OXYGEN CONCENTRATORS

OC SERIES
SERVICE MANUAL

Model OC-S
OC-S30, OC-S50,
OC-S80, OC-S100

Model OC-E
OC-E30, OC-E50,
OC-E80, OC-E100

Model OC-P
OC-P30, OC-P50
OC-P80, OC-P100

SysMed (China) Co., Ltd


11-2-3 No.17, Wensu Street,
Hunnan New District, Shenyang, China
Service Tel: +86 24 24697305
TABLE OF CONTENTS

PART I General Safety Notice ..................................................................................... 1


Special Notes
Safety Summary
1.1 Use of oxygen
1.2 Use and maintenance of the device
1.3 Standards and regulations
PART II General Introdcution...................................................................................... 3
2.1 Distributor/dealer responsibility
2.2 Important notice and symbol explanations
2.3 Specifications
Part III Working Principle and Routine Checking Operations.................................. 5
3.1 Working principle
3.2 Checking operation
3.3 Alarm system
3.4 Oxygen concentration test and parameter
Part IV User Instructions............................................................................................. 7
4.1 General instructions
4.2 Routine maintenance by the user
4.2.1 Cleaning the air inlet filter
4.2.2 Cleaning nasal tube and humidifier bottle
Part V Distributor/dealer Maintenance....................................................................... 8
5.1 Routine maintenance
5.1.1 Cabinet air inlet filter
5.1.2 Inlet filter replacement
5.1.3 Bacteria filter replacement
5.1.4 Taking record
5.2 Preparing for new user
Part VI Repair/replacement Guide............................................................................... 9
6.1 Outside, inside structure
Outside structure
Inside structure
Control board layout
6.2 Cabinet removal
6.2.1 Removal cabinet front and back
6.2.2 Main structure
6.2.3 Caster replacement
6.3 Compressor
6.3.1 Compressor replacement
6.3.2 Capacitor replacement
6.4 Pneumatic valve replacement
6.5 Pressure equalizing valve replacement
6.6 Sieve beds replacement
6.6.1 Sieve beds removal
6.6.2 Sieve beds installation
6.7 Radiator and Fan replacement
6.8 Control board replacement
6.8.1 Control board removal and installation
6.8.2 Other notes
Buzzer
Pressure switch
Timing setting key
6.9 Transformer replacement
6.10 LCD replacement
6.11 Pressure regulating valve
6.11.1 Pressure regulating valve checking and setting
6.11.2 Pressure regulating valve removal and installation
6.12 Power switch replacement
6.13 Flow meter replacement
6.14 Circuit breaker replacement
Part VII Trouble Shooting Guide............................................................................... 23
7.1 Outlet pressure test
7.2 General troubleshooting
7.3 Trouble shooting table
7.4 Tools table
PART I General Safety Notice
SPECIAL NOTES
 The procedures in this manual must be performed by supplier, dealer or an appointed maintenance
company only.
 Please have a close read of this user manual, DO NOT use this product or any available optional
equipment without first completely reading and understanding these instructions and any additional
instructional material such as Instruction Sheets supplied with this product or optional equipment,
otherwise, injure or damage may occur.
 If you are unable to understand the warnings, cautions or instructions, contact supplier before
attempting to use this equipment.
 Keep this manual for future reference.
 Do not use unauthorized parts, accessories other than those authorized by SYSMED.

NOTICE: THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO


CHANGE WITHOUT NOTICE.

SAFETY SUMMARY
1.1 Use of oxygen
 Oxygen is not a flammable gas but a combustion-supporting gas, so the machine must be far away
from fire and heat. DO NOT SMOKE while using this device, for your own safe, the user must quit
smoking. Keep all matches, lighted cigarettes or other sources of ignition out of the room.
 Keep power cord away from sources of ignition and heat.
 Turn off the power when nobody uses the machine.
 DO NOT put oxygen tubing under bedspread or chair cushion.
 Unplug the power cord when you clean the machine or change fuse, in order to avoid getting an
electric shock.
 DO NOT use any lubricants unless recommended by SysMed.
 Keep unit in a dry, well-ventilated, without dust, corruption and toxicological gas environment, away
from sunlight and no intense electromagnetic interference

1.2 Use and maintenance of the device


Use the electric cable provided and check that the voltage of the mains socket used complies with
the electrical characteristics of the appliance indicated on the manufacturers plate on the rear of the
appliance.
Do not use an extension cord or multiple sockets which can create sparks and therefore pose a fire
risk.
This equipment is not be used for life supporting or life sustaining. Use of the unit must be restricted
solely to oxygen therapy on medical prescription in compliance with flow and period set by physician.
Do not use in a specifically magnetic environment (MRI, etc.).
The concentrator has an audible alarm intended to warn the user of any problems. The user must
determine the maximum distance away from the unit, based on the sound levels in the environment, to
ensure that the alarm is always audible.
Unplug the power cord when you clean the machine or replace circuit breaker, in order to avoid
getting an electric shock.
1
DO NOT start this device when the flow meter is closed.
The Operating Environment Temperature of this concentrator is during 10-37℃, if lower 10℃, the
compressor may difficult to start to work; if higher than 37℃, the compressor may work overheat, thus it
will cause life reduced of the device.
NEVER block the air inlet of the concentrator or place it on a soft surface, such as a bed or couch,
where the air exhaust outlet may be blocked. Keep the openings free from lint, hair and the like.
Humidifier shall adopt distilled water or cold boiled water, and the water shall be kept under the
maximum scale line.
For optimum performance, SysMed recommends that each concentrator be on and running for a
minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.
Response Time: Acceptable concentration is normally achieved in about 240 seconds, allow 6 minutes to
attain full concentration.
Do NOT turn the bobber in the flow meter up to the red line, otherwise, the purity of oxygen will
reduce.
DO NOT come in contact with the concentrator while wet.
Operate the unit in an upright position, maintaining at least six inches of open space on all sides for
ventilation.

1.3 Standards & regulations


In compliance with [EN60601-1]
“The manufacturer, installer or importer are not considered to be responsible for consequences or the
safety, reliability and characteristics of a device unless,
 The assembly, repairs or replacement have been performed by persons authorized by the
manufacturer,
 The electrical installation of the corresponding premises complies with appropriate regulations and
codes,
 The device is used in accordance with the instructions for its use.
If the replacement parts used for periodic servicing by an approved technician do not comply with the
manufacturer’s specifications, the manufacturer is absolved of all liability in the event of an incident.

Do not open the device when it is powered on: risk of electrocution.


This device complies with the requirements of the EU Directive 93/42/EEC, but its operation may be
affected by the use in the surrounding area of appliances such as diathermy, high frequency
electro-surgical instruments, short wave treatment appliances, cell-phones, microwave ovens, induction
hot plates or remote control toys, and more generally, by electromagnetic interference exceeding the
levels specified in standard IEC(EN) 60601-1-2.

2
PART II General Introduction
2.1 Distributor/dealer responsibility
All Distributor/dealer of the SYSMED OC Series Oxygen Concentrator must assume responsibilities
for handling, operational check-out, user instruction, and maintenance. These responsibilities are
outlined below and throughout this manual.

WARNING SYSMED units must not be used for or with any life-supporting or life sustaining
applications. Patients unable to communicate discomfort while using this device, they may require
additional monitoring. Advise patients to immediately notify their physician(s) and/or dealer in case of an
alarm or any discomfort.

As a Distributor/dealer, you must do all of the following:


 Inspect the condition of each OC unit immediately upon delivery to your business location. Note any
sign of damage, external or internal, on the delivery receipt, and report it directly to both the freight
company and SYSMED immediately.
 Check the operation of each OC concentrator before delivery to a patient. Always operate the unit
for a reasonable length of time and check that the oxygen concentration level is within specifications
as referred to in Part III 3.4. Test the power failure alarm as described in Part III 3.3 of this manual.
 This equipment is not be used for life supporting or life sustaining. Only be used as an oxygen
supplement. Other backup oxygen supply device for the persons who are in dire need of oxygen
must be available. (Such as oxygen cylinder or liquid oxygen)
 Instruct patients how to use the OC concentrator in conjunction with the User Manual.
 Instruct each patient how to perform routine maintenance of the cooling air filter, nasal tube and
humidifier bottle. (Refer to Part IV 4.2) Be available to service each patient at any time. Maintain the
OC concentrator in accordance with Part V and Part VI.
Repair components and replace parts only as outlined in this manual. Use only SYSMED parts for
replacement in OC Oxygen Concentrators.

2.2 Important Notice and Symbol Explanations


As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages.
They identify safety guidelines or other important information.

SYMBOLS
ON
OFF
Warning
No Smoking
Type B Equipment
Upright
CE Certificate
Frangibility
Alternating Current

3
Table 1 function
Parameters
Function and setting
model no.
OC-S30 Standard function: LCD display, temperature alarm, high/low pressure alarm, power failure
OC-S50 alarm, timed setting, intelligent fault diagnosis, repair reminder alarm.

OC-S80
OC-S100
OC-E30 Standard function: LCD display, oxygen concentration indicator (Red, green, yellow bar on
OC-E50 the LCD), temperature alarm, high/low pressure alarm, low oxygen concentration alarm,
power failure alarm, timed setting, intelligent fault diagnosis, repair reminder alarm.
OC-E80
OC-E100
OC-P30 Standard function: LCD display, oxygen concentration indicator (Red, green, yellow bar &
OC-P50 realtime concentration display on the LCD), temperature alarm, high/low pressure alarm,
low oxygen concentration alarm, power failure alarm, timed setting, intelligent fault
OC-P80
diagnosis, repair reminder alarm.
OC-P100
Optional function Positive pressure outlet (for nebulizer)
Table 2 Parameter
Model no. OC-S30 OC-E30 OC-P30 OC-S50 OC-E50 OC-P50
Rating Power (W) 350 350
Rating current(A) 1.59 1.59
Electrical Requirements 220V±22V, 50HZ±1HZ 220V±22V, 50HZ±1HZ
Flow rate(L/min) 0-3 0-5
purity(3L/min) 93±3% 93±3%
purity(5L/min) —— 93±3%
Outlet pressure(Mpa) 0.05±0.005 0.05±0.005
Sound level dB(A) ≤42 ≤42
Equipment Class and Type Class II Type B Class II Type B
Net weight(kg) 25kg 25kg
Dimension (mm) 381x347x689 381x347x689

Model no. OC-S80 OC-E80 OC-P80 OC-S100 OC-E100 OC-P100


Rating Power (W) 480 480
Rating current(A) 2.25 2.25
Electrical Requirements 220V±22V, 50HZ±1HZ 220V±22V, 50HZ±1HZ
Flow rate(L/min) 0-8 0-10
purity(6L/min) 93±3% 93±3%
purity(8L/min) 93±3% 93±3%
purity(10L/min) —— 93±3%
Outlet pressure(Mpa) 0.05±0.005 0.05±0.005
Sound level dB(A) ≤50 ≤50
Equipment Class & Type Class II Type B Class II Type B
Net weight(kg) 27kg 27kg
Dimension (mm) 381x347x689 381x347x689
Notice: The information contained in this manual is subject to change without notice.

4
Part III Working Principle and Routine Checking Operations
3.1 Working principle
Oxygen Concentrator for Medical Use adopts the principle of the world-advanced pressure swing
adsorption technology. At normal temperature, the machine can continuously separate medical oxygen
with high density (93%±3% oxygen for medical use) from the air when its power is turned on.

Note: Using oxygen concentrator doesn’t influence oxygen content in the air of the room.

Air enters the OC Oxygen Concentrator through an external cabinet air intake filter. This filtered air
enters the compressor, and pressurized air then exits the compressor and passes through a radiator,
which reduces the temperature of the compressed air. Next, an pneumatic valve system directs the air
into one of two sieve beds that contain molecular sieve. The special characteristic property of molecular
sieve is that it physically attracts (adsorbs) nitrogen when air passes through this material, thus enabling
the production of high purity oxygen.
There are two sieve beds or adsorbent columns; while one produces high purity oxygen, the other is
purged of the nitrogen it adsorbed (collected) while it was producing oxygen. Each column produces
oxygen and delivers it to the oxygen tank. Oxygen exits the oxygen tank through a pressure regulator,
flow control valve, flow meter, and bacteria filter. The flow control valve, which is part of the flow meter,
controls the flow rate of oxygen delivered to the patient. The OC unit delivers up to 93%±3% oxygen
concentration at rated flow. The remaining constituents of the product gas stream are nitrogen and argon,
both of which are part of the air we breathe, are inert and are completely safe.
3.2 Checking operation
SYSMED runs each device through a burn in period and tests every OC Oxygen Concentrator
thoroughly after manufacture before releasing for shipment. As the dealer, it is your responsibility to
perform the following test to ensure that no damage occurred in shipping or handling.
(1)Open and inspect all concentrator cartons upon receipt. Unpack each unit and remove it from its
carton. Inspect the unit itself for damage. If the exterior of the carton is damaged, or the unit itself is
damaged, note it on the freight bill signed by the driver.
 Before inserting the plug of power cord into the power socket, turn on the power switch and note that
the audible alarm sounds continuously. (See Part III 3.3). And then off the power.
(2)Plug in the power cord of the unit, and set the I/0 (ON/OFF) switch to the I (ON) position. Check

5
to see that the following occurs:
 The compressor runs, listen for the sound.
 Exhaust air flows out of the bottom of the unit.
 OPTIONAL for Units Equipped with Oxygen Concentration Indicator: The green light remains flicker
until the oxygen concentration reaches 85%. (approximately six minutes).
(3)Turn the flow meter adjustment knob counterclockwise to see if it works well, If not, refer to Part
VII troubleshooting guide.
(4)Perform an oxygen concentration test, as described in Part III 3.4.
3.3 Alarm System
The OC Oxygen Concentrator is equipped with alarm system, which sounds a continuous and loud
alarm when a power failure occurs or when one or more cycle variables are not within specification. It
sounds an alarm if the high or low pressure indicators are activated. The alarm remains on until you
correct the alarm condition or you set the I/0 (ON/OFF) switch to the 0 (OFF) position. Refer to Part VII for
a list of probable alarm causes.
3.3.1 Power Failure Alarm Test
To test the power failure alarm, perform the following actions:
Unplug the power cord from the wall outlet, and set the I/0 (ON/OFF) switch to the I (ON) position.
This should immediately activate the audible alarm. If it doesn’t, refer to the troubleshooting guide in Part
VII of this manual.
3.4 Oxygen Concentration Test and Specification
To ensure that the output of oxygen from the device is within specification, you must perform an
oxygen concentration test. Test the unit upon delivery to a patient and at periodic intervals. Dealers,
based on their expertise and documentation, may establish and implement their own plans for checking
oxygen concentration. The interval established may be longer or shorter than 180 days, which is
SYSMED default time period for providers who do not choose to establish their own plan.
(1)Remove oxygen humidifier bottle from the oxygen outlet.
(2)Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
(3)Set the I/0 (ON/OFF) power switch to the I (ON) position. (It takes approximately 6 minutes for
the oxygen concentration to stabilize.) Take oxygen concentration readings over a period of several
minutes to reduce any cyclic variations
(4)Verify that the product flow rate delivered by the unit matches the patient’s prescription and
does not exceed the capacity of the unit.
(5)Disconnect the oxygen analyzer, and reconnect the humidifier bottle and any other equipment /
accessories that may be required.
(6)Adjust the flow meter adjustment knob to the prescribed flow rate.
NOTE: Do not measure oxygen concentration output after the product stream passes through a
humidifier bottle, erroneous readings will result and your oxygen concentration measuring device might
be damaged.

OC Concentration Specifications
Liter Flow Specification Within Specification
Rated flow 93%±3% 90% or higher

Part IV User Instructions


6
4.1 General instructions
It is important that patients thoroughly understand how to operate the SYSMED OC unit. This
enables proper treatment as prescribed by a qualified, licensed physician. You must explain that the
purpose of this therapy is to alleviate symptoms. If patients experience any discomfort or the unit alarms,
they must notify their Home Service Provider and/or physician immediately. You, as the Dealer, are
responsible to see that each patient receives the Patient Manual. Explain each step in the operation of
the unit to the patient in reference to this manual.

4.2 Routine maintenance by the user


To ensure accurate output and efficient operation of the unit, the patient must perform two simple
routine maintenance tasks:
 Cleaning the cabinet air inlet filter
 Cleaning nasal tube and humidifier bottle

4.2.1 Cleaning the cabinet air inlet filter


(1)In the condition of power off, make a clean for the outside case by soft wet towel with little
detergent, and then wipe it up with dry towel, once or twice per month.
(2)Air intake filter is used to prevent dust, and need to be cleaned regularly. It is a critical step for
ordinary maintenance to clean the filter, at least once a week depending on environmental conditions.
The filter may require daily cleaning if the OC unit operates in a harsh environment. Operation steps
as below:
A. Take off the two filter nets on both sides of the case, take out the sponge mesh.
B. Rinse the sponge mesh with clean water. Get ride of extra water and dry up naturally.
C. Set back after dry up for future use.
D. Or use the back up one.

NOTE: If replacement is needed, please use the same SYSMED accessories.


Do not operate the unit without the air intake filter in place.

4.2.2 Cleaning nasal tube and humidifier bottle

(1) Clean the cannula every day. First clean with detergent and then rinse with clean water
completely, and dry up naturally. Changing a new cannula once a month is recommended.
(2) The distilled water or cold boiled water in humidifier should be refreshed every day.
The humidifier should be cleaned and disinfected every 3 weeks, clean the humidifier with detergent
and then rinse with warm or hot water, dip it in household disinfectant for 20-30 minutes, clean again with
warm or hot water, and then dry up naturally. If not use, put it in a clean bag.

Part V Distributor/dealer Maintenance


5.1 Routine maintenance
7
The OC unit has three filters that require scheduled maintenance and replacement. To ensure that
the output of oxygen from the unit is within specification, you must perform an oxygen concentration test.
Test the unit upon delivery to a patient and at periodic intervals. Dealer, based on their expertise and
documentation, may establish and implement their own practices for checking oxygen concentration. The
interval established may be longer or shorter than 180 days, which is the default time period
recommended for providers who do not choose to establish their method.
SYSMED does not require preventive maintenance on the concentrator. You do not need to perform
any maintenance as long as the OC unit remains within specifications at the prescribed flow rate. (Refer
to Part III 3.4)
5.1.1 Cabinet air inlet filter
The external cabinet air filter is located on the two sides of the unit near the bottom; it is easily
removed by hand. Instruct the patient to clean this filter weekly.
(Refer to Part IV 4.2.a.)
5.1.2 Inlet filter replacement
The inlet filter requires changing every 6 months or 4,000 hours of
use. See Fig.1
(1) Turn off the power, and unplug the power cord.
(2) Remove the filter access panel to remove the intake filter.
(3) Replace with a new filter.
(4) Reinstall the access panel.
Fig.1
5.1.3 Bacteria filter replacement
The bacteria filter must be replaced after every 2 years or 16,000
hours of use. See Fig.2
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back to locate the bacteria filter.
NOTE: Observe the position of the filter before removal.
(3) Separate the silicone tubing from both sides of the filter. Fig.2
(4) Install the new filter with the inlet side in the same position as before.
(5) Reinstall the cabinet back.
5.1.4 Taking record
As the dealer, it is suggested that you record all routine maintenance and repairs performed on the
OC unit, including hours and dates of service.
5.2 Preparing for new user
With the growing concern about possible cross infection from home oxygen equipment from one
home care patient to another, appropriate cautions should be exercised.
When you remove the OC unit from a patient’s home, always dispose of the used nasal cannula and
humidifier bottle. Clean the exterior of the OC unit with a soapy water solution or commercial detergent
cleaner to remove any debris, organic or otherwise. Be careful not to get any liquid into the interior of the
unit. Next, clean the exterior with either a common chemical disinfectant and allow it to air dry. Retest the
OC unit before you return it to your inventory.
Replace the cabinet dust filter between each patient’s use or clean with warm soapy water if it is in
good condition. Clean this filter at least once per week or more frequently if operated in a dusty
environment. Change the bacteria filter every 2 years or 16,000 hours of use (under adverse conditions,
filter may need to be changed more frequently if dingy or dirty) It is recommended to change the bacteria
filter between patients.
Part VI Repair/replacement Guide
The design of the OC Oxygen Concentrator allows for easy access and removal of most components.
This allows you to perform scheduled maintenance, repair, and replacement of parts with minimal time
8
and effort. Besides OC series oxygen concentrator has the intelligent fault diagnosis, displays
malfunction code and alarm data on LCD which are convenient for Maintenance personnel to check and
repair the machine.
6.1 Exterior and Interior structure
Exterior structure

1. Humidifying Bottle
2. Connecting Nut
3. Oxygen Outlet
4. Humidifier Connector
4 7 5. Oxygen Flow-meter
2 6
6. Knob of the Flow-meter
9 7. Large LCD Display
1 5 (See Fig. 4)
3 8. Power Switch
8 9. Adjusting Panel and Keys.
See Fig. a 10. Positive Pressure Outlet
(See Fig. 3)
11. Model

10
11

Front View Fig.3

14 12. Filer Access Panal


13. Air Intake Filters
(two sides) 12
14. Silver Sticker
15. Warn Sticker 15
16. Circuit Breaker 16
17. Power Cord Socket 17
18. Caster 18 13

Back View
Side View
7.6 7.5 7.4 7.3 7.2

9
88 7.1

维 护
Maintenance
故 障
Malfunction 低Low氧 富Rich
氧 高High氧

定时时间
Timed setting

当次时间
Running time

故障代码
Malfunction Code
8 8 8:8 8 7.7

累计时间8
Elapsed time
8888 7.8

Fig.4
7.1 Oxygen purity percentage (type OC-P)
7.2 High purity sign
7.3 Rich purity sign
7.4 Low purity sign
7.5 Malfunction alarm
7.6 Maintenance alarm
7.7 Timed setting, Running time and Malfunction code display
7.8 Elapsed time
Interior structure

Pressure
Equalizing Valve

Pressure
Regulating Valve

Exhaust Muffler
Canister

Bacteria Filter
One-way valve

Transformer

Control Board

Purging Hole

10
Sieve Bed

Oxygen Tank

Fan

Radiator

Pneumatic Valve

Filter

Inlet Canister

Compressor

Capacitor

11
OC- E /OC- P Series Control Board
1 4 2 5 12

3
7
11

10 9 8

1. Power Connector 2. Pneumatic Valve Connector 3. Program Download Connector 4. Pressure Equalizing Valve
5. Pressure Transducer Connector 6. OSD Inlet 7. OSD Outlet 8. LCD Connector 9. Key Connector
10. Power Failure Alarm Connector 11. Transformer Power Connector 12. Temperature Sensor

OC- S Series Control Board


1 4 2 5 10

1. Power Connector
2. Pneumatic Valve Connector
3. Program Download Connector
4. Pressure Equalizing Valve
5. Pressure Transducer Connector
6. LCD Connector
7. Key Connector
8. Power Failure Alarm Connector
3
9. Transformer Power Connector
9 10. Temperature Sensor

8 7 6
12
CAUTION: For your safety, be sure to turn off the power and unplug the power cord before
you service the OC Oxygen Concentrator.
NOTE: Record all scheduled maintenance.

6.2 Cabinet removal


6.2.1 Removal cabinet front and back
To remove the cabinet front and back, remove the filter access panel and air intake filter grille first,
and then remove six screws which placed on the back of the handle and near the middle and bottom of
the cabinet of the two sides. See Fig. 5
Remove the cabinet back first, then disconnect tubes, wires and connector related to remove cabinet
front. And usually it’s not necessary to remove the cabinet front unless the LCD and power switch need to
be replaced.

Fig.5
NOTE: Pay attention to the force used in dismantling or assembling process in case of making the
linkage section broken.
6.2.2 Main structure
The weight and foal components rest solely on four parts: the main structure, the compressor plate,
the base-plate and the base. These parts were specially designed and formed. They should never
require replacement under normal use.
6.2.3 Caster replacement
The casters are a push in type that do not require any fasteners. Lay the device on its back to access
the casters from the bottom. Pull them straight out away from the bottom, and then insert new on.
6.3 Compressor
The compressor is the pump within the oxygen concentrator that supplies air to the sieve beds, and
forces oxygen to flow out of the top of the oxygen tank.
The compressor is the likely cause of two potential specific problems:
a. An insufficient amount of air is supplied to the process.
b. An excessive sound level.
 Air Supply
Compressor output refers to how much compressed air the compressor can produce. This depends
upon the model of the compressor, length of stroke, piston diameter, speed of rotation and condition of
seals. The cup seals form the seal between the piston and the cylinder wall. As the cup seals wear, the
output begins to gradually decrease. This reduction in compressor output results in less air, and thus less
oxygen, entering the sieve beds. Therefore, the production of oxygen decreases.

13
Because this drop in oxygen production occurs over a long period of time, preventive maintenance
on the compressor is not required.
You can continue a patient’s therapy on the OC unit as long as the oxygen concentration level at the
prescribed liter flow rate is within SYSMED specification limits. Refer to Part 3.4.
 Sound Level
The sound level is largely determined by the condition of the compressor’s bearings.
There are four bearings located within the compressor that allow the inner components of the
compressor to rotate. If the bearings wear to the point that they become loose and noisy, the compressor
becomes noticeably loud and needs servicing. The life of a compressor is determined primarily by its
operating temperature. It is extremely important that the inlet cooling air filters are cleaned and replaced
as required.
6.3.1 Compressor replacement
To remove the compressor assembly for replacement, follow the steps listed below:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Remove the two screws that connect the compressor side plate, lift and slide out the compressor
assembly.
(4) Take off the capacitor from its mount, disconnect the connector of the compressor power wires.
(5) Unscrew the bolt which fixed the yellow and green ground wires.
(6) Remove the screw which fixed the wire harness retainer.
(7) Take off the inlet filter. Remove the screw to take off the baffle plate.
(8) Turn anti-clockwise to take off the external canister of the inlet canister, remove the screw, turn
the inside canister assembly anti-clockwise to disconnect the inlet tube, while holding the inlet tube metal
head.
(9) At the opposite side of the compressor, Snip the cable tie of the outlet tube with diagonal pliers,
and pull off the outlet tube.
(10) Remove the four compressor bolts and take out the springs.
(11) Check to see all related parts disconnected, then lift and slide out the compressor from one
side.
(12) Turn the inlet tube anti-clockwise to take off. Remove the safety valve and outlet connector of
the compressor.
(13) Remove the compressor from the plate by removing the four screws with solid wrench.
(14) Snip the capacitor wire.
Refer to Fig.6-Fig.19

Fig.6 Fig.7
14
Fig.8 Fig.9

Fig.10 Fig.11

Fig.12 Fig.13

Fig.14 Fig.15
15
Fig.16 Fig.17

Fig.18 Fig.19
To install a new or rebuilt compressor assembly, follow the steps listed below:
(1) Perform the compressor removal procedure in reverse order.
(2) Leak test all connections.
Note: Before install plates and spring fittings, add some anaerobic adhesives into the screw hole.
Use Adhesive 704 for compressor safety valve and Sealing strip for outlet joint. The outlet tube need use
cable tie to tighten.
6.3.2 Capacitor replacement
The capacitor helps the compressor to start and run more efficiently. If the compressor cannot start,
the capacitor may be defective and require replacement. To replace the capacitor, take the following
steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Take off the capacitor from its mount.
(4) Disconnect the two leads to the capacitor.
(5) Install the new capacitor, connect the leads and replace the capacitor.
6.4 Pneumatic valve replacement
The OC uses a pneumatic valve assembly to control the air separation process. There is a feed port
that connects to the radiator and an exhaust port that connects to the exhaust muffler outlet.
The pneumatic valve of the OC unit requires no scheduled maintenance. If a valve does not function
as required, it is best to replace the complete sieve beds as it is probable that one or both of the beds has
been damaged. You can make a check first, if the oxygen concentrator is normal after replacement of
pneumatic valve, then it’s no need to replace the sieve beds.
To replace the pneumatic valve, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Disconnect the pneumatic valve wire connector first. Then pull off the inlet tube and outlet tube.

16
(4) Disconnect one side of the tube with sieve bed by pushing the blue washer and pull off the tube, and
disconnect the other tube in the same way. (Note: You may turn the valve to a proper position to take off
the tube.)

Fig.20
6.5 Pressure equalizing valve replacement
To remove the compressor assembly for replacement, follow the steps listed below:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Push the blue washer as the yellow arrows indicate in the Fig. 21 and pull off the tube. And take off
the tube at the other side of the equalizing valve in the same way.
(4) Disconnect the wire of the equalizing valve which inserts in the control board.
(5) Replace a new equalizing valve and perform the valve removal procedure in reverse order to install.

Fig.21
6.6 Sieve beds replacement
CAUTION: Do not expose molecular sieve (contents of bed) to air for an extended period of time.
Prolonged exposure of molecular sieve to the moisture in room air results in contamination and
permanent damage to the sieve material. Keep all openings to the sieve beds sealed during periods of
storage.
6.6.1 Sieve Beds Removal
(1) Turn off the power and unplug the power cord.
(2) Remove the cabinet back.
(3) Refer to removal method of equalizing valve, pull out the tube from the up joint of the sieve bed in
the same way.
(4) Remove the four screws which fixed the sieve bed.
(5) Pull out the tube which connected with pneumatic valve near the bottom of the sieve bed to
remove the sieve bed. Remove the other sieve bed in the same way.

17
Fig.22
6.6.2 Sieve Beds installation
To install the sieve beds, follow the sieve beds removal procedure in reverse order. It is very
important to properly connect all tubes to eliminate leaks.
To check for leaks, take the following steps.
(1) Plug in the unit.
(2) Turn on the power switch for three minutes with the flow meter closed to pressurize the system.
(3) Apply soapy water around all of the hose connections at the valve and check for leaks.
CAUTION: There is an electrical shock hazard with the Power On. Be careful that no water
contacts any of the electrical connections or components.
NOTE: Small leaks can affect a concentrator’s performance and can cause contamination of the
sieve. Careful leak testing is important.
6.7 Radiator and Fan replacement
To replace the radiator in the OC unit, take the following steps:
Turn off the power and unplug the power cord.
(1) Remove the cabinet back.
(2) Cut off the two tubes connected with radiator.
(3) Remove the four screws which fixed the radiator.
(4) Replace radiator and reinstall the radiator.
NOTE: If the tubes getting short, replace new tubes.

Fig.23
Fan replacement
Remove the radiator first as described above (Don’t need to cut off the tubes), disconnect the wire
connector of the Fan and take off it.
NOTE: There is arrow sign on the Fan which shows the direction of the wind, the wind should blow
to the compressor. The side with label should be put inside, align at the hole on the plate.
18
Fig.24
6.8 Control board replacement
The control board controls the main parts, such as power, valves, sensor and alarm system functions
etc. Consult the troubleshooting chart in Part 7.0 to determine when to replace the printed circuit board.
CAUTION: The control board contains components that are sensitive to electrostatic discharge (ESD)
that can damage the board if not handled properly. As when handling any ESD sensitive board, observe
standard ESD safety procedures. These procedures include the following:
 Handle the control board by the edges only.
 Work on a grounded ESD mat.
 Wear a grounded wrist strap.
 Store control board in anti-static bags only.
6.8.1 Control Board Removal
(Note that the Control Board on the OSD unit is different to the Control Board on the standard unit.)
(1) Turn off the power, and unplug the power cord.
(2) Remove the front and cabinet backs.
(3) Pull out the tube which connect the oxygen tank and control board, disconnect all the leads,
connectors on the control board.
(4) Use nipper pliers to clamp the board fixed PCB support, lift the control board to remove, there are
four PCB support need to disconnect one by one.

PCB Support

Fig. 25
19
6.8.2 Control Board Installation
To install the control board, follow its removal procedure in reverse order. Aim the install holes of the
control board to the fixed PC barb columns, and then press down the control board. Connect all tubes,
leads, connectors to the control board as they were before.
NOTE: Handle the new circuit board only by the edges to prevent
electrostatic damage to the unit. Remember the place of each tube,
leads and connectors to install correctly.
There is a black temperature sensor line on the back of the board,
when install the board, the sensor line mush insert in the hole of the
plate first.
Sensor line the hole
When install the control board with OSD sensor, the connect spot
of the tubes and OSD sensor should use cable tie to tighten.
6.8.3 Other tips
Buzzer The buzzer is a fixed component on the control board and is not individually replaceable.
Press Switch
The high and low pressure alarms are activated by a pressure transducer on the control board, and
is not individually replaceable.
Time Setting Key
Keypad Replacement procedures:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Disconnect the connectors from the control board. Tear off the key film to remove the keypad.
(4) Replace keypad.
6.9 Transformer replacement
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Disconnect the input connector, and then disconnect its output connector from the control board.
(4) Remove the two fixed screws, and take off the transformer.
(5) Replace a new transformer and follow the removal procedure in reverse order to install.
6.10 LCD replacement
To replace the radiator in the OC unit, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Pull out the tubes, disconnect the leads on the LCD board.
(4) Remove the four screws which fix the LCD board, and then take off the LCD.
(5) Follow the removal procedure in reverse order to install a new LCD.

Fig.26
20
6.11 Pressure regulating valve
6.11.1 Pressure regulating valve check and set
The pressure regulating valve enables you to set the maximum flow of oxygen output by the OC unit.
To check for proper adjustment of the product regulator, take the following steps:
(1) Turn on the power.
(2) Allow the unit to run for a few minutes.
(3) Remove the cabinet back.
(4) Connect a pressure gauge directly to the outlet.
(5) The pressure should read 7.345 ± 10% psi
(6) Adjust the regulator if necessary. Pull out downward the bottom of the pressure regulating valve,
turn the bottom clockwise or anticlockwise while reading the gauge to get a proper pressure.
(7) Reinstall the cabinet front and cabinet backs.
6.11.2 Pressure regulating valve replacement
If the flow meter ball fluctuates, or low oxygen concentration, the pressure regulating valve may need
check, adjust or replace. To replace the pressure regulating valve in the OC unit, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Pull out the outlet tube, and then holding tightly the pressure regulating valve, turn anti-clockwise
to remove it.
(4) Follow the removal procedure in reverse order to install a new pressure regulating valve.
(5) Reset the pressure of the valve.
Note: When install the pressure regulating valve, apply 704 glue on the thread, and the tube which
connect the outlet should be tightened with cable tie.

Fig.27

6.12 Power switch replacement


(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Remove the inlet tube which connected with the flow meter.
(4) Unscrew the two screws that hold the SD bracket, and take off the bracket.
(5) Depress top and bottom of the power switch to release the locking tabs from the cabinet, and then
push forward to disconnect the power switch with the control panel.
(6) Pull out the wire from the power switch. Be careful to note the color and position of each specific wire.
(7) Follow the removal procedure in reverse order to install a new power switch.

21
Fig.28

6.13 Flow meter replacement


6.13.1 Flow Meter Removal
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Remove the tubes from the flow meter fittings.
(4) Unscrew the flow meter nuts with a wrench.
(5) Remove the flow meter through the control panel.

Fig.29
6.13.2 Flow Meter Installation
To install a new flow meter, follow the flow meter removal procedure in reverse order. Then perform
a leak test on the connections.

22
6.14 Circuit breaker replacement
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Unscrew the nut of the circuit breaker.
(4) Disconnect the leads, replace the circuit breaker.
(5) Follow the removal procedure in reverse order to install a new circuit breaker.

Fig.30

Part VII Troubleshooting


7.1 The machine working pressure test
Testing the operating pressure is a useful diagnostic tool
when a concentrator has low purity and requires servicing.
Units functioning normally do not require operating tests.
Use the following procedure to test the operating pressure of
the unit.
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Remove the oxygen supply tube from cabinet front, connect
the tube directly to the pressure test gauge.
(4) Plug in the power cord, and turn on the power. Set the flow
meter to rated flow, and allow the unit to run at least six
minutes.
(5) Observe the readings on the pressure test gauge.
The reading should be 7.345 ± 10% psi. (0.05±0.005Mpa)
NOTE: When you turn the unit on, it will take several minutes
to reach normal operating pressures.

7.2 General Troubleshooting


Before reviewing the troubleshooting chart, the following steps may be useful to analyze any
malfunctions:
(1) Turn the concentrator on. If unit does not operate, refer to troubleshooting chart.
(2) Make sure all filters are clean.
(3). Set the concentrator at rated flow, connect test pressure gauge to tube witch connect the outlet of the
pressure regulator valve. The pressure should read approximately 7.345 ± 10% psi. If pressures are
high or low, refer to the troubleshooting chart.
(4) Make sure that the unit is leak free by testing all tubing connections and fittings with leak testing
solution. Protect circuit board from solution and start leak test at the compressor outlet, following the air
flow through the unit to the oxygen outlet. Repair all leaks by tightening connections and fittings.
(5) Review troubleshooting chart to analyze and repair any other malfunctions.
23
7.3 Troubleshooting table
PROBLEM PROBABLE CAUSE SOLUTION
Power on, the equipment is not Power cord not plug well or bad contact * Check whether the power cord is damaged or not.
working, LCD not display, or works * check whether the power cord is plug well.
un-continuously.
Compressor does not run. Constant No power to unit. Unit circuit breaker tripped or Check wall outlet for power. Reset or replace circuit
audible alarm with turn on the power. faulty. breaker.
Power on, the equipment is not Start capacity of compressor is broken or Replace capacity,
working but LCD display well compressor is not working Replace compressor
Compressor shuts down Unit overheating due to improper location. Locate unit away from heating source, providing
intermittently; Defective cooling fan. adequate ventilation on all sides. Replace fan.
Extreme cold start. Allow unit to reach room temperature.
Compressor does not start. Power on
state, intermittent alarm, and cabinet
fan turns.
Compressor runs with intermittent low Leak Leak test and repair leak.
pressure alarm and low oxygen
concentration.

Compressor runs with intermittent Defective sieve module. Replace sieve module.
high pressure alarm and low oxygen Restriction in exhaust muffler. Defective pneumatic Replace or clean muffler foam.
concentration. valve. Replace Pneumatic valve.

Compressor safety valve releases Defective pneumatic valve. Replace pneumatic valve.
(popping sound). Contaminated sieve module. Replace sieve module.
Defective safety valve. Replace safety valve.

Constant alarm with power on. Circuit Defective circuit breaker. Replace circuit breaker.
breaker repeatedly trips. Defective capacitor. Replace capacitor.
Defective compressor. Replace compressor.
Defective circuit board. Replace circuit board.
Alarm does not sound. Defective power switch. Replace power switch.
Defective buzzer. Replace control board.
Defective pressure sensor. Replace control board

Flow meter fluctuates. Improperly set or faulty pressure regulator valve. Check regulator valve or replace regulator valve.
Leak. Test for leaks.
Worn compressor. Replace compressor
Defective flow meter. Replace flow meter.
Kinked tubing Check tubing that connects the top of the sieve beds.
Note: Please do not turn the button abruptly, or the
value spool may be broken. Turn the knob slowly and
softly
Cooling fan does not turn. Defective cooling fan. Replace cooling fan.
Defective electrical connections. Check electrical connections.

Limited or Iow flow. Restriction in humidifier or tubing. Pressure Replace humidifier or tubing.
regulator valve set too low. Adjust regulator setting.
Leak. Leak test and repair leak.
Weak compressor. Check system pressure, exchange compressor.
Check filter, suction resonator for obstruction.
Air flow obstruction.
Low concentration. Compressor inlet filter is dirty or partially blocked. Replace inlet filter.
System leak
Faulty compressor Leak test and repair leak.
Unit temperature too high, Replace compressor.
24
Blocked air intake or dirty inlet filter, faulty cooling fan.
Restriction in exhaust muffler, Contaminated sieve Replace or clean exhaust muffler.
module. Check outlet pressures are within range. Replace sieve
module.
Defective pneumatic valve. Replace pneumatic valve.
Restriction of inlet air. Check inlet tubing for obstruction and remove.
* Outlet muffler taken off Reinstall outlet muffler
Exhaust sound too loud
* Outlet muffler broken Replace outlet muffler
No exhaust sound but safety valve * Pneumatic valve broken Replace pneumatic valve
has air out * control board broken Replace control board
Power switch on, the equipment * the flow knob is too tight or broken Replace flow meter
works well, but no oxygen out and * the equipment is leaking badly Leak test and repair leak.
ball of flow meter at the bottom not
rising up.
The concentration is within parameter The oxygen sensor device on the control board Replace control board
ranger by Oxygen density sensor error
test, the LCD realtime display low
concentration or alarm.
Defective Pressure equaling valve Replace pressure equaling valve.
The equipment works well, but the
oxygen concentration is low.

Defective LCD Replace LCD

LCD no display or miss words

intelligent fault diagnosis


High Pressure Alarm molecular sieve lose efficacy due to long time use, Replace sieve beds
The malfunction code is H:07 thus cause high pressure of oxygen tank
Oxygen outlet tube, pressure regulator valve Replace tube, pressure regulator valve.
broken or overlarge amount of flow that cause low Adjust flow meter and the flow should not exceed the
pressure of oxygen tank rated flow.
Low Pressure Alarm Defective pressure transducer
The malfunction code is H:06 Pneumatic valve malfunction Replace control board
Check the pneumatic valve are not leak, or replace
leak pneumatic valve.
Leak test
Defective radiator or fan Replace radiator or fan.
High Temperature Alarm
Exhaust outlet blocked Clear block
The malfunction code is H:05.
Defective control board Replace control board
Defective OSD Sensor Replace control board
The fault of OSD Sensor circuit
The malfunction code is E:03,

Low oxygen concentration alarm. The See chart below


malfunction code is H:04,

Analyze Flow Chart for Oxygen Purity below standard

Oxygen concentration
low or no oxygen
exhaust
Ⅶ. TROUBLE S OF PRE SSURE RE DUCING VALVE
25
NO

YES

YES

NO

7.4 Tools table

 The tools needed for you to properly service the OC unit are listed below, requires no special tools:
 An accurate pressure test gauge to take pressure readings on the OC unit should be kept available
at all times.

Table 3 Main Tools & Accessory Materials

26
TOOLS MATERIALS

NAME MODEL NAME MODEL


Anaerobic
"+" Screw driver 300mm ~ 250mm GY-340
adhesives

"-" Screw driver 200mm Adhesive 704

Diagonal pliers Sealing strip

Nipper pliers Wire jumper

Solid wrench 8 ~ 10 Cable tie 3.6×150mm

Solid wrench 12 ~ 14 Cable tie 2.5×100mm

Multimeter Insulating tape

"T" Wrench
4mm

"T" Wrench 3mm

Pressure gauge 0 ~ 0.4Mpa

Oxygen
concentration
analyzer

Tube cutter

Attachment I Circuit Diagram

OC-S30

27
OC-E30 \ OC-P30

QUALITY WARRANTY

28
Model Warranty Period
OC-S 30 12Months

OC-S 50 12 Months

OC-S 80 12 Months

OC-S 100 12 Months

OC-E 30 12 Months

OC-E 50 12 Months

OC-E 80 12 Months

OC-E 100 12 Months

OC-P 30 12 Months
OC-P 50 12 Months
OC-P 80 12 Months
OC-P 100 12 Months

All after-sale service commitment shall be fulfilled by the supplier, dealer or


the appointed maintenance company.
During the warranty period, if the product fails under conditions of normal
use, the damages to the equipments are not caused by man-made reasons, the
supplier, dealer or the appointed maintenance company is responsible for the
maintenance or replacement free of charge.
The warranty does not include consumables like oxygen tubing, filters etc.
The warranty does not cover the breakdown or damages caused by improper
operation, abuse of the product, accident, or to products damaged by reason of
repairs made to any components without the specific consent of the supplier. The
users should not dismantle the equipment by themselves in the warranty period.
This warranty does not include normal wear and tear or shipping charges.
SysMed and the dealer’s sole obligation and your exclusive remedy under this
warranty shall be limited to such repair or replacement. SysMed shall not be
liable for any consequential or incidental damages whatsoever.
After the guarantee period, or damages excluded in the warranty, the supplier,
dealer or the appointed maintenance company will supply customers the paid
service.