Service Information Bulletin


This Bulletin is for information only.

Device Software ARC_2402 is available

Task
Timing Category
Apply at all sites

Next service Apply at affected sites as listed below

Optional to improve functionality of product

Task Tracking
After completion of your task the following entry in your Service Report is required: DD+DIS161.17E *
* Insert the document number into the field "Comment" (SMS form).

Purpose of this document:


This document informs about the software version ARC_2402.

Affected serial number(s) / batch:


See section 1.2, Affected serial numbers.

Use, dissemination, distribution or reproduction of this bulletin by unauthorized personnel is not permitted and may be unlawful.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Copyright © 2017 Agfa HealthCare N.V. | Document ID: 62249980 | service_information_bulletin_template_v07
Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 1 of 4
Service Information Bulletin DD+DIS161.17E / No. 27

1 Introduction

Device Software ARC_2402 is released since December 2017.

1.1 Solved software issue in Digitizer software ARC_2402 compared to ARC_2401

Symptom Communication problems or slow connection.

Cause Faulty initialization of the Ethernet chip on the PMI Board.

Solution Install Digitizer software version ARC_2402.


As of software version ARC_2402 the Ethernet chip is initialized by a reset pulse during
start-up before the Ethernet communication starts.
This may improve the transfer rate of devices with a sporadic slow transmission to the NX
Workstation.

1.2 Affected serial numbers

ARC_2402 is installed in following SNs ex-factory (production site Germany):


Digitizer Type Serial Numbers
CR 10-X 5151 / 100 As of SN 9762
CR Reader 5151 / 110 As of SN 20859
CR 12-X 5151 / 200 As of SN 31172
CR Advanced Reader 5151 / 210 As of SN 40602
CR 15-X 5151 / 300 As of SN 54205,
and additionally in SN 54058, SN 54203,
CR Multiformat Reader 5151 / 310 As of SN 46116

ARC_2402 is installed in following SNs ex-factory (production site China):


Digitizer Type Serial Numbers
CR 10-X 5151 / 100 As of SN 73921
CR Reader 5151 / 110 Not applicable
CR 12-X 5151 / 200 As of SN 81565
CR Advanced Reader 5151 / 210 Not applicable
CR 15-X 5151 / 300 As of SN 90676
CR Multiformat Reader 5151 / 310 Not applicable

Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 2 of 4
Service Information Bulletin DD+DIS161.17E / No. 27

2 Prerequisites

SOFTWARE:
Download the Digitizer software ARC_2402 from the Agfa HealthCare Library:

Computed Radiography  CR Digitizers  CR 10-X  Software

Contents of the software ZIP File (approx. 8.5 MB):


 Software ARC_2402
 Checksum.md5 file
 Installation Instructions for Digitizer Software ARC_2401 and higher versions,
Document ID 34181094

REFERENCED DOCUMENTS:
 Enclosure - Installation Instructions for Digitizer Document ID 34181094
Software ARC_2401 and higher versions
 CR 10-X_CR 12-X_CR 15-X Service Manual Document ID 42866196

IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare N.V or one of its affiliates
or who are otherwise authorized by Agfa HealthCare N.V. or one of its affiliates to provide
such services.

NOTE:
 While the device is in active clinical use (image plates are scanned), logging in
through telnet and redirecting the console to this telnet session is strictly forbidden, as
it might lead to an image loss.
 Perform a shading calibration:
If the software ARC_2402 is installed on a device, which was delivered ex production
with a photomultiplier having a 5-digit serial number, the error message "22758"
(loading the Digitizer calibration data failed) may be shown during the start-up.
In this case perform a new shading calibration with the used Image Plate type(s)
PIP and/or NIP. Refer to chapter 5, Calibration, of the Service Manual.
If ARC_2402 has been installed in production, a shading calibration is not necessary.

Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader
Agfa HealthCare Company Confidential Type 5151 / 100 / 110 / 200 / 210 / 300 / 310 Page 3 of 4
Service Information Bulletin DD+DIS161.17E / No. 27

3 Instructions
REQUIRED TIME:
Approximately 15 minutes for the installation via USB flash drive
(1) Install the software ARC_2402 during preventive maintenance or to improve transfer
speed on devices with a slow transmission speed to the NX Workstation.
Follow the instructions described in the Installation Instructions for Digitizer Software
ARC_2401 and higher versions. enclosed to the software ZIP file.
(2) Inform the customer about the improvements.

4 Verification
Verify the successful software upgrade as described in the Installation Instructions for
Digitizer Software ARC_2401 and higher versions.
The instructions are enclosed to the software ZIP file on the Agfa HealthCare Library.

5 Keywords
Software upgrade, ARC_2402, Ethernet; communication problem, slow transfer rate

6 Version History
Version Change Date
1.0 Initial version 12-2017

Release date: 12-2017 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader / CR 15-X / CR Multiformat Reader