H30 Installation Guidance
Content
1 Installation Requirements
2 Electric and Hydraulic Connections
3 Reagent Prime & Start-up
4 Quality Control Test
5 Calibration (if needed)
6 QC Verification(if needed)
7 Sample Test
* To load the target values via USB drive, one has to create a new directory: Main directory:
\H30\QC. The target values files shall be under this directory.
1. Make sure the QC has not expired either in open condition or packaged condition.
2. Make sure the QC re-warm procedure is properly done.
3. Make sure there is no hemolysis of QC.
4. Make sure the instrument itself is OK via doing repeatability test, if not OK, try to solve
the instrument failure via contact Edan service engineers.
5. If the repeatability test passes, try to use new bottle of QC. If the QC test fails again,
please try to do the same QC continuously for 3 times, if deviations are at the same
level, then one can do calibration.
If the QC failed, and the repeatability is OK, one shall do calibration on H30
There are three methods as listed below:
3. Automatic calibration - Fresh blood (all parameters are calibrated once in a row)
Step 4: Obtain the previous HGB calibration factors by the following way:
Apply the equation below to calculate the new HGB calibration factor.
Where A= Fresh blood test result on standard machine, B= Fresh blood test result on
H30.
Obtain the new HGB calibration factor and then replace the previous HGB calibration factor with
the new HGB calibration factor.
Therefore, new calibration factor of HGB should be set as 1.007. Calibration factor increased by
0.007
* To load the target values via USB drive, one has to create a new directory: Main
directory: \H30\CALI. The target values files shall be under this directory.
23 ©2018 EDAN Confidential
2018/10/10
H30 Installation Guidance
Calibration - Automatic calibration (with calibrator) ( if needed)
Step 5: Navigate to RESULTS to go to test interface.
Step 6: Press START KEY to aspirate calibrator, test for at least 7 times.
Caution: calibrator shall be used within 7 days after it is open, rewarm the calibrator for at
least 15 minutes at room temperature before doing test. Mix the calibrator thoroughly.
Step 1: Collect one fresh blood sample. ( storage time no more than 4 hours )
Step 2: Apply 5 to 7 measurements on the comparison hematology analyzer and collect the average values of the
parameters that are out of control.
e.g. if you have five parameters that are out of control, collect the average values as follows:
Step 3: Use the average values as target values and go to the calibration screen to fill in the corresponding blanks.
e.g.
Limits calculation
1. Take the standard deviation of the results. One can turn to the link below:
https://www.easycalculation.com/statistics/standard-deviation.php
2. The limit is then 3 times the standard deviation calculated in the first step.
e.g.
For WBC, the test results for 5 times on comparable hematology analyzer are: 10.5, 10.6, 10.3, 10.2, 10.5. Then
standard deviation is 0.16. ( use the link above ). Three times 0.16 is 0.48. One has to take half of this value to make
the limit. So, the limit for WBC is 0.24. Similar calculations can be applied on the other parameters.
Step 5: Press START KEY to aspirate calibrator, test for at least 7 times.
Step 6: Click Calibrate and choose the desired parameters. Press and finish calibration.
Caution: If there are abnormal calibration results, please unselect these results then click calibrate. Unselect the parameters that
don’t need to calibrate.
Step 2: Press to go into the following screen and edit patient sample information .
Step 3: After inputing patient information, press to save changes, then the system will return to previous screen and the sampling
needle will move to the sampling position automatically
Step 5: Press the Start key located behind the sample needle and it will aspirate 9.6ul sample.
Step 6: Remove the sample device when the sample needle has risen up. Then the system will
automatically analyze the sample and display the test results on the current screen.
Step 7: On the result display interface, click TOOLS then select printer icon, the result will be
printed.