Dualstream 3 – Operating Principles  

0  Initial Version  A Collins  21/12/16  M Tudge  21/12/16 


REV  MODIFICATION  PREPARED BY  DATE  CHECKED BY  DATE 

© SOLARTRON ISA 2018  Page 1 of 5 
This document and its contents are the exclusive property of Solartron ISA and may not  Revision 0 
be copied, transmitted or communicated without prior permission.  This document and its  
contents do not constitute any contractual obligation unless expressly agreed. 
 
 
Dualstream 3 – Operating Principles 
 
 

CONTENTS 
1  INTRODUCTION ....................................................................................................................... 3 
1.1  DUALSTREAM 3 ......................................................................................................................... 3 
2  DUALSTREAM 3 ALGORITHM ................................................................................................... 4 
2.1  WET GAS CORRECTION .............................................................................................................. 4 
2.2  WATER FRACTION METER ........................................................................................................... 5 
2.3  PVT ........................................................................................................................................ 5 
2.4  SALINITY .................................................................................................................................. 5 
 
CONTENTS  
FIGURE 1 – DUALSTREAM 3 CROSS SECTION ..................................................................................................... 3 
FIGURE 2 ‐ DUALSTREAM 3 CALCULATION FLOW CHART .................................................................................... 4 
FIGURE 3 ‐ EFFECT OF WATER (AND SALINE) ON RESONANT FREQUENCY ................................................................ 5 
FIGURE 4 ‐ PERMITTIVITY OF FLUID COMPONENTS (RELATIVE PERMITTIVITY PROPORTIONAL TO AREA) .............. ERROR! 
BOOKMARK NOT DEFINED. 
 
   

 
 
© SOLARTRON ISA 2018    Page 2 of 5 
This document and its contents are the exclusive property of Solartron ISA and may not  Revision 0 
be copied, transmitted or communicated without prior permission.  This document and its  
contents do not constitute any contractual obligation unless expressly agreed. 
Dualstream 3 – Operating Principles 

1  INTRODUCTION 
1.1 Dualstream 3 
The Dualstream 3 flow meter integrates Solartron ISA’s patented microwave water fraction device 
together  with  the  Dualstream  wet  gas  Venturi  to  provide  the  operator  with  high  accuracy,  on‐line 
three phase measurement of the gas, condensate and water phases.  The current meter design also 
provides  an  indication  of  changes  in  salinity  (for  water  continuous  flow)  to  differentiate  between 
condensed water (low salinity) and formation water (high salinity). 

Figure 1 – Dualstream 3 cross section 
The Venturi meter employs gauge pressure, differential pressure (DP) and temperature transmitters, 
and  returns  the  indicated  mass  flow  rate.    The  patented  water  fraction  device  uses  a  microwave 
resonant  cavity  with  non‐wetted  probes  to  measure  the  mixture  permittivity,  which  is  particularly 
sensitive to  water content.  This information is combined with  PVT data, including  the component 
densities  and  the  condensate  to  gas  mass  ratio  (CGMR).    An  iterative  sequence  of  calculations  is 
performed to determine the water and condensate flow rates, and apply a correction to the gas rate 
due to the presence of the liquids. 

Please note that the Dualstream 3 has previously been marketed as the Dualstream Elite. 

© SOLARTRON ISA 2018  Page 3 of 5 
This document and its contents are the exclusive property of Solartron ISA and may not  Revision 0 
be copied, transmitted or communicated without prior permission.  This document and its  
contents do not constitute any contractual obligation unless expressly agreed. 
Dualstream 3 – Operating Principles 

2 DUALSTREAM 3 ALGORITHM 
2.1 Wet Gas Correction 
The Dualstream 3, in a similar manner to most Venturi‐based flow meters, determines the wet gas 
flow rates by applying a wet gas correction to the standard mass flow equation,  , such that: 

2 ∆   (1)
1 4

where   and   are temperature corrected measurements taken from the flow meter, the discharge 


coefficient   is either taken from ISO 5167:4 [1] or measured at a calibration laboratory, the fluid 
density    is calculated from the pressure and temperature, and the gas expansibility   is calculated: 

1 1
  (2)
1 1 1

∆ (3)

(4)

(5)

where   is the wet gas correction term.  The exact form of the wet gas correction term   used in the 


Dualstream 3 algorithms is part of Solartron ISA’s intellectual property; however, it conforms to the 
dry gas boundary condition ( 1 for dry gas) and is specifically correlated to the wet gas calibration 
data for the flow meter. 

Figure 2 ‐ Dualstream 3 Calculation Flow Chart 

© SOLARTRON ISA 2018  Page 4 of 5 
This document and its contents are the exclusive property of Solartron ISA and may not  Revision 0 
be copied, transmitted or communicated without prior permission.  This document and its  
contents do not constitute any contractual obligation unless expressly agreed. 
Dualstream 3 – Operating Principles 
2.2 Water Fraction Meter 
The Dualstream 3 incorporates a water fraction meter (based on a microwave resonant cavity) with a 
wet gas Venturi meter.  The cavity resonates at a particular frequency dependent upon the properties 
of the fluids present, in particular the permittivity.  As the relative permittivity of water is significantly 
higher  than  for  gas  or  condensate  (see  Error!  Reference  source  not  found.),  the  flow  meter  is 
particularly sensitive to the water content. 

Figure 3 ‐ Effect of water (and saline) on resonant frequency 
2.3 PVT 
The PVT (Pressure, Volume and Temperature) calculations carried out within the flow metering system 
characterise  the  total  hydrocarbon  composition  using  Equation  of  State  (EoS)  software  (typically 
GasVLe  by  DNV.GL).    This  data  provides  the  component  densities  at  line  and  base  conditions,  and 
quantifies the split between gas and condensate at line and base conditions. 
2.4 Salinity 
The presence of salt alters the electrical properties of the water component, which is detected as a 
characteristic signature response in the microwave cavity signals.  By using this to indicate changes in 
the salinity, the Dualstream Elite algorithm continues to provide accurate flow rates for all phase flow 
rates. 