SPECTROMAXx
2 Safety................................................................................................................ 12
2.1 User information................................................................................................................ 13
2.2 Intended use ..................................................................................................................... 14
2.2.1 Samples ............................................................................................................... 14
2.2.2 Preparing the Sample .......................................................................................... 14
2.2.3 Sample material................................................................................................... 15
2.2.4 Sample weight ..................................................................................................... 16
2.2.5 Spark stand ......................................................................................................... 16
2.2.6 Sample clamp ...................................................................................................... 16
2.2.7 Gas supply ........................................................................................................... 17
2.3 Prohibited operating conditions ........................................................................................ 17
2.4 Residual risks.................................................................................................................... 18
3 Technical data................................................................................................... 20
4 Transportation/Setting-up ................................................................................. 22
4.1 Space requirements .......................................................................................................... 23
4.2 Storage ............................................................................................................................. 23
4.3 Gas supply ........................................................................................................................ 24
6 Operation .......................................................................................................... 26
6.1 Overview of controls ......................................................................................................... 26
6.2 Switching on the instrument .............................................................................................. 27
6.2.1 Argon saving module ........................................................................................... 27
6.3 Switching off the instrument .............................................................................................. 28
6.4 Measuring procedure ........................................................................................................ 29
6.4.1 Preparing the measuring procedure .................................................................... 30
• iCALising the measuring device (iCAL) ............................................................... 31
• Changing the base (multi-base instruments only) ............................................... 31
6.4.2 Start the measurement ........................................................................................ 32
SPECTROMAXx — 21.02.2020 —3
7 Maintenance ..................................................................................................... 33
7.1 Maintenance Reminder ..................................................................................................... 33
7.2 Maintenance schedule ...................................................................................................... 34
7.3 Cleaning the UV optics light inlet window ......................................................................... 35
7.4 Cleaning the UV optics light inlet window ......................................................................... 37
7.5 Cleaning the spark stand .................................................................................................. 40
7.6 Exhaust filter system - Change filter insert ....................................................................... 45
7.7 Exhaust filter system - Inspecting the gas-washing bottle ................................................ 47
7.8 Replacing the filter mats.................................................................................................... 49
7.8.1 Housing of air filter ............................................................................................... 49
7.8.2 Optic air filter ........................................................................................................ 50
7.9 Replacing the ICALization sample .................................................................................... 51
11 De-commissioning ............................................................................................ 57
—4 SPECTROMAXx — 21.02.2020
Description of instrument
1 Description of instrument
1.1 Description
Instrument: SPECTROMAXx
Model: LMX08
Type/Part No.: 76004144
1.2 Function
Sample material is evaporated by spark discharge inside the unit. In this process,
the released atoms and ions are excited and emit light. This light is directed into
the optical systems and measured using sensors.
Calibration data are already stored in the unit's memory. The measured values are
compared with these data. The measured values are converted into
concentrations and then displayed on the screen.
SPECTROMAXx — 21.02.2020 —5
Description of instrument
1.3 Overview
3 PC 6 Air filter
—6 SPECTROMAXx — 21.02.2020
Description of instrument
Fig. 2
SPECTROMAXx — 21.02.2020 —7
Description of instrument
Fig. 3
3 On/Off switch
spark excitation 7 Access to exhaust filter system
(plasma generator) (gas-washing bottle and filter
bottle)
—8 SPECTROMAXx — 21.02.2020
Description of instrument
Fig. 4
SPECTROMAXx — 21.02.2020 —9
Description of instrument
Fig. 5
1 Replaceable window (in the light 6 Holder for fiber-optics to air optics
channel to the UV optics)
— 10 SPECTROMAXx — 21.02.2020
Description of instrument
Fig. 6
SPECTROMAXx — 21.02.2020 — 11
Safety
2 Safety
Every user of the instrument must have read and understood these operating
instructions. This applies in particular to the safety instructions.
Notice!
This sign warns of activities or conditions that may adversely affect the function of
the instrument.
Information
Displays information that is important for the correct and safe use of the
instrument.
— 12 SPECTROMAXx — 21.02.2020
Safety
SPECTROMAXx — 21.02.2020 — 13
Safety
2.2.1 Samples
Only dry metal samples must be measured.
— 14 SPECTROMAXx — 21.02.2020
Safety
SPECTROMAXx — 21.02.2020 — 15
Safety
— 16 SPECTROMAXx — 21.02.2020
Safety
SPECTROMAXx — 21.02.2020 — 17
Safety
— 18 SPECTROMAXx — 21.02.2020
Safety
Information
A high-voltage pulse ignites the plasma. Subsequently, the burning voltage is
applied to the electrode.
SPECTROMAXx — 21.02.2020 — 19
Technical data
3 Technical data
Description Value
— 20 SPECTROMAXx — 21.02.2020
Technical data
Description Value
Argon inlet pressure on the instrument 4.8 to 5.0 bar (69.6 to 72.5 psi)
SPECTROMAXx — 21.02.2020 — 21
Transportation/Setting-up
4 Transportation/Setting-up
Notice! Damage to instrument!
Improper transport may cause damage to the instrument or to other property within
the surroundings.
Only specifically trained personnel shall be permitted to transport the
instrument in its transport packaging.
If the usage site must be changed, please contact our Service Center.
Consider the dimensions and the weight of the instrument.
Ensure that the transportation route is free of obstacles and bottlenecks
are avoided.
Ensure that the static properties of the transport route and the installation
location correspond to the weight of the unit (approx. 78 kg for the
benchtop or 132 kg for the floor instrument).
Only service personnel authorized by the manufacturer are permitted to
unpack, install, set up, and connect the instrument.
— 22 SPECTROMAXx — 21.02.2020
Transportation/Setting-up
4.2 Storage
Storage conditions:
The instrument must be stored in a dry place
Room temperature -15°C – +45°C (5°F – 113°F)
SPECTROMAXx — 21.02.2020 — 23
Transportation/Setting-up
— 24 SPECTROMAXx — 21.02.2020
Initial start-up
5 Initial start-up
SPECTROMAXx — 21.02.2020 — 25
Operation
6 Operation
Fig. 7
— 26 SPECTROMAXx — 21.02.2020
Operation
Information
To operate the instrument in compliance with the specification, it should be
stabilized for at least 2 hours after it has been switched on.
This may take longer because this procedure is dependent on the temperature
during instrument transport and/or storage.
Before all other activities, ensure that the operating temperature of the UV optics
and air optics (33 °C; 91.5 °F) has been reached (see chapter 9.2).
To switch on the instrument, proceed as follows:
1. Connect the instrument to the power supply.
2. In order to switch on the plasma generator, press the following buttons:
SPECTROMAXx — 21.02.2020 — 27
Operation
Information
If the instrument was completely switched off, restarting may take some time as
the optics need to reach the right temperature and the system must be flushed with
argon.
It is recommended to leave the instrument connected to the power supply and to
leave the argon supply line open all the time.
To switch off the instrument, proceed as follows:
— 28 SPECTROMAXx — 21.02.2020
Operation
Information
A flawed initial start-up, operation or sample preparation can lead to measurement
errors.
Therefore, ensure all necessary requirements for a correct measurement are
met.
SPECTROMAXx — 21.02.2020 — 29
Operation
Information
An optimum measurement requires a good argon atmosphere inside the spark
stand. Therefore, the opening of the spark stand must always be covered. This will
make it more difficult for ambient air to enter the spark stand. A flat sample is
highly recommended.
Oxygen and humid air inside the spark stand may result in poor burn spots (see
Chapter 9 „Troubleshooting“).
To remove oxygen from the spark stand, use the "Flush" button on the computer to
invoke the flush function.
Before starting the measurement, proceed as follows:
1. Ensure that
the instrument is ICALized, see page 31,
the correct and base-relevant parts have been installed, see page 31.
2. Check the spark stand and the sample for any moisture.
It is important that no moisture is visible on the outside.
If necessary, dry the sample.
3. If working with toxic samples, ensure to verify the safety precautions.
The measurement is now prepared.
— 30 SPECTROMAXx — 21.02.2020
Operation
SPECTROMAXx — 21.02.2020 — 31
Operation
Information
Use the electrode brush in order clean the electrode tip after each measurement.
— 32 SPECTROMAXx — 21.02.2020
Maintenance
7 Maintenance
Information
In order to ensure the availability and operational safety of the equipment, we
recommend having the system checked for correct operation and performance at
least once per year by specialized technicians. To find contact addresses of
service organizations authorized by SPECTRO go to www.spectro.com
SPECTROMAXx — 21.02.2020 — 33
Maintenance
1. Check the argon supply. The min. pressure in the bottle is 10 bar.
3. Check the water level in the gas- Top up/replace, see page 47.
washing bottle.
Clean the electrode. Use the small wire brush to clean the
electrode after each spark test.
Clean the air optics light inlet window. See page 37.
As required
— 34 SPECTROMAXx — 21.02.2020
Maintenance
Fig. 8
4. Unscrew the fixing ring (3) of the
light entry window (2) using the
auxiliary tool (1).
Fig. 9
Fig. 10
SPECTROMAXx — 21.02.2020 — 35
Maintenance
Fig. 11
7. Refit the light inlet window.
Cleaning of the UV optics light inlet window is now complete.
— 36 SPECTROMAXx — 21.02.2020
Maintenance
Fig. 12
3. Remove the spacer (1) from the
spark stand.
SPECTROMAXx — 21.02.2020 — 37
Maintenance
Fig. 13
6. Loosen the union nut (1).
Fig. 14
7. Remove the fiber optic holder
(1) from the spark stand.
8. Loosen the window holder (2)
from the spark stand.
The quartz window is
located between the holder
and the O-ring.
Direct contact with the optical
surfaces must be avoided!
Fig. 15
9. Clean the light inlet window (1),
using optical paper or a lint-free
soft cloth.
If the light inlet window is
too dirty, it must be
replaced.
Fig. 16
— 38 SPECTROMAXx — 21.02.2020
Maintenance
Fig. 17
11. If necessary, use a grease-free
cloth to clean the O-ring (1).
Fig. 18
12. Reassemble components in reverse order.
Cleaning of the air optics light inlet window is now complete.
SPECTROMAXx — 21.02.2020 — 39
Maintenance
— 40 SPECTROMAXx — 21.02.2020
Maintenance
The sparking process soils the spark stand with a black precipitate (metal
condensate). This metal condensate may create a conductive connection between
the electrode and the metal structure of the spark stand wall.
Fig. 19
Fig. 20
SPECTROMAXx — 21.02.2020 — 41
Maintenance
Fig. 21
6. Loosen the fixing screw of the
electrode (1), using the Allen
key (AF 2) until the electrode
can be removed.
When loosening the
electrode, hold firmly on to
it. After loosening the
screw, a spring presses the
screw upward. Loosen the
screw only until the
electrode can be removed
easily.
7. Remove and clean the
Fig. 22
electrode.
If the electrode is worn, it
must be replaced.
Fig. 23
— 42 SPECTROMAXx — 21.02.2020
Maintenance
Fig. 24
Fig. 25
12. Install the spark stand plate (1).
13. Insert the electrode (2) into the
spark stand opening.
Fig. 26
SPECTROMAXx — 21.02.2020 — 43
Maintenance
Fig. 27
Fig. 28
— 44 SPECTROMAXx — 21.02.2020
Maintenance
1. Floor instrument:
Open the door on the right side of
the instrument. 1
Benchtop instrument:
Turn the opening of the filter
cartridge container to the front. 2
Fig. 29
SPECTROMAXx — 21.02.2020 — 45
Maintenance
— 46 SPECTROMAXx — 21.02.2020
Maintenance
1. Floor instrument:
Open the right door of the instrument.
Benchtop instrument:
Turn the opening of the filter cartridge container to the front.
Fig. 30
3. For refilling, grab the filter cartridge (2) with both hands and unscrew it.
4. Refill water or change it as required.
SPECTROMAXx — 21.02.2020 — 47
Maintenance
Fig. 31
— 48 SPECTROMAXx — 21.02.2020
Maintenance
Fig. 32
3. Remove the filter mat (1). 1
Fig. 33
4. Install the new filter mat.
5. Reinstall the cover.
The housing of the air filter is replaced.
SPECTROMAXx — 21.02.2020 — 49
Maintenance
Fig. 34
3. Remove the filter mat (1).
Fig. 35
4. Install the new filter mat.
5. Tighten the screws.
6. Install the cover.
The optic air filter is cleaned / replaced.
— 50 SPECTROMAXx — 21.02.2020
Maintenance
Information
Do not discard the old ICALization sample before the intensities have been
matched. Matching is not possible without using the old samples.
SPECTROMAXx — 21.02.2020 — 51
Status messages
8 Status messages
The status information is displayed in the footer of the software. These indications
provide important information about the status of the instrument.
The following information is available:
— 52 SPECTROMAXx — 21.02.2020
Troubleshooting
9 Troubleshooting
Malfunction Cause Remedy
The instrument cannot be The power plug is not Check if the power plug was
switched on. plugged in properly. properly plugged in.
The burn spot is The argon outlet (exhaust) Clean the spark stand including
incorrect. is clogged with the argon outlet.
condensate.
SPECTROMAXx — 21.02.2020 — 53
Troubleshooting
Low intensities in UV Light inlet window soiled. Clean light inlet window if
optics. required, see page 37.
Low intensities in the air Light inlet window soiled. Clean light inlet window if
optics. required, see page 37.
— 54 SPECTROMAXx — 21.02.2020
Troubleshooting
When starting the “Spark Communication Exit the “Spark Analyzer Pro
Analyzer Pro MAXx” MAXx” software.
software, the instrument Disconnected the
switches into DEMO mode. spectrometer from the power
for a short time. After a
waiting period of approx. 60
sec., restart the “Spark
Analyzer Pro MAXx”
software.
Safety check. The safety circuit has not Acknowledge the message.
been activated within the Move the clamp up for
last 5 minutes. approx. 1 second.
An error has occurred whilst Plasma generator is not Switch on the Plasma
attempting to access the switched on. generator.
spectrometer (Plasma Confirm by pressing “OK”.
generator is switched off,
measurement could not be Insufficient argon supply. Check the argon supply:
started). Pre-pressure (4.8 – 5.0 bar
Check the condition of the (69.6 – 72.5 psi)) and
spectrometer. available quantity.
An error has occurred during Safety circuit not closed. The function circuit "Clamp –
spectrometer access (clamp Sample – Spark stand" is
up, measurement could not not closed. Ensure electric
be started). Check the contact is provided.
condition of the
spectrometer.
SPECTROMAXx — 21.02.2020 — 55
Troubleshooting
Fig. 36
When the instrument is ready for operation, the green LED symbol is lit. If an error
occurs, the LED symbol changes to red. For a detailed description, refer to the
online help of the instrument.
— 56 SPECTROMAXx — 21.02.2020
Spare parts and consumables
11 De-commissioning
Only organizations or persons authorized by the instrument manufacturer shall be
permitted to de-commission the instrument. Please contact your Service Center.
12 Manufacturer address
SPECTRO Analytical Instruments GmbH
Boschstr. 10
D-47533 Kleve
Germany
Tel.: +49.2821.8 92-0
Fax: +49.2821.8 92-22 00
http://www.spectro.com
SPECTROMAXx — 21.02.2020 — 57