Professional Documents
Culture Documents
MAintenance Manual of Spectro Xepos 05
MAintenance Manual of Spectro Xepos 05
MAintenance Manual of Spectro Xepos 05
Möglicherweise wurde die Datei v erschoben, umbenannt oder gelöscht. Stellen Sie sicher, dass die Verk nüpfung auf die
k orrek te Datei und den k orrek ten Speicherort zeigt.
SPECTRO XEPOS
2 Safety................................................................................................................ 12
2.1 Symbols used ................................................................................................................... 12
2.2 User information................................................................................................................ 13
2.3 Intended use ..................................................................................................................... 13
2.4 Prohibited operating conditions ........................................................................................ 14
2.5 Residual risks.................................................................................................................... 14
2.6 Radiation protection advice .............................................................................................. 16
2.6.1 Other countries .................................................................................................... 16
1 Description of instrument
1.1 Description
Designation: SPECTRO XEPOS
Model: XEP05
Type: 76004852
1.2 Function
X-rays excite the atoms of the sample, causing them to emit radiation. This
radiation is measured by a semiconductor detector.
To achieve a higher sensitivity, the exciting radiation can be optimized by using X-
ray optical elements or filters.
The instrument comes with a prepared analysis method. Data are already stored
in the instrument's memory. The measured values are compared with these data.
After the measurement is complete, the results of a sample relating to an unknown
composition are displayed on the screen.
The sample material may be solid, liquid or in powder form, see Chapter "How to
measure different samples".
1.3 Overview
1.3.1 Configuration
The instrument is available in various configurations. Depending on the
application/requirement, the instrument is equipped with the following X-ray optics:
Direct excitation
Direct excitation and single polarization channel
Direct excitation and dual polarization channel
3 3
No. Component
1 2
Fig. 2
No. View
When pressing slightly against the handle, the sample chamber lid can be pushed
back and locked into place. In order to open the cover, slightly press against the
handle and pull on the sample chamber lid.
3 4 5 6 7
Fig. 3: Rear view
Note
Set up the instrument to ensure the power isolators (switches and power cable)
can be easily reached.
3
5
4
Fig. 4
No. Component
4 Tray handle
5 Tray screw
3
2 4
5
6
Fig. 5
No. Component
2 Discharge reservoir
2 Safety
Every user of the instrument must have understood these operating instructions.
This applies in particular to the safety regulations.
Notice!
Warns of activities or conditions that may adversely affect the safety or function of
the instrument.
Note
Displays information that is important for the correct and safe use of the instrument
Note
If the instrument is in its normal operating mode (standby or measuring mode), it
produces ionizing radiation (X-rays).
The safety circuits in the instrument are designed to ensure the sample chamber
lid is always locked during the measurement.
For your safety, the safety circuits are checked automatically. Therefore, when
starting the software, a message is displayed requesting the opening and closing
of the cover (referred to as Start-up test, see Chapter 5.3).
Automatic safety circuit checks must be carried out at least once a month. If the
device has not been switched off and switched on again at the main switch for one
month, the software prompts you to do so (see also chapter 5.3.4). Three days
before the deadline, a first reminder appears in the status display of the software.
3 Technical data
Description Value
Fuse 16 A (slow-blow)
Vacuum (option)
4 Transportation/Setting-up
Only instructed personnel may unpack, install, and set up the instrument.
4.1 Storage
Storage conditions:
The instrument must be stored in a dry place
Room temperature -10 to +50°C (14°F to 122°F)
Avoid storing the instrument at very low temperatures.
Fig. 6
3. Install the oil mist filter (1).
To do this, use the clamp to
fasten the oil mist filter.
Fig. 7
4. Use the vacuum hose to
connect the vacuum pump and
the instrument.
To do this, fasten the
vacuum hose with the
clamps at both connecting
points.
Fig. 8
5. Use the power cable to connect
the vacuum pump and the
instrument.
Fig. 9
In case of a power failure, the UPS (Uninterrupted Power Supply) sends a signal
to the instrument. The instrument is then shut down in a controlled manner. In the
case of extended power failures, the computer is shut down after a given amount
of time.
1 2 3 4 5 6
Fig. 10
5 Operation
5.1 Displays
2
3
Fig. 11
1 Status line Display for information from the program and the
instrument
Note
Ensure that the ambient conditions specified in Chapter "Technical data" are
maintained. If the temperature and/or relative humidity does not comply with the
specifications, the software of the instrument will initiate a warning.
If the connection is successful, the software will first check the condition of the
safety circuits. The software checks the condition of the sample chamber lid
(open/closed). If the sample chamber lid is closed, you will be prompted to open it:
If the sample chamber was open when the instrument was switched on, this step
will be omitted and the software will skip this procedure.
Subsequently, the software will prompt you to close the sample chamber lid:
5.3.2 Checking the safety circuit on the filter changer ("Beam Stop")
Since an electric motor is used to move the assembly, the safety switches on the
Beam Stop are checked autonomously by the instrument's software. During this
procedure, the software displays an appropriate message:
After the safety test was completed successfully, the stepping motors that are
integrated into the spectrometer, are initialized.
Alternatively, if the start-up test fails based on a malfunction, the user will be
informed accordingly:
Note
If the start-up test fails, the safety circuits are faulty.
The high-voltage supply of the X-ray tube cannot be switched on.
In this case, please contact your Service Center.
If the startup test and the initialization were successful, the high-voltage supply of
the X-ray tube can be activated.
The X-ray tube should be switched to standby mode.
On the screen dialog click on "Enter HV-Standby".
3. If at that point the X-ray tube is not switched on, subsequently, the high-
voltage supply will be shut down gradually and automatically.
5. Ensure that only such samples are present on the sample tray that are
measured under the same conditions (air, helium, vacuum).
The measurement is now prepared.
Always measure the following samples with gas flushing or in a diluted condition:
gassing samples,
highly volatile samples,
flammable samples,
samples with a Cl content of more than 2%.
After completing the measurement of liquid and powdery samples, the
spectrometer will sound a signal. Furthermore, a message on the screen will
prompt you to remove the sample trays with the cuvettes from the sample
chamber.
Note
Glass pallets must always be measured on their "own" tray. Prior to the
measurement, all dust must be cleaned off the tray. A contaminated sample tray
will falsify the analysis.
Fig. 12
4. If possible, use the single
1
sample holder (1) that is
included with the delivery.
Fig. 13
Three single sample holders are included with each delivery. Additional holders
can be ordered from our SPECTRO Shop. The part number can be found on page
55.
The installation of the sample tray and the sample tray support takes place in
reverse order. During the installation of the sample tray support, tighten the tray
screw until the sample tray support can be rotated.
Note
Measure a reference sample at regular intervals. This will allow you to identify
changes of the measurement results in good time.
Fig. 14
2. Carefully remove the sample
tray.
Fig. 15
3. Install the new sample tray.
Ensure the red dot of the
sample tray and the red dot
of the tray support are
aligned (2).
4. Close the locking device of the
sample tray by rotating it
clockwise and simultaneously 2
lifting the device slightly.
Fig. 16
The maintenance intervals specified in the following table are guidelines which
should be observed as a minimum requirement. The maintenance intervals may
vary, depending on the use of the instrument.
To ensure the flawless function of the instrument, proceed with the following
maintenance tasks:
Clean the sample chamber. Use a soft cloth and, if possible, a mild
cleaning agent or pure alcohol, e.g.,
isopropanol, see Chapter 6.5.
Shutting down and rebooting the computer. See Chapter 5.2 and 5.3.
Shutting down and restarting the See note on page 15 and chapter 5.3.4
spectrometer and repeating the start-up test
(testing the safety circuits).
Clean the sample rotation drive (do not Use a soft brush.
lubricate!).
Vacuum system maintenance (optional). Changing the vacuum oil and cleaning the
oil mist filter. See Chapter 4.3.
2 1 2
Fig. 18
2. Remove the new tray pads and
stick them carefully and
accurately onto the marked
surfaces.
3. Subsequently, push the tray pads
downward against the sample
chamber floor.
Fig. 19
Now, the tray pads are replaced.
6.8 Insert the film for the "liquid protect" sample cup
Parts required for inserting the film:
XRF film (2)
Sample cup holder (4)
Clamping ring (3)
Tool for removing the clamping
ring (1)
Fig. 20
To insert the film, proceed as follows:
1. Place the parts of the sample cup
holder on a clean surface.
2. Place the film on the narrow end
of the sample cup holder.
3. Place the clamping ring on the
film.
Fig. 21
4. Push the clamping ring down
until the sample cup holder and
the clamping ring are on the
same level.
5. Ensure the film is not crinkled.
Fig. 22
Fig. 24
The filter mat is included with each delivery. Additional mats can be ordered from
our SPECTRO Shop. The part number can be found on page 55.
To replace the filter mat, proceed as follows:
1. Loosen the cover on the
magnetic locks (1) by pulling 1 1
downward.
2. Remove the old filter mat.
6.11 Troubleshooting
The spectrometer cannot The power plug is not Insert the power plug properly.
be switched on. inserted properly.
The computer cannot be The power plug is not Insert the power plug properly.
switched on. inserted properly.
The monitor cannot be The power plug is not Check if the power plug was
switched on. inserted properly. properly inserted.
The keyboard does not The keyboard is not Check the keyboard connection
react. properly connected. to the computer.
The mouse does not The mouse is not properly Check the mouse connection to
convert any entries. connected. the computer.
Incorrect start-up test The sample chamber lid is Check if the sample chamber lid
The X-ray tube not closed. is properly closed.
cannot be switched
on. Faulty safety switches or Please contact your Service
defective X-ray Center.
tube/generator.
When the measurement Faulty safety switches or Please contact your Service
starts, the high-voltage is defective X-ray tube. Center.
shut down again
No spectrum is displayed There is no sample at the Check the sample position on the
when the measurement is measurement position. sample tray or place the sample
started. onto the measurement position.
During the MCA There is no proper sample Place the MCA sample onto the
calibration, error at the measurement correct position of the sample
messages appear. position or no sample at tray.
all.
The analysis results There is no proper sample Place a reference sample at the
of the reference at the measurement measurement position.
sample show great position or no sample at
deviations. all.
The helium supply was not Connect the helium supply and start
connected when the the procedure again.
helium flushing was used.
The sample chamber The measurement Wait several seconds until the end of
lid does not open. procedure is not yet the measurement the sample
complete or there is still chamber lit is unlocked or the
residual vacuum in the sample chamber is completely
sample chamber. aerated.
The dust protection The shutter is deactivated. Click the Diagnostic Tools of the
shutter / beam "XRF Analyzer Pro" software. Select
stabilization vacuum the menu item "Shutter" and click the
shutter does not close option "Re-initialize shutter".
the excitation
chamber after The drive mechanism of Please contact your Service Center.
measurement. the shutter is defective.
8 De-commissioning
Only organizations or persons authorized by the instrument manufacturer shall be
permitted to de-commission the instrument. Please contact your Service Center.
9 Manufacturer address
SPECTRO Analytical Instruments GmbH
Boschstr. 10
D-47533 Kleve
Germany
Tel.: +49.2821.8 92-0
Fax: +49.2821.8 92-22 00
http://www.spectro.com