Professional Documents
Culture Documents
Guide To The Tests, Frequency, Requirements and Paragraph For In-Use Materials
Guide To The Tests, Frequency, Requirements and Paragraph For In-Use Materials
1
Concentration levels based upon suppliers specifications.
Weekly Water-Based Penetrant 7.8.2.2 Using a refractometer, check the concentration
Water Concentration levels based upon suppliers specifications.
Monthly Non-Water-Based 7.8.2.2 For water-washable (Method A) penetrants, the
Penetrant Water Content water content should never exceed 5%.
of Method A
Monthly Penetrant Removability 7.8.4.2 Using a test piece specified in AMS-2644
(4.4.11.1.1) and following the normal wash
parameters, compare the removability of the in-use
penetrant with new un-unused penetrant. If the in-
use material has a noticeable inferior performance
the penetrant should be replaced.
Monthly Emulsifier Removability 7.8.4.4 Using a test piece specified in AMS-2644
(4.4.11.1.1) and following the normal wash
parameters, compare the removability of the in-use
emulsifier with unused penetrant with un-unused
emulsifier also with unused penetrant. If the in-use
material has a noticeable inferior performance the
emulsifier should be modified or replaced.
Monthly Lipophilic Emulsifier 7.8.2.3 Water content should not exceed 5%, if so the
Water Content material should be modified or replaced
Quarterly Fluorescence Brightness 7.8.4.1 The in-use penetrant must be at least 90% of the
unused penetrant. If lower the in-use penetrant
must be modified to reach the brightness level of
the unused material or replaced.
Quarterly Drying Oven Calibration 7.8.5.5 Drying ovens should not exceed 160oF (71oC). The
controls should maintain the oven temperature at ±
15oF of the temperature the oven is set. The
temperature indicator should be accurate to ± 10oF.
Semiannually Light Meter Calibration 7.8.5.2 Both black and visible light meters should be
calibrated in accordance to MIL-STD-45662