Professional Documents
Culture Documents
Centaur - Release Notes V 3.3.3 Rev 2
Centaur - Release Notes V 3.3.3 Rev 2
Centaur - Release Notes V 3.3.3 Rev 2
3.3.3 2016-11-08
Software Summary
Firmware version 3.3.3 fixes a critical SD archive recording bug where continuous data
archiving to the removable SD card may stop unexpectedly until the operator reboots the unit.
This bug was present in firmware versions 3.x.x prior to version 3.3.3. This release also fixes an
issue where, on very rare instances, an unexpected reboot (such as a reboot due to a power
outage) could hang the unit until the power was cycled again. This release also significantly
improves the write lifetime of SD cards, and fixes a rare issue where the unit boots up with a
persistent timing offset of up to 1 second. It is highly recommended that you upgrade to this
release.
www.nanometrics.ca
techsupport@nanometrics.ca 1 of 11
January 20, 2017
(Release note revision 2)
Data would continue to be recorded to the internal Store ringbuffer. If the Centaur was also
configured for streaming, data would continue to stream properly, however no further data
would be recorded on the removable SD card until the instrument was rebooted. Firmware
version 3.3.3 fixes this issue.
With this release the continuous data archive on the removable SD card is updated a minimum
of every five to six minutes, for aggregate sample rates of 100 Hz or less. For higher aggregate
sample rates, the SD card is updated more frequently. For example, if the configured aggregate
sample rate is 1000 Hz, the SD card is updated approximately once per minute.
Fixed issue where Centaur may hang indefinitely after an unexpected system
reboot
In extremely rare cases, an unexpected system reboot (for example, a system reboot caused by
a temporary power loss or recovery from a system error), may hang indefinitely until power to
the Centaur is removed and reapplied. Until the power is cycled, the Centaur is not responsive
to external commands or button presses and does not digitize or stream data. Firmware version
3.3.3 fixes this issue.
www.nanometrics.ca
techsupport@nanometrics.ca 2 of 11
January 20, 2017
(Release note revision 2)
Workaround: To ensure that all desired configuration changes have taken effect:
● After clicking the A pply button, wait 2 minutes before clicking the C
ommit
button.
● Observe the instrument behaviour. For example, view the W aveform page to
ensure that a the newly configured sample rate is in use.
If this does not resolve the error, reboot the instrument.
A blank static default gateway cannot be selected once a non-blank gateway has
been configured
In the Ethernet settings, in the Configuration > Communications menu, if the Static Default
Gateway setting has been configured, or entered but not applied, to a non-blank value, this field
cannot be cleared to remove the Static Default Gateway setting. When this is attempted, a
warning appears indicating that the blank value is invalid, and the new setting cannot be
applied. This could prevent configuring the Centaur to dynamic IP mode. Fixed in firmware
version 3.4.
Workaround: Ensure a valid IP address exists in the Static Default Gateway setting
before configuring the instrument for dynamic IP mode. The instrument will ignore the IP
address in the Static Default Gateway setting if it is a valid IP address.
www.nanometrics.ca
techsupport@nanometrics.ca 3 of 11
January 20, 2017
(Release note revision 2)
Operating without a GPS signal for a prolonged period could result in an erroneous
time change
If a Centaur has been operating for more than a week since a power-cycle, reboot or
configuration change without the GPS receiver ever locking, the instrument time could jump to
the year 1999. The instrument will revert to the correct time once it locks to a valid GPS signal.
Fixed in firmware version 3.4.
Response files for Titan sensors cannot be downloaded for sensitivities less
than 1 g
,
When attempting to download response files for a Titan sensor with range set to less than ±1 g
an error will be displayed indicating that the response cannot be found. This issue will be
addressed in an upcoming firmware release.
Incomplete information on the Web Interface Sensor page for custom sensor
configurations
When configuring a sensor using the custom sensor configuration form in the Sensor Library,
the Web Interface Sensor page needs to be manually refreshed to display the following
information:
● for custom sensors connected to port B of a 6-channel Centaur, the C ontrol B panel is
not displayed
● for custom sensors connected to port A of a Centaur (both 3- and 6- channel), the
Sensor page is completely blank
ensor page will not update automatically.
● After a configuration change, the S
www.nanometrics.ca
techsupport@nanometrics.ca 4 of 11
January 20, 2017
(Release note revision 2)
Workaround: If using sample rates of 5000 Hz, configure the archive file length to be 12
hrs or less
Maximum of two simultaneous users can display waveform page in web interface
Centaur allows a maximum of two web interfaces to display waveforms simultaneously. If more
than 2 web interfaces attempt to display the Waveform or Summary pages at the same time,
an error indicating that there is a problem with receiving waveform data will be displayed.
www.nanometrics.ca
techsupport@nanometrics.ca 5 of 11
January 20, 2017
(Release note revision 2)
Firmware upgrade will cause the Linux root password to revert to the factory
default password
If upgrading the firmware from version 3.1.3 or earlier to version 3.2.8 or later, the passwords for
logging in to the Linux root or the calibration file upload accounts will revert to factory default
passwords. The passwords to log on to the Web user interface are preserved.
Workaround: If you have previously changed the Linux passwords, you must use the
factory default password to login and manually update them to the desired passwords.
Search for “Calibration Password” in the on-line help or in the PDF User Guide.
Users upgrading from a firmware version that is between 1.1.15 and 1.1.24
Most users can use the standard firmware upgrade process detailed in the Centaur on-line help
or user guide. For users who have pre-release firmware versions between 1.1.15 and 1.1.24,
follow the instructions below.
www.nanometrics.ca
techsupport@nanometrics.ca 6 of 11
January 20, 2017
(Release note revision 2)
Firmware versions 1.1.15 through 1.1.24 are susceptible to a potential memory corruption issue
that can result in unstable behaviour. This issue was resolved in firmware version 1.1.32.
Centaur-6 instruments are more susceptible than Centaur-3 instruments.
When upgrading from a firmware version between 1.1.15 and 1.1.24, follow these instructions:
Cannot login to Linux root after downgrading from firmware version 3.x.x to
version 1.x.x/2.x.x
If you downgrade an instrument from firmware version 3.x.x or later back to versions 1.x.x/2.x.x,
the Linux root password will no longer work. To login to the Linux root account, you must
upgrade again to a 3.x.x version. The normal web UI user and admin logins are not affected.
Most users never need to use the Linux root account.
Other Considerations
Maximum Archiving and Streaming Rates
The Centaur supports a maximum “aggregate sample rate” of 5010 Hz. This aggregate rate is
the sum of the sample rates of the primary and secondary channels from all sensors, and can
involve a combination of streaming and SD archiving. Examples:
● a 3-channel Centaur with a sample rate of 5000 Hz archiving the primary channels and
10 Hz streaming the secondary channels has an aggregate rate of 5010 and is
supported.
www.nanometrics.ca
techsupport@nanometrics.ca 7 of 11
January 20, 2017
(Release note revision 2)
● a 3-channel Centaur with a primary rate of 100 sps that is being archived and also
streamed simultaneously to two destinations has an aggregate rate of 300 and is
supported.
● a 6-channel Centaur with both sensor A and B are sampled at 5000 Hz has an
aggregate rate of 10,000 and is not supported.
● a 6-channel Centaur with Sensor A and B each having a primary sample rate of 1000 Hz
being archived and a secondary rate of 100 Hz being streamed has an aggregate rate of
2200 and is supported.
Note that at higher sample rates, a larger frames per packet setting may be required. The web
interface will indicate if the frames per packet setting is too low for a particular sample rate.
Minimum swap time for gap-free archives when exchanging SD removable media
The Centaur buffers data to be written to the SD card archive when the SD media is being
exchanged to preserve a contiguous archive record between the old and new cards. However,
the time interval permitted for gap-free media exchange can be quite short, depending on the
sample rates and frames-per-packet selections. The following formula can be used to determine
the time (T) within which the SD media can be exchanged and preserve a contiguous archive:
S S
T = 2850 / ( F1 + F2 + ...) ,
1 2
where S1,S2 are the primary and secondary sample rates of sensor A and/or B, and F1,F2 are
the corresponding frames-per-packet settings.
For example, a 6-channel Centaur with primary channels configured with 100 Hz, 7 fpp for
sensor A and 500 Hz, 42 fpp for sensor B, both with secondary channels configured for 10 Hz, 1
fpp would allow one minute to swap SD cards and still have a contiguous archive:
T = 2850 / (100
7 + 42 + 1 + 1 ) = 61s
500 10 10
Likewise, a 3-channel Centaur with a single 100 Hz channel at 7 fpp would allow more than 3
minutes to swap cards, but the card exchange for a single 5000 Hz channel at 42 fpp would
have to be done within 24 s to ensure a contiguous record. This does not affect the internal
Store which will always have a continuous record.
www.nanometrics.ca
techsupport@nanometrics.ca 8 of 11
January 20, 2017
(Release note revision 2)
Workaround: If the newly recreated store has been corrupted, deleting the Store and
restarting the instrument will allow normal operation. Contact
techsupport@nanometrics.ca for a procedure to delete the Store.
No waveforms displayed on Summary page for Centaur with greater than 6 active
channels (Centaur-6 only)
Waveforms may not display on the S ummary page on 6-channel Centaur instruments that are
configured to have more than 6 active channels (primary and secondary channels combined).
aveform page as expected.
All active waveforms will display on the W
Workaround: On the Waveform page, select the channel(s) of interest and observe
ummary
them being rendered. Close the browser tab and re-open it, or navigate to the S
page and force a browser reload. The channels that were selected on the W aveform
ummary page.
page will now be displayed on the S
Waveforms may not display correctly on the Web Interface for older versions of
web browsers
The Web Interface requires websocket support to display waveforms correctly. Current versions
of all major web browsers support websockets, however older versions of web browsers may
not support this feature. If you encounter display difficulties, ensure that you have installed the
minimum required version for your web browser.
Workaround: If you encounter display difficulties, ensure that your web browser meets
the version requirements below, and upgrade your browser software if required. Note
that Internet Explorer and Microsoft Edge are not supported at this time.
www.nanometrics.ca
techsupport@nanometrics.ca 9 of 11
January 20, 2017
(Release note revision 2)
Firefox 11
Chrome 16
Safari 7
Note: The firmware supports the operation of the high-gain digitizer option, however the Web
Interface and configuration settings do not yet display the correct high gain ranges and
sensitivities. See the workaround below for details.
1 40 0.4 4 10 1.6
2 20 0.8 8 5 3.2
10 4 4 40 1 16
20 2 8 80 0.5 32
40 1 16 160 0.25 64
The Sensor B inputs on Centaur model CTR2-6S/H-8, and all sensor inputs on Centaur models
CTR2-3H-8 and CTR2-6H-8 are high-gain digitizers that have 4 times the gain of the standard
www.nanometrics.ca
techsupport@nanometrics.ca 10 of 11
January 20, 2017
(Release note revision 2)
channels. The Web Interface does not yet reflect this. Therefore, the displayed sensitivity for
those channels on the Health page is one-quarter of the actual value, and the selected I nput
range [V] is 4 times greater than the actual value. For example, if the selected Input range is 40
Vpp, the actual input range will be 10 Vpp. See the S ensitivity table above.
Workaround: Select an input range that is 4 times greater than the actual range that
you want. For example, if you want to use gain of 80 (0.5 Vpp range), choose the 2 Vpp
ealth page by 4. Note that response files
range. Divide the sensitivity displayed on the H
provided by Nanometrics for these gains have the correct sensitivities. It is simply that at
this time, the Centaur Web Interface and response files downloaded from the Web
Interface have not been updated to reflect the new gains.
www.nanometrics.ca
techsupport@nanometrics.ca 11 of 11