Dymind Global Customer Service Center QA-F00XX

0x63006040-Positive pressure abnormal (high)


Use Background
When analyzer gives a false alarm - HGB background voltage abnormal, and we can’t remove it.

Alarm principle: The host queries the pressure in the positive-pressure chamber. The current
pressure is over the upper limit (9KPa) of the attached parameter.

Declaration
20/8/2020, V1.0, For reference only.

Reference Standard
Module Possibility Solution

1.Air pump driver


chip is in the
on-state of
For DH36/
breakdown, and the
DH36X/
air pump keeps
DF50/
working;
DF50CRP

2.Air pump is
damaged

For DH76 N/A N/A

Dymind Global Customer Service Center Jersey


Dymind Global Customer Service Center QA-F00XX

Relative spare part code


Part Name P/N Picture

DH31/DH33/
Mainboard with SD card(without
DH36/ 22.01.0023A
battery)
DH36VET

Mainboard with SD card(without


DH36X 22.01.0199A
battery)

DF50/DF51/
DF52/ Mainboard with SD card(without
22.01.0021C
DF50VET battery)
&DF50CRP

DH31/DH33/
DH36/
DH36VET
Vacuum Pump 23.99.0008A
&DF50/DF51
/DF52/
DF50VET

DH36X
Air Pump 23.99.0012A
&DF50CRP

If you have any problem with this file, please contact our engineer ASAP, thank you!