PACB108 Operation Manual

PONOVO POWER CO., LTD

No. 139 Jinghai Third Road, BDA, Beijing, China, 100176


Office TEL. +86 (10) 59089666
E-Mail Info@relaytest.com
Website www.relaytest.com / www.ponovo.com.cn

PACB108
OPERATION MANUAL

VERSION: PACB108-AE-1.10
DATE: AUGUST 2020

This manual is the publication of PONOVO POWER CO., LTD. Any form of copy should
obtain the consent of it.

This manual represents the technical status for the moment of publishing. The product
information, description and specifications mentioned in the manual do not have any
contact binding force and PONOVO POWER CO., LTD remains the right to make
modifications to the technical specifications and configurations without prior notice.
PONOVO POWER does not take responsibility to the possible error/mistakes in this
manual.

1
PACB108 Operation Manual

PACB08 consists of two parts: optical scanning head and pulse convertor box.

The functions of optical scanning head:


1. Convert optical signal into pulse signal on meter;
2. Convert the marker line of turntable into pulse signal.

Pic 1: Scanning Head

The functions of pulse convertor box:


1. Provide power to optical scanning head;
2. Change the pulse signal into different amplitude pulse signal to provide for other test
equipments such as PW460 and PW336i, etc.
3. Convert pulse output method: single 5m pulse and synchronization optical scanning
head pulse.
4. Extend an output port, type: RJ11. It is used for other optical scanning head input.

2
PACB108 Operation Manual

Pic 2: Convertor Box

Pic 3: Sensitivity Adjustment turn at scanning head

3
PACB108 Operation Manual

Pic 4: Optical output and optical sensor

Pic 5: Power of optical output

4
PACB108 Operation Manual

Pic 6: Focus OK

Pic 7: Focus BAD

5
PACB108 Operation Manual

Method of applications:
1. Convert optical signal into pulse signal
1.1 Switch on the optical output in the optical scanning head and adjust the focal length to
LED position in the meter.
1.2 Switch off the optical signal output.
1.3 Adjust the sensitivity parameter until the optical scanning head synchronizes with
meter’s signal.
1.4 Adjust the “Pulse output method” and “Output level” on the optical convertor according
to requirements.

2. Convert the marker line of turntable into pulse signal.


2.1 Switch on the optical output and adjust the focal length to the turntable position in the
meter.
2.2 Adjust the scanning head’s sensitivity parameters until the marker lines of scanning
head and meter synchronize.
2.3 Adjust the “Pulse output method” and “Output level” on the optical convertor box
according to requirements.