Professional Documents
Culture Documents
Integrated Autosampler (I-AS) : Agilent ICP-MS
Integrated Autosampler (I-AS) : Agilent ICP-MS
2 Setup 21
Unpacking the I-AS 22
Setting up the I-AS 24
Preparation of the Arm 24
Installing the Support Table or Bracket 26
Tubing and Wiring 27
Installing the Sample Tray 32
Installing the Tray Cover 33
Single Sample Position Setup 34
Software Settings 35
G7201C (ICP-MS MassHunter 4.1 or later) Configuration 35
G7201A/B Configuration 38
G7200B Configuration 42
Settings 44
Checking and Adjusting Operations 46
Checking Operations 46
Installing and Adjusting the Needle 53
3 Operation 63
Measuring with Single Sample Position
(Using the G7200B ICP-MS Software) 64
Measuring with Single Sample Position 64
Single Sample Position Run Dialog Box 65
4 Maintenance 67
Daily Maintenance 68
Power Cable Fuse 69
Features 8
Operations Overview 8
Rinse System 9
System Configuration 10
Main Unit Components 10
Specifications 15
Main Unit 15
External Bottle 15
Single Sample Position 15
Utility Requirements 16
Electromagnetic Compatibility 17
Operating Environment 17
Operating Conditions 17
Storage Conditions 17
Serial Number 18
Safety Information 19
Overview
The I-AS is designed to automate analysis and improve productivity for chemical
research and industrial testing facilities.
The I-AS has six types of sample trays. Replacing the sample tray allows the
instrument to handle sample vials of different sizes. Also, using the Single
Sample Position feature enables the use of a single bottle for analysis.
OPM070-2
OPM070
Features
Equipped with the I-AS, the ICP-MS is able to introduce a sample from a vial on
the I-AS sample tray, or introduce a sample directly from Single sample position.
The I-AS is equipped with a rinse port and an external bottle for cleaning the
needle. Depending on the option selected, up to three external bottles can be
used.
Needle
Three external
bottles
Tray cover
Rinse port
Figure 3. I-AS
Operations Overview
Rinse System
The overflow rinse system (rinse port) uses the peristaltic pump for rinse and
waste fluids.
The internal diameter of the peristaltic pump tubing outlet is larger than inlet.
This prevents rinse fluids from overflowing from the rinse port.
System Configuration
In addition to the sample tray driver and other peripheral equipment, the main
unit is equipped with a needle, arm, and a rinse port for cleaning the needle.
The I-AS is mounted on the ICP-MS and connected to MassHunter Workstation.
The sample tray mounts onto the I-AS.
Arm
Needle
Tray cover
Three external
Sample tray
bottles
Main unit
Reset switch
Needle
The needle is installed on the arm of the main unit. The needle tube is connected
to the sample introduction tube.
Arm
The needle is installed on the arm. The arm can rotate and move up and down to
position the needle in the necessary location.
Reset switch
On the right side of the main unit is a red reset switch. Push this switch to
interrupt or stop the instrument in an emergency situation.
Pushing the reset switch will cancel Autosampler operation, return the needle to
home position, and initialize the I-AS. It will also cancel any analyses in process.
Rinse Port
The rinse port uses an overflow feature to supply rinsing fluid to clean the tip of
the needle.
External Bottle
In addition to cleaning with the rinse port, the external bottle is used when
cleaning with other types of cleaning fluids. Up to three external bottles can be
used. These may also be filled with standard tuning solution.
Sample Tray
The sample tray contains slots for holding sample vials. Insert a sample vial into
a slot on the sample tray.
The following six types of sample trays are available as options. Select the
sample tray that matches the sample vials to be used.
Table 1 Sample Tray Types
Type A 89 6 mL
Type B 53 18 mL
Type C 18 50 mL
Type E 89 6 mL, 18 mL
Single Sample
Position stand
Power switch
Power cable
connector
Power Switch
I-AS power switch.
Figure 8.
Specifications
Main Unit
Dimensions: 290 (W) x 360 (D) x 220 (280, probe included) (H) mm
Weight: 5 kg
External Bottle
The client must provide this bottle. Bottles up to 100 mL may be used. The
allowable bottle dimensions are shown below.
Dimensions: A diameter of 55 mm or less and a height of: 126 mm or less.
Utility Requirements
Electromagnetic Compatibility
EN55011
Group 1 ISM equipment: Group 1 contains all ISM equipment in which there is
intentionally generated and/or used conductively coupled radio-frequency energy
which is necessary for the internal functioning of the equipment itself.
Class B equipment is equipment suitable for use in domestic establishments and
in establishments directly connected to a low voltage power supply network
which supplies buildings used for domestic purposes.
ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
The “Installation category” implies the regulation for impulse withstand voltage. It
NOTE is also called the “Over voltage category.” “II” applies to electrical equipment.
“Pollution degree” describes the degree to which a solid, liquid, or gas which
deteriorates dielectric strength is adhering. “2” applies to a normal indoor
atmosphere.
Operating Environment
Operating Conditions
Ambient temperature: 5 - 35 °C (41 - 95 °F)
Relevant humidity: 20 - 80 % (no condensation)
Storage Conditions
Ambient temperature: -35 - 65 °C (-31 - 149 °F)
Relevant humidity: 15 - 85 %
To avoid personal injury, death, or damage to the instrument, the operator must refer to an
explanation in the Manual.
Alternating current.
ON (power)
OFF (power)
Serial Number
Safety Information
The I-AS has been designed and tested in accordance with IEC publication
1010-1, Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use.
Only connect the I-AS to a power source with a protective earth contact. Use
WARNING only a certificated power cord to make a connection. Connecting the I-AS to a
power source that is not equipped with a protective earth contact creates a
shock hazard for the operator and can damage the instrument. Likewise,
interrupting the protective conductor inside or outside the I-AS or defeating the
power cord ground creates a shock hazard for the operator and can damage
the instrument.
Never use the I-AS in a way not described in the manual. Improper use may
WARNING result in danger to the user.
If rinse fluid, sample fluid, standard fluid or other liquids spill, wipe up
WARNING immediately. These liquids can short circuit or corrode the instruments. Highly
alkaline or acidic fluids can injure the skin and mucus membranes. Exercise
caution when handling these fluids.
Do not cover the air circulation outlet holes on the right side of the I-AS.
CAUTION
Do not remove or install the tray on to the I-AS while the needle is moving. The
CAUTION I-AS can be damaged if the needle is touched during operation.
If you change the tray type used, you must change the tray type in the I-AS
Configuration dialog box.
• Always make sure there is at least 10mm or more distance between the
CAUTION surface of the sample fluid to the top edge of the vial. This is to keep the
sample fluid lower than the rinse port fluid, and ensure the entire needle is
thoroughly cleaned.
• If rinse fluid, sample fluid, standard fluid or other liquids spill, wipe up
immediately. These liquids can short circuit or corrode the instruments.
Highly alkaline or acidic fluids can injure the skin and mucus membranes.
Be very careful when handling these fluids.
Software Settings 35
G7201C (ICP-MS MassHunter 4.1 or later) Configuration 35
G7201A/B Configuration 38
G7200B Configuration 42
Settings 44
The I-AS arrives packaged in multiple boxes. Below is a packing list of the
contents. Check to make sure all the necessary components are included.
The number of sample vials and sample trays will differ depending on the option
ordered.
Be careful not to drop or hit the components against anything when removing
CAUTION them from their packaging. Many parts are heavy or can be easily damaged.
Name Quantity
Main Unit 1
Tray Cover 1
Name Quantity
Needle 1
Power Cable 1
Tube Clip(s) 2
Type A 6 mL
Type B 18 mL
Type C 50 mL
Type E 6 mL, 18 mL
50 mL Vial
18 mL Vial
6 mL Vial
2 mL Vial
1.5 mL Vial
The number of boxes will differ depending on the options ordered. Make sure to
NOTE verify that all the necessary components are present.
Figure 9.
3 Loosen the two fixed arm block screws and slide the arm block with screws to
the right.
Do not remove the two fixed arm block screws. This will cause the arm block to
CAUTION be removed from the bracket.
Figure 11.
OPM075-2
Screw
Screw
Screw
Support Bracket
Screw Leg
OPM075
1 To install the tubing and wiring, place the I-AS on the bracket installed on the
ICP-MS with its power display LED facing front.
2 Connect the Peristaltic Pump inlet (sample uptake) and outlet (drain) tubes.
Connect the rinse fluid inlet and outlet tubes to the IN and OUT ports. Tighten
the connectors by hand.
Do not connect sample tubing for Outlet and drain tubing for Inlet. The internal
NOTE diameter of the drain tubing is larger than the sample tubing to prevent rinse fluid
from overflowing.
When sample tubing is connected to Outlet and drain tubing is connected to Inlet,
rinse fluid could overflow from the rinse port.
Figure 14.
Communications
cable
Figure 15.
• G7200B Connection
Connect the other end of the communications network cable to the PC
RS-232C connector.
When ISIS is used with I-AS, connect the communication network cable to
the ISIS controller (COM1).
5 Make sure the I-AS power switch is in the OFF position.
Make sure to check the following prior to connecting the power cable. Skipping
CAUTION this procedure can result in damage to the instrument.
• Before connecting the power cable, make sure it is designated for this
purpose.
• Make sure the protective grounding is secure.
• Only use the provided cables and properly connect them in their designated
locations. Do not use unauthorized cords or cables. Do not uses these
cables on instruments other than those listed in this manual.
Power cable
Figure 17.
7 Connect the other end of the power cable to a properly grounded wall outlet.
Leave the I-AS power switch in the OFF position.
In the case of leakage (liquid under the I-AS), disconnect the power cable and
WARNING contact your local Agilent office.
8 Connect the rinse fluid outlet tube to the rinse port waste bottle.
Use the tube clip to secure the tube to the bottle.
Tube
Tube clip
Bottle
Figure 18.
9 Connect the rinse fluid inlet tube to the rinse port rinse fluid bottle.
Use the tube clip to secure the tube to the bottle.
1 Slide the sample tray onto the shaft of the main unit and set into place.
Line up the sample tray mating pin with the mating grove on the shaft. After
inserting the sample tray, rotate it slightly until the pin and groove line
interlock.
Mating pin
Mating groove
Figure 19.
Sample tray
Main unit
Figure 20.
Figure 21.
Tray cover
Figure 22.
The tray cover is acrylic. Organic solvents and some sample materials may
CAUTION cause the cover to become discolored.
Use the Single Sample Position with the sample tray removed from the
NOTE instrument.
The Single Sample Position sits in the designated stand. Make sure the mouth of
the bottle is centered on the Single Sample Position stand.
The Single Sample Position is not included with the peripheral equipment. An
appropriate bottle measuring 50 mm or less in diameter and 126 mm or less in
height must be provided.
When using the Single Sample Bottle, the lowest needle position is the upper
surface of the top plate of the Single Sample Position stand. The needle will not
reach the bottom of the bottle.
Bottle
Lowest needle
position
Screw
The mouth on the Single Sample Position bottle should be lower than the rinse
port. If the mouth is higher than the rinse port, remove the two screws holding
the lower plate and re-attach the stand it in a lower position.
Software Settings
1 Launch MassHunter, and then click the Settings gadget on the dashboard.
The Settings dialog box appears.
2 On MassHunter 4.2 or later, click System, and then click at the side of
Hardware.
On MassHunter 4.1, skip this step.
3 Click Sample Introduction, and then select Use Autosampler in Properties.
Figure 25.
6 Click the Rack tab, and then select the rack that will be used from the
Autosampler Rack list box.
Figure 26.
You need MassHunter 4.4 or later to use the 28-vial sample tray (28 Samples
Rack). If the software installed is earlier than MassHunter 4.4, update to
MassHunter 4.4 or later.
The I-AS needle does not move to the correct position if the Autosampler Rack
CAUTION is not configured correctly.
7 Click the COM Port tab, and then select the COM port number for the
Autosampler.
You should normally select the highest port number from the COM port
setting. (for example: if COM 1 to COM 3 are available, usually use COM 3)
Figure 27.
8 Click the Probe Depth tab, and then set the appropriate probe depth.
Figure 28.
Be very careful when setting the values for the Rinse Port, Tray/Single, and
CAUTION standard/Rinse to prevent the needle from coming into contact with anything
else. Start with smaller values and gradually increase until an appropriate
setting is achieved. If the value is too high, the tip of the needle makes contact
and may bend or break.
Figure 29.
10 Click OK.
11 Click Close to close the Settings dialog box.
G7201A/B Configuration
1 Launch MassHunter, and then select Hardware from the task menu.
2 Click on the Mainframe illustration and select Communication... from the
popup menu.
Select Online from the Set Communication dialog box, check Use ALS and click
OK.
3 When the Autosampler (ALS) illustration appears, select the Autosampler from
Autosampler Type.
4 Select the number of samples from each available setting in the sample rack
list box.
Figure 31.
The I-AS needle does not move to the correct position if the Autosampler Rack
CAUTION is not configured correctly.
5 Click on the I-AS illustration and select Set Communication... from the popup
menu.
You should normally select the highest port number from the COM port
setting in the Set ALS Communication dialog box.
6 Click on the I-AS illustration and select Set Sampling Depth of I-AS... from the
popup menu.
Figure 32.
7 Configure each setting in the Set Sampling Depth of I-AS dialog box.
Be very careful when setting the values for the Rinse Port, Tray/Single, and
CAUTION standard/Rinse to prevent the needle from coming into contact with anything
else. Start with smaller values and gradually increase until an appropriate
setting is achieved. If the value is too high, the tip of the needle makes contact
and may bend or break.
Figure 34.
G7200B Configuration
Setup button
Figure 35.
4 Click the Setup button located to the right of the Autosampler menu.
The I-AS Configuration dialog box will be displayed.
5 Set the menu items in the I-AS Configuration dialog box.
Be very careful when setting the values for the Rinse Port, Tray/Single, and
CAUTION STD/Rinse to prevent the needle from making contact. Start with smaller
values and gradually increase until an appropriate setting is achieved. If the
value is too high, the tip of the needle will make contact and may bend or
break.
6 After all menu items have been correctly input, click the OK button.
The I-AS Configuration dialog box will close, returning to the ICP-MS
Configuration dialog box.
7 Click the Save button in the ICP-MS Configuration dialog box.
The ICP-MS Configuration dialog box will close and save the new settings.
Settings
COM port for Set the connection port for the I-AS communications
Autosampler: network cable. If the software dialog box shows that
ISIS is using the port, this port cannot be used for the
I-AS.
Sample Tray: Select the sample tray type that will be used.
Single Sample
89 Samples: Type A, D, E, F
53 Samples: Type B
18 Samples: Type C
28 Samples: Type G (Available on MassHunter 4.4 or
later)
Sampling Depth: In the positions below, set the travel distance for the
I-AS arm from its highest to lowest positions. The drive
motor moves in increments of 0.125 mm. Any smaller
settings will be automatically rounded up to this value.
Rinse Port: Set the distance the arm will lower from its highest
point in the rinse port position. Start with small values
and slowly increase (approx. 10 mm) to avoid contact.
The maximum setting is 55.00 mm.
Tray/Single: Set the distance the arm will lower from its highest
point in the sample vial or Single Sample Position. Start
with small values and slowly increase (approx. 10 mm)
to avoid contact. The maximum setting is 63.75 mm.
STD/Rinse: Set the distance the arm will lower from its highest
point in the external bottle (for standard and cleaning
fluid) position. Start with small values and slowly
increase (approx. 10 mm) to avoid contact. The
maximum setting is 63.75 mm.
Enable ESC Mode: Check this to enter escape mode. In escape mode, to
move the needle over the sample tray, the sample tray
is moved to a pre-determined position before moving
the needle. The position of the needle in escape mode
will be the last (highest) sample number on the inner
radius of the sample tray, as illustrated in the following
table. Samples in this row cannot be measured in
escape mode.
15-Vial Sample Tray (Type F) 57, 89 (At 89-Vial sample tray setting)
This section explains how to check the operation of the I-AS and install the
needle.
Checking Operations
Buttons
Home button: Moves the arm to the home position (directly over the rinse
port) and to the highest possible position without lowering
the arm.
1 button: Moves the arm to the first external bottle and lowers the
arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
2 button: Moves the arm to the second external bottle and lowers
the arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
3 button: Moves the arm to the third external bottle and lowers the
arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
Sampler Rack Displays the type of sample tray configured in the
Type: Autosampler Rack drop down menu.
Vial No. [ ]: Click on the desired sample position number icon or enter
the sample position number in the edit box to move
sample position numbers.
The configurable sample numbers are displayed in the
Select Vial dialog box. External bottles are numbered 1-3.
Sample vials are numbered sequentially with four digits
starting from 1000 (1001, 1002,...). The Single Sample
Position is numbered 1000.
Go to button: Moves the arm to the sample position configured in Vial
No. and lowers the arm to the setting in the dialog box. It
then closes this dialog box.
Move button: Moves the arm to the sample position configured in Vial
No. and maintains the high position of the arm. The dialog
box remains open.
Close button: Closes the dialog box without performing any operations.
Pressing Enter key after inputting the Vial No.[ ] in the Edit Box will have the same
NOTE effect as clicking the Go to button.
4 From the Select Vial dialog box designate an optional vial, press the Move
button and verify that the I-AS moves properly.
5 Press the Go to button on the Select Vial dialog box and verify that the arm
lowers.
Each operation can be initiated from the pull-down menu of the Autosampler
NOTE icon on the toolbar.
Buttons
Home button: Moves the arm to the home position (directly over the rinse
port) and to the highest possible position without lowering
the arm.
1 button: Moves the arm to the first external bottle and lowers the
arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
2 button: Moves the arm to the second external bottle and lowers
the arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
3 button: Moves the arm to the third external bottle and lowers the
arm to the setting in the Select Vial dialog box. It then
closes this dialog box.
Sampler Rack Displays the type of sample tray configured in the
Type: Autosampler Rack drop down menu.
Vial No. [ ]: Click on the desired sample position number icon or enter
the sample position number in the edit box to move
sample position numbers.
The configurable sample numbers are displayed in the
Select Vial dialog box. External bottles are numbered 1-3.
Sample vials are numbered sequentially with four digits
starting from 1000 (1001, 1002,...). The Single Sample
Position is numbered 1000.
Go to button: Moves the arm to the sample position configured in Vial
No. and lowers the arm to the setting in the dialog box. It
then closes this dialog box.
Move button: Moves the arm to the sample position configured in Vial
No. and maintains the high position of the arm. The dialog
box remains open.
Close button: Closes the dialog box without performing any operations.
Pressing Enter key after inputting the Vial No.[ ] in the Edit Box will have the same
NOTE effect as clicking the Go to button.
4 From the Select Vial dialog box designate an optional vial, press the Move
button and verify that the I-AS moves properly.
5 Press the Go to button on the Select Vial dialog box and verify that the arm
lowers.
Each operation can be initiated from the ALS icons on the toolbar.
NOTE
Buttons
HOME button: Moves the arm to the home position (directly over the rinse
port) and to the highest possible position without lowering
the arm.
1 button: Moves the arm to the first external bottle and lowers the
arm to the setting in the I-AS Configuration dialog box. It
then closes this dialog box.
2 button: Moves the arm to the second external bottle and lowers
the arm to the setting in the I-AS Configuration dialog box.
It then closes this dialog box.
3 button: Moves the arm to the third external bottle and lowers the
arm to the setting in the I-AS Configuration dialog box. It
then closes this dialog box.
Sample Tray Type: Displays the type of sample tray as selected in the I-AS
Configuration dialog box.
Vial No. [ ]: The number in brackets represents the sample number
that can be set. External bottles are displayed with the
numbers 1 to 3. Sample vials are displayed in four-digit
numbers, represented by 1000 and the sequential number
of the vial. (1001, 1002...). The Single Sample Position is
displayed as 1000.
GoTo button: Moves the arm to the location set for the Vial No. and
lowers it to the position in the I-AS Configuration dialog
box. This then closes the dialog box.
Move button: Moves the arm to the location set for the Vial No. and
keeps the arm at the upper position. The dialog box
remains open.
Cancel button: Closes the dialog box without performing any operations.
Help button: Displays the help screen.
Pressing Enter key after inputting the Vial No.[ ] in the Edit Box will have the same
NOTE effect as clicking the GoTo button.
3 Operate the buttons in the ALS GoTo dialog box and verify that the I-AS is
operating normally.
Installation on G7201A/B/C
1 Set the sample tray or Single Sample Position to match the setting in the
software.
2 Click on the Autosampler illustration and select Set Sampling Depth of I-AS...
G7201A/B
G7201C
Figure 44. [Set Sampling Depth of I-AS...] dialog box
3 Input the following settings in the Set Sampling Depth of I-AS dialog box.
The location of the sample vial on the sample tray is identified as a four digit
NOTE number consisting of 1000 and the number displayed on the sample tray (1001,
1002...).
Arm
Holder
Knob
Needle
Figure 47.
The needle is made of a pliable material and can bend easily. Handle the
CAUTION needle with care. Pay careful attention to the following when determining the
height of the needle tip:
• The tip of the needle should be completely submerged in the sample fluid.
• The tip of the needle should not touch the bottom of the sample vial.
• The needle should not touch anything when the arm is rotating.
If the needle is positioned too low, it can hit something when the arm is moving
and break, bend, or cause liquid to spurt or leak from the instrument.
Use the Micro Flow nebulizer needle when using a Micro Flow nebulizer.
NOTE
The lowest position of the needle should be the upper surface of the Single
Sample Position stand when using the Single Sample Position feature (See the
figure below).
If the tip of the needle is set lower than the lowest position, it will hit the tray
CAUTION cover and other objects when the arm is rotating.
Arm
Needle
The lowest
position of the
needle when using
the Single Sample
Position feature
Figure 48.
10 Display the Select Vial dialog box and click the Home button.
The arm will move to the home position. Verify that the needle is directly over
the rinse port.
Needle
(Home position)
Rinse port
Figure 49.
60.00 mm is the standard setting. Input an appropriate value so that the tip of
CAUTION the needle does not touch the bottom of the external bottle.
13 Display the Select Vial dialog box and click the 1 button.
The arm will move to the first external bottle location and the needle will
lower.
Verify that the needle properly lowers into the external bottle.
14 Connect the peristaltic pump tube.
• This step is not necessary if using a Micro Flow nebulizer as the Micro
Flow nebulizer needle is used. Continue to the next step.
• If a Micro Flow nebulizer is not used, attach the end of the tube coming
from the needle to the peristaltic pump tube.
Installation on G7200B
1 Set the sample tray or Single Sample Position to match the setting in the I-AS
Configuration dialog box.
2 Display the I-AS Configuration dialog box.
The location of the sample vial on the sample tray is identified as a four digit
NOTE number consisting of 1000 and the number displayed on the sample tray (1001,
1002...).
8 Turn the knob located on the end of the arm counterclockwise to loosen.
Arm
Holder
Knob
Needle
Figure 52.
The needle is made of a pliable material and can bend easily. Handle the
CAUTION needle with care. Pay careful attention to the following when determining the
height of the needle tip:
• The tip of the needle should be completely submerged in the sample fluid.
• The tip of the needle should not touch the bottom of the sample vial.
• The needle should not touch anything when the arm is rotating.
If the needle is positioned too low, it can hit something when the arm is moving
and break, bend, or cause liquid to spurt or leak from the instrument.
Use the Micro Flow nebulizer needle when using a Micro Flow nebulizer.
NOTE
The lowest position of the needle should be the upper surface of the Single
Sample Position stand when using the Single Sample Position feature (See the
figure below).
If the tip of the needle is set lower than the lowest position, it will hit the tray
CAUTION cover and other objects when the arm is rotating.
Arm
Needle
The lowest
position of the
needle when using
the Single Sample
Position feature
Figure 53.
10 Display the ALS GoTo dialog box and click the HOME button.
The arm will move to the home position. Verify that the needle is directly over
the rinse port.
Needle
(Home position)
Rinse port
Figure 54.
60.00 mm is the standard setting. Input an appropriate value so that the tip of
CAUTION the needle does not touch the bottom of the external bottle.
13 Display the ALS GoTo dialog box and click the 1 button.
The arm will move to the first external bottle location and the needle will
lower.
Verify that the needle properly lowers into the external bottle.
14 Connect the peristaltic pump tube.
• This step is not necessary if using a Micro Flow nebulizer as the Micro
Flow nebulizer needle is used. Continue to the next step.
• If a Micro Flow nebulizer is not used, attach the end of the tube coming
from the needle to the peristaltic pump tube.
Verify the following before beginning measurement with the Single Sample
Position feature:
• Is the sample tray type field set to “Single Sample” in the I-AS Configuration
dialog box?
Figure 55.
• Has the sample tray been removed and the Single Sample Position properly
set up?
• Is the sample vial number set at “1000”? (The vial number for Single Sample
Position is “1000”)
• Is the LED on the I-AS illuminated? (If not, make sure the power switch is ON.)
In addition to setting the sample vial location to “1000” when using the Single
Sample Position feature for measurement, make sure the other settings are the
same as when using the sample tray.
The Single Sample Position Run dialog box can be used to run single samples.
Use the following procedure to display the Single Sample Position Run dialog box.
1 From the ICP-MS Top window, select Acquire Data >> Main Panel.
The ICP-MS Acquisition window will be displayed on the screen.
2 In the ICP-MS Acquisition window, select Acquire Data >> Single Sample
Position Run.
The Single Sample Position Run dialog box will be displayed on the screen.
Figure 56.
The following procedure explains how to use the Run Options in the Single
Sample Position Run dialog box.
PreRun Uptake/ Select this option to move the needle to the single sample
Stabilization: position before measuring (collecting data). If the needle is
already in the single sample position, it will not move.
When the PreRun Uptake/Stabilization check box is selected,
the system will follow the settings in the Peristaltic Pump
Program dialog box and uptake (replace) the sample
(increase the peristaltic pump speed) and then stabilize
(return the pump to its original speed and wait a specified
amount of time) prior to measuring.
Unselect the PreRun Uptake/Stabilization check box and the
system will bypass the sample uptake and stabilization
processes, and proceed directly to measuring the sample
(collecting data).
PostRun Rinse: Select this option to rinse the needle (includes rinsing at the
rinse port, and cleaning external bottles and sample vials)
according to the Peristaltic Pump Program dialog box
settings after the measurement (data collection) is
complete.
Unselect this check box to bypass the rinse process. The
needle will remain in the bottle after measurement is
complete.
Data Analysis: Select this option for analysis by method.
Daily Maintenance
To replace the fuse for the power cable, open the fuse holder and replace.
Use a 2AT (time delay) fuse.
Confirm that the power cable has been removed from the wall outlet before
CAUTION replacing a fuse.
Fuse holder
Power switch
Power cable
connector
www.agilent.com
© Agilent Technologies, Inc. 2018
Printed in USA
October 2018
*G3160-90027*
G3160-90027 Revision A