CHECK LIST FOR ELECTRONIC MEMORY GAUGE

Tool Type: Date:

Serial Number: Client / Well No:

CONDITION RESULT

Check for any visual cracks or damage on the gauge housing. √ if correct

Check all threads of the gauge (top, bottom, internal). √ if correct


Mechanical Ensure that the bellow/diaphragm is clean and protector
housing ports are not blocked by any kind of debris or dirt. √ if correct
Checks
Ensure that the bottom sub assembly housing is clean and the
√ if correct
fluid entry ports are not blocked by any kind of debris or dirt.
Remove dirt and old grease from all housing threads and ‘O’
√ if correct
rings and replace with fresh.
Perform a visual inspection of the electronics for debris,
√ if correct
damaged components and wires.
Ensure that electrical connectors are clean, dry and
√ if correct
undamaged.
Electrical Connect to gauge software and check for correct data values
Checks for pressure and temperature. Apply some gentle vibration to √ if correct
expose potential failure.
Carry out bench test by programming with required sample rate
by connecting battery and check if the recorded values are √ if correct
correct by downloading the data.
Calibration Check the calibration date and ensure that the gauge has been
√ if correct
Check calibrated as per its specifications.

Safety TRANSDUCER PROTECTION


Precaution The transducer will not be vibration-damaged under normal down hole conditions, but
high frequency (i.e. Metal to metal) shock should be avoided

REMARKS:

CHECKED BY:

F-704
28-05-06