Partial Flux - Measurement Reliability of Lensed LEDs

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Partial Flux - Measurement Reliability of Lensed LEDs

Application Note

Introduction Baffles to
The majority of LED manufacturers use units avoid reflections on Tube to cover
inner tube surface from abient light
of luminous intensity (cd) for the
measurement and classification of LED Device
under test
brightness. For numerous LED types, this
procedure is reliable and reproducible.
However, for LEDs with narrow emission Detector with V(λ)
angle characteristics, this method is response round 1mm²

insufficient. This application note describes optical distance


the procedures for measuring luminous 100mm
intensity, partial flux and luminous flux, along Figure 1: Luminous intensity
with the advantages and disadvantages measurement
relating to LED design.
which can negatively influence a luminous
Luminous Intensity Measurement intensity measurement include:
Luminous intensity is measured in the units - Component tolerances (e.g.: variations
of cd. The luminous intensity Iv is calculated in the emission angle, squint angle)
from the luminous flux dφv, which passes - Positioning tolerance during
through a solid angle dΩ: measurement (e.g.: tipping the
component during contact)
dΦ v  lm 
Iv =  cd =  (Formula 1) The tolerances described above can be
dΩ  sr 
neglected for components with a large
emission angle (e.g.: 120°) and without
Osram Opto Semiconductors follows the primary optics, for example, TOPLED,
recommendations of the International Power TOPLED, SIDELED and
Commission on Illumination (CIE) for MiniTOPLED.
luminous intensity measurements of LEDs
as described in Publication 127-1997, in For components with a narrow emission
which the detector is positioned a distance angle (< ±20°), the described measurement
of 100 mm from the tip of the LED to be difficulties can arise. The tolerances can
measured (see Figure 1). The light-sensitive prevent reproducible measurements from
surface must be round and possess an area being carried out for these LEDs. The nature
of 1 cm². This geometry results in a solid of this problem is not dependent on the
angle of 0.01 sr (Steradiant) through which construction or manufacturer of the LED.
the luminous intensity is measured. The tolerance is most pronounced for radial
In the 2D projection, the solid angle of 0.01 components whose leads are inserted
sr corresponds to an aperture angle of through the board. Because of their
±3.2° (see Figure 2). mounting technique, SMT LEDs provide an
advantage in this area.
In order for luminous intensity to be
measured in a reproducible manner, the
distribution of luminous intensity within the
solid angle must remain constant. Factors

April 22, 2003 page 1 of 5


Luminous Flux Measurement If this requirement were to be extended to
In contrast to the luminous intensity the production environment, a large number
measurement which is used to measure the of calibrations would be necessary for the
intensity of light passing through a narrow various types of LEDs. The logistical
angle, the luminous flux measurement is problems which arise are not in keeping with
used to measure the total power of light the cost effectiveness gained through the
emitted in all directions. The unit of mass production of LEDs. The positional
measurement for luminous flux is the Lumen requirements for such a measurement also
(lm). From a technical standpoint, this impact the highly-automated cost-effective
measurement can be carried out in two production of LEDs. Together, these two
different ways: either by means of a difficulties lead to enormous expenditures
goniometer or by utilizing a so-called which are required to carry out a "true" flux
integrating sphere. measurement under production conditions,
and can only be implemented in an
With a goniometer, the light intensity is extremely limited number of cases such as
recorded as a function of the emission the new high power LEDs (Dragon).
angle. By using the inverse of Formula 1, the
luminous flux can be ascertained: Due to their robust characteristics and ever
increasing brightness, LEDs are continually
Φ v = ∫ Iν dΩ (lm) (Formula 2) found in new applications. Along with the
4π capture of new applications, new demands
are placed on the LEDs. In order to be able
Although this method provides the most to better address the demands, Osram Opto
precise measurement of the luminous flux Semiconductors has developed a new
emitted from an LED, it is extremely time- measurement technique for LEDs with
consuming, and is therefore not well suited narrow emission angles. The partial flux
for production environments. measurement technique takes into
consideration the numerous applications for
When measuring luminous flux by means of
narrow-angle LEDs while smoothing out the
an integrating sphere, the device under test
tolerances created by errors during
is placed inside a hollow sphere which inner
measurement.
surface is coated with a high-grade diffuse
material. The multiple reflections of light
created by the inner surface of the sphere partial flux luminous intensity
measurement measurement
serve to create a distribution of illumination
which is proportional to the luminous flux
emitted from the LED. By recording the
luminous intensity

illumination intensity present at an opening


in the sphere, the entire luminous flux can
be measured. The difficulty in measuring
luminous flux with this procedure arises from
the type-specific calibration which is
necessary, along with the requirement that
the device under test must be completely
enclosed within the sphere. Type-specific
calibration of the integrating sphere implies -90° -60° -30° 0° 30° 60° 90°
that in order to carry out accurate emission angle
measurements, a reference light source
must be used which corresponds to the form Figure 2: Comparison of luminous
and emission characteristics of the device intensity and partial flux measurement
under test.

April 22, 2003 page 2 of 5


Partial Flux Measurement Tolerance Considerations –
In comparison to luminous intensity Luminous Intensity and Partial
measurements, the partial flux measurement Flux Measurements
extends the solid angle used for the The following section addresses errors in
measurement from 0.01 sr to approximately measurement for narrow-angled LEDs.
0.38 sr; the aperture angle is thus increased Figures 3 and 4 show the influence of
from ±3.2° to ±20° (see Figure 2). measurement techniques on mechanical
In order to achieve the partial flux tolerances. This can arise, for example, from
measurement geometry described above, squinting of the device, or from the positional
the tip of the LED to be measured is accuracy of the device.
positioned at a distance of 19.2 mm from the For intensity measurements, a deviation of
measurement opening of an integrating 4° leads to an error of 7.2%. In comparison,
sphere. The measurement opening has a the error introduced with the partial flux
diameter of 14 mm, which results in the measurement is only 1.4%. The
desired measurement angle of ±20°. This reproducibility of the measurement is
allows only the luminous flux from the LED significantly improved.
which lies within the acceptance angle
created by the integrating sphere to be Figure 5 shows 3 emission patterns which
applied to the detector. each result in a flux of 1 lm. In spite of the
identical luminous flux, the luminous
This angular range significantly increases intensity at 0° varies from 1.9 cd to 2.6 cd;
the measurement reproducibility for LEDs this corresponds to a variance of 31%. Here,
with narrow emission angles. In addition, the the partial flux measurement proves much
new measurement geometry is better suited more stable, exhibiting only a variance of
for current applications employing narrow- 14.2% . Since a majority of applications for
angled LEDs than that used for classical narrow-angle LEDs utilize almost the entire
LED luminous intensity measurements. For width of the angular emission characteristic,
these new applications, a secondary optic the advantages of the partial flux
with a limited aperture angle is often used, measurement are clearly visible in
but which always is greater than the ±3.2° comparison to Iv measurements for narrow-
angle used for luminous intensity angle LEDs. The observation of
measurements. measurement accuracy is shown in Table 1.
In order to emphasize the new nature of the Why isn't a flux measurement just carried
partial flux measurements as well as avoid out? A flux measurement requires that the
possible confusion between the units of component is completely enclosed in an
measurement for luminous flux, luminous integrating sphere. For a majority of LED
intensity and partial flux measurements, housings, this is very difficult to implement in
partial flux measurements are expressed in mass production, due to their construction. If
terms of illumination. The unit of the LED is not completely enclosed,
measurement, as a result, is the Lux (lx). compromises will be introduced, and the flux
will not be measured in its entirety. In the
dΦ v  lm 
Ev =  lx = 2  (Formula 3) strictest sense, this should not be declared
dA  m  as a flux measurement. Furthermore, the
entire diffused light from a component is
measured. For narrow-angled LEDs, the
diffused light is normally not significant.

April 22, 2003 page 3 of 5


Figure 3: Positional accuracy (e.g.: squinting, alignment failures)

Figure 4: Error analysis – positional accuracy

April 22, 2003 page 4 of 5


Figure 5: Emission characteristics

Table 1: Measurement accuracy of emission characteristics Figure 5

Summary advantages. By means of this development


The three measurement techniques process, Osram Opto Semiconductors
described – luminous intensity, partial flux employs a measurement technique which is
measurement and flux measurement each well suited for mass production and whose
have their legitimacy. Due to the design and reproducibility is considerably greater than
emission characteristics of narrow-angle that provided by luminous intensity
LEDs, the partial flux measurement measurements.
technique provides considerable

Author: Dr. Werner Marchl, Joachim Reill

About Osram Opto Semiconductors


Osram Opto Semiconductors GmbH, Regensburg, is a wholly owned subsidiary of Osram GmbH,
one of the world’s three largest lamp manufacturers, and offers its customers a range of solutions
based on semiconductor technology for lighting, sensor and visualisation applications. The
company operates facilities in Regensburg (Germany), San José (USA) and Penang (Malaysia).
Further information is available at www.osram-os.com

All information contained in this document has been checked with the greatest care. OSRAM Opto
Semiconductors GmbH can however, not be made liable for any damage that occurs in connection
with the use of these contents.

April 22, 2003 page 5 of 5

You might also like