Professional Documents
Culture Documents
Lod 1970
Lod 1970
Most methods of quantitative chemical analysis use standard The proposed method for calculating the detection limit D
series and calibration curves in order to convert the apparatus is based on the relation:
response into the actual sample concentration.
Around the blank however, the problem of sensitivity =
*/-?_ (t\ + ti) (1)
arises: this comes down to the ability to discern a true dif- 1
the blank value, e.g., x times the standard deviation (70). But
when this value is being considered as the difference between It should be emphasized that formula 1, based on previous
the blank and the smallest detectable quantity, the variability statistical work (13, 14), is approximate in nature but quite
of both has to be taken into account (77,12). A i-test can satisfactory for most practical purposes down to 10. =
then be carried out on the difference between the mean of the The actual calculation of the detection limit is as follows:
blank and the mean of a sample with a very low content of the if we have, for a blank of a certain substance or element to be
substance. analyzed, n replications with a mean Xh we can calculate the
The aim of this study was to develop a method for calcu- variance and the standard deviation s, the latter representing an
lating the detection limit covering any chosen concentration estimate of the true standard error per unit. A sample with
range. Especially in routine multi-element analysis, one ele- a very low content, also n times analyzed, results in a mean j?2
ment is found in the concentration range of e.g., 0-2 %, while and a standard deviation which we assume to be equal to the
another is only likely to be detected using a 0-2 ppm standard same s. The sample will contain a detectable quantity when
series, depending on the element level in the sample. It is
evident that the latter case requires the highest sensitivity Xi- }
conditions with regard to analytical equipment. Here, and in
which has probability P of being detected by means of a /-test
trace analysis in general, it is not unusual to find a detection
with significance level ai. Now it is also evident that
limit which is not lower than the one calculated for a standard
series with a larger range, e.g., 0 to 20 ppm, because of the it- ^ Xi +
photomultiplier dark current or other background elevating
phenomena which generally result in lower precision and de- Hence, the minimal detectable quantity D is given by
creased signal-to-noise ratios.
D =
Xi + (2)
Because we are only interested in detecting differences in
(1) “Limitations of Detection in Spectro-chemical Analysis,” one direction with regard to the blank (one-tailed test), the
Hilger and Watts Ltd., London, 1964. value of h has to be read corresponding to twice the signifi-
(2) P. W. J. M. Boumans and F. J. M. J. Maessen, Z. Anal. Chem.,
cance probability. Here, the test was done at the 5% sig-
220, 241 (1966).
(3) H. Delaffolie, Mikrochim. Acta, 1, 304 (1955). nificance level, so that the value of / in the standard tables is
(4) K. Doerffel, “Reinststoffprobleme, II Reinststoffanalytik,” to be read at 0.10 or 10%.
Akademie Verlag, Berlin, 1966. If one is satisfied with a 90% chance (P 0.90) of detecting =
(5) R. Gerbatch and H. Scholtze, Rev. Yena, (2) 89 (1963). D and by combining Equations 1 and 2
(6) H. Kaiser, Z. Anal. Chem., 209, 1 (1965).
(7) Ibid., 216, 80 (1966).
(8) O. G. Koch and G. A. Koch-Dedic, “Handbuch der Spuren- D =
Xl + í
" (f0.10; n—1 4" /0.20; J!_])
analyse,” Springer Verlag, Berlin, 1964. n
(9) . H. Ruessmann and K. R. Brooks, Acta Chim. Acad. Sci.
Hung., 42, 1 (1964).
(10) ASTM “Methods for Emission Spectrochemical Analysis,” (13) W. G. Cochran and G. M. Cox, “Experimental Designs,” 2nd
4th Ed., p 227, Philadelphia, Pa., 1964. ed., J. Wiley and Sons, Inc., New York, 1957, p 19.
(11) L. A. Currie, Anal. Chem., 40, 586 (1968). (14) J. Neyman and B. Tokarska, J. Amer. Statist. Ass., 31, 318
(12) F. Scharf, Experientia, 20 (8), 470 (1964). (1936).
1 + Í0.20 ; n-l) n -
1 + io.20; n-l) become small, it is customary, where justified, to pool the
1 9.39 21 0.92 estimates sft1 and s*22 by calculating
2
2 3.92 22 0.90
3 2.82 23 0.88 , S*12 + S*22
0.86 Sx
-
2.32
—
4 24 *
2
5 2.02 25 0.84
6 1.81 26 0.82
7 1.65 27 0.81 This increases the degrees of freedom from ( — 1) to 2(n —
1),
8 1.54 28 0.79 and the formula becomes
9 1.44 29 0.78
10 1.36 30 0.76 D =
Xl + Si /2 [/o.lO; 2(n—1> + .20; 2<n-l)]
11 1.29 35 0.71
12 1.23 40 0.66 When planning experiments it is sometimes useful also to
13 1.18 45 0.62
know the number of replications required to obtain a detection
14 1.13 50 0.59
15 1.09 55 0.56 limit which is, for example, defined as 1 X s, an ASTM rec-
16 1.06 60 0.54 ommendation (10). From Table I it is easily seen that in this
17 1.02 70 0.50 case 19 replications are needed.
18 0.99 80 0.46
19 0.97 90 0.44
20 0.94 100 0.42 ACKNOWLEDGMENT
The methods normally used for analyzing aged propellant Packed with 2.5% OV-17 and 2.5% QF-1 on 60-80 mesh
Gas Chrom Q
samples for residual stabilizers and other components are a Column temp, 70-230 °C
combination of infrared and thin-layer chromatography as Column temp, program rate, 10 °C/min
reported by Macke (1) or a gas chromatography method pre- Detector temp, 250 °C
viously reported by the author (2). Injection port temp, 70 °C
Helium flow, 65 cc/min
A recent attempt to apply the gas chromatography method
Hydrogen flow, 28 cc/min
to accelerated aged samples in which essentially all of the Air flow, 500 cc/min
stabilizers had been nitrated, was not successful because of Sample size, 0.004 cc
interference of 1,3-dinitroglycerine with the measurement of
low concentrations of resorcinol. Therefore, a new gas chro-
matography method for the analysis of propellant components
was investigated. investigation. The instrument was modified by placing an
injection port inside the oven compartment to allow true on-
EXPERIMENTAL column injection. Detailed instrument parameters are
Apparatus. A commercial gas chromatograph, F&M summarized in Table I.
5754B with dual flame ionization detectors was used in this Column packings were prepared using the Applied Science
Laboratories Hi-EfF Fluidizer. The columns were packed
(1) G. F. Macke, J. Chromatogr., 38, 47 (1968). under vacuum with vigorous vibration. The ends of the
(2) J. M. Tro well and M. C. Philpot, Anal. Chem., 41, 166 (1969). column were then plugged with silanized glass wool and