NATIONAL CONTROL LABORATORY FOR BIOLOGICALS

ELECTRIC LOAD ESTIMATES


CELL CULTURE AND MICROBIOLOGY BLOCK

Unit Electric Total


Sr. # Description Quantity
Load work load
Air conditioner / HVAC 08/02 KW 2.0 16.0/ 20
1.
for HVAC
2. Microbiological Cabinets 03 KW 1.875 5.625
3. Hot Oven 03 KW 3.5 10.5
4. Incubators 03 KW 1.2 3.6
5. ELISA Systems 02 KW 0.8 1.6
6. PCR System 01 KW 1.3 1.3
7. Gel Documentation System 01 KW 1.6 1.6
8. Deep Freezer 02 KW 0.1 0.2
9. Centrifuge Machine 03 KW 0.4 1.2
10. Inverted Microscope 01 KW 0.25 0.25
11. Automatic Cell Counter 01 KW 0.1 0.1
12. Vortex Mixer 04 KW 0.3 1.2
13. Refrigerator 01 KW 0.4 0.4
14. pH Meter 02 KW 0.1 0.2
15. Karl Fisher 01 KW 0.3 0.3
16. Refrigerated Centrifuge 01 KW 3.0 3.0
17. Ice Machine 01 KW 3.5 3.5
18. Tube Lights 100 KW 0.04 4.0
19. Computers 12 KW 1.2 12
Total Number of Equipment 0 KW 21.97

Kilowatts to KVA Conversion

Total KW= 71
Power factor = 0.8
Total KVA= 88.75
Requirement for Generator= 90 KVA for one block of Laboratory