Professional Documents
Culture Documents
AR-6300, AR-6400 Hematology Analyzer User Manual
AR-6300, AR-6400 Hematology Analyzer User Manual
AR- 6300
AR- 6400
Operation Manual
For In Vitro Diagnostic Use
About this manual
Statement
This manual will help you understand the operation and maintenance
of the product better. It is reminded that the product shall be used
strictly complying with this manual. User’s operation failing to comply
with this manual may result in malfunction or accident for which the
manufacturer cannot be held liable.
The user shall understand that nothing in this manual grants him,
expressly or implicitly, any right or license to use any of the intellectual
properties of the manufacturer.
The manufacturer reserves the right to modify and update this manual
without prior notice.
Warning
Make sure to use the instrument under the provisions of this Operation
Manual. Otherwise, the instrument may not work normally, the test
result is unreliable and it may damage the instrument parts and
endanger people’s safety.
User should read the Safety Notice and Operation Notice carefully.
There are several safety warning signs in this manual to help operators
avoid hurting themselves or making the instrument damaged, resulting
in incorrect measuring results:
Safety Notice
Avoid hand touch the sample, reagent and waste directly. When
operate the instrument, be sure to wear gloves to avoid being infected.
Use the reagent carefully to avoid hand and clothes touch directly.
Once hand or clothes touch, flush the touched area with soap and
plenty of water immediately.
If the reagent enters into the eyes incautiously, flush with plenty of
water at once, and inquire a doctor for further treatment.
Operation Notice
Purpose
Table of Contents
1.5.2 Menu................................................................................... 14
i
Table of Contents
ii
Table of Contents
iii
Table of Contents
iv
Table of Contents
v
Table of Contents
8.10 Maintenance............................................................................ 94
9.4.1 Clean the dust of the instrument every month ................. 100
9.6 Maintenance before Transport or for the Instrument that will not
be used for a long time.................................................................. 101
vi
Table of Contents
vii
Table of Contents
viii
Instrument Introduction
1.1 Composition
Auto Hematology Analyzer consists of two parts: the host and the
accessories.
1.1.1 Host
Control the process of sample collection, dilution and analysis,
including the following units:
1.1.2 Accessory
Including the following accessories:
(1) Mouse
1
Instrument Introduction
1.2 Purpose
This analyzer is used for detecting the parameters of RBC, WBC, HGB
and differential counting. The following 20 parameters and 3
histograms of blood can be obtained from the analysis.
Hematocrit HCT %
2
Instrument Introduction
Plateletcrit PCT %
WBC
White Blood Cell Histogram
Histogram
RBC
Red Blood Cell Histogram
Histogram
3
Instrument Introduction
1.3 Specifications
Standard Classification of the Instrument
According to anti-electrical shock hazard: Grade II, Pollution 2
According to the defence for deleterious liquid: General Device
(Closing device of non-defensive liquid)
According to the recommended disinfection/sterilization method of the
manufactory: Disinfectant device recommended by manufactory
According to safety degree in condition that using gas mixture of
flammable anesthetic gas and air or gas mixture of oxygen and nitrous
oxide: Do not use the equipment in condition that using gas mixture of
flammable anesthetic gas and air or gas mixture of oxygen and nitrous
oxide.
According to the working status: Continuous running equipment
Principles of Measurement
4
Instrument Introduction
Parameters of Measurement
Basic Parameters:
Table 1-3
Derived from Histograms:
Mean Corpuscular
MCV fL
(erythrocyte) Volume
5
Instrument Introduction
Table 1-4
Calculated Parameters:
Hematocrit HCT %
Plateletcrit PCT %
Table 1-5
Sampling Features:
Sample Volume:
6
Instrument Introduction
WBC/HGB 1:285
RBC/PLT 1:21667
WBC/HGB 1:500
RBC/PLT 1:43333
Working mode:
Display:
Interface:
7
Instrument Introduction
Recorder:
Repeatability:
Parameter Repeatability
WBC ≤ 2.0%
RBC ≤ 1.9%
HGB ≤ 1.9%
MCV ≤ 0.4%
PLT ≤ 4.0%
Table 1-6
Linear Range:
8
Instrument Introduction
0~0.99×1012/L (±0.05×1012)/L
RBC
1.0~9.99×1012/L ±5%
0~99g/L ±2.0g/L
HGB
100~300g/L ±2%
0~99×109/L (±8.0×109)/L
PLT
100~999×109/L ±10%
Table 1-7
Parameter
Parameter Parameter Parameter Range
Range
WBC 0.0-99.9x103/μL NEUT# 0-99.9x103/μL
9
Instrument Introduction
Parameter
Parameter Parameter Parameter Range
Range
NEUT% 0-100% MPV 0.0-30.0fL
Table 1-8
Operation Condition
Temperature: 18℃-35℃
Humidity: 10%RH-90%RH
Temperature: -10℃-55℃
Humidity: 10%RH-90%RH
Electric Specifications
Power Supply:
10
Instrument Introduction
Fuse: F6.3AL250V
1.4 Structure
1.4.1 Front Panel
11
Instrument Introduction
2
3
12
Instrument Introduction
1.5 Operation
Designed according to human engineering and transferring information
to users by software windows and mouse, both are convenient for your
operating.
13
Instrument Introduction
1 2
4
1.5.2 Menu
You can operate the instrument by button. In the ‘Measure’ menu,
14
Instrument Introduction
press “Menu” button, the main menu window pops up, as shown in
Figure 1-3.
Figure 1-3
(2) Auto: Enter the auto calibration window to calibrate the instrument
15
Instrument Introduction
automatically.
QC: Enter the quality control window to carry out quality control
operation.
(2) Run: Enter the QC run window to carry out the selected QC files.
(3) L-J Graph: Enter the QC graphs window to browse the selected
QC graphs.
(4) List: Enter the QC table window to browse the selected QC tables.
(1) Language: Enter the language setting window to set the language
of the instrument.
(2) Time: Enter the time setting window to set the time of the
instrument.
(3) Print: Enter the printout window to set the functions such as print
or report.
(4) Control: Enter the config window to set the functions such as print,
standby, clean, and display etc.
16
Instrument Introduction
(7) Doctor Info: Enter the operator window to set the department,
abbreviation, department and operator.
(3) Cleaning: Run the normal rinse process to rinse the measurement
unit.
(4) Prime Chamber Outside: Inject the solutions into the counter
chambers.
(6) Drain All: Run the Drain All program to rinse and empty the tubing.
(7) Prime Diluent: Inject the diluent into the appointed tubing.
17
Instrument Introduction
(11) Mechanic: Enter the Mechanic window to verify the functions of the
instrument.
18
Instrument Introduction
Figure 1-4
(2) When the diluent passes through the aperture sensor, electrodes
19
Instrument Introduction
(3) When cells pass through the aperture, the resistance between the
electrodes increases as the cell volume increases, as shown in
Figure 1-4. According to the Ohm’s law: U=RI (U=Voltage,
I=Current, R=Resistance). If I is a constant, U increases as the cell
volume increases.
(4) Passing through the magnification circuit, the voltage signal will be
magnified and the noise will be filtered, then you will gain the
analytical results, as shown in Figure 1-5.
(5) One counting chamber and the detection circuit count the WBC.
Another counting chamber and the detection circuit count the RBC
and PLT. The microprocessor of the instrument calculates and
analyzes the cells (WBC, RBC, PLT), and then gives out the
histograms.
20
Instrument Introduction
Diluent
Voltage Analytical
circuit
Electrode
Constant
current source
Time
Pulse
Figure 1-5
21
Instrument Introduction
WBC: 20-1000fL
RBC: 82-98fL
PLT: 2-35fL
LYM: 35-90fL
MID: 90-160fL
NEUT: 160-450fL
22
Instrument Introduction
PCT.
23
Instrument Introduction
24
Installation
Chapter 2 Installation
2.1 Package
If packing damage is discovered after receiving the instrument, or the
instrument is badly damaged, contact with the freight agent
immediately to file a claim according to the damage level. At the same
time contact with your supplier to make sure the packing is complete.
Then unpack and install the instrument according to the following
steps.
2.2 Unpack
Take out the instrument and accessories from the packing box carefully.
Preserve the packing material for future transportation or storage.
If there is any problem, please recover the package and contact with
your supplier immediately.
25
Installation
Installation Environment
26
Installation
Power Requirement
The power requirement is as follows:
100-240V~, 50/60Hz
The maximum power consumption is 150VA
Warning:
The instrument requires a single well-grounded power socket. Its
grounding voltage is no more than 0.5V.
A grounded outlet is required to connect with the grounding pole
on the rear panel. Make sure of the reliability of the power
grounding at working site.
Caution:
Fluctuating electric will badly decrease the performance and
reliability of the instrument.
Proper action such as the installation of UPS (self-preparation)
should be taken before use.
First Installation Notice
(1) Open the right door, cut the ties binding to the sampling module
and remove the ties.
(2) Remove the protective film on the injection ports of the chambers.
27
Installation
28
Installation
secure.
Caution:
After completing all the tubing installation, keep the tubing natural
state, without distortion, folding, and twist.
All the tubing connectors must be installed manually. Forbid using
any tool.
The reagent must be supplied with the instrument. Otherwise it
may easily cause inaccurate measurement results and incorrect
classification of the blood cell, or make the tubing system badly
damaged.
Keep the reagent from direct sunlight.
After replacing the reagent container, discard the rest of the
reagent in the former container. Forbid putting the rest of the
reagent into the replacing container. Avoid polluting the new
replacing reagent.
When replacing the reagent, avoid making the plastic tube
inserted into the reagent container or contacted with other things.
Avoid polluting the new replacing reagent.
Avoid using frozen reagent.
Avoid using the reagent out of its expiration date.
Warning:
Handle and dispose of the waste according to acceptable laboratory,
local state and national standards.
29
Installation
1) Tear down the gummed paper from the door of the recorder.
3) Insert the new paper into the paper entry and make the printing
side towards the thermal head.
4) When the paper juts out from the other side, pull it out and keep it
straight.
Plug keyboard cable into the interface marked with . Plug mouse
cable into the interface marked with on the rear panel of the
instrument.
The keyboard and mouse can be placed where it is convenient for your
operating. We suggest that the keyboard placed under the display
screen of the instrument, the mouse cushion at the right side of the
keyboard, and the mouse on the mouse cushion.
30
Installation
31
Sample Analysis
(1) Whether the diluent and waste tubing are distorted, folded or
twisted, and whether the connection is secure.
3.2 Startup
If external printer is equipped, turn on the printer power and ensure
they are in ready status.
Press the power button on the rear panel of the instrument, then the
power lamp on the front panel of the instrument lights on. The
instrument automatically performs initialization program.
After completing initialization, the instrument will enter into the self-test
window. It tests the working conditions of each part and if there are
enough diluent and lyse. Simultaneously prime and clean the tubing.
After self-test, the instrument will enter into the ‘Measure’ menu, as
shown in Figure 3-1.
32
Sample Analysis
Figure 3-1
33
Sample Analysis
3.2.1 Function
In the ‘Measure’ menu, press “Func” button to enter into the “Function”
window, as shown in Figure 3-2.
Figure 3-2
34
Sample Analysis
Figure 3-3
35
Sample Analysis
Figure 3-4
Diluent: Inject the diluent quantitatively by sample needle. This menu is
used only in Prediluted mode, and “Diluent” button will not display in
other modes.
Histogram: Enter the histogram adjustment window of the current
sample. You can manually adjust the histograms of the current
sample.
36
Sample Analysis
(1) In the ‘Measure’ menu, press “Info” button, the information editing
window pops up. Click the item textbox to change the ID as 999,
then press “Ok” to return to the ‘Measure’ menu.
(2) In venous or capillary mode, press the “START” key below the
sample needle directly for background count. In prediluted mode,
the sample needle aspirates some non-contaminated diluent from
the sample cup for background count. The diluent pick-up method
is as follows. Analyzing sample in prediluted mode is also
necessary to dilute sample with the diluent which is previously
picked-up in this method:
(3) Place a clean empty sample cup or tube under the sampling
needle, press “START” key to dispense the diluent into the
sample cup. Press the “START” key once, the diluent is
dispensed into the sample cup or tube once. The quantity can
be used for background count in the second prediluted mode.
Press “Exit”, the instrument returns to the main operation
area.
37
Sample Analysis
Parameter Data
WBC ≤ 0.2109/L
RBC ≤ 0.021012/L
HGB ≤ 1g/L
HCT ≤ 0.5%
PLT ≤ 10109/L
Table 3-1
38
Sample Analysis
(2) Pinch the puncturing part, and puncture it with aseptic sample
needle quickly. The puncturing depth is about 2-3mm.
(3) Wipe off the first drop of blood, and then start the collection.
39
Sample Analysis
(4) Press the wound with tampon for a moment after finishing the
collection.
Notice:
The capillary blood collection should follow the professional capillary
blood collection standard. The general method is partial centesis. The
typical collection is puncturing from the tip of the finger.
Caution:
If the blood flows not freely, press far away from the wound, not around
the wound. Avoid making the histiocyte mixed into the blood, which
may lead to incorrect analytical result.
(1) In the ‘Measure’ menu, press the “Func” button, select “Sample
Type”, set “Prediluted” as the sample type, then press “Function”
button. In menu operation area, select “Diluent”.
(2) When the “Diluent” pops up, place a clean sample cup obliquely
under the sampling needle. Press “START” key. The instrument
starts to add diluent quantitatively. Scrape the drops at the top of
the sample needle into the sample cup, collect 20μL capillary
sample in pipette. Wipe the blood attached on the outer layer of
the pipette with clean tissues. Immediately mix up the sample in
the diluent sample cup.
40
Sample Analysis
Caution:
When adding the diluent, the sample cup must be placed obliquely
under the sample needle, which allows the diluent to run down the cup
wall without forming air bubbles.
Notice:
If there is a need to prepare samples in quantities in prediluted mode,
users can use the “Diluent” function to continuously prepare the
diluent.
41
Sample Analysis
42
Sample Analysis
ID: The input range is: 001~998. The instrument owns an intelligent
number manager. The final number derives from the comprehensive
information that the instrument adds the information of year, month
and day in front of the input number according to the measured time. If
user does not input the serial number, the instrument will accumulate
the number from 001 everyday according to the measured time.
Department: You can also choose the recorded department
information from the right optional textbox, or choose automatically by
the instrument after inputting the code of the department in the left
input textbox.
Doctor: You can also choose the recorded doctor information from the
right optional textbox, or choose automatically by the instrument after
inputting the code of the doctor in the left input textbox.
Operator: You can also choose the recorded doctor information from
the right optional textbox, or choose automatically by the instrument
after inputting the code of the doctor in the left input textbox.
Checker: You can also choose the recorded doctor information from
the right optional textbox, or choose automatically by the instrument
after inputting the code of the doctor in the left input textbox.
Limits: Choose from General, Man, Woman, Child, Newborn, Define 1,
Define 2, and Define 3 automatically. Set the low and high limits of the
corresponding parameters.
43
Sample Analysis
Notice:
The number 999 is a special number for background test. Please
don't input this number in blood sample test.
If there is a need to input the sample test information in quantities,
users can concentrate on inputting the sample test information in
the review window after the current sample test finished.
(1) Place a sample cup under the sample needle, press “START” key,
the instrument starts to aspirate the liquid.
(3) After the analysis finished, the result will be displayed behind the
corresponding parameters on ‘Measure’ menu of the instrument,
with histograms WBC, RBC, and PLT.
If “Auto” item is “ON”, the recorder or printer will output the test results
automatically.
If there are clogs during the count and analysis process, the
information section would display “Clog”.
44
Sample Analysis
45
Sample Analysis
(1) In the ‘Measure’ menu, select “Func”, then press “Histogram”, the
instrument enters into the histogram adjustment window, as
shown in Figure 3-5. Press “Para” button to adjust the histogram.
46
Sample Analysis
Figure 3-5
(2) After selecting the histogram, press “Line” button, and select the
sorting line you want to adjust.
(3) Press “Left” or “Right” button, you can move the sorting line to the
left or right, the data of the line will be displayed at the top right
corner of the screen.
47
Sample Analysis
Caution:
Unnecessary and incorrect manual adjustment will cause unreliable
analytical result. Make sure the necessity of the operation.
Notice:
If the result of WBC is less than 0.5x109/L, the system will not perform
leukocyte differential automatically.
3.8 Shutdown
Shutdown routine must be performed before turning off the power
every day.
During shutdown procedure, the instrument will perform daily
maintenance and clean the tubing automatically.
The procedures are as follows:
Figure 3-6
48
Sample Analysis
Figure 3-7
4) Turn off the printer power (if equipped), clean the workbench and
dispose the waste.
Notice:
Do not turn off the power of the instrument directly while performing the
shutdown procedures.
49
Quality Control
50
Quality Control
select “Edit” in the menu. The instrument enters into the quality
control edit window, as shown in Figure 4-1.
Figure 4-1
3) Input lot number and expiration date of the control, the assay and
limit of the quality control parameters.
6) Press “Back” button to save the data of the current QC file and
return to the quality control edit window.
51
Quality Control
Notice:
If the parameter is zero, the system will take the assay and limit as
no assignment. If the parameter exceeds the display range, the
system directly displays the maximum value in the display range.
It needs to re-input. Or else the quality control will only act on
other parameters.
If the lot number or expiration date is invalid, the system will take
them as no assignment. It needs to re-input. Or else the QC file
cannot run quality control program.
52
Quality Control
Figure 4-2
3) Place the controls under the sampling needle. Press “START” key,
the indicator lamp starts to glitter, the instrument aspirates the
controls, move the controls away until the indicator lamp stops
glittering.
4) When the test finished, the result will be displayed on the column
of the current window. If alarm occurs during the test, the current
result may be incorrect.
5) Press “Del” button to delete the result and test again after the
problem is solved.
53
Quality Control
menu.
Notice:
Each QC file can store up to 31 QC data.
Figure 4-3
54
Quality Control
The screen can display QC graph of one parameter. The graph shows
the lot number, date, serial number, QC mode and distribution of the
QC data. Click the scroll bar on the right side of the screen, and
continuously select the QC graph of different parameters.
The seven parameters at the left side of the QC graph from top to
bottom:
Assay + 3limit
Assay + 2limit
Assay + limit
Assay
Assay - limit
Assay - 2limit
Assay - 3limit
The three parameters at the right side of the QC graph from top to
bottom:
Mean: average value
Diff: standard deviation
CV: coefficient of variation
The vertical line in the middle of the QC graph indicates the current
chosen number. Click the scroll bar on the left side of the screen to
continuously select the different number.
The chosen number is displayed in the data box below the
corresponding parameter. The test time for the chosen number is
55
Quality Control
56
Quality Control
Figure 4-4
57
Calibration
Chapter 5 Calibration
The instrument has been calibrated strictly at the factory. You may
need to perform calibration procedures when you replace any
component that involves the primary measurement characteristics, or
when the shift occurs during controls. Calibration of the instrument is to
provide the assurance that the instrument is providing results with
accuracy as design. The purpose of calibration is to ensure the
accuracy of the measurement result to meet the requirement of the
design all the time.
To ensure the measurement accuracy of the instrument and obtain
reliable results, it is necessary to calibrate the instrument in the
following situations:
4) Replace reagent.
Warning:
Calibrators should be stored and used properly according to the
calibrator manual.
58
Calibration
Warning:
Make sure the instrument is in a normal state before calibration.
Warning:
Avoid using the measurement result on medical test or clinical before
the calibration is finished accurately.
Check the instrument and reagents carefully before calibration. Make
sure the instrument is in normal status, and the required sample mode
of the measurement is selected.
2) Record the data of WBC, RBC, HGB, MCV and PLT. Calculate
the CV value according to the following formula. Only results in
59
Calibration
(X
i 1
i X )2
CV n 1 100%
X
Parameters CV (%)
WBC ≤ 2.0
RBC ≤ 1.9
HGB ≤ 1.9
MCV ≤ 0.4
PLT ≤4
Table 5-1
60
Calibration
Figure 5-1
61
Calibration
2) Select the needed calibration data from the calibration data menu
box.
4) Press “Ok” button to save the current calibration result and return
to the main menu window.
62
Calibration
Figure 5-2
63
Calibration
6) Press “Del” button to delete the result and test again after the
problem is solved.
Notice:
Calibration can be performed 5 times at the most.
The calibration result will be displayed after testing 3 times.
If the parameter contains invalid data, the system will not perform
calibration. It needs to re-input valid data.
64
Setting
Chapter 6 Setting
The instrument has accomplished all the settings at the factory. For
customer's convenience, most of the system parameters can be set by
operators so as to meet various requirements.
Notice:
Operations in this chapter will change the running state of the
instrument. Make sure of the necessity of changing before operating.
65
Setting
Figure 6-1
6.2 Printout
In the ‘Measure’ menu, press “Menu” button, select “Setting” in the
menu, then select “Print". The instrument will enter into the print
window, as shown in Figure 6-2.
66
Setting
Figure 6-2
Click the menu box you want to adjust, the option window pops up.
Click the item you want to set.
Press “Ok”, the system will save the current parameters and return to
the setting window. Press “Cancel” button, the system will not save the
current parameters and return to the main menu window.
Print select
Select recorder, the data will only be transported to the recorder.
Select serial port, the data will only be transported to the net.
Select recorder and serial port, the data will be transported to the
recorder and net at the same time.
67
Setting
Select Laser Jet, the data will be transported to a printer such as the
HP Laser Jet1018 printer or HP Laser Jet1008 printer.
Select Desk Jet, the data will be transported to a printer such as the
HP desk Jet1668 printer or HP Laser Jet 2668printer.
Figure 6-3
Click “Ok” to enter the laser print setting interface, select the print font
(format).
68
Setting
(1) A4 Print
Figure 6-4
When the paper size is A4, it defaults to 24 fonts. Adjust the vertical
distance of the printing layout by setting the distance and size of the
title. The contents occupy almost the entire layout of A4 paper, as
shown in Figure 6-5.
69
Setting
Figure 6-5
70
Setting
(2) A5 Print
Figure 6-6
When the paper size is A5, you can select 12 or 16 fonts. Adjust the
vertical distance of the printing layout by setting the distance and size
of the title. The contents occupy almost the entire layout of A5 paper
and about half of the A4 paper, as shown in Figure 6-7 (take 16 fonts
for example).
71
Setting
72
Setting
73
Setting
Notice:
Print caption
Print type
Notice:
The selected printer type must be accord with the used one, otherwise
it will not be able to print normally.
6.3 Control
In the ‘Measure’ menu, press “Menu” button, select “Setting” in the
menu, then select “Control”, the instrument will enter into the control
setting window, as shown in Figure 6-8.
74
Setting
Figure 6-8
Click the menu box you want to adjust, the option window pops up.
Click the item you want to set.
Press “Save”, the system will save the current parameters and return
to the hematology window.
Press “Cancel” button, the system will not save the current parameters
and return to the main menu window.
Auto
Set “Print” as “OFF”, the system will not print the test report
automatically after the sample test finished.
Set “Standby” as “OFF”, the instrument will keep working status all
75
Setting
along. Set “Standby” as data. When the time, which the user does not
perform any operation, exceeds the setting time, the instrument will
access standby state. During standby time, the instrument will turn off
part of the circuit power, and take the sample needle back. Standby
function can greatly save electrical energy and extend the life of the
instrument. Press the left key of the mouse or any key on the keyboard,
standby will be canceled.
Set “Start test” as “OFF”, the system will not perform background test
after booting.
Set “Shutdown immersion” as “OFF”, the system will not perform probe
cleanser soaking after shutdown.
Set “Enhanced Flush” as number. The sample test times interval of the
enhanced flush that controls the automatic maintenance of the
instrument is selected by the pull-down menu.
Sound
Set “Alarm” as “OFF”. When the instrument appears alarm, it will not
make alarm sound, but display the alarm information in the Information
Section.
Set “Alarm” as “ON”. When the instrument appears alarm, it will make
alarm sound and display the alarm information in the Information
Section.
76
Setting
Set “Prompt” as “OFF”. When the sample test finished, the instrument
makes no sound.
Set “Prompt” as “ON”. When the sample test finished, the instrument
will make a prompt sound like “di”.
Display
Set “Para” as “English”. In parameter area of the hematology analyzer,
the parameters will be displayed in abbreviations.
Set “Limits” as “No Display”. In parameter area of the hematology
analyzer, the parameters will not display reference value of the current
sample.
Set “Limits” as “Low-High”. In parameter area of the hematology
analyzer, the reference values will be displayed from low to high.
Set “Limits” as “High-Low”. In parameter area of the hematology
analyzer, the reference values will be displayed from high to low.
6.4 Limits
In the ‘Measure’ menu, press “Menu” button, select “Setting” in the
menu, then select “Limits” to enter into the limits setting window, as
shown in Figure 6-9.
The 8 group reference values are displayed at the left side of the
screen. The dot in front of the reference value represents that this is
the current selected reference group. The limits of the reference value
are displayed in the limits boxes on the screen.
77
Setting
Click the dot in front of the group to select the group required for
setting.
Figure 6-9
Click “Default” button, the instrument will select the default reference
values to replace the current reference values.
Click the adjustable bar of the limits box to adjust the upper and lower
limit of the reference value.
Function of menu buttons:
Save: Save the reference values of the current group.
Print: Print the reference values of the current group.
Back: Exit from current window and return to main menu window.
78
Setting
Notice:
The default reference value may not suitable for the local conditions.
Users can modify it according to the local specific circumstances.
Caution:
The changing of the reference value will cause variation of abnormal
prompt on hematology target. Please make sure the necessity of the
variation.
79
Setting
Figure 6-10
Click the scroll bar at the right side of the window, the instrument can
record the serial number continuously.
Click “Back” button, the system exits from the current window and
returns to the ‘Measure’ menu.
80
Setting
recommended.
Click “Ok” button to save the current input information. The system
returns to department setting window.
Click “Cancel” button, the current input information will be canceled,
and the system returns to department setting window.
Figure 6-11
81
Setting
Figure 6-12
Click the scroll bar at the right side of the window to display
continuously the records of different numbers.
Click “Back” button, the system will exit the current window and return
to the main menu window.
82
Setting
Figure 6-13
83
Setting
Notice:
The coding method of the abbreviation should be easy to remember. In
most cases, abbreviation of the department name is recommended.
84
Review
Chapter 7 Review
The instrument can automatically store the results after each sample
analysis. The operator can review, search, modify, print, and delete the
data.
Figure 7-1
85
Review
86
Review
2) Select “Delete” in the menu, a dialog box pops up, and choose to
delete the data or not. Press “Ok” button, the data will be deleted.
Press “Cancel” button, the operation will be cancelled.
1) Click “All” button, the data will be selected. Click again, the
selection will be canceled.
2) Click “Delete” button, a dialog box pops up, and choose to delete
the data or not. Press “Ok” button, the data will be deleted. Press
“Cancel” button, the operation will be cancelled.
1) Click “All” button at the top of the screen, the data will be selected.
Click again, the selection will be canceled.
87
Review
1) Click “Query” in the menu, the search window pops up, as shown
in Figure 7-2.
2) Click the button on the left of the item to be searched, input the
keyword you want to search.
Figure 7-2
88
Review
89
Review
3) Click “Right” or “Left” button to move the line to the right or left.
Figure 7-3
90
Review
91
Service
Chapter 8 Service
Figure 8-1
8.1 Flush
This function is used to solve general gem hole clog issue. The
92
Service
instrument makes a fixed pressure and voltage act on the gem hole for
a few seconds to eliminate dirt at the holes.
8.3 Cleaning
This function is used to clean the sampling needle and tubes.
93
Service
8.9 Prime
Prime diluents and lyse to the related tube.
8.10 Maintenance
Perform the weekly maintenance according to the prompt information
on the screen. Prepare probe cleanser before maintenance.
8.11 Mechanic
Users can detect the valve, motor, circuit and interface in the
“Mechanic” menu. It is helpful to the detection of malfunctions.
94
Service
Figure 8-2
Valve:
“ON” represents the valve is on. “OFF” represents the valve is off.
Motor:
"OK" represents that the motor is in normal status.
"FAIL" represents the motor has failure.
Recorder:
Click this button, the recorder will print a test paper.
Printer:
Click this button, the external print will print a test paper.
Back:
95
Service
Figure 8-3
8.12 Special
The instrument enters special maintenance window. User can adjust
the running control data. This function should be performed in
guidance of the engineers authorized by the manufacturer.
96
Maintenance
Chapter 9 Maintenance
Warning:
97
Maintenance
Caution:
Avoid using corrosive acids, alkali and volatile organic solvent,
such as acetone, aether, chloroforms, to wipe the appearance of
the instrument. Only neutral detergent can be used.
Avoid wiping the inner of the instrument.
Warning:
98
Maintenance
99
Maintenance
1) Turn off the power of the instrument, and pull out the power cord.
Open the right side door and brush away the internal dust of the
instrument gently.
Note: We must prevent the dust fall into the measuring cup to
contaminate it and make an abnormal test result.
2) Wipe the outer wall of the sample probe with a cotton swab
dipped in alcohol and gently wipe the outer wall of the sample
probe. Do not bend it hard.
3) Wipe the bottom and top of the swab with a cotton swab
dampened with alcohol.
100
Maintenance
2) Wipe the outer wall of the shielding box with a cotton swab dipped
in alcohol and gently wipe the outer wall of the shielding box.
101
Maintenance
2) Attach the pipe plugs unplugged in the first installation to the line
interfaces on the rear panel of the instrument, according to the
colors’ one to one correspondence.
4) Clean the power cable with a clean cloth dampened with neutral
detergents, place it in a shade place and put it in a plastic bag.
5) Put the instrument and plastic bags with components into the
packaging box
After the instrument is not used for two weeks, the maintenance of the
boot:
1) Connect the lyse and the diluent, and select the whole machine
for infusion in the main menu service;
102
Maintenance
103
Troubleshooting
Chapter 10 Troubleshooting
104
Troubleshooting
1) Replace diluent.
1) Replace lyse.
2) If the upper method could not solve the problem, perform as the
following procedures.
105
Troubleshooting
2) If the upper method could not solve the problem, perform as the
following procedures.
2) If the upper method could not solve the problem, perform as the
following procedures.
2) Insert the new paper into the paper entry, and make the printing
106
Troubleshooting
3) When the paper juts out from the other side, pull it out and keep it
straight.
2) If the above method could not solve the problem, repeat the
procedure 3 times. If the problem still cannot be solved, perform
as the following procedures.
4) If the above method could not solve the problem, repeat the
107
Troubleshooting
108
Help
Chapter 11 Help
Figure 11-1
This window shows in chapter mode. User can search help information
conveniently. Click the scroll bar up and down to read the previous or
next information.
109
Symbol
Appendix 1 Symbol
Equipotential
COM 2 Interface
Terminal
High
Keyboard Interface
Temperature
In vitro
Date of
diagnostic
manufacture
medical device
110
Symbol
CE marking. The
device is fully in
Authorized
conformance with
representative in
the Directive
the European
98/79/EC on in
Community
vitro diagnostic
medical devices
Catalogue number
111
Symbol
package.
Note: The illustrations are for reference only, subject to pictures on the
outer package.
112
Specification
Appendix 2 Specification
A 2.1 Reagent
Diluent, lyse and probe cleanser
Item Unit
WBC 109/L
Neu# 109/L
Lym# 109/L
Mid# 109/L
Neu% %
Lym% %
Mid% %
RBC 1012/L
HGB g/L
HCT %
113
Specification
Item Unit
MCV fL
MCH pg
MCHC g/L
RDW-SD fL
RDW-CV %
PLT 109/L
MPV fL
PDW None
PCT %
P-LCR %
114
Specification
Venous 10μL
Capillary 10μL
Prediluted 20μL
Item Range
WBC 0~200.0×109/L
RBC 0~20.00×1012/L
HGB 0~300g/L
115
Specification
Item Range
PLT 0~2000×109/L
HCT 0%~80%
Item Range
WBC ≤ 0.2×109/L
RBC ≤ 0.02×1012/L
HGB ≤ 1.0g/L
HCT ≤ 0.5%
PLT ≤ 10.0×109/L
116
Specification
0~0.99×1012/L (±0.05×1012)/L
RBC
1.0~9.99×1012/L ±5%
0~99g/L ±2.0g/L
HGB
100~300g/L ±2%
0~99×109/L (±8.0×109)/L
PLT
100~999×109/L ±10%
Item CV Range
117
Specification
Item CV Range
LYM%≥15.0%,
LYM% ≤8.0%
WBC≥4.0×109/L
MID%≥5.0%,
MID% ≤15.0%
WBC≥4.0×109/L
NEUT%≥30.0%,
NEUT% ≤8.0%
WBC≥4.0×109/L
RBC ≤1%
HGB ≤2%
PLT ≤2%
A.2.9 Accuracy
Deviation
Parameter Range
Requirement
WBC 3.5×109/L~9.5×109/L ≤±10%
118
Specification
2-8 Accuracy
Voltage Frequency
119
P0 1 . 9 1 . 3 0 0 2 5 4 - 0 4