Gage R&R - ANOVA Method - Measurements
Gage R&R - ANOVA Method - Measurements
Operators: Operators
Parts: Parts
Measurements: Measurements
ANOVA: crossed
The StatAdvisor
Based on a study involving 2 operators, each measuring 10 parts 3 times, the estimated standard deviation of the measurement process
equals 8.40057. Including parts,the total variation (TV) equals 62.1052. You should compare this total variation to your desired
performance. In addition, you should be sure that the percent total variation due to R&R is relatively small. In this case, the value equals
13.5264%. General rules of thumb classify a measurement system as acceptable if the percent total variation from R&R is less than 10%,
although up to 30% may be acceptable for some situations.
The number of distinct categories (ndc) that can reliably be distinguished by the measurement system analyzed in this study equals 10.
Usually, this number should be at least 5.
Of the total variance, 0.0741234% is due to differences between operators (Reproducibility) while 99.9259% is due to the instrument
(Repeatability). You can use this breakdown to help determine where the measurement system might need improvement.
Run Chart
180 Operators
AE
150 DC
Measurements
120
90
60
30
0
1 2 3 4 5 6 7 8 9 10
Parts
Gage Measurements by Operators
180 Operators
AE
150 DC
120
Average
90
60
30
0
1 2 3 4 5 6 7 8 9 10
Parts
______________________________________________________________________
______________________________________________________________________
Gage R&R - ANOVA Method - Col_6
Operators: Operators
Parts: Parts
Measurements: Col_6
ANOVA: crossed
The StatAdvisor
Based on a study involving 2 operators, each measuring 10 parts 3 times, the estimated standard deviation of the measurement process
equals 11.5828. Including parts,the total variation (TV) equals 11.5828. You should compare this total variation to your desired
performance. In addition, you should be sure that the percent total variation due to R&R is relatively small. In this case, the value equals
100.0%. General rules of thumb classify a measurement system as acceptable if the percent total variation from R&R is less than 10%,
although up to 30% may be acceptable for some situations.
The number of distinct categories (ndc) that can reliably be distinguished by the measurement system analyzed in this study equals 0.
Usually, this number should be at least 5.
Of the total variance, 0.0% is due to differences between operators (Reproducibility) while 90.7237% is due to the instrument
(Repeatability). You can use this breakdown to help determine where the measurement system might need improvement.
Run Chart
170 Operators
AE
160 DC
150
Col_6
140
130
120
110
1 2 3 4 5 6 7 8 9 10
Parts
170 Operators
AE
160 DC
150
Average
140
130
120
110
1 2 3 4 5 6 7 8 9 10
Parts