CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Task
Timing Category
Apply at all sites
Task Tracking
After completion of your task the following entry in your Service Report is required: DD+DIS161.17E *
* Insert the document number into the field "Comment" (SMS form).
Use, dissemination, distribution or reproduction of this bulletin by unauthorized personnel is not permitted and may be unlawful.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Copyright © 2017 Agfa HealthCare N.V. | Document ID: 62249980 | service_information_bulletin_template_v07
Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 1 of 4
Service Information Bulletin DD+DIS161.17E / No. 27
1 Introduction
Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 2 of 4
Service Information Bulletin DD+DIS161.17E / No. 27
2 Prerequisites
SOFTWARE:
Download the Digitizer software ARC_2402 from the Agfa HealthCare Library:
REFERENCED DOCUMENTS:
Enclosure - Installation Instructions for Digitizer Document ID 34181094
Software ARC_2401 and higher versions
CR 10-X_CR 12-X_CR 15-X Service Manual Document ID 42866196
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare N.V or one of its affiliates
or who are otherwise authorized by Agfa HealthCare N.V. or one of its affiliates to provide
such services.
NOTE:
While the device is in active clinical use (image plates are scanned), logging in
through telnet and redirecting the console to this telnet session is strictly forbidden, as
it might lead to an image loss.
Perform a shading calibration:
If the software ARC_2402 is installed on a device, which was delivered ex production
with a photomultiplier having a 5-digit serial number, the error message "22758"
(loading the Digitizer calibration data failed) may be shown during the start-up.
In this case perform a new shading calibration with the used Image Plate type(s)
PIP and/or NIP. Refer to chapter 5, Calibration, of the Service Manual.
If ARC_2402 has been installed in production, a shading calibration is not necessary.
Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 3 of 4
Service Information Bulletin DD+DIS161.17E / No. 27
3 Instructions
REQUIRED TIME:
Approximately 15 minutes for the installation via USB flash drive
(1) Install the software ARC_2402 during preventive maintenance or to improve transfer
speed on devices with a slow transmission speed to the NX Workstation.
Follow the instructions described in the Installation Instructions for Digitizer Software
ARC_2401 and higher versions. enclosed to the software ZIP file.
(2) Inform the customer about the improvements.
4 Verification
Verify the successful software upgrade as described in the Installation Instructions for
Digitizer Software ARC_2401 and higher versions.
The instructions are enclosed to the software ZIP file on the Agfa HealthCare Library.
5 Keywords
Software upgrade, ARC_2402, Ethernet; communication problem, slow transfer rate