OC SERIES
SERVICE MANUAL
Model OC-S
OC-S30, OC-S50,
OC-S80, OC-S100
Model OC-E
OC-E30, OC-E50,
OC-E80, OC-E100
Model OC-P
OC-P30, OC-P50
OC-P80, OC-P100
SAFETY SUMMARY
1.1 Use of oxygen
Oxygen is not a flammable gas but a combustion-supporting gas, so the machine must be far away
from fire and heat. DO NOT SMOKE while using this device, for your own safe, the user must quit
smoking. Keep all matches, lighted cigarettes or other sources of ignition out of the room.
Keep power cord away from sources of ignition and heat.
Turn off the power when nobody uses the machine.
DO NOT put oxygen tubing under bedspread or chair cushion.
Unplug the power cord when you clean the machine or change fuse, in order to avoid getting an
electric shock.
DO NOT use any lubricants unless recommended by SysMed.
Keep unit in a dry, well-ventilated, without dust, corruption and toxicological gas environment, away
from sunlight and no intense electromagnetic interference
2
PART II General Introduction
2.1 Distributor/dealer responsibility
All Distributor/dealer of the SYSMED OC Series Oxygen Concentrator must assume responsibilities
for handling, operational check-out, user instruction, and maintenance. These responsibilities are
outlined below and throughout this manual.
WARNING SYSMED units must not be used for or with any life-supporting or life sustaining
applications. Patients unable to communicate discomfort while using this device, they may require
additional monitoring. Advise patients to immediately notify their physician(s) and/or dealer in case of an
alarm or any discomfort.
SYMBOLS
ON
OFF
Warning
No Smoking
Type B Equipment
Upright
CE Certificate
Frangibility
Alternating Current
3
Table 1 function
Parameters
Function and setting
model no.
OC-S30 Standard function: LCD display, temperature alarm, high/low pressure alarm, power failure
OC-S50 alarm, timed setting, intelligent fault diagnosis, repair reminder alarm.
OC-S80
OC-S100
OC-E30 Standard function: LCD display, oxygen concentration indicator (Red, green, yellow bar on
OC-E50 the LCD), temperature alarm, high/low pressure alarm, low oxygen concentration alarm,
power failure alarm, timed setting, intelligent fault diagnosis, repair reminder alarm.
OC-E80
OC-E100
OC-P30 Standard function: LCD display, oxygen concentration indicator (Red, green, yellow bar &
OC-P50 realtime concentration display on the LCD), temperature alarm, high/low pressure alarm,
low oxygen concentration alarm, power failure alarm, timed setting, intelligent fault
OC-P80
diagnosis, repair reminder alarm.
OC-P100
Optional function Positive pressure outlet (for nebulizer)
Table 2 Parameter
Model no. OC-S30 OC-E30 OC-P30 OC-S50 OC-E50 OC-P50
Rating Power (W) 350 350
Rating current(A) 1.59 1.59
Electrical Requirements 220V±22V, 50HZ±1HZ 220V±22V, 50HZ±1HZ
Flow rate(L/min) 0-3 0-5
purity(3L/min) 93±3% 93±3%
purity(5L/min) —— 93±3%
Outlet pressure(Mpa) 0.05±0.005 0.05±0.005
Sound level dB(A) ≤42 ≤42
Equipment Class and Type Class II Type B Class II Type B
Net weight(kg) 25kg 25kg
Dimension (mm) 381x347x689 381x347x689
4
Part III Working Principle and Routine Checking Operations
3.1 Working principle
Oxygen Concentrator for Medical Use adopts the principle of the world-advanced pressure swing
adsorption technology. At normal temperature, the machine can continuously separate medical oxygen
with high density (93%±3% oxygen for medical use) from the air when its power is turned on.
Note: Using oxygen concentrator doesn’t influence oxygen content in the air of the room.
Air enters the OC Oxygen Concentrator through an external cabinet air intake filter. This filtered air
enters the compressor, and pressurized air then exits the compressor and passes through a radiator,
which reduces the temperature of the compressed air. Next, an pneumatic valve system directs the air
into one of two sieve beds that contain molecular sieve. The special characteristic property of molecular
sieve is that it physically attracts (adsorbs) nitrogen when air passes through this material, thus enabling
the production of high purity oxygen.
There are two sieve beds or adsorbent columns; while one produces high purity oxygen, the other is
purged of the nitrogen it adsorbed (collected) while it was producing oxygen. Each column produces
oxygen and delivers it to the oxygen tank. Oxygen exits the oxygen tank through a pressure regulator,
flow control valve, flow meter, and bacteria filter. The flow control valve, which is part of the flow meter,
controls the flow rate of oxygen delivered to the patient. The OC unit delivers up to 93%±3% oxygen
concentration at rated flow. The remaining constituents of the product gas stream are nitrogen and argon,
both of which are part of the air we breathe, are inert and are completely safe.
3.2 Checking operation
SYSMED runs each device through a burn in period and tests every OC Oxygen Concentrator
thoroughly after manufacture before releasing for shipment. As the dealer, it is your responsibility to
perform the following test to ensure that no damage occurred in shipping or handling.
(1)Open and inspect all concentrator cartons upon receipt. Unpack each unit and remove it from its
carton. Inspect the unit itself for damage. If the exterior of the carton is damaged, or the unit itself is
damaged, note it on the freight bill signed by the driver.
Before inserting the plug of power cord into the power socket, turn on the power switch and note that
the audible alarm sounds continuously. (See Part III 3.3). And then off the power.
(2)Plug in the power cord of the unit, and set the I/0 (ON/OFF) switch to the I (ON) position. Check
5
to see that the following occurs:
The compressor runs, listen for the sound.
Exhaust air flows out of the bottom of the unit.
OPTIONAL for Units Equipped with Oxygen Concentration Indicator: The green light remains flicker
until the oxygen concentration reaches 85%. (approximately six minutes).
(3)Turn the flow meter adjustment knob counterclockwise to see if it works well, If not, refer to Part
VII troubleshooting guide.
(4)Perform an oxygen concentration test, as described in Part III 3.4.
3.3 Alarm System
The OC Oxygen Concentrator is equipped with alarm system, which sounds a continuous and loud
alarm when a power failure occurs or when one or more cycle variables are not within specification. It
sounds an alarm if the high or low pressure indicators are activated. The alarm remains on until you
correct the alarm condition or you set the I/0 (ON/OFF) switch to the 0 (OFF) position. Refer to Part VII for
a list of probable alarm causes.
3.3.1 Power Failure Alarm Test
To test the power failure alarm, perform the following actions:
Unplug the power cord from the wall outlet, and set the I/0 (ON/OFF) switch to the I (ON) position.
This should immediately activate the audible alarm. If it doesn’t, refer to the troubleshooting guide in Part
VII of this manual.
3.4 Oxygen Concentration Test and Specification
To ensure that the output of oxygen from the device is within specification, you must perform an
oxygen concentration test. Test the unit upon delivery to a patient and at periodic intervals. Dealers,
based on their expertise and documentation, may establish and implement their own plans for checking
oxygen concentration. The interval established may be longer or shorter than 180 days, which is
SYSMED default time period for providers who do not choose to establish their own plan.
(1)Remove oxygen humidifier bottle from the oxygen outlet.
(2)Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
(3)Set the I/0 (ON/OFF) power switch to the I (ON) position. (It takes approximately 6 minutes for
the oxygen concentration to stabilize.) Take oxygen concentration readings over a period of several
minutes to reduce any cyclic variations
(4)Verify that the product flow rate delivered by the unit matches the patient’s prescription and
does not exceed the capacity of the unit.
(5)Disconnect the oxygen analyzer, and reconnect the humidifier bottle and any other equipment /
accessories that may be required.
(6)Adjust the flow meter adjustment knob to the prescribed flow rate.
NOTE: Do not measure oxygen concentration output after the product stream passes through a
humidifier bottle, erroneous readings will result and your oxygen concentration measuring device might
be damaged.
OC Concentration Specifications
Liter Flow Specification Within Specification
Rated flow 93%±3% 90% or higher
(1) Clean the cannula every day. First clean with detergent and then rinse with clean water
completely, and dry up naturally. Changing a new cannula once a month is recommended.
(2) The distilled water or cold boiled water in humidifier should be refreshed every day.
The humidifier should be cleaned and disinfected every 3 weeks, clean the humidifier with detergent
and then rinse with warm or hot water, dip it in household disinfectant for 20-30 minutes, clean again with
warm or hot water, and then dry up naturally. If not use, put it in a clean bag.
1. Humidifying Bottle
2. Connecting Nut
3. Oxygen Outlet
4. Humidifier Connector
4 7 5. Oxygen Flow-meter
2 6
6. Knob of the Flow-meter
9 7. Large LCD Display
1 5 (See Fig. 4)
3 8. Power Switch
8 9. Adjusting Panel and Keys.
See Fig. a 10. Positive Pressure Outlet
(See Fig. 3)
11. Model
10
11
Back View
Side View
7.6 7.5 7.4 7.3 7.2
9
88 7.1
维 护
Maintenance
故 障
Malfunction 低Low氧 富Rich
氧 高High氧
定时时间
Timed setting
当次时间
Running time
故障代码
Malfunction Code
8 8 8:8 8 7.7
累计时间8
Elapsed time
8888 7.8
Fig.4
7.1 Oxygen purity percentage (type OC-P)
7.2 High purity sign
7.3 Rich purity sign
7.4 Low purity sign
7.5 Malfunction alarm
7.6 Maintenance alarm
7.7 Timed setting, Running time and Malfunction code display
7.8 Elapsed time
Interior structure
Pressure
Equalizing Valve
Pressure
Regulating Valve
Exhaust Muffler
Canister
Bacteria Filter
One-way valve
Transformer
Control Board
Purging Hole
10
Sieve Bed
Oxygen Tank
Fan
Radiator
Pneumatic Valve
Filter
Inlet Canister
Compressor
Capacitor
11
OC- E /OC- P Series Control Board
1 4 2 5 12
3
7
11
10 9 8
1. Power Connector 2. Pneumatic Valve Connector 3. Program Download Connector 4. Pressure Equalizing Valve
5. Pressure Transducer Connector 6. OSD Inlet 7. OSD Outlet 8. LCD Connector 9. Key Connector
10. Power Failure Alarm Connector 11. Transformer Power Connector 12. Temperature Sensor
1. Power Connector
2. Pneumatic Valve Connector
3. Program Download Connector
4. Pressure Equalizing Valve
5. Pressure Transducer Connector
6. LCD Connector
7. Key Connector
8. Power Failure Alarm Connector
3
9. Transformer Power Connector
9 10. Temperature Sensor
8 7 6
12
CAUTION: For your safety, be sure to turn off the power and unplug the power cord before
you service the OC Oxygen Concentrator.
NOTE: Record all scheduled maintenance.
Fig.5
NOTE: Pay attention to the force used in dismantling or assembling process in case of making the
linkage section broken.
6.2.2 Main structure
The weight and foal components rest solely on four parts: the main structure, the compressor plate,
the base-plate and the base. These parts were specially designed and formed. They should never
require replacement under normal use.
6.2.3 Caster replacement
The casters are a push in type that do not require any fasteners. Lay the device on its back to access
the casters from the bottom. Pull them straight out away from the bottom, and then insert new on.
6.3 Compressor
The compressor is the pump within the oxygen concentrator that supplies air to the sieve beds, and
forces oxygen to flow out of the top of the oxygen tank.
The compressor is the likely cause of two potential specific problems:
a. An insufficient amount of air is supplied to the process.
b. An excessive sound level.
Air Supply
Compressor output refers to how much compressed air the compressor can produce. This depends
upon the model of the compressor, length of stroke, piston diameter, speed of rotation and condition of
seals. The cup seals form the seal between the piston and the cylinder wall. As the cup seals wear, the
output begins to gradually decrease. This reduction in compressor output results in less air, and thus less
oxygen, entering the sieve beds. Therefore, the production of oxygen decreases.
13
Because this drop in oxygen production occurs over a long period of time, preventive maintenance
on the compressor is not required.
You can continue a patient’s therapy on the OC unit as long as the oxygen concentration level at the
prescribed liter flow rate is within SYSMED specification limits. Refer to Part 3.4.
Sound Level
The sound level is largely determined by the condition of the compressor’s bearings.
There are four bearings located within the compressor that allow the inner components of the
compressor to rotate. If the bearings wear to the point that they become loose and noisy, the compressor
becomes noticeably loud and needs servicing. The life of a compressor is determined primarily by its
operating temperature. It is extremely important that the inlet cooling air filters are cleaned and replaced
as required.
6.3.1 Compressor replacement
To remove the compressor assembly for replacement, follow the steps listed below:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Remove the two screws that connect the compressor side plate, lift and slide out the compressor
assembly.
(4) Take off the capacitor from its mount, disconnect the connector of the compressor power wires.
(5) Unscrew the bolt which fixed the yellow and green ground wires.
(6) Remove the screw which fixed the wire harness retainer.
(7) Take off the inlet filter. Remove the screw to take off the baffle plate.
(8) Turn anti-clockwise to take off the external canister of the inlet canister, remove the screw, turn
the inside canister assembly anti-clockwise to disconnect the inlet tube, while holding the inlet tube metal
head.
(9) At the opposite side of the compressor, Snip the cable tie of the outlet tube with diagonal pliers,
and pull off the outlet tube.
(10) Remove the four compressor bolts and take out the springs.
(11) Check to see all related parts disconnected, then lift and slide out the compressor from one
side.
(12) Turn the inlet tube anti-clockwise to take off. Remove the safety valve and outlet connector of
the compressor.
(13) Remove the compressor from the plate by removing the four screws with solid wrench.
(14) Snip the capacitor wire.
Refer to Fig.6-Fig.19
Fig.6 Fig.7
14
Fig.8 Fig.9
Fig.10 Fig.11
Fig.12 Fig.13
Fig.14 Fig.15
15
Fig.16 Fig.17
Fig.18 Fig.19
To install a new or rebuilt compressor assembly, follow the steps listed below:
(1) Perform the compressor removal procedure in reverse order.
(2) Leak test all connections.
Note: Before install plates and spring fittings, add some anaerobic adhesives into the screw hole.
Use Adhesive 704 for compressor safety valve and Sealing strip for outlet joint. The outlet tube need use
cable tie to tighten.
6.3.2 Capacitor replacement
The capacitor helps the compressor to start and run more efficiently. If the compressor cannot start,
the capacitor may be defective and require replacement. To replace the capacitor, take the following
steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Take off the capacitor from its mount.
(4) Disconnect the two leads to the capacitor.
(5) Install the new capacitor, connect the leads and replace the capacitor.
6.4 Pneumatic valve replacement
The OC uses a pneumatic valve assembly to control the air separation process. There is a feed port
that connects to the radiator and an exhaust port that connects to the exhaust muffler outlet.
The pneumatic valve of the OC unit requires no scheduled maintenance. If a valve does not function
as required, it is best to replace the complete sieve beds as it is probable that one or both of the beds has
been damaged. You can make a check first, if the oxygen concentrator is normal after replacement of
pneumatic valve, then it’s no need to replace the sieve beds.
To replace the pneumatic valve, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Disconnect the pneumatic valve wire connector first. Then pull off the inlet tube and outlet tube.
16
(4) Disconnect one side of the tube with sieve bed by pushing the blue washer and pull off the tube, and
disconnect the other tube in the same way. (Note: You may turn the valve to a proper position to take off
the tube.)
Fig.20
6.5 Pressure equalizing valve replacement
To remove the compressor assembly for replacement, follow the steps listed below:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Push the blue washer as the yellow arrows indicate in the Fig. 21 and pull off the tube. And take off
the tube at the other side of the equalizing valve in the same way.
(4) Disconnect the wire of the equalizing valve which inserts in the control board.
(5) Replace a new equalizing valve and perform the valve removal procedure in reverse order to install.
Fig.21
6.6 Sieve beds replacement
CAUTION: Do not expose molecular sieve (contents of bed) to air for an extended period of time.
Prolonged exposure of molecular sieve to the moisture in room air results in contamination and
permanent damage to the sieve material. Keep all openings to the sieve beds sealed during periods of
storage.
6.6.1 Sieve Beds Removal
(1) Turn off the power and unplug the power cord.
(2) Remove the cabinet back.
(3) Refer to removal method of equalizing valve, pull out the tube from the up joint of the sieve bed in
the same way.
(4) Remove the four screws which fixed the sieve bed.
(5) Pull out the tube which connected with pneumatic valve near the bottom of the sieve bed to
remove the sieve bed. Remove the other sieve bed in the same way.
17
Fig.22
6.6.2 Sieve Beds installation
To install the sieve beds, follow the sieve beds removal procedure in reverse order. It is very
important to properly connect all tubes to eliminate leaks.
To check for leaks, take the following steps.
(1) Plug in the unit.
(2) Turn on the power switch for three minutes with the flow meter closed to pressurize the system.
(3) Apply soapy water around all of the hose connections at the valve and check for leaks.
CAUTION: There is an electrical shock hazard with the Power On. Be careful that no water
contacts any of the electrical connections or components.
NOTE: Small leaks can affect a concentrator’s performance and can cause contamination of the
sieve. Careful leak testing is important.
6.7 Radiator and Fan replacement
To replace the radiator in the OC unit, take the following steps:
Turn off the power and unplug the power cord.
(1) Remove the cabinet back.
(2) Cut off the two tubes connected with radiator.
(3) Remove the four screws which fixed the radiator.
(4) Replace radiator and reinstall the radiator.
NOTE: If the tubes getting short, replace new tubes.
Fig.23
Fan replacement
Remove the radiator first as described above (Don’t need to cut off the tubes), disconnect the wire
connector of the Fan and take off it.
NOTE: There is arrow sign on the Fan which shows the direction of the wind, the wind should blow
to the compressor. The side with label should be put inside, align at the hole on the plate.
18
Fig.24
6.8 Control board replacement
The control board controls the main parts, such as power, valves, sensor and alarm system functions
etc. Consult the troubleshooting chart in Part 7.0 to determine when to replace the printed circuit board.
CAUTION: The control board contains components that are sensitive to electrostatic discharge (ESD)
that can damage the board if not handled properly. As when handling any ESD sensitive board, observe
standard ESD safety procedures. These procedures include the following:
Handle the control board by the edges only.
Work on a grounded ESD mat.
Wear a grounded wrist strap.
Store control board in anti-static bags only.
6.8.1 Control Board Removal
(Note that the Control Board on the OSD unit is different to the Control Board on the standard unit.)
(1) Turn off the power, and unplug the power cord.
(2) Remove the front and cabinet backs.
(3) Pull out the tube which connect the oxygen tank and control board, disconnect all the leads,
connectors on the control board.
(4) Use nipper pliers to clamp the board fixed PCB support, lift the control board to remove, there are
four PCB support need to disconnect one by one.
PCB Support
Fig. 25
19
6.8.2 Control Board Installation
To install the control board, follow its removal procedure in reverse order. Aim the install holes of the
control board to the fixed PC barb columns, and then press down the control board. Connect all tubes,
leads, connectors to the control board as they were before.
NOTE: Handle the new circuit board only by the edges to prevent
electrostatic damage to the unit. Remember the place of each tube,
leads and connectors to install correctly.
There is a black temperature sensor line on the back of the board,
when install the board, the sensor line mush insert in the hole of the
plate first.
Sensor line the hole
When install the control board with OSD sensor, the connect spot
of the tubes and OSD sensor should use cable tie to tighten.
6.8.3 Other tips
Buzzer The buzzer is a fixed component on the control board and is not individually replaceable.
Press Switch
The high and low pressure alarms are activated by a pressure transducer on the control board, and
is not individually replaceable.
Time Setting Key
Keypad Replacement procedures:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Disconnect the connectors from the control board. Tear off the key film to remove the keypad.
(4) Replace keypad.
6.9 Transformer replacement
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Disconnect the input connector, and then disconnect its output connector from the control board.
(4) Remove the two fixed screws, and take off the transformer.
(5) Replace a new transformer and follow the removal procedure in reverse order to install.
6.10 LCD replacement
To replace the radiator in the OC unit, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back and front.
(3) Pull out the tubes, disconnect the leads on the LCD board.
(4) Remove the four screws which fix the LCD board, and then take off the LCD.
(5) Follow the removal procedure in reverse order to install a new LCD.
Fig.26
20
6.11 Pressure regulating valve
6.11.1 Pressure regulating valve check and set
The pressure regulating valve enables you to set the maximum flow of oxygen output by the OC unit.
To check for proper adjustment of the product regulator, take the following steps:
(1) Turn on the power.
(2) Allow the unit to run for a few minutes.
(3) Remove the cabinet back.
(4) Connect a pressure gauge directly to the outlet.
(5) The pressure should read 7.345 ± 10% psi
(6) Adjust the regulator if necessary. Pull out downward the bottom of the pressure regulating valve,
turn the bottom clockwise or anticlockwise while reading the gauge to get a proper pressure.
(7) Reinstall the cabinet front and cabinet backs.
6.11.2 Pressure regulating valve replacement
If the flow meter ball fluctuates, or low oxygen concentration, the pressure regulating valve may need
check, adjust or replace. To replace the pressure regulating valve in the OC unit, take the following steps:
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Pull out the outlet tube, and then holding tightly the pressure regulating valve, turn anti-clockwise
to remove it.
(4) Follow the removal procedure in reverse order to install a new pressure regulating valve.
(5) Reset the pressure of the valve.
Note: When install the pressure regulating valve, apply 704 glue on the thread, and the tube which
connect the outlet should be tightened with cable tie.
Fig.27
21
Fig.28
Fig.29
6.13.2 Flow Meter Installation
To install a new flow meter, follow the flow meter removal procedure in reverse order. Then perform
a leak test on the connections.
22
6.14 Circuit breaker replacement
(1) Turn off the power, and unplug the power cord.
(2) Remove the cabinet back.
(3) Unscrew the nut of the circuit breaker.
(4) Disconnect the leads, replace the circuit breaker.
(5) Follow the removal procedure in reverse order to install a new circuit breaker.
Fig.30
Compressor runs with intermittent Defective sieve module. Replace sieve module.
high pressure alarm and low oxygen Restriction in exhaust muffler. Defective pneumatic Replace or clean muffler foam.
concentration. valve. Replace Pneumatic valve.
Compressor safety valve releases Defective pneumatic valve. Replace pneumatic valve.
(popping sound). Contaminated sieve module. Replace sieve module.
Defective safety valve. Replace safety valve.
Constant alarm with power on. Circuit Defective circuit breaker. Replace circuit breaker.
breaker repeatedly trips. Defective capacitor. Replace capacitor.
Defective compressor. Replace compressor.
Defective circuit board. Replace circuit board.
Alarm does not sound. Defective power switch. Replace power switch.
Defective buzzer. Replace control board.
Defective pressure sensor. Replace control board
Flow meter fluctuates. Improperly set or faulty pressure regulator valve. Check regulator valve or replace regulator valve.
Leak. Test for leaks.
Worn compressor. Replace compressor
Defective flow meter. Replace flow meter.
Kinked tubing Check tubing that connects the top of the sieve beds.
Note: Please do not turn the button abruptly, or the
value spool may be broken. Turn the knob slowly and
softly
Cooling fan does not turn. Defective cooling fan. Replace cooling fan.
Defective electrical connections. Check electrical connections.
Limited or Iow flow. Restriction in humidifier or tubing. Pressure Replace humidifier or tubing.
regulator valve set too low. Adjust regulator setting.
Leak. Leak test and repair leak.
Weak compressor. Check system pressure, exchange compressor.
Check filter, suction resonator for obstruction.
Air flow obstruction.
Low concentration. Compressor inlet filter is dirty or partially blocked. Replace inlet filter.
System leak
Faulty compressor Leak test and repair leak.
Unit temperature too high, Replace compressor.
24
Blocked air intake or dirty inlet filter, faulty cooling fan.
Restriction in exhaust muffler, Contaminated sieve Replace or clean exhaust muffler.
module. Check outlet pressures are within range. Replace sieve
module.
Defective pneumatic valve. Replace pneumatic valve.
Restriction of inlet air. Check inlet tubing for obstruction and remove.
* Outlet muffler taken off Reinstall outlet muffler
Exhaust sound too loud
* Outlet muffler broken Replace outlet muffler
No exhaust sound but safety valve * Pneumatic valve broken Replace pneumatic valve
has air out * control board broken Replace control board
Power switch on, the equipment * the flow knob is too tight or broken Replace flow meter
works well, but no oxygen out and * the equipment is leaking badly Leak test and repair leak.
ball of flow meter at the bottom not
rising up.
The concentration is within parameter The oxygen sensor device on the control board Replace control board
ranger by Oxygen density sensor error
test, the LCD realtime display low
concentration or alarm.
Defective Pressure equaling valve Replace pressure equaling valve.
The equipment works well, but the
oxygen concentration is low.
Oxygen concentration
low or no oxygen
exhaust
Ⅶ. TROUBLE S OF PRE SSURE RE DUCING VALVE
25
NO
YES
YES
NO
The tools needed for you to properly service the OC unit are listed below, requires no special tools:
An accurate pressure test gauge to take pressure readings on the OC unit should be kept available
at all times.
26
TOOLS MATERIALS
"T" Wrench
4mm
Oxygen
concentration
analyzer
Tube cutter
OC-S30
27
OC-E30 \ OC-P30
QUALITY WARRANTY
28
Model Warranty Period
OC-S 30 12Months
OC-S 50 12 Months
OC-S 80 12 Months
OC-E 30 12 Months
OC-E 50 12 Months
OC-E 80 12 Months
OC-P 30 12 Months
OC-P 50 12 Months
OC-P 80 12 Months
OC-P 100 12 Months
29