Service Report Maclab Gorontalo

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Service Request Type: Service Request Number:

Service Report Corrective Repair 2022-02/02034

Customer Details

Customer Name: RSUD.Prof Aloei Saboe Location: Jakarta-Indonesia


PO #: Contact Phone: +6281356604725
Contact Nome: Mr.Fahmi Sjawie Email Address:
Initial Problem Description: EKG Artefact

Equipment Details

System ID: - Initial Equipment Status: Hemodinamik monitoring


Serial Number: - Entitlement: -
Maclab ( Marquet GE )
Product Description:

Activity Description

Activity #: Activity Close Dote:


Activity Type: Field Support Engineer Name: Sukrisdianto
Tube Slice Reading:
Problem Description: Hemodinamik not available

Couse/ Diagnosis:
Feb,10 Feb 2022
- EKG result Artefact
- Presure ABP intermithen
- Spo2 malfuction
- Create Charting report from hemodinamik monitoring
- Zerro On ABP malfuction

Resolution/ Service Performed: Feb 10-Feb-2022:


- Check and inspect tramp module hemodinamik OK
- Check and repaired lead cable EKG OK
- Check and calibration pressure ABP ( Arterial Blood Presure ) OK
- Check and test sensor transduser ABP OK
- Check and calibration finger sensor Spo2 OK
- Check and inspect pressure bag OK
- Creat and seting report from hemodinamik OK

Recommendotion :

“ Maintenance according to schedule and provide stock Ekg cable “

Page No. 1 of 2
Service Request Type: Corrective Repair
Service Report Request Number: 232202-20

Date Time Hours Labour Type


Feb 10, 2022 08:30-17:00 Onsite Repair
Nov 10,2022 17:00-00:00 Zona Travel

Ports Used Details

Port Number 0tY Description Asset Serial No

Expense Tracker

Expense Dote Type 0tY Comments

Engineer Signoture: Customer Signoture:


10-Feb-2022 17:00
SUKRISDIANTO
Page No.2 of 2