Verification of Analytical Measurement Ranage on HORIBA YUMIZEN H550

Dated: 10 March,2022 1:16 1:8 1:4 1:2


Specimen concentration -> 6.25% 12.5% 25.0% 50.0% 100%
0.0% ### ### ###
###
###
### ###
###
###
###
###
###
###
###
### ###
###
###
###
###
###
###
###
###
###
###
###
###
###
###
###
###
###
###
Range Parameters Unit Value obtained after dilution of concentrated specimen
0.5 -500 TLC /cumm 1,980.00 3920.00 7,900.00 15,150.00 31120.00
3-250 HGB mill/cumm 1.9 2.90 5.70 10.80 21.10
PLATLET gm/ dl 43.00 77.00 139.00 295.00 550.00

AMR Graph TLC Linear (TLC)

35000.0

30000.0

25000.0

20000.0

15000.0

10000.0

5000.0

0.0
7.5

100
0.0

2.5
6.3
5.0

17.
20.
22.

27.

32.

42.
45.

50.

55.
57.
60.

65.
67.

72.

77.

82.
85.
87.

95.
12.
15.

25.

30.

35.
37.
40.

47.

52.

62.

70.

75.

80.

90.
92.

97.
AMR Graph HGB Linear (HGB)

25.0

20.0

15.0

10.0

5.0

0.0
0.0

5.0

100
2.5
6.3

7.5
12.
15.

20.
22.

27.
30.

35.
37.

42.
45.

52.

57.
60.
62.

67.
70.

75.
77.

82.
85.

90.
92.

97.
17.

25.

32.

40.

47.
50.

55.

65.

72.

80.

87.

95.

AMR Graph PLATLET


Linear (PLATLET)
600.0

500.0

400.0

300.0

200.0

100.0

Remarks:
0.0
1

As per the graphs above it seems that the instrument is giving the value in the linear range as per the equipment specification.
The straight line drawn from '0' to the maximum value is more or less touching all points obtained by dilution of the specimen.
This proves that the linear range as specified by the equipment manufacturer is OK and acceptable.

Tech.sign Approved By