Output Error

Percent of Output Current


Input Pressure Current Accuracy
Range (Percent of
(Ideal)
(Measured) Span)

0% 0 PSI 4 mA 4.02 mA 0.125% 99.88%


25% 50 PSI 8 mA 6.65 mA -8.44% 99.94%
50% 100 PSI 12 mA 11.55 mA -2.81% 97.19%
75% 150 PSI 16 mA 18.22 mA 13.88% 86.12%
100% 200 PSI 20 mA 20.1 mA 0.625% 99.38%

ERROR:
4.02−4
1. x 100=0 . 125 %
16

6.65−8
2. x 100=−8.44 %
16

11.55−12
3. x 100=−2.81 %
16

18.22−16
4. x 10 0=13.88 %
16

20.1−20
5. x 100=0.625 %
16
ACCURACY:
1. 100% - 0.125% = 99.88%

2. 100% -|−8.44 %| = 99.94%

3. 100% - |−2.81 %| = 97.19%

4. 100% - 13.88% = 86.12%


5. 100% - 0.625% = 99.38%

GRAPH:

HYSTERISIS ERROR
Output (IDEAL) Output (MEASURED)
20

16

12

0
0 50 100 150 200