Professional Documents
Culture Documents
Keenan Miller2010Aninexpensivelow-maintenanceriverinesystems
Keenan Miller2010Aninexpensivelow-maintenanceriverinesystems
net/publication/250020055
CITATION READS
1 3,159
2 authors:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Steven J. Miller on 21 November 2014.
593
594 KEENAN AND MILLER
FIGURE 1.—Schematic diagram of the sensor signal conditioning circuit using LM1077 operational amplifiers (picofarad).
cycle; (4) the circuit should not interfere with other circuit layout file used in our work can be obtained by
sensors connected to the data logger; (5) the circuit contacting either author. This copy can be submitted
should be stable over long periods and large temper- directly to ExpressPCB, Inc., to order finished circuit
ature ranges; (6) the circuit should be small enough to boards ready for parts placement. The completed
fit inside the data logger’s weather-resistant case; and circuit boards are tested by measuring output voltage
(7) the circuit should be easily assembled out of readily with an input of approximately 40 mV and with the
available, inexpensive parts. inputs shorted. Because of the small power require-
An integrated circuit operational amplifier (Op Amp) ments, the voltages needed to power the circuit and run
is an appropriate choice to begin design of a stable, the tests can be developed using a rheostat and a 3-V
linear, floating differential amplification circuit. We battery to power the circuit and another rheostat to
chose a precision Op Amp LT-1077 (Linear Technol- produce the 40-mV input. The 40-mV input should
ogy Corp., Milpitas, California) for various suitable read approximately 2 V or about 50 times the actual
characteristics, particularly its ability to be powered by input voltage. The output should read approximately 16
the data logger’s 2.5-V sensor read signal and its ability mV with the input shorted. If either reading is
for output to swing from nearly 0 to 2.5 V, almost the incorrect, check for solder tracks between component
entire measurement range of the data logger. Two Op solder joints, poorly soldered joints, and misplaced
Amps were used to meet the impedance and amplifi- components. These problems are the cause of nearly
cation requirements. A schematic of the amplification every failure and are easily repaired.
circuit is supplied (Figure 1). All resistors were metal- The sensor wires were sealed by using the watertight
film–type rated at 0.125 W (accuracy, 61%). All parts compression glands provided with the data logger
were readily available by order from Digi-key enclosure. The DO sensor-circuits were connected to
Corporation (www.digikey.com). channel 4 of the data loggers. This can be hardwired
The circuit design is relatively simple and layout of a directly or by way of a 2.5-mm stereo plug, the wiring
circuit board from the schematic should be easy for order being trigger signal, positive output, and ground
anyone with moderate electronics experience. We used from tip to base of plug. Detailed color pictures of
free layout software from ExpressPCB, Inc. (www. hardwire attachment points for channel 4 can be
expresspcb.com), which greatly simplified the task. obtained from either author by e-mail. The data loggers
This software also allowed direct Internet ordering of were also set to record temperature on channel 1, using
custom-printed circuit boards with predrilled holes for temperature sensors strapped to the side of each DO
component mounting, an epoxy solder mask, and silk- sensor housing (TMx-HD Soil/Water temperature
screened labeling for parts placement. Using a custom sensor; Onset Computer Corp.).
layout and a solder mask allowed the entire circuit to be For field applications, DO sensors equipped with
constructed on a compact 1.1 cm 3 6 cm board that fit signal conditioning circuits and connected to data
within the data logger’s weather-resistant housing on loggers were deployed as anchored floating units in the
top of the its circuit board. An electronic copy of the upper reaches of the St. Johns River, Florida, USA.
NOTE 595
tures ranging from 58C to 338C with units previously approximately 0.1 mg/L higher than the constructed
calibrated at 218C. units did. With only mild mixing using the air pump,
The sensitivity of the DO sensors to flow was measured DO differed by less than 0.1 mg/L between
assessed by operating four units simultaneously with a any of the instruments. Although flows in the mild
YSI multiparameter sonde (Yellow Springs Instru- mixing regime were so low they were hard to observe
ments, Yellow Springs, Ohio) in the laboratory. Tests or accurately measure, this was apparently enough to
were conducted with the DO and temperature sensors allow accurate DO measurements, even though the
and the YSI sonde suspended in a 113-L aquarium. The manufacturer stipulates a 2 cm/s minimum flow. The
sensitivity to flow was compared in three different flow DO probes may be more sensitive to low flows with
regimes: a vigorous mixing regime in the tank low DO concentrations or high temperatures but this
maintained by a 120-V AC submersible fountain pump was not tested in the laboratory.
rated for 150 L/min; a mild mixing regime using only
an aquarium air pump and an air stone; and a stagnant, Field Tests
no-flow regime. The vigorous flow regime had Measurements between constructed units, YSI
minimum observed flows between DO sensors esti- recording sondes, and spot checks during service visits
mated at 5 cm/s and maximum flows near the bottom generally showed good agreement. Constructed units
of the tank greater than 30 cm/s. The mild mixing had a operational period of at least 4 months 82% of
regime had flows visually estimated at 4 cm/s in the the time with approximately 93% of the data passing
immediate vicinity of the bubble stream but only initial quality assurance requirements. Units rarely had
between 0.5 and 0.1 cm/s next to the DO probes. differences larger than 0.25 mg/L DO for data that
Visual measurement of the velocity of particulates in passed quality assurance.
the water could not reliably discriminate by better than
0.5–0.1 cm/s. In the static flow test, no water Discussion
movement was visible in the tank. Dissolved oxygen measurements over wide temper-
Field tests.—Several long-term deployments were ature ranges generally are not recommended by meter
completed between 2003 and 2007, generally in pairs manufacturers because of inaccuracy under those
at seven major locations. All deployments used 15-min conditions. Calibration near the expected sample
intervals for measurements. Several included simulta- temperature is recommended. Still, the results for our
neous deployment of YSI recording sondes. Data were units indicate that they are broadly, if not universally,
downloaded and the probe’s membrane was wiped applicable used with simple linear calibrations tech-
clean every 2 weeks. During each service visit, the DO niques. Because less than a 1% difference from
was measured independently with a freshly calibrated expected values was observed between 118C and
YSI sonde. The units were retrieved after 4 to 6 months 338C in the temperature tests, we found that application
to replace the desiccant and battery in the data logger of a simple linear conversion factor at local observed
and the membrane and electrolyte in the DO probe. water temperatures from 138C to 348C provides a good
prediction of DO levels in our system. Still, the use of
Results this design in rivers where temperatures fall below
Temperature and Flow Regime Tests 108C may require additional calibration.
Less than a 1% difference from expected values was The sensitivity of the DO probe to flow varies with
observed between for temperatures between 118C and temperature and oxygen concentration. The effective-
338C. At temperatures below 148C, DO measurements ness of bulk water flow past a probe is affected by the
began to underestimate the expected values, differing probe’s shape and orientation and by nearby structures.
by more than 3% at 88C and by more than 5% at 68C. Our trials indicated probe results in very low flow
Differences less than 0.5% were observed as temper- could be effective, but this might be very site-specific.
atures exceeded 308C Furthermore, we believe that having the units moored
In all flow regimes, the constructed DO units had as a floating platform allowed even small surface
nearly identical responses. However, in the stagnant waves to reduce the sensitivity to low flow by
regime, the YSI with its flow-insensitive pulsed DO providing vertical movement.
probe, measured DO concentrations nearly 2 mg/L A 4–6-month service interval is likely to be similar
higher than the constructed units. This probably for all situations when 15-min readings are taken;
resulted from the constant consumption of oxygen by however, the membrane cleaning and data downloads
the galvanic probes and the development of an oxygen- are more site-specific. Our units, which were in waters
depleted zone at the membrane surface of the probes. with high organic loads, would build up organic
With vigorous mixing, the YSI measured DO at coating on the Teflon membranes over a period of
NOTE 597
several weeks. Additionally, given the potential effects and Bond 1999). Limitations include the need for
of alligator bites, boat strikes, the occasional hurricane, complicated temperature compensation techniques in
and rafts of floating vegetation, we were reluctant to coldwater systems and the addition of a stirrer or
chance losing more than 2 weeks worth of data. Service agitator for very-low-flow or still waters, which would
intervals could be much longer under less demanding entail significantly greater costs.
circumstances. Because the units were fairly inexpen-
sive to construct, they could be deployed in pairs; References
problems with a unit were immediately obvious when APHA (American Public Health Association), American
the measurements from the two units differed from Water Works Association, and Water Environment
each other or from the spot check measurement taken at Federation. 1998. Standard methods for the examination
the service interval. of water and wastewater, 20th edition. APHA, Wash-
In conclusion, this design provides an inexpensive ington, D.C.
(,$700) alternative to commercial units for collecting Mackereth, F. J. H. 1964. An improved galvanic cell for
frequent and remote recorded DO measurements. It is determination of oxygen concentrations in fluids. Journal
easy to construct Even persons with little electronics of Scientific Instruments 41:38–41.
Phillips, N., and B. J. Bond. 1999. A micropower precision
knowledge can easily construct this device, which is
amplifier for converting the output of light sensors to a
easy to calibrate, rugged, and dependable and requires
voltage readable by miniature data loggers. Tree
little maintenance during long deployments in harsh Physiology 19:547–549.
environments. In addition, because two A–D channels Weiss, R. 1970. The solubility of nitrogen, oxygen, and argon
remain unused on the data logger after DO and in water and seawater. Deep-Sea Research 17:721–35.
temperature sensors are attached, use of additional Wetzel, R. G. 2003. Limnology, lakes, and river ecosystems,
sensors such as light sensors may be possible (Phillips 3rd edition. Academic Press, San Diego, California.