Professional Documents
Culture Documents
PB92971EN CO2RE User Manual - English - ECO-4784
PB92971EN CO2RE User Manual - English - ECO-4784
PB92971EN
Copyright © 2017, Syneron Medical Ltd. All rights reserved.
Syneron Medical Ltd. reserves the right to make changes to its products or specifications to improve performance,
reliability, or manufacturability. Information furnished by Syneron Medical Ltd. is believed to be accurate and
reliable. However, Syneron Medical Ltd. assumes no responsibility for its use. No license is granted by its implication
or otherwise under any patent or patent rights of Syneron Medical Ltd.
No part of this document may be produced or transmitted in any form or by any means, electronic or mechanical, for
any purpose, without the express written permission of Syneron Medical Ltd.
Data is subject to change without notification.
Syneron Medical Ltd. has patents and pending patent applications, trademarks, copyrights, or other intellectual
property rights covering subject matter in this document. The furnishing of this document does not give you any
license to these patents, trademarks, copyrights, or other intellectual property rights except as expressly provided in
any written agreement from Syneron Medical Ltd.
Specifications are subject to change without notice.
Manual Catalog Part Number: PB92971EN
Revision Release Date: March 2017
● Before using the device, please check local regulations. If any local legislation is violated,
use cannot be authorized.
● In the United States, federal law restricts prescription medical devices to be sold by or on
the order of a physician, or properly licensed practitioner. Syneron-Candela makes no
representations regarding federal, state, or local laws or regulations that might apply to
the use and operation of this device.
Table of Contents
CHAPTER 1: Safety and Regulatory Page
1.1. Introduction ....................................................................................................................... 1-1
1.2. Burn Hazard ...................................................................................................................... 1-1
1.3. Reflected and Direct Eye Exposure Hazard ...................................................................... 1-2
1.4. Laser Safety Eyewear ........................................................................................................ 1-2
1.5. Explosion and Fire Hazard ................................................................................................ 1-2
1.6. Electrical Safety ................................................................................................................ 1-3
1.6.1. High Voltage Hazard ............................................................................................... 1-3
1.6.2. Grounding the System ............................................................................................. 1-3
1.6.3. Using the Proper Power Receptacle and Plug ......................................................... 1-3
1.7. Safety Precautions ............................................................................................................. 1-3
1.7.1. Electrical and Mechanical Safety ............................................................................ 1-3
1.7.2. Ocular Safety ........................................................................................................... 1-4
1.7.3. Smoke Evacuation and Laser Plume Pollution Hazards ......................................... 1-4
1.7.4. Surgical Safety ........................................................................................................ 1-5
1.7.5. Operating Safety Precautions .................................................................................. 1-5
1.8. System Safety Features ..................................................................................................... 1-5
1.8.1. Keyswitch ................................................................................................................ 1-5
1.8.2. Safety Shutter .......................................................................................................... 1-5
1.8.3. Footswitch ............................................................................................................... 1-6
1.8.4. Double-Tiered Laser Emission Security ................................................................. 1-6
1.8.5. System Self-Testing ................................................................................................ 1-6
1.8.6. Emergency Shut-Off Knob...................................................................................... 1-6
1.8.7. Safety Fuses............................................................................................................. 1-6
1.8.8. Laser Beam Emission Indicators ............................................................................. 1-7
1.8.9. Remote Interlock Connector ................................................................................... 1-7
1.8.10. Fault Detection ........................................................................................................ 1-7
1.9. Compliance with International Standards ......................................................................... 1-8
1.10. Glossary of the Symbols Used in the System ................................................................... 1-9
1.11. System Labels ................................................................................................................. 1-10
PB92971EN TOC-1
Table of Contents User Manual
TOC-2 PB92971EN
User Manual Table of Contents
PB92971EN TOC-3
Table of Contents User Manual
APPENDIX B: Clinical Guide - Fractional Vaginal, Vulval and Introitus Treatments Page
B.1. CO2 Vulvovaginal Laser Treatment ...................................................................................... B-1
B.2. Indications for Use................................................................................................................. B-1
B.3. Contraindications ................................................................................................................... B-1
B.4. Realistic Expectations............................................................................................................ B-2
B.5. Possible Side Effects ............................................................................................................. B-3
B.6. Pre-Treatment ........................................................................................................................ B-3
B.7. Test spots ............................................................................................................................... B-4
B.8. Internal Treatment Procedure ................................................................................................ B-5
B.8.1. Internal Treatment ........................................................................................................ B-5
B.8.2. External Treatment ....................................................................................................... B-6
B.9. Treatment Parameters ............................................................................................................ B-7
B.10. Post-Treatment Care ............................................................................................................ B-7
B.11. Follow-Up ............................................................................................................................ B-8
B.12. Treatment Conclusion.......................................................................................................... B-8
B.13. After the Treatment Session ................................................................................................ B-8
C.1. Electromagnetic Emissions.................................................................................................... C-1
C.2. Electromagnetic Immunity .................................................................................................... C-2
C.3. Recommended Separation Distances..................................................................................... C-4
TOC-4 PB92971EN
User Manual Table of Contents
List of Figures
Figure 1-1: System Labels Locations .......................................................................................... 1-10
Figure 2-1: Physical Dimensions .................................................................................................. 2-3
Figure 2-2: Unpacked Articulated Arm ........................................................................................ 2-5
Figure 2-3: Articulated Arm Installation ....................................................................................... 2-5
Figure 2-4: Lens Assembly Connection ........................................................................................ 2-6
Figure 2-5: Service Panel .............................................................................................................. 2-7
Figure 2-6: Remote Interlock Connector ...................................................................................... 2-8
Figure 3-1: CO2RE Laser System ................................................................................................. 3-4
Figure 3-2: CO2RE Main Treatment Screen (sample) .................................................................. 3-5
Figure 3-3: Beam Scanning Principle (Hexagon Scan Pattern Example
with Ring 1 Ablation Spot Option) ............................................................................ 3-6
Figure 3-4: CO2RE Handpieces and Lenses.................................................................................. 3-8
Figure 4-1: Lens Assembly Connection ........................................................................................ 4-2
Figure 4-2: Handpiece Connection ............................................................................................... 4-2
Figure 4-3: Main Menu Screen – Fractional/Full Resurfacing ..................................................... 4-3
Figure 4-4: Handpiece and Lens Confirmation Screen – Fractional/Full Resurfacing ................ 4-4
Figure 4-5: Treatment Screen Elements – CO2RE Aesthetics Mode ............................................ 4-5
Figure 4-6: Treatment Screen – CO2RE Surgical Mode ............................................................... 4-7
Figure 4-7: Main Menu Screen – Intima ....................................................................................... 4-8
Figure 4-8: Handpiece and Lens Confirmation Screen – Intima External and Internal ............... 4-8
Figure 4-9: Treatment Screen – CO2RE Intima Modes (left: External / right: Internal).............. 4-9
Figure 4-10: Utilities Screen Elements ....................................................................................... 4-10
Figure 4-11: Password Entry Keypad ......................................................................................... 4-11
Figure 5-1: Laser Beam Alignment Check Results ....................................................................... 5-8
Figure 5-2: Replacing the Fuses .................................................................................................... 5-9
Figure 6-1: Sample System Error Message ................................................................................... 6-2
Figure B-1: Handpiece Rotation by 90° ........................................................................................ B-6
PB92971EN TOC-5
Table of Contents User Manual
TOC-6 PB92971EN
CHAPTER 1
Safety and Regulatory
1.1. Introduction
The CO2RE system is a carbon dioxide laser device operating at a
wavelength of 10,600nm. The CO2 laser delivers pulses to create a distinct
pattern of affected tissue.
Caution
● The CO2RE system should be operated only by a physician
trained to operate the CO2RE system. Training should be in
accordance with applicable local regulations. Please contact
Syneron-Candela for additional information regarding training.
● Improper use or adjustment of this system may invalidate the
service warranty agreement. Please contact your authorized
Syneron-Candela representative before attempting to use the
system in any manner other than those specified in this
manual.
Warning
Use of controls or adjustments, or performance of procedures
other than those specified herein may result in hazardous
radiation exposure. Therefore personnel operating or servicing this
system must be thoroughly familiar with all safety requirements
and operating procedures
PB92971EN 1-1
Safety and Regulatory User Manual
Warning
Do not stare into the laser beam or allow it to be reflected from
any reflective surface. Even rough or black metal can reflect
laser radiation.
1-2 PB92971EN
User Manual Safety and Regulatory
The CO2RE system generates high electrical voltage within the main
console. To avoid injury, do not operate the system before ensuring that all
panels are properly installed. Do not remove or disassemble any panels.
● Use only a power receptacle and plug that are in good condition and are
specified for the system.
● Use only a hospital grade plug and a correctly matched power receptacle.
● To remove the power cable from the receptacle, hold it by the plug.
Never pull on the power cable to remove the plug from the receptacle.
● The system's cables are equipped with isolation coating and the
electronic circuitry is also isolated from the mains power supply. In
addition, the system’s plastic covers serve as a means of isolation.
PB92971EN 1-3
Safety and Regulatory User Manual
● Guard against accidental exposure to the laser beam by ensuring that all
personnel and patients wear protective eyewear whenever the system is
in use.
● Clearly identify the room in which the laser system is located, and post
the warning sign (supplied with the system) in a visible location.
● Make sure that all treatment room personnel are familiar with the
system's controls and know how to shut down the system instantly.
● Never look directly into the laser beam emitted from the articulated arm's
endjoint – with or without the lens assembly and handpiece – even when
wearing protective eyewear.
● Never direct the laser beam at anything other than the targeted treatment
site.
Caution
Laser plume may contain viable tissue particulates.
● A commercial smoke evacuator designed for use with surgical lasers may
be used; these are usually most effective when the plume is extensive.
The vacuum tubing or probe used to evacuate the laser plume should not
be used to suction blood or fluids unless it is specifically designed and
set up to perform both functions simultaneously.
1-4 PB92971EN
User Manual Safety and Regulatory
● Never use surgical accessories that have not been sterilized. Use of non-
sterilized accessories creates a potential risk of infection which may each
lead to significant medical complications.
● Do not treat patients with the CO2RE system if the aiming beam is not
coincident with the treatment beam (see Section 5.5.3).
● Do not treat patients with the CO2RE system if the aiming beam is not
operating.
● Always turn off the system when it is not in use (after a 10-minute
interval of inactivity, the system automatically goes into Idle mode).
● Never press the footswitch without first verifying that the handpiece is
safely oriented.
1.8.1. Keyswitch
The keyswitch ensures that unauthorized persons do not use the laser system.
When set to On the system is enabled. When set to Off the system is
disabled. To prevent unsafe and unauthorized use of the system when it is
not being used, remove the key and keep it in a secure location.
A spring-loaded, software driven safety shutter inside the console blocks the
laser beam. The safety shutter will open only when the system is in Ready
mode.
PB92971EN 1-5
Safety and Regulatory User Manual
1.8.3. Footswitch
The footswitch is the only means of emitting the CO2 laser beam, and it is
disabled except when the system is in Ready mode.
CO2 laser beam emission is enabled only when the system is set to Ready
mode and the operator presses the footswitch; accidental laser beam emission
may only occur due to a double error condition.
A self-test of the electrical circuits occurs after the system is switched on.
The test circuits continuously monitor system operation during treatment.
The CO2RE system is surge-protected by a pair of safety fuses that will de-
activate if the current consumption exceeds 7.1 A.
1-6 PB92971EN
User Manual Safety and Regulatory
The system features two laser emission indicators: one amber LED and one
red LED located on the control panel.
PB92971EN 1-7
Safety and Regulatory User Manual
● Proper labeling
Warning
No modifications to this equipment are allowed.
1-8 PB92971EN
User Manual Safety and Regulatory
Symbol Description
Caution!
CE Compliance Symbol
0344
Pushing Prohibited
Stepping Prohibited
Fuse
| Power On
Power Off
Alternating Current
Storage temperature minimum &
maximum limits
Waste of Electrical and Electronic
Equipment (WEEE) compliance
symbol
RF Interference may occur in the
vicinity of equipment with the
following symbol
Type B Equipment
Manufacturer
PB92971EN 1-9
Safety and Regulatory User Manual
FG70411
Z11371239
Series No: 01
100-240V~7.1A
POWER: 50-60 Hz
0344
1-10 PB92971EN
CHAPTER 2
Installation
2.1. Introduction
The CO2RE system is designed for installation in a clinical environment and
requires minimal site preparation. The customer carries out the installation
by doing the following:
● Unpack the system and position it in its pre-selected location.
● Verify the integrity of the system and its components.
Refer to the current revisions of the following standards regulating the safe
use of laser systems:
● In the USA: the American National Standards Institute (ANSI) Standard
Z136.3: Safe Use of Lasers in Health Care Facilities and Z136.1: Safe
Use of Lasers
● In Canada: the Canadian Standards Association (CSA) Standard
CAN/CSA-Z386-01: Laser Safety in Health Care Facilities
1
Optional Fractional handpiece may be ordered from Syneron-Candela.
PB92971EN 2-1
Installation User Manual
Input power lines should be free of transients, voltage and current spikes,
sags and surges. Consequently, the system power line should not be shared
with other heavy variable loads such as elevators, air conditioning systems,
large motors, etc.
Warning
● For continued protection against fire, replace the fuse only
with one of the same type and rating.
● Proper grounding is essential for safe operation.
● Do not replace any electric part of the system except fuses.
● Only authorized Syneron technicians are qualified to service
the system.
2-2 PB92971EN
User Manual Installation
Space should be allocated with adequate ventilation and free air flow. The
working area for the system should be prepared according to the dimensions
shown in Figure 2-1.
Note that power cable and switch are situated on the rear panel of the system.
Ensure easy and quick access to them.
”
.3
: 17 )
pth cm
De (44
Caution
● Do not leave the console unattended on a ramp with an
incline greater than 5 degrees. Wheel locks are intended to
secure the console's location on a level floor.
● The wheel locks may not be able to maintain the position of
the console on certain surfaces. When not moving, all four
wheels should be locked.
PB92971EN 2-3
Installation User Manual
Air Quality:
The system should operate in a non-corrosive atmosphere. Corrosive
materials such as acids can damage electrical wiring, electronic components
and the surfaces of optical components.
Temperature:
For optimal operation of the system, maintain room temperature between 10-
30ºC (59-86ºF) and relative humidity of up to 80%.
2.4. Installation
2.4.1. Unpacking and Inspection
The system has passed full quality assurance testing before shipment and
should be operational upon delivery.
Unpack the system's console, carefully lift it out of the shipping box and set
it on the floor.
Warning
● Any damage to the packaging or to the system found prior to
opening the packaging, or during unpacking and installation
of the system, should be immediately reported to Syneron-
Candela and to the insurance carrier.
● Severe damage to the system may occur if it is knocked
over or is allowed to fall. If such an occurrence happens
contact Syneron- Candela Service immediately.
2-4 PB92971EN
User Manual Installation
Distal Endjoint
Proximal Connector
3. Carefully set the proximal connector (A) into the system's connection
port (B).
4. Engage and tighten the threaded knob until it is firmly hand-tight (C).
A C
B
Caution
Do not power up the CO2RE system or connect it to the wall
electrical outlet if the articulated arm is not assembled on the
system.
PB92971EN 2-5
Installation User Manual
Caution
The CO2RE system should be operated only with original
Syneron- Candela accessories. Attaching any other type of
accessory may damage the system and may invalidate the
service warranty agreement.
2. Screw the proximal end of the lens assembly (A) into the articulated
arm's endjoint (B).
A B
Note
Connection of the handpiece will be discussed in Chapter 4 –
Operating Instructions.
2-6 PB92971EN
User Manual Installation
The service panel is located on the lower section of the system's rear panel
and incorporates the following (see Figure 2-5):
Connect the supplied power cable to its connection port in the system's
service panel (see Figure 2-5).
Connect the other side of the cable to the main power supply.
PB92971EN 2-7
Installation User Manual
To connect the footswitch, place it on the floor, plug its connector into the
footswitch connection port located in the service panel (see Figure 2-5) and
turn the connector clockwise to lock it.
Caution
Do not power up the CO2RE system or connect it to the wall electrical
outlet if the footswitch is not connected to the system.
The remote interlock connection on both sides – the system's connector and
the treatment room's door – should be performed by a qualified electrician.
2-8 PB92971EN
User Manual Installation
Read the entire operating manual before turning on the system. Of particular
importance is Chapter 1 – Safety.
The CO2RE system may not be operated by any person who has not
received training from a Syneron-Candela clinical representative.
Before system testing, ensure that the laser operating area is safe and secure.
As when performing any laser procedure, flammable materials should be
moistened or beyond contact of the laser beam. All personnel should wear
protective eyewear.
If at any point during the test the system does not perform as described,
discontinue use and contact Customer Service.
Before beginning the initial system test, familiarize yourself with the system
controls described below.
The main control panel is located at the top of the console and incorporates
the touch-screen LCD panel.
PB92971EN 2-9
Installation User Manual
1. Ensure that the emergency shutoff knob is not engaged (pressed in). If it
is, turn it clockwise until it releases and pops out.
2. Turn the keyswitch to the horizontal position; the system will be enabled.
3. Turn the system's main On/Off switch – located on the service panel (see
Figure 2-5) – to the On position; the software will initialize and the
system will perform a series of internal power-up routines. When these
have completed satisfactorily, the Treatment Menu screen will appear
on the LCD (see Chapter 4).
4. If the system does not perform as described, discontinue use and contact
Syneron-Candela Service. If the system performs as described, continue
on to the footswitch connection check (see Section 2.5.3).
Note
Attempting to start the system without turning the keyswitch will
generate an error message pop-up on the LCD. Follow the on-
screen instructions to clear the error.
2. Unplug the footswitch; the system should display the following error
message: Footswitch Not Connected.
3. If the system does not display the error message and remains in Ready
mode, discontinue use and contact Syneron-Candela Service.
2-10 PB92971EN
User Manual Installation
2. If the system does not display the error message and remains in operative
mode, discontinue use and contact Syneron-Candela Service.
Laser beam emission is disabled when the remote interlock plug is not
connected or is improperly connected to the service panel, even if it is not
wired to an actual door interlock. To check this:
2. If the system does not display the error message and remains in operative
mode, discontinue use and contact Syneron-Candela Service.
PB92971EN 2-11
Installation User Manual
2. Slowly push or pull the system using the handles built into the front and
rear panels, to the desired location. Use the same handles to lift the
system off the floor.
3. When moving the system, do not leave it unattended. The wheel locks
may not be able to maintain the position of the console on certain
surfaces. When not moving, all four wheels should be locked.
Caution
● Never push or pull the system by the articulated arm; the
system's optical alignment may be seriously impaired by
doing so.
● Do not push or pull the system by the touch-screen control
panel.
Warning
To avoid injury, do not transport on a ramp with an incline
greater than 5 degrees. Do not leave the console unattended on
a ramp with an incline greater than 5 degrees. Wheel locks are
intended to secure console location on a level floor.
2-12 PB92971EN
CHAPTER 3
System Description
The high degree of collimation and coherence enable the focusing of the
beam to small spot sizes.
The active (lasing) medium of a laser can be either gas, liquid or solid. Most
gas lasers consist of atoms, molecules, or mixtures of both. Solid-state lasers
consist of atoms or ions "doped" in some solid matrix. Liquid lasers consist
of higher molecular weight molecules dissolved in liquids.
Under specific pumping conditions, all these materials can undergo the
unnatural phenomenon of "population inversion" that results in stimulated
emission of radiation at a wavelength characteristic of the active medium.
PB92971EN 3-1
System Description User Manual
3-2 PB92971EN
User Manual System Description
● Cooling System – forced air fans to ensure that the laser assembly is
kept at a constant, optimal operating temperature.
PB92971EN 3-3
System Description User Manual
Articulated
Arm Assembly
Handpiece
Touch-Screen
Control Panel
Handles
System
Controls
Console
Footswitch
3-4 PB92971EN
User Manual System Description
PB92971EN 3-5
System Description User Manual
By scanning the focused beam in both X and Y axis directions, the system
will emit pin-point laser beams adjacent to each other that form a laser spot.
The energy distribution within each spot is even.
The X and Y axis scanning enables laying down several spots to form a pre-
defined pattern, where the operator can control the distance between the
spots (fractional percentage setting), as shown in Figure 3-3:
Non-ablated
Area
Figure 3-3: Beam Scanning Principle (Hexagon Scan Pattern Example with
Ring 1 Ablation Spot Option)
3.3.5. Footswitch
Laser emission occurs only when the footswitch is pressed. The footswitch is
enabled only when the system is in Ready mode. The footswitch is a water-
tight pedal with a metal guard, to prevent accidental activation.
3-6 PB92971EN
User Manual System Description
Three color-coded lens assembles are supplied with the system (see Figure
3-4). The lens is permanently fixed inside an assembly designed to keep the
lens safe and easy to handle when connecting, disconnecting and cleaning.
● The silver lens assembly delivers a pin-point laser beam diameter of
150 microns.
● The purple lens assembly delivers a pin-point laser beam diameter of
150 microns.
● The gold lens assembly delivers a pin-point laser beam diameter of
120 microns.
3.3.6.2. Handpieces
The design of the handpieces maintain the proper distance between the lens
assembly and the surface of the patient's treated site, in order to deliver
scanned patterns according to the requested energy, scan shape and size.
PB92971EN 3-7
System Description User Manual
Standard Handpiece
Surgical Handpiece
Fractional
Handpiece
(optional)
3-8 PB92971EN
User Manual System Description
Laser Type
● CO2
Wavelength
● 10,600 nm
Output Power
● Max. peak: 60 Watts
Pulse Duration
● 2 – 1166 µsec
Power Stability
● ±1.5 watts (±5%)
Beam Size
● Ø 1.8 ±0.5 mm
Beam Divergence
● 7.5 ±0.5 mrad (full angle)
Polarization
● >100:1 linear (orthogonal to mounting pads)
1
30 – 80 mJ outside the United States
PB92971EN 3-9
System Description User Manual
Deep fractional
CO2RE Deep resurfacing 30 – 80 2 170 – 4532 1–5 500 – 750
CO2RE Mid and Deep 46.3 – 129 8.9 – 24.8 100 – 200
20 – 40
Fusion modes combined 30 – 802 170 – 4532 500 – 750
Full thickness
Classic resurfacing
1 – 10 5.6 – 56.4 100 100 – 200
0.2 mm x
Incision and selected line length
Surgical excision of lesions
5 – 40 28.1 – 226 N/A
0.2 mm2 freehand
5-15W
(*) The energy settings for Light and Mid modes (30-60 mJ and 46.3-257.6 mJ,
respectively) are provided per ablated spot (and are shown on the control panel), while the
energy settings for Deep mode (50-70 mJ) are provided per pin point (per pixel).
Caution
Proceed with caution when performing advanced off-the-face
treatments, reduce parameters accordingly.
2
70 mJ / 170 – 396 J/cm2 in the United States
3-10 PB92971EN
User Manual System Description
Aiming Laser
● 5mW, 650nm (red) diode laser
Articulated Arm
● Lightweight, aluminum fiber, 7-joint, pneumatically balanced
● Treatment radius at full arm extension: 85 cm / 33.5 inches
● Horizontal rotation: 360°
Lens Assemblies
● Silver 150 µm spot size
● Purple 150 µm spot size
● Gold 120 µm spot size
PB92971EN 3-11
System Description User Manual
Handpieces
● Standard Fractional
● Intima – Internal and External
● Surgical
● Short Fractional (optional)
User Interface
● 10.4", high-resolution, color liquid crystal display (LCD)
● Touch-screen technology
System Turn-On
● On/Off switch on service panel
● Keyswitch on front panel
● Password-protected access
System Turn-Off
Electrical Requirements
● 100-240 VAC self-sensing, 50-60 Hz, 7.1 A, single phase
3-12 PB92971EN
User Manual System Description
Dimensions
● Width: 15.75" (40 cm)
● Depth: 17.3" (44 cm)
● Height at top of control panel: 46.1" (117 cm)
● Height at top of articulated arm: 70" (177 cm)
Weight
● 58 Lbs. (26 Kg)
Treatment Room
● Temperature: 15 – 30°C [59 – 86°F]
● Relative humidity: Up to 80%
● Atmospheric pressure range: 70 – 106 kPa
Storage/Transportation
● Temperature: 10 – 55°C [50 – 131°F]
● Relative humidity: 10 – 80%
● Atmospheric pressure range: 70 – 106 kPa
3.4.7. Classifications
Electromagnetic Compatibility
● Class A
Laser Classification
● Class IV
PB92971EN 3-13
System Description User Manual
3-14 PB92971EN
CHAPTER 4
Operating Instructions
4.1. Introduction
This chapter describes in detail the operating instructions for the CO2RE laser
system.
Warning
● Use of controls or adjustments, or performance of procedures
other than those specified herein may result in hazardous
radiation exposure.
● Do not operate the system without first familiarizing yourself with
all sections of this manual. Special attention should be paid to the
safety precautions in Chapter 1.
● CO2 laser radiation is invisible to the human eye and is capable of
inflicting third degree burns, of igniting flammable materials and of
being reflected by metallic and other reflecting objects.
● A sign warning of the danger of laser radiation should be placed
at the entrance to the treatment room whenever the laser is used.
● Everyone in the treatment room should wear protective eyewear.
● Never treat patients with the CO2RE system if the aiming beam is
not operating.
PB92971EN 4-1
Operating Instructions User Manual
2. Refer to Figure 4-1: connect the lens assembly to the articulated arm's
endjoint by screwing the proximal end of the lens assembly (A) into the
articulated arm's endjoint (B).
A B
3. Refer to Figure 4-2: prepare the handpiece and connect it to the lens
assembly by screwing the proximal end of the handpiece (A) onto the
lens assembly (B).
B
A
Warning
● Connection of the laser accessories not according to these
instructions can cause injury.
● Inspect the handpiece prior to use for any damage following
sterilization. Do not use handpiece if any corrosion, warping, or
corruption of the metal surface is evident.
4-2 PB92971EN
User Manual Operating Instructions
2. Turn the system's main On/Off switch – located on the service panel – to
the On position; the software will initialize and the system will perform a
series of internal power-up routines.
3. Upon turn-on, a self-test routine starts that checks out the system. During
this time the laser system will automatically run an initialization process
of its components which takes approximately one minute to complete. If
a fault is detected, the system issues an appropriate error message (see
Chapter 6). Once the system satisfactorily completes the self-test routine,
it is ready for operation.
4. After the self-test routines are complete the system will display the Main
Menu screen (see Figure 4-3). Select the CO2RE mode of operation –
CO2RE Aesthetics, CO2RE Surgical or CO2RE Intima – by pressing
the appropriate button on the left side of the screen; the selected mode's
button becomes colored.
If you selected CO2RE Aesthetics the Resurfacing menu will appear on the
right side of the screen, where you may select the desired treatment mode by
pressing the appropriate button: Light, Mid, Deep, Fusion or Classic.
PB92971EN 4-3
Operating Instructions User Manual
Mid Fractional
Fractional CO2RE Mid
Resurfacing (External)
60 – 90 Any 20 – 40
Deep Fractional
Fractional CO2RE Deep 50 – 802 Any 1–5
Resurfacing (External)
CO2RE Mid and Deep Combined 60 – 90
Fractional Any 50 – 40
Fusion (External) 50 – 802
Caution
Proceed with caution when performing advanced off-the-face
treatments, reduce parameters accordingly.
● After you have selected the desired CO2RE treatment mode the
Handpiece and Lens Confirmation screen will appear (see Figure 4-4).
4-4 PB92971EN
User Manual Operating Instructions
Figure 4-5 presents the main Treatment screen. The numbered tags in the
illustration correspond to the following numbered explanations and
definitions of the screens elements:
1. Treatment Mode Selector – the default mode when accessing this screen
is the one selected in the Treatment Menu screen (see Figure 4-3); note
the Fusion button in Figure 4-5 as an example. To change the mode press
any of the other mode buttons. To change to Classic or Surgical mode
you will have to go back to the Treatment Menu screen by pressing the
Home button.
2. Pattern Shape Selector – press the appropriate button to select the
desired scanned pattern shape; the pressed button will change color to
express its being selected.
13
6 1
4
8
3
2
15
5
16
14
9
12
PB92971EN 4-5
Operating Instructions User Manual
4-6 PB92971EN
User Manual Operating Instructions
If you selected CO2RE Surgical on the Main Menu screen (see Figure 4-3)
the screen will transition to a Lens Selection screen. Press the return-arrow
button to toggle and select the 120 µm or 150 µm lens, and then press the
button to confirm that the displayed Surgical handpiece and selected lens are
connected to the system's articulated arm; the Main Treatment screen will
appear (see Figure 4-6).
The CO2RE Surgical mode is operated basically in the same manner as the
CO2RE Aesthetics mode, but with only the parameter options appropriate
for surgical mode: Pattern Size, output energy (Core mJ) and Repeat.
The selectable shapes are: Square (■), Line ( ), Ring () and
Freehand ( ).
PB92971EN 4-7
Operating Instructions User Manual
If you selected CO2RE Intima the Women's Intimate Wellness menu will
appear on the right side of the screen, where you may select the desired
treatment mode by pressing the appropriate button: Internal or External.
Refer to Figure 4-8: press the return-arrow button to toggle and select the
External or Internal modes, confirm that you have the correct handpiece and
lens assembly connected to the system (as shown in the selected mode) and
then press the button to finalize the selection; the External or Internal
Intima Treatment screen will appear (see Figure 4-9).
4-8 PB92971EN
User Manual Operating Instructions
The CO2RE Intima mode is operated basically in the same manner as the
CO2RE Aesthetics mode, but with only the parameter options appropriate
for the Intima modes: Pattern Size, fractional coverage, output energy
(Core mJ) and Repeat.
PB92971EN 4-9
Operating Instructions User Manual
The Utilities screen may be accessed by pressing the Utilities button on the
Main Menu screen (see Figure 4-3).
1. Update Firmware – Operating this option should be done only under the
direct advice of a Syneron-Candela service representative. This requires
the use of the system's administrator password (see Figure 4-11).
2. Factory Settings – this area of the software is designed for the exclusive
use of Syneron-Candela-qualified service personnel, and is protected by a
proprietary service password.
4. License – press this button to access the operating license code input
screen (consult with your Syneron distributor).
5
1
6 2
7 8
4-10 PB92971EN
User Manual Operating Instructions
8. Home – press the Home button to return to the Treatment Menu screen
(see Figure 4-3).
2. Turn off and remove the key from the keyswitch, and store it in a secure
location to prevent unauthorized use of the laser system.
PB92971EN 4-11
Operating Instructions User Manual
4-12 PB92971EN
CHAPTER 5
Maintenance
5.1. Introduction
This chapter contains the maintenance instructions for the CO2RE laser
system. Routine maintenance may be performed by clinic staff unless
otherwise specified. Any maintenance procedure not mentioned in this
chapter must be performed only by Syneron-Candela authorized technical
personnel.
The system is designed to operate reliably without the need for operator
maintenance. However, the outer surfaces of the system should be kept clean
for medical reasons (see Section 5.3.2.).
Warning
● Unauthorized servicing or modification of this system, not
described in this manual, may expose the operator/patient to
potential electrical and laser radiation hazards.
● Improper use or adjustment of this system may invalidate the
service warranty agreement.
Warning
● The CO2RE system generates high voltages and laser energy
radiation.
● The interior of the system may be serviced only by Syneron-
Candela authorized technical personnel.
PB92971EN 5-1
Maintenance User Manual
The exterior of the system should be inspected once a week to ensure that
there are no loose cable connections, that there is no damage to the system
and that the touch-screen panel is intact. Contact Syneron-Candela Service if
any damage is found.
The external surfaces of the system (console and articulated arm) and the
footswitch should be cleaned when the system is received, and thereafter as
required by the facility's cleaning protocol.
The outer surfaces of the system and the lens assembly may be wiped clean
with a soft, lint-free cloth dipped in 70% isopropyl alcohol, or a hospital-
grade disinfectant solution such as Cidex * or equivalent.
The optical lens housed in the lens assembly may be cleaned with a soft, lint-
free cloth dipped in 99% isopropyl alcohol, or hospital-grade acetone.
Visually inspect the outer surfaces of the system and the lens to verify that it
is completely free of soil and there are no signs of degradation (e.g.
discoloration, pitting or cracked seals).
Caution
Do not allow the lens to come into contact with water or any water-
based product; dry water stains can become hot-spots during laser
emission, damaging the lens' optical coating.
● The handpiece must be cleaned with soap and water to remove the bulk
of extraneous material.
● The handpiece should then be wiped with an alcohol wipe (70% alcohol)
or soaked for two minutes in sodium hypochlorite at 500 ppm,
OR:
*
Cidex® is a product of ASP®, a Johnson & Johnson company.
5-2 PB92971EN
User Manual Maintenance
Warning
Caution
The following cleaning and sterilization instructions are designed
for the CO2RE reusable aesthetic or surgical handpieces.
PB92971EN 5-3
Maintenance User Manual
Caution
Always clean the handpiece immediately after use before stains
dry.
5. Rinse well under running water, holding the handpiece such that the
water runs through it.
†
Enzol® Enzymatic Detergent is a product of ASP®, a Johnson & Johnson company.
5-4 PB92971EN
User Manual Maintenance
Caution
Do not use flash sterilization - this may damage the handpiece
surfaces.
Warning
Ensure that all steam sterilization vessels are valid before use.
2. Place the handpiece in an autoclave pouch, and put the pouch in the
autoclave.
4. Remove the handpiece from the autoclave and allow it to cool to room
temperature.
5. Store the handpiece in a clean environment. If the sterile pouch has been
compromised in any way, re-sterilize just prior to use in order to
minimize the chance of contamination.
The CO2RE handpiece parts are allowed to be subjected to no more than ten
sterilization cycles. After the tenth use the handpiece must be properly
discarded.
PB92971EN 5-5
Maintenance User Manual
During the periodic service the system's protective panels must be opened.
Therefore, this is performed only by Syneron-Candela service personnel.
5-6 PB92971EN
User Manual Maintenance
Laser beam emission is disabled when the door interlock plug is not
connected or is improperly connected to the service panel, even if it is not
wired to an actual door interlock. To check this:
Warning
Laser beam alignment checks are extremely important for the safe
operation of your laser equipment. Do not use the laser if aiming
and treatment beams are not coincident; contact Syneron-Candela
service immediately. Misalignment of aiming and treatment beams
may result in laser exposure to non-target tissues and possible
injury.
1. Verify that all persons in the treatment room are wearing appropriate
laser safety eyewear.
6. Position the distal end of the handpiece's tip on the tongue depressor in a
perpendicular orientation.
PB92971EN 5-7
Maintenance User Manual
7. Direct the aiming beam at the center of the “X,” and press the laser
footswitch. Observe the burned pattern on the tongue depressor; it
should appear sharp and non-diffused.
8. Verify that the burn spot lies within the area of the aiming beam's red
outline (see Figure 5-1).
9. If the burn is not within the aiming beam, if the burn is unacceptable, or
if the aiming beam is not visible:
● Verify that the lens assembly is securely attached to the articulated
arm.
● It may be helpful to move the articulated arm or rotate the articulated
arm knuckles closest to the endjoint. Sometimes changing the
orientation of the articulated arm knuckles can affect aiming beam
transmission, particularly if the articulated arm is extended or moved
during a procedure.
● If the brightness of the aiming beam fluctuates greatly or if the
aiming beam is not visible, contact Syneron-Candela Service.
10. Repeat the beam alignment procedure. If the beam alignment is still
unacceptable, contact Syneron-Candela Service.
11. Press the Standby button to set the laser to Standby mode until ready for
use.
5-8 PB92971EN
User Manual Maintenance
1. Turn off the system and disconnect the power cable from the outlet.
2. Unplug the power cable from the connection port on the service panel.
4. Remove the burnt fuses from the housing and replace them with the new
ones.
5. Make sure that the fuses fit snugly within the fuse housing.
6. Insert the fuse housing into its receptacle and push it in until it snaps into
place.
7. Connect the power cable to the system and to the mains outlet.
Caution
● Make sure that the type and rating of the new fuses and the
replaced fuses match.
● Always replace both fuses at the same time.
PB92971EN 5-9
Maintenance User Manual
The CO2RE laser system can easily be moved between rooms in a medical
practice by rolling the console on the base-mounted wheels using the top-
and bottom-mounted maneuvering handles. Disconnect the footswitch and
power cable prior to moving the system between rooms.
4. Disconnect the articulated arm from the system (see Chapter 2).
5. When moving the system, do not leave it unattended. The wheel locks
may not be able to maintain the position of the console on certain
surfaces. When not moving, all four wheels should be locked.
Caution
Do not ship the system without the factory packaging materials.
Doing so may result in damage to the components during shipping
and void the warranty. Contact Syneron-Candela if packaging
materials or repacking instructions are required.
Warning
To avoid injury, do not transport on a ramp with an incline greater
than 5 degrees. Do not leave the console unattended on a ramp
with an incline greater than 5 degrees. Wheel locks are intended to
secure console location on a level floor.
5-10 PB92971EN
CHAPTER 6
Troubleshooting
6.1. Introduction
The CO2RE laser system is equipped with self-testing software that
continuously monitors system operation. If a system malfunction is detected,
an error message will appear on the LCD display screen. Should a
malfunction occur, consult the troubleshooting guides in Table 6-1 and
Table 6-2.
Caution
Improper use or adjustment of this system may invalidate the
service warranty agreement. Please contact your authorized
Syneron-Candela representative before attempting to troubleshoot
this system in any manner other than those specified in this
manual.
Warning
The system generates high electrical voltage and laser radiation in
the main cabinet and laser aperture. Only Syneron-Candela-
authorized technical personnel are qualified to service the interior
of the system.
PB92971EN 6-1
Troubleshooting User Manual
A system error message will be displayed on the control panel's LCD screen
as a popup, shown in the sample in Figure 6-1:
Table 6-2 lists some possible symptoms that indicate malfunctions that do
not appear on the LCD panel. If the corrective action listed in the table does
not solve the problem, contact Syneron-Candela Service.
6-2 PB92971EN
User Manual Troubleshooting
PB92971EN 6-3
Troubleshooting User Manual
6-4 PB92971EN
APPENDIX A
Clinical Guide
Aesthetic & Surgical Applications
A.3. Contraindications
● Pregnant and/or breastfeeding.
● Permanent implant in the facial skin area, such as an injected chemical
substance.
● History of diseases stimulated by heat, such as recurrent Herpes
Simplex in the treated area, unless treatment is conducted following a
prophylactic regimen.
● Patients with vitiligo and melasma are at risk; ablative fractional laser
application may aggravate the skin condition.
● Current active acne.
● Use of non-steroidal anti-inflammatory drugs (NSAIDS, e.g.,
ibuprofen- containing agents) one week before and after each treatment
session.
● Use of retinoids, antioxidants or therapeutic skin nourishing
supplements within 2 months of treatment or during the study and oral
retinoids within 6 months of treatment.
● Having received a facial dermabrasion or chemical peel treatment
within 3 months of treatment or during the study.
● Having received treatment with light, RF or other devices in the treated
area within 6 months of treatment.
PB92971EN A-1
Clinical Guide – Aesthetic & Surgical Applications User Manual
A-2 PB92971EN
User Manual Clinical Guide – Aesthetic & Surgical Applications
Caution
In case of uncertainty regarding potential side effects, have the
patient consult his/her physician and bring a written consent for
treatment.
Note
Asian skin-type patients – after spot testing enforce a longer
"waiting period" than apparent skin type.
PB92971EN A-3
Clinical Guide – Aesthetic & Surgical Applications User Manual
Warning
Results of test spots cannot predict possible long-term adverse
effects.
Warning
CO2RE's fully ablative Classic and Surgical modes carry the same
risks as other fully ablative systems present including the potential
for scarring and hyper/hypopigmentation. Use of these modes
should be confined to small areas or lesions. Caution should be
exercised to reduce the potential for side effects.
A-4 PB92971EN
User Manual Clinical Guide – Aesthetic & Surgical Applications
● Inform the patient about the treatment protocol, typical treatment results
and possible adverse effects and discomfort.
● The patient should arrive for treatment with clean skin. There should be
no lotion, make-up, perfume, powder or bath/shower oil present on the
skin in the area to be treated.
● It is suggested to take a photo of the treated area before the treatment and
at the follow-up visit to assess clinical response.
Warning
To reduce the risk of infection and cross contamination, a recently
sterilized treatment handpiece should be used.
If a system malfunction occurs during treatment and CO2 laser
emission cannot be stopped, aim the handpiece to a neutral target
and press the emergency stop button.
PB92971EN A-5
Clinical Guide – Aesthetic & Surgical Applications User Manual
8. The operator may choose between the available ablative modes (CO2RE
Light, CO2RE Mid, CO2RE Deep, and CO2RE Fusion) used for
fractional skin resurfacing and treatment of lines and wrinkles, and the
fully ablative Classic resurfacing and Surgical incision/excision modes.
9. The operator can further choose between five treatment patterns
(hexagon, triangle, square, rectangle or ring) for the fractional treatment
modes and a smaller number of shapes lines and freehand patterns for the
fully ablative Classic and Surgical modes.
10. Prior to applying laser pulses to the patient, confirm spot pattern and
laser alignment. Perform a test spot using a tongue depressor moistened
with water. Make sure that pattern is full and the same as on the screen
(complete laser alignment test instructions may be found in Chapter 5 –
Maintenance).
11. The operator’s decisions are determined when taking into consideration
not only the severity of the treated condition, but also the specific
anatomical location and proximity to bones, nose, eyelids, etc.
Warning
CO2RE's digital SuperPulse laser easily produces the desired
treatment effect without the post-pulse heat deposition and other
characteristics common to prior generation SuperPulse and CW
mode lasers. To prevent the occurrence of unwanted side effects,
do not attempt to reproduce with CO2RE the profound redness,
tissue char or other treatment endpoints common to these older
systems.
A-6 PB92971EN
User Manual Clinical Guide – Aesthetic & Surgical Applications
13. Treatment consists of a single pass over the designated area. A second
pass may be performed on crow’s feet, wrinkles or shoulders of rhytids.
14. Unless otherwise instructed, the energy parameters can be modified
between the treatment visits, as per the physician's discretion.
Deep fractional
CO2RE Deep 30 – 70 * 170 – 396* 1–5 500 – 750
resurfacing
CO2RE Combination of Mid 46.3 – 129 8.9 – 24.8 100 – 200
20 – 40
Fusion and Deep modes 30 – 70* 170 – 396* 500 – 750
Full thickness
Classic resurfacing
1 – 10 5.6 – 56.4 100 100 – 200
0.2 mm x
Incision and selected line length
Surgical excision of lesions
5 – 40 28.1 – 226 N/A
0.2 mm2 freehand
5-15W
(*) The energy settings for Light and Mid modes (30-60 mJ and 46.3-257.6 mJ,
respectively) are provided per ablated spot (and are shown on the control panel), while the
energy settings for Deep mode (50-70 mJ) are provided per pin point (per pixel).
Caution
Proceed with caution when performing advanced off-the-face
treatments, reduce parameters accordingly.
Proceed with caution when treating skin types IV and V, keeping fractional
density below 20%.
*
30 – 80 mJ / 170 – 453 J/cm2 outside the United States
PB92971EN A-7
Clinical Guide – Aesthetic & Surgical Applications User Manual
discomfort longer than that, cool/cold compresses (wet cold gauzes) may
be applied to relieve it. Gauzes should be kept cold using cold saline
solution. Post-treatment emollient cream may be applied over the
treatment area.
● On the night following treatment, patients should wash the treated area
gently with lukewarm water and avoid hot water in these areas. The skin
should be kept very clean to avoid infection; any mechanical or thermal
damage to the area must be avoided.
● Moisturizer can be applied 1–2 days after each treatment and then should
be applied regularly throughout the course of the treatment. Make-up may
be applied only 3–5 days after each treatment, if desired, unless an
adverse event occurs in the area. Generally patients may use regular
soaps, but not scrub soaps or exfoliants. The patient should wait 1–2 days
or until the scabs shed off before shaving. At the beginning of the
treatment, the physician should review the patient's routine skin care
regimen and may suggest modifications if the patient uses products that
are not recommended for use during the course of the treatment.
● Tanning of any sort (sun exposure, tanning beds, and artificial sunless
tanning lotions) is not allowed in the treated areas during the entire course
of the treatment since it might cause hyperpigmentation.
● Patients should be instructed to use a high factor sunscreen with SPF of at
least 30, and to protect the treated area from direct sunlight for the entire
course of the treatment.
A.11. Follow-Up
The following are recommendations for post-treatment follow-up. They may
serve as a basis for defining your treatment regimen.
● The patient should return for follow up visits 4-7 days after treatment, 4-6
weeks and 12 weeks after treatment for evaluation of the treatment area.
● Examine the treatment area and evaluate the results.
● Update the patient history while being mindful of new medications, etc.
● If the patient has experienced adverse events, they should be treated
according to physician discretion.
A-8 PB92971EN
APPENDIX B
Clinical Guide – Fractional Vaginal,
Vulval and Introitus Treatments
The Intima is intended for the treatment of vulvovaginal atrophy (VVA) and
vaginal rejuvenation.
B.3. Contraindications
● Pregnant and/or breastfeeding.
● Abnormal cell cytology.
● Evidence of dysplasia and occult or active infection.
● Injury or bleeding in external vaginal area.
● Undiagnosed vaginal bleeding.
● History of diseases stimulated by heat, such as recurrent Herpes Simplex
in the treated area, unless treatment is conducted following a prophylactic
regimen.
● Active malignancy or history of malignancy in the past five years.
● Active electrical implant anywhere in the body, such as a pacemaker or
an internal defibrillator.
● Suffering from significant concurrent illness, such as cardiac disorders,
diabetes (type I or II), lupus, porphyria or pertinent neurological
disorders (i.e., any disease state that in the opinion of the physician
would interfere with the treatment or healing process).
● Having a known anti-coagulative or thromboembolic condition or taking
anticoagulation medications one week prior to and during the treatment
PB92971EN B-1
Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments User Manual
course (to allow inclusion, temporary cessation of use as per the patient
physician's discretion).
● History of immunosuppression/immune deficiency disorders (including
HIV infection or AIDS) or currently using immunosuppressive
medications.
● Suffering from hormonal imbalance, whether related to thyroid, pituitary,
or androgen.
● History of significant lymphatic drainage problems.
● History of cancer which required lymph node biopsy or dissection.
● Suffering from significant conditions in the treated areas or inflammatory
conditions, including but not limited to open lacerations or abrasions of
the treatment area prior to treatment or during the treatment course.
● History of keloid scarring, abnormal wound healing and/or prone to
bruising.
● History of epidermal or dermal disorders (particularly if involving
collagen or microvascularity), including collagen vascular disease or
vasculitic disorders.
● Use of isotretinoin (Accutane®) within six months of treatment or during
treatment.
● On systemic corticosteroid therapy six months prior to and throughout
the course of treatment.
● Dysplastic nevi in the area to be treated.
● Surgery or any other procedure for Vaginal Tightening in the last
12 months.
Warning
Do not treat the patient if there are signs of injury, laceration or
bleeding in the external or internal vaginal area.
B-2 PB92971EN
User Manual Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments
These are the side effects that may appear, but are not limited to, in the
treatment area:
● Ecchymosis
● Purpura
● Bruising
● Swelling
● Twinge (pain)
● Numbness
● Infection
● Blisters
● Itching
● Scarring
● Burns
● Vulvodynia
B.6. Pre-Treatment
● Determine if the patient can be treated and has no contraindications for
treatment.
● Inform the patient about the treatment protocol, typical treatment results
and possible adverse effects and discomfort.
PB92971EN B-3
Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments User Manual
● The patient should arrive for treatment with clean skin. There should be
no lotion, make-up, perfume, powder or bath/shower oil present on the
area to be treated.
● The patient should be fully undressed from the waist down and be
comfortable on the treatment table in a dorsal lithotomy position to
provide clear exposure to the treatment area.
● The internal treated area (vestibule, introitus and vaginal canal) should be
thoroughly washed with disinfecting solution.
● The disinfecting solution should be carefully dried from the treated area
and removed from the mucosa. This can be done with folded gauze and
ringed forceps.
● Test spot results will indicate if the patient can tolerate the treatment
without developing adverse effects. According to patient tolerance to the
test spot, you may determine treatment parameters and if anesthesia is
required (generally not). In addition, it is advisable to perform a test spot
whenever changing parameters.
B-4 PB92971EN
User Manual Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments
● With use of the different handpieces, both the vulva (vestibule and
introitus – external) and the vagina (vaginal canal – internal) can be
treated during the same procedure.
● The handpiece must be cleaned with soap and water to remove the bulk
of extraneous material.
● The handpiece should then be wiped with an alcohol wipe (70% alcohol)
or soaked for two minutes in sodium hypochlorite at 500 ppm,
OR:
PB92971EN B-5
Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments User Manual
● Press the footswitch to release a single pulse (pulse #1). Retract the
handpiece 1 cm posterior towards the vulva (external) area and release an
additional single pulse.
● Rotate the handpiece by 45-90° and reinsert in the vaginal canal. Release
a pulse and continue with a single pulse after each 1 cm retraction.
● Repeat with 4 – 8 rotations until handpiece has been rotated a full 360°.
● After completion of the final pass, fully retract the probe and cage.
● The distal part of the applicator should be clean and dry. The applicator
should be held perpendicular to and in close contact with the skin. In the
same manner, the distal end of the applicator should be moved to the
designated spot.
● The choice of treatment settings should take into consideration the
patient's skin type, tendency to bruise, the specific anatomical location
(vulva) and the necessary ablation depth (which is influenced from the
amount of water in the vaginal tissue).
● Use a cloth to gently clean the exterior surfaces with medical grade
alcohol. Allow the handpiece to air-dry until all the alcohol evaporates.
● Select the ablation mode to CO2RE Deep.
● Select the desired pattern shape, size, energy and density.
● Once the laser operating parameters have been selected and the
handpiece positioned at the proper treatment area, press the footswitch to
activate the laser treatment.
● Treatment should be single-pass over the designated area and non-
overlapping.
B-6 PB92971EN
User Manual Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments
PB92971EN B-7
Clinical Guide – Fractional Vaginal, Vulval and Introitus Treatments User Manual
B.11. Follow-Up
The following are recommendations for post-treatment follow-up. They may
serve as a basis for defining the treatment regimen.
● The patient should return for follow-up visits 4-7 days after treatment,
4-6 weeks and 12 weeks after treatment for evaluation of the treatment
area.
● Examine the treatment area and evaluate the results.
● Update the patient history while being mindful of new medications, etc.
● If the patient has experienced adverse events, they should be treated
according to physician discretion.
B-8 PB92971EN
APPENDIX C
Electromagnetic Compliance
The CO2RE system must emit electromagnetic energy in order to perform its
intended function. Nearby electronic equipment may be affected.
The CO2RE system is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes, provided the following warning
statement is heeded:
Warning
This system is intended for use by health care professionals only.
This system may cause interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigating measures,
such as re-orienting or relocating the system or shielding the
location.
PB92971EN C-1
Electromagnetic Compliance User Manual
Note
UT is the AC mains voltage prior to application of the test level.
C-2 PB92971EN
User Manual Electromagnetic Compliance
IEC 60601
Immunity Test Compliance Guidelines
Test Level
d =1.17 * SQRT(P) 80 MHz to 800 MHz
d =2.33 * SQRT(P) 800 MHz to 2.5 GHz
(1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CO2RE system is
used exceeds the applicable RF compliance level above, the CO2RE system should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the CO2RE system.
(2) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Note
● At 80 and 800 MHz, the higher frequency range applies.
● These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
PB92971EN C-3
Electromagnetic Compliance User Manual
For transmitters rated at a maximum output power not listed below, the
recommended separation distance (d) in meters (M) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to
the transmitter's manufacturer.
Note
At 80 and 800 MHz, the higher frequency range applies.
C-4 PB92971EN