Professional Documents
Culture Documents
Uji Kesesuaian Standar Nilai HVL Filter Aluminium Pada Pesawat Sinar-X Mammografi: Studi Kasus Di Ruang Instalasi Radiologi Rs. Siloam Makassar
Uji Kesesuaian Standar Nilai HVL Filter Aluminium Pada Pesawat Sinar-X Mammografi: Studi Kasus Di Ruang Instalasi Radiologi Rs. Siloam Makassar
ABSTRACT
The quality control of X-ray radiation in a radiology device plays an important role in
determining the clinical use feasibility. The HVL value test of a filter is very important because
it determines the effective of X-ray energy. HVL values indicate the thickness of the absorbent
material to absorb X-ray intensity up to half of its original intensity. This research is devoted
to testing the feasibility of Mammography equipment in the radiology installation room of
Siloam Makassar hospital by comparing two methods of determinations the HVL value in the
variation of voltage 23, 25, 27, 29, 31 and 33 kVp. The HVL value that determined from direct
methods without filters is used as a base value that to be compared with the HVL value which
determined by indirect methods through calculations after the addition of aluminum filter thicks
to gradually. HVL values by indirect measurements ranged from 0.268−0.401 mmAl, while by
direct measurements between 0.291−0.338 mmAl with a deviation between 1.39−7.90%. The
results obtained indicate that the HVL value for all X-ray tube voltage variations from the two
methods in use still on ranges the tolerance standard so that it can be reported that the X-ray
mammography at the Siloam hospital was declared feasible.
Keywords: mammography, HVL value, the quality of radiation
ABSTRAK
Kontrol kualitas radiasi sinar-X dalam perangkat radiologi memainkan peran penting dalam
menentukan kelayakan penggunaan klinis. Uji nilai HVL filter sangat penting karena
menggambarkan energi efektif sinar-X. Nilai HVL menunjukkan ketebalan bahan penyerap
untuk menahan intensitas radiasi sinar-X hingga setengah dari intensitas mula-mula. Penelitian
ini memiliki tujuan untuk menguji kelayakan peralatan Mammografi di ruang instalasi radiologi
rumah sakit Siloam Makassar dengan membandingkan dua metode penentuan nilai HVL dalam
variasi tegangan 23, 25, 27, 29, 31 dan 33 kVp. Nilai HVL yang ditentukan dari metode
langsung tanpa filter digunakan sebagai nilai dasar yang akan dibandingkan dengan nilai HVL
yang ditentukan dengan metode tidak langsung melalui perhitungan setelah penambahan tebal
filter aluminium secara bertahap. Nilai HVL dengan pengukuran tidak langsung berkisar antara
0,268−0,401 mmAl, sedangkan dengan pengukuran langsung antara 0,291−0,338 mmAl
dengan deviasi antara 1,39−7,90%. Hasil yang diperoleh menunjukkan bahwa nilai HVL untuk
semua variasi tegangan tabung sinar-X dari dua metode yang digunakan masih pada rentang
17
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
standar toleransi sehingga dapat dilaporkan bahwa mamografi sinar-X di rumah sakit Siloam
dinyatakan layak.
Kata Kunci: mamografi, nilai HVL, kualitas radiasi
18
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
19
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
20
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
HVL secara tidak langsung dimana nilai- untuk satu nilai tegangan tabung. Setengah
nilai Da, Db, ta dan tb selanjutnya akan diolah dari dosis mula-mula (D0/2) merupakan
untuk mendapatkan nilai HVL berdasarkan dasar perkiraan bahwa nilai HVL dari suatu
persamaan (2). Tahap ketiga adalah filter akan mampu melewatkan intensitas
pengukuran nilai HVL secara langsung radiasi yang setara dengan nilai D0/2
tanpa menggunakan filter untuk tersebut.
memperoleh nilai HVL patokan yang Tabel 1 menunjukkan terjadinya
terbaca langsung melalui multimeter. Pada peningkatan nilai dosis mula-mula setiap
tahap kedua, penetapan nilai dosis di atas peningkatan tegangan tabung sinar-X.
dan di bawah nilai dosis D0/2 disesuaikan Demikian juga untuk nilai D0/2, perubahan
dengan kombinasi ketebalan filter dosis setara dengan kenaikan faktor
aluminium yang disusun untuk setiap variasi tegangan eksposi.
tegangan tabung.
Hasil pengukuran dosis atas (Da) dan
Pengaturan tebal filter yang dipilih dosis bawah (Db) dari nilai setengah dosis
sebagai ta dan tb pada setiap variasi tegangan mula-mula (D0/2) ditunjukkan pada Tabel 2.
pada tahap kedua disesuaikan dengan hasil Tahapan penelitian yang kedua ini
pembacaan dosis yang menghasilkan nilai merupakan metode penentuan nilai HVL
sedikit di atas dan di bawah nilai dosis D0/2.
secara tidak langsung. HVL diperoleh
Selisih nilai di atas dan di bawah dosis D0/2 melalui perhitungan berdasarkan nilai Da,
yang paling mendekati secara pasti tidak Db, ta dan tb yang terukur.
dapat ditentukan karena pengaturan
penambahan filter hanya disesuaikan dengan Untuk pengukuran secara tidak
ketersediaan tebal pelat filter yang ada. langsung menggunakan filter aluminium
pada ketebalan berbeda untuk
memperkirakan nilai dosis di atas dan dosis
HASIL DAN PEMBAHASAN di bawah nilai setengah dosis mula-mula.
Selain itu, ditunjukkan juga batas nilai
Nilai dosis mula-mula pada setiap variasi
minimum dan maksimum dari toleransi nilai
tegangan tabung dapat dilihat dalam Tabel 1.
HVL yang memenuhi untuk setiap faktor
Dosis mula-mula (D0) sebagai dasar untuk
tegangan tabung.
mendefenisikan nilai dosis yang terukur
Tabel 1. Data nilai dosis mula-mula (D0) dan nilai setengah dosis mula-mula (D0/2).
Arus-waktu D0 (mGy)
No Tegangan (kVp) (mAs) Rata-rata D0/2(mGy)
1 2 3
1 23 4,120 4,121 4,121 4,121 2,061
2 25 5,272 5,272 5,272 5,272 2,636
3 27 6,596 6,595 6,596 6,596 3,298
50
4 29 50 8,017 8,018 8,019 8,018 4,009
5 31 9,561 9,562 9,563 9,562 4,781
6 33 11,15 11,14 11,15 11,15 5,575
21
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
Tabel 2. Data nilai Da, Db, ta, tb serta hasil perhitungan nilai HVL pada metode tidak langsung.
Teganga 𝐷0 𝐷0⁄2 𝐷𝑎 𝐷𝑏 𝑡𝑎 𝑡𝑏 HVL Batas nilai
N HVL
n (mGy (mGy (mGy (mGy (mmAl (mmAl (mmAl
o
(kVp) ) ) ) ) ) ) ) Min Maks
1 23 4,121 2,061 2,292 1,860 0,216 0,318 0,268 0,26 0,35
2 25 5,272 2,636 2,802 2,541 0,271 0,326 0,305 0,28 0,37
3 27 6,596 3,298 3,332 3,253 0,318 0,326 0,321 0,30 0,39
4 29 8,018 4,009 4,190 3,771 0,318 0,373 0,341 0,32 0,41
5 31 9,562 4,781 5,147 4,659 0,318 0,373 0,359 0,34 0,43
6 33 11,15 5,575 6,018 5,174 0,373 0,428 0,401 0,36 0,45
Penjelasan dari penetapan nilai ta dan tb nilai setengah dosis mula-mula sebesar
pada Tabel 2 adalah sebagai berikut. 1,860 mGy, dibutuhkan tebal filter (𝑡𝑏 )
Misalkan untuk data tegangan 23 kVp sebesar 0,318 mmAl melalui penumpukan 3
dengan nilai D0/2 sebesar 2,061 mGy, pelat aluminium ukuran 0,106 mmAl.
dibutuhkan dosis di atas (Da) nilai setengah
Demikian seterusnya untuk variasi
dosis mula-mula (D0/2) sebesar 2,292 mGy,
tegangan eksposi 25, 27, 29, 31 dan 33 kVp,
sedangkan dosis di bawah (Db) nilai
dilakukan prosedur seperti di atas (data
setengah dosis mula-mula (D0/2) sebesar
kalkulasi tidak ditampilkan). Dari hasil
1,860 mGy. Untuk mendapatkan estimasi
perhitungan nilai HVL yang ditunjukkan
kedua dosis tersebut (Da dan Db) diperoleh
pada Tabel 2, rata-rata hasil tidak jauh dari
dengan menggunakan dua atau tiga pelat
penetapan nilai batas minimum range
aluminium yang ditumpuk. Nilai Da
toleransi HVL. Semua data hasil
diperoleh melalui penggabungan filter
perhitungan masuk dalam range tersebut
aluminium yaitu 1 pelat ukuran 0,106 mmAl
sehingga dapat dinyatakan bahwa pesawat
dan 2 pelat ukuran 0,055 mmAl dengan tebal
mammografi di Instalasi Radiologi RS.
keseluruhan filter (𝑡𝑎 ) sebesar 0,216 mmAl. Siloam Makassar dalam kondisi layak.
Sedangkan untuk nilai dosis di bawah (Db)
22
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
Tabel 4. Perbandingan nilai HVL pada metode langsung dan tidak langsung.
Pada tahap ini, penentuan nilai HVL langsung untuk semua variasi tegangan
dilakukan dengan cara langsung membaca selengkapnya disajikan dalam Tabel, 4
nilai HVL yang terukur pada display sedangkan Grafik perbandingan nilai HVL
multimeter X-ray Raysafe untuk setiap dari kedua metode disajikan dalam Gambar
variasi tegangan tabung. 3.
Tabel 3 menunjukkan hasil pengukuran Perbandingan nilai HVL dari kadua
langsung nilai HVL untuk setiap nilai metode pengukuran sebagaimana dalam
tegangan tabung. Dengan nilai batas Tabel 4 tidak jauh berbeda dengan deviasi
minimal dan maksimum toleransi range <10%. Nilai HVL yang diperoleh melalui
yang dipersyaratkan, hasil pengukuran nilai dua metode yang dilakukan untuk semua
HVL untuk semua variasi tegangan tabung variasi tegangan tabung sebesar
pada pengukuran langsung berada pada nilai 0,291−0,388 mmAl (untuk pengukuran
tengah range. langsung) dan sebesar 0,268−0,401 mmAl
(pengukuran tidak langsung) dengan deviasi
Hal ini yang menyebabkan dalam
sebesar 3,35−7,90 %.
analisis penelitian nilai HVL hasil
pengukuran dengan metode langsung dapat Pada tegangan 25 kVp didapatkan nilai
dijadikan sebagai nilai patokan. Selain itu, HVL sebesar 0,305 mmAl untuk
nilai HVL dari pengukuran langsung dapat pengukuran tidak langsung dan
diasumsikan merupakan nilai HVL yang 0,311 mmAl pada pengukuran secara
sebenarnya dari setiap nilai tegangan tabung langsung dengan deviasi yang menurun
karena alat ukur multimeter X-ray Raysafe cukup signifikan dibanding tegangan 23 kVp
yang digunakan sensitif dan akurat. Setiap yaitu menurun sebesar 5,97 % karena adanya
kenaikan tegangan 2 kVp diperoleh sebaran dari variabel data yang kecil atau
peningkatan nilai HVL sebesar 0,02 mmAl. tidak adanya kesenjangan yang cukup besar
terhadap kedua pengukuran.
Perbandingan nilai HVL dari kedua metode
pengukuran secara langsung dan tidak
23
Berkala Fisika ISSN: 1410 - 9662
Vol. 23, No. 1, Januari 2020, Hal. 17-25
0,45
Pengukuran Tidak Langsung
Pengukuran Langsung
0,4 y = 0,2776e0,0583x
R² = 0,9883
Nilai HVL (mm Al)
0,35
0,3
y = 0,2552e0,0733x
R² = 0,9743
0,25
0,2
23 25 27 29 31 33
Tegangan (kVp)
Gambar 3. Hubungan tegangan terhadap nilai HVL dari dua metode pengukuran secara
langsung dan tidak langsung.
[5] Richard B. The impact of increased Al [11] Muslim G. Penentuan half value layer
filtration on X-ray tube loading and (HVL) pada computed radiography
image quality in diagnostic radiology. (CR) dengan metode pengolahan citra
Medical Physics. 2003;30:69–78. digital. Makassar: Skripsi,
[6] Papp J. Quality management in the Departemen Fisika, Fakultas Ilmu
imaging sciences. Mosby Elsevier; Pengetahuan Alam, Universitas
2006. Hasanuddin; 2017.
[7] Tung CJ, Lin MT, Hsu FY, Lee JH, [12] Yudi P. Penentuan nilai HVL (half
Chu CH, Tsai HY. Half-value layer value layer) dari pelat aluminium
determination using dengan variasi ketebalan filter
thermoluminescent dosimeters for menggunakan pesawat linac (linear
digital mammography. Radiation accelerator) dengan metode
Measurement. 2015;45:729-732. ekstrapolasi. Makassar: Skripsi,
[8] Lincewati S, Afrina SH, Martha R, Departemen Fisika, Fakultas Ilmu
Timbangen S, Alemin BD. Pengetahuan Alam, Universitas
Determination of half value layer Hasanuddin: 2017.
(HVL) value on X-rays radiography [13] Cicilia A, Suryono, Evi S. Penentuan
with using aluminum, copper and lead nilai tebal paruh (HVL) pada citra
(Al, Cu, and Sn) attenuators. Journal digital computed radiography.
of Physics: Conference Series. Youngster Physics Journal.
2018;1116:032032. 2015;4:55–60.
[9] Aswad M, Abdullah B, Tahir D. [14] Devi YH, Evi S, Choirul A. Evaluasi
Determination of the half-value layer metode penentuan half value layer
by using filter aluminum sheets for (HVL) mengguakan multi purpose
digital mammography. International detector (MPD) Barracuda pada
Journal of Engineering and Science pesawat sinar-X mobile. Youngster
Applications. 2018;15-22. Physics Journal. 2014;3113–118.
[10] Hendee WR, Ritenour ER. Medical
imaging physics.
New York: Wiley-Liss, Inc; 2002.
25