Percent of Output Current Output Current


Input Pressure (Percent of Accuracy
Range (Ideal) (Measured)
Span)

0% 0 PSI 4 mA 4 mA 0% 100%
25% 50 PSI 8 mA 8.78 mA 4.88% 95.12%
50% 100 PSI 12 mA 13.25 mA 7.81% 92.19%
75% 150 PSI 16 mA 17.2 mA 7.50% 92.5%
100% 200 PSI 20 mA 20.5 mA 3.13% 96.87%

ERROR:
4−4
1. x 100=0 %
16

8.78−8
2. x 100=4.88 %
16

13.25−12
3. x 100=7.81%
16

17.2−16
4. x 10 0=7.50 %
16

20.5−20
5. x 100=3.13 %
16

ACCURACY:
1. 100% - 0% = 100%
2. 100% - 4.88% = 95.12%
3. 100% - 7.81% = 92.19%
4. 100% - 7.50% = 92.5%
5. 100% - 3.13% = 96.81%