cm² 2 - - - - -

pr at zpii pr at zpii MPa 3.41 - - - - -

Mode Components - B* B B B
CFM CFM* CFM* CFM*
Table 2-124 RNA5-9-D, Color Flow Doppler Mode (CFM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 18.0 18.0 18.0 18.0

B-Image sector angle Degree 115.7 25.0 25.0 115.7

B-Image sector width cm - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-75
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

CFM-Box start cm 0.0 2.5 2.5 0.0

CFM-Box end cm 2.7 5.3 5.3 2.7

CFM-Box width cm - - - -

CFM-Box angle Degree 115.7 12.0 12.0 115.7

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 21

Line Density - 1 1 1 9

Flow Res - high high high high

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm low

CFM-Frequency - mid high high high


Table 2-125 Operating Control Conditions : RNA5-9-D, Color Flow Doppler Mode (CFM)

2.15.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.2 1.0

Index component value - 0.01 0.01 0.01 0.01 -


0.00 0.00 0.00 0.01
0.71* 0.50* 0.54* 1.17*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 31.1 36.1 38.0

P1x1 W1x1 mW - 27.1 24.9 -

zs z1 cm - - 1.3 - - -

zb zsp cm - - - - 1.6 -

zMI zsp cm 0.8 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.51 5.55 4.57 5.55

Other Information

prr PRF Hz 1650 - - - - -

srr - Hz 165 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 260.0 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-76 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 31 - - - - -

Ispta at zpii ITA at zpii mW/cm² 50 - - - - -

pr at zpii pr at zpii MPa 3.41 - - - - -

Mode Components - B* B B B
M M M M
cM cM* cM* cM*
Table 2-126 RNA5-9-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 18.0 18.0 18.0 18.0

B-Image sector angle Degree 10.0 10.0 10.0 10.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - norm norm norm norm

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 16

Flow Res - high high high high

PRF KHz 0.1 0.9 0.9 0.9

Speed - 6 6 6 6

cM-Frequency - Low High Mid High


Table 2-127 Operating Control Conditions : RNA5-9-D, Color Motion Doppler Mode (cM)

2.15.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.0 0.2 0.8 0.6

Index component value - 0.20 0.17 0.20 0.83 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.08 - - - - -

P W0 mW - 9.8 9.8 9.8

P1x1 W1x1 mW - 9.8 9.8 -

zs z1 cm - - 0.6 - - -

zb zsp cm - - - - 0.6 -

zMI zsp cm 0.6 - - - - -

zpii,α zpii,3 cm 0.6 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-77
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

fawf fc MHz 4.31 4.31 4.31 4.31

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 209 - - - - -

Ispta at zpii ITA at zpii mW/cm² 253 - - - - -

pr at zpii pr at zpii MPa 0.09 - - - - -

Mode Components - CW CW CW CW
Table 2-128 RNA5-9-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-129 Operating Control Conditions : RNA5-9-D, Continuous Wave Doppler Mode (CW)

2.15.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 0.7 0.8

Index component value - 0.57* 0.57* 0.34 0.34 -


0.04 0.03 0.19* 0.35*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.93 - - - - -

P W0 mW - 23.0 7.1 39.4

P1x1 W1x1 mW - 21.7 7.1 -

zs z1 cm - - - - - -

zb zsp cm - - - - 2.1 -

zMI zsp cm 0.8 - - - - -

zpii,α zpii,3 cm 0.8 - - - - -

fawf fc MHz 5.51 5.51 5.51 5.51

Other Information

prr PRF Hz 3900 - - - - -

srr - Hz 325 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 260.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 74 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-78 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Ispta at zpii ITA at zpii mW/cm² 119 - - - - -

pr at zpii pr at zpii MPa 3.41 - - - - -

Mode Components - B* B* B B*
M M M* M
Table 2-130 RNA5-9-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 13.5 0.0

B-Image sector end cm 1.8 1.0 15.3 1.8

B-Image sector angle Degree 10.0 25.0 63.6 115.7

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - norm norm norm norm

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 1.0 8.0 4.0 2.5


Table 2-131 Operating Control Conditions : RNA5-9-D, Motion Mode (M)

2.15.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.8 1.1 1.0

Index component value - 0.77 0.54 0.61 1.14 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.85 - - - - -

P W0 mW - 33.5 32.5 33.5

P1x1 W1x1 mW - 29.1 28.3 -

zs z1 cm - - 1.3 - - -

zb zsp cm - - - - 1.6 -

zMI zsp cm 2.0 - - - - -

zpii,α zpii,3 cm 2.1 - - - - -

fawf fc MHz 5.20 5.55 4.54 5.55

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-79
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 312.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 93 - - - - -

Ispta at zpii ITA at zpii mW/cm² 202 - - - - -

pr at zpii pr at zpii MPa 4.09 - - - - -

Mode Components - PW PW PW PW
Table 2-132 RNA5-9-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 0.40 1.50

Gate depth cm 0.04 5.70 5.48 5.70

PRF KHz 0.9 0.9 1.3 0.9

Power % 100 100 100 100


Table 2-133 Operating Control Conditions : RNA5-9-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-80 5504378 Revision 3
Acoustic Output & Index Determination Tables

2.16 Probe RSP6-16-D

2.16.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 0.9 0.8

Index component value - 0.92 0.92 0.92 0.92 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 4.02 - - - - -

P W0 mW - 20.2 20.2 25.3

P1x1 W1x1 mW - 20.2 20.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.0 - - - - -

zpii,α zpii,3 cm 1.0 - - - - -

fawf fc MHz 10.32 9.58 9.58 9.58

Other Information

prr PRF Hz 2284 - - - - -

srr - Hz 54 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 470.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 18 - - - - -

Ispta at zpii ITA at zpii mW/cm² 41 - - - - -

pr at zpii pr at zpii MPa 5.77 - - - - -

tMode Components - B B B B
u
e
v
Table 2-134 RSP6-16-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 7.9 1.9 1.9 1.9

B-Image sector angle Degree - - - -

B-Image sector width cm 3.7 3.7 3.7 3.7

B-Line Dens. - high high high low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - resol norm norm norm

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-81
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.0 0.0 0.0

B-Zoom Box end cm 1.0 1.0 1.0 1.0

B-Zoom Box angle Degree - - - -

B-Zoom Box width cm 0.8 0.8 0.8 0.8

Mode Type - B B B B

Focal Depth cm 0.9 3.2 3.2 5.5


Table 2-135 Operating Control Conditions : RSP6-16-D, Brightness Mode (2D)

2.16.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 0.9 0.8

Index component value - 0.00 0.00 0.00 0.00 -


0.00 0.00 0.00 0.00
0.86* 0.86* 0.86* 0.86*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.15 - - - - -

P W0 mW - 25.4 25.4 25.5

P1x1 W1x1 mW - 21.2 21.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.9 - - - - -

zpii,α zpii,3 cm 0.9 - - - - -

fawf fc MHz 6.36 8.56 8.56 8.40

Other Information

prr PRF Hz 3916 - - - - -

srr - Hz 131 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 268.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 35 - - - - -

Ispta at zpii ITA at zpii mW/cm² 53 - - - - -

pr at zpii pr at zpii MPa 3.84 - - - - -

tMode Components - B B B B
u B B B B
e
CFM* CFM* CFM* CFM*
v
Table 2-136 RSP6-16-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-82 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 7.9 7.9 7.9 7.9

B-Image sector angle Degree - - - -

B-Image sector width cm 3.7 3.7 3.7 3.7

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

CFM-Box start cm 0.0 6.8 6.8 6.8

CFM-Box end cm 1.2 7.9 7.9 7.9

CFM-Box width cm 3.7 0.6 0.6 0.6

CFM-Box angle Degree - - - -

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 10 10 10

Flow Res - low high high low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - low high high high


Table 2-137 Operating Control Conditions : RSP6-16-D, Color Flow Doppler Mode (CFM)

2.16.3 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.9 1.0 0.8

Index component value - 0.85* 0.85* 0.73* 0.73* -


0.07 0.04 0.18 0.25

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 4.02 - - - - -

P W0 mW - 18.6 16.1 23.2

P1x1 W1x1 mW - 18.6 16.1 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 1.0 - - - - -

zpii,α zpii,3 cm 1.0 - - - - -

fawf fc MHz 10.32 9.58 9.58 9.58

Other Information

prr PRF Hz 9901 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-83
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

srr - Hz 236 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 470.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 79 - - - - -

Ispta at zpii ITA at zpii mW/cm² 177 - - - - -

pr at zpii pr at zpii MPa 5.77 - - - - -

tMode Components - B* B* B* B*
u M M M M
e
v
Table 2-138 RSP6-16-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 1.7 6.7 0.0

B-Image sector end cm 1.0 2.7 7.7 1.0

B-Image sector angle Degree - - - -

B-Image sector width cm 0.8 0.8 0.8 0.8

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - resol norm norm norm

M-Power % 100 100 100 100

Mode Type - M M M M

Focal Depth cm 0.9 3.2 3.2 5.5


Table 2-139 Operating Control Conditions : RSP6-16-D, Motion Mode (M)

2.16.4 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 1.2 0.7

Index component value - 0.64 0.38 0.50 1.16 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.15 - - - - -

P W0 mW - 19.4 16.5 19.7

P1x1 W1x1 mW - 18.7 16.5 -

zs z1 cm - - 1.1 - - -

zb zsp cm - - - - 1.0 -

zMI zsp cm 1.0 - - - - -

zpii,α zpii,3 cm 1.0 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-84 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

fawf fc MHz 6.36 7.14 6.31 6.31

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 268.3 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 245 - - - - -

Ispta at zpii ITA at zpii mW/cm² 387 - - - - -

pr at zpii pr at zpii MPa 3.96 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-140 RSP6-16-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 1.50 0.10 0.10

Gate depth cm 3.18 5.26 2.40 3.18

PRF KHz 0.9 20.0 20.0 0.9

Power % 100 100 100 100


Table 2-141 Operating Control Conditions : RSP6-16-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-85
Acoustic Output & Index Determination Tables

2.17 Probe S4-10-D

2.17.1 Brightness Mode (2D)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 0.6 0.8

Index component value - 0.58 0.58 0.58 0.58 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.11 - - - - -

P W0 mW - 26.2 26.2 14.4

P1x1 W1x1 mW - 26.2 26.2 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 0.6 - - - - -

zpii,α zpii,3 cm 0.7 - - - - -

fawf fc MHz 6.19 4.68 4.68 4.68

Other Information

prr PRF Hz 3147 - - - - -

srr - Hz 105 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 273.1 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 144 - - - - -

Ispta at zpii ITA at zpii mW/cm² 179 - - - - -

pr at zpii pr at zpii MPa 3.56 - - - - -

tMode Components - B B B B
u
e
v
Table 2-142 S4-10-D, Brightness Mode (2D)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 13.7 13.7 13.7 13.7

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - high high high high

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet Mid Mid Mid

S/PRI - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-86 5504378 Revision 3
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Power % 100 100 100 100

B-Zoom Box start cm 0.0 0.0 0.0 7.2

B-Zoom Box end cm 1.4 1.4 1.4 8.6

B-Zoom Box angle Degree 15.0 17.5 17.5 20.0

B-Zoom Box width cm - - - -

Mode Type - B HI HI HI

Focal Depth cm 1.0 8.0 8.0 1.0


Table 2-143 Operating Control Conditions : S4-10-D, Brightness Mode (2D)

2.17.2 Color Flow Doppler Mode (CFM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 0.7 0.8

Index component value - 0.02 0.02 0.02 0.02 -


0.68* 0.68* 0.68* 0.68*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.95 - - - - -

P W0 mW - 25.5 25.5 25.5

P1x1 W1x1 mW - 25.5 25.5 -

zs z1 cm - - - - - -

zb zsp cm - - - - - -

zMI zsp cm 2.0 - - - - -

zpii,α zpii,3 cm 2.1 - - - - -

fawf fc MHz 5.58 5.58 5.58 5.58

Other Information

prr PRF Hz 863 - - - - -

srr - Hz 288 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 337.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 149 - - - - -

Ispta at zpii ITA at zpii mW/cm² 266 - - - - -

pr at zpii pr at zpii MPa 4.41 - - - - -

tMode Components - B B B B
u CFM* CFM* CFM* CFM*
e
v
Table 2-144 S4-10-D, Color Flow Doppler Mode (CFM)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-87
Acoustic Output & Index Determination Tables

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 13.7 13.7 13.7 13.7

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

CFM-Box start cm 0.0 6.3 6.3 6.3

CFM-Box end cm 2.1 8.4 8.4 8.4

CFM-Box width cm - - - -

CFM-Box angle Degree 12.0 12.0 12.0 12.0

CFM-Power % 100 100 100 100

Ensemble - 22 31 31 31

Line Density - 1 1 1 1

Flow Res - low low low low

PRF KHz 0.1 0.1 0.1 0.1

CFM-Quality - low norm norm norm

CFM-Frequency - mid mid mid mid


Table 2-145 Operating Control Conditions : S4-10-D, Color Flow Doppler Mode (CFM)

2.17.3 Color Motion Doppler Mode (cM)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.7 1.1 0.8

Index component value - 0.06 0.06 0.10 0.10 -


0.01 0.01 0.02 0.02
0.59* 0.36* 0.30* 1.00*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.13 - - - - -

P W0 mW - 22.2 11.1 11.1

P1x1 W1x1 mW - 22.2 11.1 -

zs z1 cm - - 1.3 - - -

zb zsp cm - - - - 0.7 -

zMI zsp cm 0.6 - - - - -

zpii,α zpii,3 cm 0.6 - - - - -

fawf fc MHz 6.25 5.59 5.59 5.59

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-88 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 150 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 334.8 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 45 - - - - -

Ispta at zpii ITA at zpii mW/cm² 58 - - - - -

pr at zpii pr at zpii MPa 3.55 - - - - -

tMode Components - B B B B
u M M M M
e
cM* cM* cM* cM*
v
Table 2-146 S4-10-D, Color Motion Doppler Mode (cM)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 0.0 0.0 0.0

B-Image sector end cm 13.7 13.7 13.7 13.7

B-Image sector angle Degree 20.0 20.0 20.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

Frequency - resol resol resol resol

B-Power % 100 100 100 100

cM-Power % 100 100 100 100

Ensemble - 16 16 16 16

Flow Res - high high high high

PRF KHz 0.1 1.3 1.8 1.8

Speed - 6 6 6 6

cM-Frequency - High Mid Mid Mid


Table 2-147 Operating Control Conditions : S4-10-D, Color Motion Doppler Mode (cM)

2.17.4 Continuous Wave Doppler Mode (CW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 0.0 0.3 0.9 0.9

Index component value - 0.25 0.21 0.25 0.92 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 0.08 - - - - -

P W0 mW - 12.1 12.1 12.1

P1x1 W1x1 mW - 12.1 12.1 -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-89
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zs z1 cm - - 0.6 - - -

zb zsp cm - - - - 0.6 -

zMI zsp cm 0.5 - - - - -

zpii,α zpii,3 cm 0.5 - - - - -

fawf fc MHz 4.34 4.34 4.34 4.34

Other Information

prr PRF Hz - - - - - -

srr - Hz - - - - - -

npps - - - - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 0.2 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 218 - - - - -

Ispta at zpii ITA at zpii mW/cm² 254 - - - - -

pr at zpii pr at zpii MPa 0.09 - - - - -

tMode Components - CW CW CW CW
u
e
v
Table 2-148 S4-10-D, Continuous Wave Doppler Mode (CW)

Operating Control Conditions Unit MI TIS TIB TIC

Power % 100 100 100 100


Table 2-149 Operating Control Conditions : S4-10-D, Continuous Wave Doppler Mode (CW)

2.17.5 Motion Mode (M)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 0.7 0.8

Index component value - 0.49* 0.49* 0.33 0.33 -


0.09 0.06 0.13* 0.38*

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 3.11 - - - - -

P W0 mW - 22.1 6.0 12.0

P1x1 W1x1 mW - 22.1 6.0 -

zs z1 cm - - - - - -

zb zsp cm - - - - 2.1 -

zMI zsp cm 2.4 - - - - -

zpii,α zpii,3 cm 2.4 - - - - -

fawf fc MHz 6.19 4.68 4.68 4.68

Other Information

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-90 5504378 Revision 3
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

prr PRF Hz 3450 - - - - -

srr - Hz 157 - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 307.7 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 85 - - - - -

Ispta at zpii ITA at zpii mW/cm² 188 - - - - -

pr at zpii pr at zpii MPa 5.24 - - - - -

tMode Components - B* B* B B*
u M M M* M
e
v
Table 2-150 S4-10-D, Motion Mode (M)

Operating Control Conditions Unit MI TIS TIB TIC

B-Image sector start cm 0.0 2.0 11.3 0.0

B-Image sector end cm 1.4 3.4 12.7 1.4

B-Image sector angle Degree 15.0 15.0 15.0 20.0

B-Image sector width cm - - - -

B-Line Dens. - low low low low

B-Zoom - 1.0 1.0 1.0 1.0

B-Foc. Zones - 1 1 1 1

Frequency - penet norm norm norm

M-Power % 100 100 100 100

Mode Type - M HI HI HI

Focal Depth cm 3.5 8.0 4.0 1.0


Table 2-151 Operating Control Conditions : S4-10-D, Motion Mode (M)

2.17.6 Pulsed Wave Doppler Mode (PW)

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

Maximum Index Value 1.3 0.6 1.1 0.8

Index component value - 0.64 0.40 0.30 1.13 -

Associated Acoustic Parameters

IEC FDA Unit -

pr,α at zMI pr.3 at zsp MPa 2.96 - - - - -

P W0 mW - 24.1 16.1 24.1

P1x1 W1x1 mW - 24.1 16.1 -

zs z1 cm - - 1.2 - - -

zb zsp cm - - - - 1.8 -

zMI zsp cm 2.1 - - - - -

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 2-91
Acoustic Output & Index Determination Tables

Index Label MI TIS TIB TIC

at surface below surface at surface below surface

zpii,α zpii,3 cm 2.2 - - - - -

fawf fc MHz 5.59 5.56 3.87 5.56

Other Information

prr PRF Hz 903 - - - - -

srr - Hz - - - - - -

npps - - 1 - - - - -

Ipa,α at zpii,α IPA.3 at zpii,3 W/cm² 350.0 - - - - -

Ispta,α at zpii,α ITA.3 at zpii,3 mW/cm² 259 - - - - -

Ispta at zpii ITA at zpii mW/cm² 624 - - - - -

pr at zpii pr at zpii MPa 4.54 - - - - -

tMode Components - PW PW PW PW
u
e
v
Table 2-152 S4-10-D, Pulsed Wave Doppler Mode (PW)

Operating Control Conditions Unit MI TIS TIB TIC

Gate size cm 0.07 0.20 0.30 0.20

Gate depth cm 0.04 4.15 0.15 4.15

PRF KHz 0.9 12.0 1.3 12.0

Power % 100 100 100 100


Table 2-153 Operating Control Conditions : S4-10-D, Pulsed Wave Doppler Mode (PW)

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


2-92 5504378 Revision 3
Chapter 3

Accuracies

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 3-1
Accuracies

3.1 Introduction
The possible accuracy of geometric, flow speed or other measurements with this ultrasound
system is a result of various parameters that shall be equally considered. The used images
shall be optimized and scaled to provide the best view of the examined structures. To ensure
this, the correct choice of the ultrasound probe and imaging mode for a certain application
plays an essential role.
It is also important to be aware of increased inaccuracies caused by the ultrasound beam
travelling through inhomogeneous human tissue with layers of various sound speeds.
Therefore differences between operators shall be minimized by standardization of
measurement procedures.

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


3-2 5504378 Revision 3
Accuracies

3.2 Geometric measurements


To obtain good results with distances, areas and volumes the examined structure shall be
placed in the center of the image and zoomed as much as possible. Geometric
measurements, particularly in 3D data, shall only be done in regions where the structures are
clearly resolved. Otherwise the specified accuracy can drop significantly.
Considering a sound speed of 1540 m/s (average for human tissue), following inaccuracies
shall be taken into account for distances >33 mm:

Measurement Accuracy

Distance +/- 3 %

Area +/- 6 %

Circumference +/- 3 %

Volume +/- 9 %

For distances <= 33 mm an absolute inaccuracy <=1 mm can apply.

Note Area and volume accuracies are derived from the values for distance (only distances > 33 mm
are assumed):

• Inaccuracy for areas: <= +/- 6% (Distance 1 x Distance 2)

• Inaccuracy for volumes: <= +/- 9% (Distance 1 x Distance 2 x Distance 3)

In modes where additional information is overlaid to the B-mode (e.g. CFM, PD, Elasto), the
caliper measurements shall be taken with reference to the B-mode image to meet the stated
accuracy.

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 3-3
Accuracies

3.3 Flow speed measurements


Operating mode Max. rel. deviation

CW ≤ ±6%

PW ≤ ±8%

These values can be obtained in the full range provided that a good adjustment of the
ultrasound image according to the Basic User Manual has been done, including a proper angle
correction. Furthermore it shall be considered that these uncertainties are valid for a sound
speed of 1540m/s (average of soft tissue). In heterogeneous tissue, particularly with high fat
content, an additional error of about 5% can occur and shall be taken into account (for more
explicit values respective literature should be studied).

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


3-4 5504378 Revision 3
Accuracies

3.4 Other measurements


Following accuracies apply for other measured values

Measurement Accuracy

Time +/- 3 %

Heart rate +/- 3 %

Strain ratio +/- 32 %

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 3-5
Accuracies

3.5 Accuracies of acoustic parameters


The parameters MI, and the thermal indices TIs, TIb and TIc are calculated by the ultrasound
system depending on the connected probe and the chosen application setting. For their
interpretation and further explanations please refer to the Basic User Manual of the console.
The mechanical index MI and the thermal indices TIs, TIb and TIc are displayed with at least 1
decimal place, starting from 0.1. From a value of 0.4 the deviation from measurements is less
than ±25% for MI and less than ±50% for TIs, TIb and TIc.

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


3-6 5504378 Revision 3
Chapter 4

Probe Settings

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 4-1
Probe Settings

4.1 Probe Settings Voluson™ E6 BT15


Note Please note that not all listed probes may be available at the time of release of this Manual.

All Modes Doppler Modes B-Mode

Probe Designation Center Image Doppler Frequency Depth cm 2D Angle


Frequency MHz
(Probe specific) Low Mid High max. max.

11L-D 7.3 5.26 5.88 7.14 11 -

3Sp-D 3.2 1.85 2.17 2.50 24 90°

4C-D 3.1 2.00 2.86 3.23 36 81°

9L-D 5.5 3.70 4.55 5.26 14 -

C1-5-D 3.4 2.00 2.86 3.23 30 113°

C4-8-D 4.4 3.03 3.45 4.00 26 95°

IC5-9-D 5.8 5.00 5.56 6.25 16 179°

ML6-15-D 9.0 6.25 7.69 9.09 16 -

P2D 2.0 - - - - -

P6D 6.0 - - - - -

RAB2-5-D 3.2 2.33 2.94 3.23 30 98°

RAB6-D 4.7 3.03 4.00 5.00 26 90°

RIC5-9-D 6.6 5.00 5.56 6.25 16 179º

RNA5-9-D 6.1 3.70 4.55 5.56 18 144°

RSP6-16-D 11.5 6.25 7.14 8.33 8 -

S4-10-D 6.3 3.85 5.56 6.25 14 90°

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


4-2 5504378 Revision 3
Probe Settings

4.2 Probe Settings Voluson™ E8 BT15


Note Please note that not all listed probes may be available at the time of release of this Manual.

All Modes Doppler Modes B-Mode

Probe Designation Center Image Doppler Frequency Depth cm 2D Angle


Frequency MHz
(Probe specific) Low Mid High max. max.

11L-D 7.3 5.26 5.88 7.14 11 -

3Sp-D 3.2 1.85 2.17 2.50 24 90°

4C-D 3.1 2.00 2.86 3.23 36 81°

9L-D 5.5 3.70 4.55 5.26 14 -

C1-5-D 3.4 2.00 2.86 3.23 30 113°

C4-8-D 4.4 3.03 3.45 4.00 26 95°

IC5-9-D 5.8 5.00 5.56 6.25 16 179°

ML6-15-D 9.0 6.25 7.69 9.09 16 -

P2D 2.0 - - - - -

P6D 6.0 - - - - -

RAB2-5-D 3.2 2.33 2.94 3.23 30 98°

RAB6-D 4.7 3.03 4.00 5.00 26 90°

RIC5-9-D 6.6 5.00 5.56 6.25 16 179º

RNA5-9-D 6.1 3.70 4.55 5.56 18 144°

RSP6-16-D 11.5 6.25 7.14 8.33 8 -

S4-10-D 6.3 3.85 5.56 6.25 14 90°

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


5504378 Revision 3 4-3
Probe Settings

This page was intentionally left blank.

Voluson™ E6 BT15 / Voluson™ E8 BT15 Advanced Acoustic Output References


4-4 5504378 Revision 3
GE Healthcare Austria GmbH & Co OG
Tiefenbach 15
4871 Zipf
Austria
www.gehealthcare.com