Professional Documents
Culture Documents
FP 528 Highlight
FP 528 Highlight
Combustion System
Seamlessly manage data and graphs Perform calibrations and drift corrections
LECO Corporation
3000 Lakeview Avenue • St. Joseph, MI 49085 • Phone: 800-292-6141 • Fax: 269-982-8977
info@leco.com • www.leco.com • ISO-9001 • No. FM 24045 • LECO is a registered trademark of LECO Corporation.
How it Works
An encapsulated sample is placed into the loading Sample Holder
head of the FP-528, where it is sealed and purged of Holds up to
250 mg sample
any atmospheric gases that have entered during
sampling loading. The sample is then dropped into a
hot furnace and flushed with pure oxygen for very
rapid combustion.
(7)
By-products of combustion—CO2, H2O, NOx, and Collection System
N2—pass through the furnace filter and thermoelectric Ballast collects all
evolved gases;
cooler for subsequent collection in a ballast apparatus. Ensures complete
(7) These collected gases in the ballast are equilibrated, homogenization prior
to determination
and a small aliquot dose is then used for further
conversion of the gases. The remaining aliquot that
has been reduced is measured by the thermal
conductivity cell for nitrogen.
Gas Required
Carrier Gas: 99.99% Helium 40 psi (2.8 bar), ±10%
Combustion Gas: 99.99% Oxygen 40 psi (2.8 bar), ±10%
Pneumatic Gas: Compressed Air, source must be oil and water free; 40 psi (2.8 bar), ±10%
Gas Flow Rates
Carrier Gas: Analysis: 200 mL/min (measure), 30 mL/min (reference)
Conservation: 30 mL/min (measure), 30 mL/min (reference)
Combustion Gas: 1.3 to 6 L/min (user programmable oxygen profile)
o
Furnace Range Up to 975 C
Weight 150 lb (68 kg)
Physical Dimensions 28 in. H x 21 in. W x 23 in. D (71 cm x 53 cm x 58 cm)
Electrical Power Requirements 230 V~ (±10%; at max load), 50/60 Hz, single phase, 10 A, 7,900 BTU/hr
Part Numbers
FP528C Nitrogen/Protein Determinator with PC and Windows-based operating software
Optional Accessories
621-434-110 Printer
751-300-160 4-Place Balance
V~ denotes VAC.
Mixing Ballast
Thermoelectric Volume
Cooler
Expanded features including automation,
O2
Carrier
multiple elements, external PC control,
Gas
expanded sample size ranges, Windows®-
based operating software that supports
Resistance compliance to 21 CFR Part 11, and SmartLine®
(3) Furnace
Remote Diagnostics are available on the
Exhaust
Exhaust
CHN628 Series. For more information,
Lecosorb Anhydrone
Aliquot
Solenoid Lecosorb Anhydrone
N2
request form no. 209-218.
O2 CO2 H2O TC Cell
He H2O Reduces NOx
CO2
Carrier Reference Flow
Gas Exhaust
Thermal Conductivity
Specifications and part numbers may change. Consult LECO for latest information.
LECO Corporation
3000 Lakeview Avenue • St. Joseph, MI 49085 • Phone: 800-292-6141 • Fax: 269-982-8977
info@leco.com • www.leco.com • ISO-9001:2008 HQ-Q-994 • LECO is a registered trademark of LECO Corporation. Delivering the Right Results
Form No. 209-137 7/12-REV5 © 2012 LECO Corporation
TUBE
Figure 6-10
Tube Removal
608-379
QUARTZ WOOL
STRIPS
2"
502-310
STEEL WOOL (5.1)
2"
601-390
FILTER
TUBE
Figure 6-11
Packing the Furnace Filter Tube
Maintenance FP-528 6 21
Theory of Operation
Leak check is a diagnostic test that automatically checks for gas leaks in the
instrument. There is a separate leak check for oxygen and helium.
All leak checks will first pressurize the system and then monitor the pressure for 60
seconds. If the pressure does not change more than ±5 mmHg, then the leak check
will pass.
1. From the Diagnostics menu, click Leak Check. The Leak Check dialog box
will appear.
2. Click on the gas system in the actions box that should be leak checked.
Choices are Oxygen Leak Check, Ballast, Cooler, Furnace, and Helium
Leak Check. The leak check will start and the result will be displayed in
the results box. Refer to Leak Check Definitions, page 8 9, for additional
information.
3. If a leak check determines there is a leak in a system, click View to
display possible leak points within the instrument.
4. Click Close to exit leak check.
8 8 FP-528 Diagnostics
(13.1)
1. From the Samples menu, click Login. The Sample dialog box will appear.
2. Click the drop down arrow to the right of sample name and enter the
desired sample name from the drop down list. The sample name,
method comments, and description will automatically be entered into the
spreadsheet.
3. Continue with sample login and enter the other necessary information.
Operation FP-528 5 27
Balance Configuration .................................................................. 4 19
Balance Configuration Definitions ............................................. 4 19
Barometric Calibration ................................................................. 4 20
Theory of Calibration
Calibration compensates for differences between the total element measured and the
actual element concentration. A programmed equation will correlate the response
between the observed and theoretical measurement of analyte, permitting accurate
measurements to be made.
The Windows® software provides several types of calibration curves. The calibration
curves include single standard calibration, linear, quadratic, and cubic. First perform
a blank calibration, then once a standard calibration has been defined re-calibration
is not necessary unless a cell or flow has been changed. The daily routine will consist
of determining the blank and performing a drift correction.
The first step to be performed is the determination of a method specific blank.
Several blanks should be analyzed. The linearization table is applied to the peak
during the analysis. The area under the peak is calculated and stored as a raw area
(Ar).
4 4 4 4
3 3 3 3
2 2 2 2
1 1 1 1
0 0 0 0
1 9 17 25 33 1 10 19 28 37 1 10 19 28 37 1 10 19 28 37
Time Time Time Time
Raw Area = 19 Raw Area = 18 Raw Area = 19 Raw Are a = 18
If the four blanks displayed above were selected in the software to calculate a blank,
the result would be that the average of the four would be stored as the blank area
(Ab).
19+18+19+18
Ab = - 18.5
4
The blank area, Ab will be subtracted from every sample analyzed after it is set.
After the blank is defined, a calibration curve needs to be defined. Several standards
need to be selected that will cover the operating range of analyte concentration for
the samples of unknown analyte concentration. At least three replicates of each
standard should be analyzed. Care must be taken to accurately compensate for
trapped atmosphere, atmospheric blank, with each analyzed standard. If the
calibration curve is constructed using more than one sample type, the atmospheric
blank must be determined for each standard. This can be accomplished by analyzing
the same sample mass encapsulated and in an open container, dissolving in water,
or palletizing the sample. The difference in the results is the atmospheric blank.
Ti me Time Time
Diagnostics
The Diagnostics chapter explains how to monitor and check the operation of the
instrument hardware. This can help the operator or service technician determine if
the instrument is operating properly. Through the Diagnostics menu, you can
monitor and change the solenoid valve states, test the serial communications port,
view the ambient monitor and electrical switch states, and leak check the system.
Diagnostics FP-528 8 1
(13.7)
Software License
By installing the Licensed Software, the Licensee acknowledges that Licensee has
read it, understands it and agrees to be bound by its terms, and further agree that
this is the complete and exclusive statement of the Software License between the
parties, which supersedes and merges all prior proposals, understandings and other
agreements oral and written, between the parties relating to the subject matter of
this Software License. In the event of any conflict between the terms of this Software
License and the terms of purchase orders submitted by Licensee, the terms of this
Software License shall control.
I. Definitions
"Licensed Software" shall mean the software supplied with this Software License,
which is software developed and distributed by LECO.
"Instruction Manual" shall mean a guide describing the use and operation of the
Licensed Software, together with any related supporting documentation.
"Licensee" shall mean and include any of the Licensee's divisions and departments
within Licensee's organization, but shall not include wholly or partially owned
subsidiaries, affiliates, successors or assigns, or independent third parties.
1 16 FP-528 Introduction
Return to Index (18)