Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

T 296 METHODS OF SAMPLING AND TESTING

Standard Method of Test for

Unconsolidated, Undrained Compressive Strength of Cohesive Soils in Triaxial Compression

AASHTO DESIGNATION: T 296-94

1. SCOPE T 207 Method for ThinWalled Tube 3.2.3 failure-the stress condition at
Sampling of Soils failure for a test specimen. Failure is
T 208 Method for Unconfined often taken to correspond to the
1.1 This test method covers the maximum principal stress difference
determination of unconsolidated strength Compressive Strength of
Cohesive Soil (maximum deviator stress) attained or the
and stress-strain relationships for a principal stress difference (deviator
cylindrical specimen of either an T 297 Method for Consolidated- stress) at 15-percent axial strain,
undisturbed or remolded cohesive soil Undrained Triaxial Compre- whichever is obtained first during the
sheared undrained in compression at a ssion Test on Cohesive Soil performance of a test. Depending on soil
constant rate of axial deformation (strain 2.2 ASTM Standards: behavior and field application, other
controlled). suitable failure criteria may be defined,
D 422 Method for ParticleSize
1.2 The test method provides for the Analysis of Soils such as maximum effective stress
calculation of total and effective stresses obliquity, σ1/σ3, or the principal stress
on, and axial compression of, the test D 653 Terminology Relating to Soil,
Rock, and Contained Fluids difference (deviator stress) at a selected
specimen by measurement of axial load, axial strain other than 15 percent.
axial deformation, and pore-water D 2216 Method for Laboratory
Determination of Water 3.2.4 neutral stress (pore pressure)
pressure.
Content of Soil, Rock, and (pore water pressure); µ-stress
1.3 The test provides data useful in transmitted through the pore water filling
determining strength and deformation SoilAggregate Mixtures
the voids of a soil mass.
properties of cohesive soils much as D 2487 Test Method for
Classification of Soils for 3.2.5 null indicator-manufactured
Mohr strength envelopes and Young's
Engineering Purposes device used in conjunction with a
modulus. Generally, three specimens are
pressure gage to measure pore pressure in
tested at different effective confining D 4220 Practice for Preserving and a closed hydraulic system.
pressures to define a strength envelope. Transporting Soil Samples
3.2.6 principal stress difference
1.4 The determination of strength 2.3 Other Standards and References: (deviator stress)-The difference in
envelopes and the development of
2.3.1 USBR 5754 Procedure for magnitude between the major principal
relationships to aid in interpreting and
Unconsolidated-Undrained Triaxial Shear stress and minor principal stress; in a
evaluating test results are left to the
Testing of Soils (see EARTH MANUAL, triaxial compression test, the stress due to
engineer or office requesting the test.
PART 2). the axial load which is applied in excess
1.5 The values stated in either SI or non- of the confining pressure.
2.3.2 Joseph E. Bowles, Engineering
SI units shall be regarded separately as
Properties of Soils and their Measure- 3.2.7 total stress, σ the total force per
standard. The values in each system may
ment, McGraw-Hill Book Company, unit area acting within a soil mass. It is
not be exact equivalents, therefore, each
New York, pp. 158, 3d Edition. the sum of neutral and effective stresses
system must be used independently of the
other, without combining values in any (σ’ + µ).
way. 3. TERMINOLOGY
1.6 This standard may involve hazardous 4. SIGNIFICANCE AND USE
materials, operations, and equipment. 3.1 Definitions-The definitions of terms
This standard does not purport to used in this test method shall be in 4.1 The strength in this test is measured
address all of the safety problems accordance with Terminology D 653. under undrained conditions and is
associated with its use. It is the
3.2 Description of Terms Specific to this applicable to field conditions where soils
responsibility of the user of this standard
Standard: are subjected to a change in stress
to establish appropriate safety and health
3.2.0 Backpressure-A pressure applied without time for consolidation to take
practices and determine the applicability
to the specimen potawater to cause air in place (unconsolidated condition), and the
of regulatory limitations prior to use.
the ph space to pass into solution in the field stress conditions are similar to those
potawater, that is, to saturate the in the tests.
2. REFERENCED DOCUMENTS specimen. 4.2 Using the pore-water pressure
3.2.1. differential pressure transducer-a measured during the test, the shear
2.1 AASHTO Standards: commercially manufactured, direct strength determined from this test can be
reading device used to measure pore expressed in terms of effective stress.
T 89 Method for Determining the
liquid Limit of Soils pressure in a closed hydraulic system. 4.3 The shear strength determined from
3.2.2 effective stress (effective the test expressed in terms of total
T 90 Method for Determining the stresses or effective stresses is commonly
Plastic Limit and Plasticity pressure) (intergranular pressure), σ-the
average normal force per unit area used in embankment stability analyses,
Index of Soils earth pressure calculations, and
transmitted from grain to grain of a soil
T 100 Method for Specific Gravity foundation design.
mass. The stress that is effective in
of Soils
mobilizing internal friction (σ - µ).

710
T 296 METHODS OF SAMPLING AND TESTING

5. APPARATUS 0.1 percent of the axial load at failure and V = the total volume of the specimen,
so there is negligible lateral bending of in.3 (mm3), and
the piston during loading. ∆u = the change in pore pressure, psi
5.1 The requirements for equipment
needed to perform satisfactory tests are (kPa).
given in the following sections. NOTE 2-The use of two linear ball bushings
to guide the piston is recommended to
5.2 Axial Loading Device-The axial minimize friction and maintain alignment.
NOTE 4-To meet the rigidity requirement,
compression device may be a screw jack tubing between the specimen and the
NOTE 3-A minimum piston diameter of 1/6 measuring device should be short and thick
driven by an electric motor through a
the specimen diameter has been used walled with small bores. Thermoplastic,
geared transmission, a hydraulic or successfully in many laboratories to minimize copper, and stainless steel tubing have been
pneumatic loading device, or any other lateral bending. used successfully in many laboratories.
compression device with sufficient
capacity and control to provide the rate of
axial strain (loading) described in Section 5.6 Chamber Pressure Control Device- 5.9 Deformation Indicator-The vertical
8.2.6. The rate of advance of the loading The chamber pressure control device deformation of the specimen is usually
device should not deviate by more than ± shall be capable of applying and determined from the travel of the piston
1 percent from the selected value. controlling pressure to within ± 0.25 psi acting on the top of the specimen. The
Vibration due to the operation of the (2 kPa) for chamber pressures less than piston travel shall be measured with an
loading device shall be sufficiently small 28 psi (200 kPa) and to within_* 1 accuracy of at least ± 0.02 percent of the
to not cause dimensional changes in the percent for chamber pressures greater initial specimen height. The deformation
specimen or to produce changes in pore- than 28 psi (200 kPa). The device may indicator shall have a travel range of at
water pressure when the drainage valves consist of self-compensating mercury least 20 percent of the initial height of the
are closed. pots, pneumatic pressure regulator, or specimen and may be a dial indicator,
any other device capable of applying and linear variable differential transformer
controlling pressures to the required (LVDT), extensiometer, or other
NOTE 1-A loading device may be judged to tolerances.
produce sufficiently small vibrations if there measuring device meeting the
are no visible ripples in a glass of water placed 5.7 Chamber Pressure Measurement requirements for accuracy and range.
on the loading platform when the device is Device-The chamber pressure measuring 5.10 Specimen Cap and Base:
operating. device shall be capable of measuring
5.10.1 For Total Stress Determination-
pressures to the tolerances given in
An impermeable rigid cap and base shall
Section 5.6. It may be a Bourdon gage,
5.3 Axial Load-Measuring Device-The be used to prevent drainage of the
pressure manometer, electronic pressure
axial load-measuring device may be a specimen. The specimen cap and base
transducer, or any other device capable of
load ring, electronic load cell, hydraulic shall be constructed of a noncorrosive
measuring pressures to the stated
load cell, or any other load-measuring impermeable material, and each shall
tolerances.
device capable of the accuracy prescribed have a circular plane surface of contact
in this paragraph and may be a part of the 5.8 Pore-Water Pressure Measurement with the specimen and a circular cross
axial loading device. The axial load- Device-The specimen pore-water section. The weight of the specimen cap
measuring device shall be capable of pressure shall be measured to the shall produce an axial stress on the
measuring the axial load to an accuracy tolerances given in Section 5.6. During specimen of less than 0.145 psi (1
of within ± 1 percent of the axial load at undrained shear, the pore-water pressure kN/m2). The diameter of the cap and base
failure. If the load-measuring device is shall be measured in such a manner that shall be equal to the initial diameter of
located inside the triaxial compression as little water as possible is allowed to go the specimen. The specimen base shall be
chamber, it shall be insensitive to into or out of the specimen. To achieve coupled to the triaxial compression
horizontal forces and to the magnitude of this requirement, a very stiff electronic chamber so as to prevent lateral motion
the chamber pressure. pressure transducer or nullindicating or tilting and the specimen cap shall be
device must be used. With an electronic designed to receive the piston such that
5.4 Triaxial Compression Chamber-The
pressure transducer the porewater the pistonto-cap contact area is
triaxial chamber must be able to
pressure is read directly. With a null- concentric with the cap. The specimen
withstand a chamber pressure equal to the
indicating device a pressure control is cap during shear shall not tilt more than
sum of the effective confining pressure
continuously adjusted to maintain a 5°. The cylindrical surface of the
and back pressure. It shall consist of a top
constant level of the water/mercury specimen base and cap that contacts the
plate and a baseplate separated by a
interface in the capillary bore of the membrane to form a seal shall be smooth
cylinder. The cylinder may be
device. The pressure required to prevent and free of scratches.
constructed of any material capable of
movement of the water is equal to the
withstanding the applied pressures. It is
porewater pressure. Both measuring
desirable to use a transparent material or NOTE 5-The stress produced by the specimen
devices shall have a rigidity of all the
have a cylinder provided with viewing cap can exceed 0.145 psi (1 kN/m2) provided
assembled parts of the pore-water
ports so the behavior of the specimen the test data is corrected for the effects of that
pressure measurement system relative to stress.
may be observed. The top plate shall
the total volume of the specimen,
have a vent valve such that air can be
satisfying the following requirement:
forced out of the chamber as it is filled. 5.10.2 For Effective Stress
The baseplate shall have an inlet through (∆V / V ) < 2.2 x10−5 in.2 / lb(3.2 x10−6 m2 / kN )
Determination:
which pressure liquid is supplied to the ∆u
chamber. 5.10.2.1 Specimen Cap and Base-The
where: specimen cap and base shall be designed
5.5 Axial Load Piston-The piston passing to provide drainage from both ends of the
through the top of the chamber and its ∆V = change in volume of the
porewater measurement system specimen. They shall be constructed of a
seal must be designed so the variation in rigid, noncorrosive, impermeable
axial load due to friction does not exceed due to a pore pressure change,
in.3 (mm3), material, and each shall, except for the
711
T 296 METHODS OF SAMPLING AND TESTING

drainage provision, have a circular plane 5.12 Valves-Changes in volume of the 5.18 Testing Environment-The shear
surface of contact with the porous discs system due to opening and closing valves portion of the test shall be performed in
and a circular cross section. The weight may result in inaccurate pore-water an environment where temperature
of the specimen cap and top porous disc pressure measurements. For this reason, fluctuations are less than ± 7.2°F (± 4°C)
shall be less than 0.5 percent of the valves in the system shall be of the type and there is no direct contact with
applied axial load at failure or less than that produce minimum volume changes sunlight.
0.1 lb (50 g). The diameter of the cap and due to their operation. A valve may be 5.19 Miscellaneous Apparatus-
base shall be equal to the initial diameter assumed to produce minimum volume Specimen trimming and carving tools
of the specimen. The specimen base shall change if opening or closing the valve in including a wire saw, steel straightedge,
be connected to the triaxial compression a closed, saturated pore-water pressure miter box, and vertical trimming lathe,
chamber to prevent lateral motion or system does not induce a pressure change apparatus for preparing compacted
tilting, and the specimen cap shall be of greater than ± 0.1 psi (0.7 kPa). All specimens, membrane and 0-ring
designed such that eccentricity of the valves must be capable of withstanding expander, water content cans, and data
piston-to-cap contact relative to the applied pressures without leakage. sheets shall be provided as required.
vertical axis of the specimen does not
exceed 0.005 in. (1.3 mm). The end of
NOTE 7-Ball valves have been found to
the piston and specimen cap contact area provide minimum volume-change
6. TEST SPECIMEN PREPARA-
shall be so designed that tilting of the characteristics, however, any other type of
TION
specimen cap during the test is minimal. valve having suitable volume-change
The cylindrical surface of the specimen characteristics may be used.
6.1 Specimen Size-Specimens shall be
base and cap that contacts the membrane
cylindrical and have a minimum diameter
to form a seal shall be smooth and free of
5.13 Specimen-Size Measurement of 1.3 in. (3.3 cm). The height-to-
scratches.
Devices-Devices used to determine the diameter ratio shall be between 2 and 2.5.
5.10.2.2 The specimen shall be height and diameter of the specimen shall The largest particle size shall be smaller
separated from the specimen cap and measure the respective dimensions to than 1/6 the specimen diameters. If, after
base by rigid porous discs of a diameter within ±0.1 percent of the total dimen- completion of a test, it is found based on
equal to that of the specimen. The sion and shall be constructed such that visual observation that oversize particles
coefficient of permeability of the discs their use will not disturb the specimen. are present, indicate this information in
shall be approximately equal to that of the report of test data (Section 11.1.20).
fine sand (4 X 10-3 in./s (1 X 10-4 cm/s)).
The discs shall be regularly checked to NOTE 8- Circumferential measuring tapes are
determine whether they have become recommended over calipers for measuring the NOTE 9-If oversize particles are found in the
diameter. specimen after testing a particle-size analysis
clogged. may be performed in accordance with ASTM
D 422 to confirm the visual observation and
NOTE 6-Filter-paper discs of a diameter equal 5.14 Recorders-Specimen behavior the results provided with the test report
to that of the specimen may be placed between may be recorded manually or by (Section 11.1.4).
the porous discs and specimen to avoid electronic digital or analog recorders. If
clogging of the porous discs when accurate electronic recorders are used, it shall be
moduli values are not required.
6.2 Undisturbed Specimens-Prepare
necessary to calibrate the measuring
undisturbed specimens from large
devices through the recorder using
undisturbed samples or from samples
known input standards.
5.11 Rubber Membrane-The rubber secured in accordance with Practice T-
membrane used to encase the specimen 5.15 Sample Extruder-The sample 207 or other acceptable undisturbed tube
shall provide reliable protection against extruder shall be capable of extruding the sampling procedures and preserved and
leakage. To check a membrane for soil core from the sampling tube at a transported in accordance with the
leakage, the membrane shall be placed uniform rate in the same direction of practices for Group C samples in ASTM
around a cylindrical form, sealed at both travel as the sample entered the tube and D 4220. Specimens obtained by tube
ends with rubber 0-rings, subjected to a with minimum disturbance of the sample. sampling may be tested without trimming
small air pressure on the inside, and If the soil core is not extruded vertically, except for cutting the end surfaces plane
immersed in water. If air bubbles appear care should be taken to avoid bending and perpendicular to the longitudinal axis
from any point on the membrane it shall stresses on the core due to gravity. of the specimen, provided soil
be rejected. To offer minimum restraint Conditions at the time of sample removal characteristics are such that no significant
to the specimen, the unstretched may dictate the direction of removal, but disturbance results from sampling.
membrane diameter shall be between 90 the principal concern is to minimize the Handle specimens carefully to minimize
and 95 percent of that of the specimen. degree of disturbance. disturbance, changes in cross section, or
The membrane thickness shall not exceed 5.16 Weighing Device-The specimen change in water content. If compression
1 percent of the diameter of the weighing device shall determine the mass or any type of noticeable disturbance
specimen. The membrane shall be sealed of the specimen to an accuracy of within t would be caused by the extrusion device,
to the specimen cap and base with rubber 0.05 percent of the total mass of the split the sample tube lengthwise or cut
0-rings for which the unstressed inside specimen. the tube in suitable sections to facilitate
diameter is between 75 and 85 percent of 5.17 Water De-aeration Device-The removal of the specimen with minimum
the diameter of the cap and base, or by amount of dissolved gas (air) in the disturbance. Prepare trimmed specimens,
other means that will provide a positive closed test system may be decreased by in an environment such as a controlled
seal. An equation for correcting the boiling, by heating and spraying into a high-humidity room where soil water
principal stress difference (deviator vacuum, or by any other method that will content change is minimized. Where
stress) for the effect of the stiffness of the satisfy the requirement for producing removal of pebbles or crumbling
membrane is given in Section 10.4.3. deaired water having less than 6 ppm resulting from trimming causes voids on
dissolved oxygen. the surface of the specimen, carefully fill

712
T 296 METHODS OF SAMPLING AND TESTING

the voids with remolded soil obtained the longitudinal axis, remove the mold thickness of the membrane, and the combined
from the trimmings. Where the sample and determine the mass and dimensions height of the cap, base, and porous discs so
condition permits a vertical trimming of the specimen using the devices that the appropriate values may be subtracted
from the measurements.
lathe that will accommodate the total described in Sections 5.13 and 5.16.
sample may be used as an aid in Perform one or more water content
preparing the specimen to the required determinations on excess material used to 7.2.1 Wet Mounting Method:
diameter. After obtaining the desired prepare the specimen in accordance with 7.2.1.1 Fill the specimen drainage lines
diameter, place the specimen in a miter ASTM D 2216. and the pore-water pressure measurement
box and cut the specimen to the final device with de-aired water.
length with a wire saw or other suitable
NOTE 10-It is common for the unit weight of 7.2.1.2 Saturate the porous discs by boi-
device. Trim the surfaces with the steel the specimen after removal from the mold to
straightedge. Perform one or more water ling them in water for at least 10 minutes
be less than the value based on the volume of
content determinations on material and allow to cool to room temperature.
the mold. This occurs as a result of the
trimmed from the specimen in specimen swelling after removal of the lateral 7.2.1.3 Place a saturated porous disc on
accordance with ASTM D 2216. confinement due to the mold. the specimen base and after wiping away
Determine the mass and dimensions of NOTE 11-Experience indicates that it is all free water on the disc, place the
the specimen using the devices described difficult to compact, handle, and obtain valid specimen on the disc. Next, place another
in Sections 5.13 and 5.16. A minimum of results with specimens that have a degree of porous disc and the specimen cap on top
three height measurements (120° apart) saturation that is greater than about 90 percent. of the specimen. Check that the specimen
and at least three diameter measurements cap, specimen, and porous discs are
at the quarter points of the height shall be 7. MOUNTING SPECIMEN centered on the specimen base.
made to determine the average height and 7.2.1.4 Proceed with Section 7.3.
diameter of the specimen.
7.1 Preparations-Before mounting the 7.2.2 Dry Mounting Method:
6.3 Remolded Specimens-Prepare the
specimen by first thoroughly working the specimen in the triaxial chamber, make 7.2.2.1 Dry the specimen drainage
previously undisturbed specimen, which the following preparations: system. This may be accomplished by
has been tested and is still encased in the 7.1.1 If deemed necessary, check the allowing dry air to flow through the
rubber membrane, with the fingers. Then rubber membrane for leaks (see Section system prior to mounting the specimen.
reform the specimen by forming within a 5.11). 7.2.2.2 Dry the porous discs in the oven
mold having dimensions such that the 7.1.2 Place the membrane on the overnight prior to mounting the specimen
remolded specimen dimensions will be membrane expander or, if it is to be and place in a desiccator while allowing
equal to those of the undisturbed rolled onto the specimen, roll the the discs to cool to room temperature.
specimen. Exercise care to avoid membrane on the cap or base. 7.2.2.3 Place a dry porous disc on the
entrapping air in the specimen. This will specimen base and place the specimen on
7.1.3 If effective stress is to be
aid in obtaining a uniform unit weight, in the disc. Next, place a dry porous disc
determined check that the porous discs
remolding to the same void ratio as the and the specimen cap on the specimen.
and specimen drainage tubes are not
undisturbed specimen, and in preserving Check that the specimen cap, porous
obstructed by passing air or water
the natural water content of the soil. discs, and specimen are centered on the
through the appropriate lines.
6.4 Compacted Specimens-Prepare specimen base.
7.1.4 Attach the pressure-control and if
specimens using the compaction method, 7.3 Place the rubber membrane around
appropriate a pore-pressure measurement
predetermined water content, and unit the specimen and seal it at the cap and
device to the chamber base.
weight prescribed by the individual ass base with two rubber 0-rings or other
baiting the test. Compacted specimens positive seal at each end. A thin coating
may be prepared by compacting material NOTE 12-This procedure assumes pore of silicon grease on the vertical surfaces
in at least six layers, using a pressing or pressure is to be measured using a direct of the cap and base will aid in sealing the
kneading action, into a split mold of reading electronic pressure transducer. If a
membrane.
circular cross section having dimensions commercially manufactured null-indicator is
used, consult the manufacturer literature 7.4 Attach the top drainage line and
meeting the requirements of Section 6.1.
regarding indicator adjustments required check the alignment of the specimen and
Material required for the specimen shall during testing (see Section 5.8). For an the specimen cap. If the dry mounting
be batched by thoroughly mixing soil example of a locally manufactured null- method has been used, apply a partial
with sufficient water to produce the indicator with operating instructions, see vacuum of approximately 5 psi (35 kPa)
desired water content. After batching, Section 2.3.2.
(effective confining pressure) to the
store the material in a covered container
specimen through the top drainage line
for at least 16 hours prior to compaction.
7.2 Depending on whether the saturation prior to checking the alignment. If there
Specimens may be molded to the desired
portion of the test will be initiated with is any eccentricity, release the partial
density by either: (1) kneading or
either a wet or dry drainage system, vacuum, realign the specimen and cap,
tamping each layer until the accumulative
mount the specimen using the appropriate and then reapply the partial vacuum. If
weight of the soil placed in the mold is
method, as follows in either Section 7.2.1 the wet mounting method has been used,
compacted to a known volume or (2) by
or 7.2.2. the alignment of the specimen and the
adjusting the number of layers, the
specimen cap may be checked and adjus-
number of tamps per layer, and the force
ted without the use of a partial vacuum.
per tamp. Scarify the top of each layer NOTE 13-It is recommended that the dry
prior to the addition of material for the mounting method be used for specimens of
next layer. The tamper used to compact soils that swell appreciably when in contact 8. PROCEDURE
the material shall have an area in contact with water. If the wet mounting method is
used for such soils, it will be necesary to
with the soil equal to or less than 1/2 the
obtain the specimen dimensions after the 8.1 For total stress determination
area of the mold. After a specimen is specimen has been mounted. In such cases, it
formed, with the ends perpendicular to (without pore pressure measurement).
will be necessary to determine the double
713
T 296 METHODS OF SAMPLING AND TESTING

8.1.1 With the specimen encased in the is located inside the chamber, it will not the recorded "pore pressure" will be a
rubber membrane, which is sealed to the be necessary to correct or compensate for combination of pore air and pore water
specimen cap and base and positioned in the uplift force acting on the axial pressure existing within the voids of the tested
soil mass. See Section 2.3.1 for detailed
the chamber, assemble the triaxial loading device or for piston friction. In
instructions for the separate determination of
chamber. Bring the axial load piston into both cases record the initial reading on pore air and pore water pressure in a partially
contact with the specimen cap several the deformation indicator when the piston saturated sample.
times to permit proper seating and contacts the specimen cap.
alignment of the piston with the cap. 8.1.4 Apply the axial load to produce
When the piston is brought into contact 8.2.2 Saturation-The objective of the
axial strain at a rate of approximately 1
the final time, record the reading on the saturation phase of the test is to fill all
percent/min for plastic materials and 0.3
deformation indicator. During this voids in the specimen with water without
percent/min for brittle materials that
procedure, take care not to apply an axial undesirable prestressing of the specimen
achieve maximum deviator stress at
stress to the specimen exceeding or allowing the specimen to swell.
approximately 3 to 6-percent strain. At
approximately 0.5 percent of the Saturation is usually accomplished by
these rates, the elapsed time to reach
estimated compressive strength. If the applying back pressure to the specimen
maximum deviator stress will be
weight of the piston is sufficient to apply pore water to drive air into solution after
approximately 15 to 20 minutes.
an axial stress exceeding approximately either: (1) applying vacuum to the
Continue the loading to 15 percent axial
0.5 percent of the estimated compressive specimen and dry drainage system (lines,
strain, except loading may be stopped
strength, lock the piston in place above porous discs, pore-pressure device, and
when the deviator stress has peaked then
the specimen cap after checking the discs) and allowing de-aired water to
dropped 20 percent or the axial strain has
seating and alignment and keep locked saturate the system while maintaining the
reached 5 percent beyond the strain at
until application of the chamber pressure. vacuum; or (2) saturating the drainage
which the peak in deviator stress
system by boiling the porous discs in
8.1.2 Place the chamber in position in occurred.
water and allowing water to flow through
the axial loading device. Be careful to 8.1.5 Record load and deformation the system prior to mounting the
align the axial loading device, the axial values at about 0.1, 0.2, 0.3, 0.4, and 0.5- specimen. It should be noted that time is
load-measuring device, and the triaxial percent strain; then at increments of required to place air into solution.
chamber to prevent the application of a about 0.5-percent strain to 3 percent; and, Accordingly, removing as much air as
lateral force to the piston during testing. thereafter at every 1 percent. Take possible prior to applying back pressure
Attach the pressure-maintaining and sufficient readings to define the will decrease the amount of air that will
measurement device and fill the chamber stressstrain curve; hence, more frequent have to be placed into solution and will
with the confining liquid. Adjust the readings may be required in the early also decrease the back pressure required
pressure-maintaining and measurement stages of the test and as failure is for saturation. In addition, air remaining
device to the desired chamber pressure approached. in the specimen and drainage system just
and apply the pressure to the chamber
prior to applying back pressure will go
fluid. If it is determined necessary to
NOTE 17-Alternate intervals for the readings into solution much more readily if
stabilize the specimen under the chamber
may be used provided sufficient points are deaired water is used for saturation. The
pressure prior to application of the axial
obtained to define the stressstrain curve. use of de-aired water will also decrease
load, wait approximately 10 minutes after
the time and back pressure required for
applying chamber pressure before
saturation. Many procedures have been
continuing the test. 8.1.6 For completion of the total stress developed to accomplish saturation. The
determination, proceed to Section 9 and following are suggested procedures:
NOTE 14-In some cases the chamber will be applicable portions of Section 10.
8.2.2.1 Starting with Initially Dry
filled and the chamber pressure applied before 8.2 For effective stress determination Drainage System-Increase the partial
placement in the axial loading device. (with port, pressure measurement). vacuum acting on top of the specimen to
NOTE 15-Make sure the piston is locked or 8.2.1 Prior to Saturation-After the maximum available vacuum. If the
held in place by the axial loading device assembling the triaxial chamber, perform
before applying the chamber pressure.
effective confining pressure under which
the following operations: the strength is to be determined is less
NOTE 16-The purpose of the waiting period
8.2.1.1 Bring the axial load piston into than the maximum partial vacuum, apply
is to allow the specimen to stabilize under the
chamber pressure prior to application of the contact with the specimen cap several a lower partial vacuum to the chamber.
axial load. times to permit proper seating and The difference between the partial
alignment of the piston with the cap. vacuum applied to the specimen and the
During this procedure, take care not to chamber should never exceed the
8.1.3 If the axial load-measuring apply an axial load to the specimen effective confining pressure for the test
device is located outside of the triaxial exceeding 0.5 percent of the estimated and should not be less than 5 psi (35 kPa)
chamber, the chamber pressure will axial load at failure. When the piston is to allow for flow through the sample.
produce an upward force on the piston brought into contact, record the reading After approximately 2 hours, allow
that will react against the axial loading of the deformation indicator. desired water to percolate from the
device. In this case, start the test with the bottom to the top of the specimen under a
piston slightly above the specimen cap, 8.2.1.2 Fill the chamber with the
differential vacuum of less than 3 psi (20
and before the piston comes in contact chamber liquid, being careful to avoid
kPa).
with the specimen cap, either: (1) trapping air or leaving an air space in the
chamber. There should always be a positive
measure and record the initial piston
effective stress of at least 2 psi (13 kPa)
friction and upward thrust of the piston
at the bottom of the specimen during this
produced by the chamber pressure and NOTE 18-This procedure assumes the sample part of the procedure. When water
later correct the measured axial load, or is to be tested in a fully saturated condition (S
appears in the burette connected to the
(2) adjust the axial load-measuring = 100 percent). A sample can be tested at
conditions other than S = 100 percent by top of the specimen, close the valve to
device to compensate for the friction and
omitting Section 8.2.2 below, in which case the bottom of the specimen and fill the
thrust. If the axial load-measuring device
714
T 296 METHODS OF SAMPLING AND TESTING

burette with de-aired water. Next, reduce percent when the pressures are constant. or values of ∆u greater than ∆σ3 indicate
the vacuum acting on top of the specimen To check for equalization after a leak of chamber fluid into the
through the burette to atmospheric application of a back pressure increment specimen. Decreasing values of ∆u with
pressure while simultaneously increasing or after the full value of back pressure time may indicate a leak in that part of
the chamber pressure by an equal has been applied, close the specimen the pore pressure measurement system
amount. During this process, the drainage valves and measure the change located outside of the chamber.
difference between the pore pressure in pore-pressure over a 1-minute interval. 8.2.4.3 Reapply the same effective
measured at the bottom of the specimen If the change in pore-pressure is less than confining stress as existed prior to the
and the chamber pressure should not be 1 percent of the difference between the Bvalve measurement by reducing the
allowed to exceed the past chamber pressure increment pressure and chamber pressure by 10 psi (70 kPa) or
preconsolidation pressure. When the pore the back pressure, another back pressure by alternatively, increasing the back
pressure at the bottom of the specimen increment may be added or a pressure by 10 psi (70 kPa). If B is
stabilizes, proceed with back pressuring measurement may be taken of the pore continuing to increase with increasing
of the specimen pore-water as described pressure parameter B (see Section 8.2.4) back pressure, continue with back
in Section 8.2.3. To check for to determine if saturation is completed. pressure saturation. If B is equal or
equalization, close the drainage valves to Specimens shall be considered to be greater than 0.95 or if a plot of B versus
the specimen and measure the pore saturated if the value of B is equal to or back pressure indicates no further
pressure change over a 1-minute interval. greater than 0.95, or if B remains increase in B with increasing back
If the change is less than 1 percent of the unchanged with addition of back pressure pressure, saturation is complete.
chamber pressure, the pore pressure may increments.
be assumed to be stabilized. 8.2.5 Shear-During shear, the chamber
pressure shall be kept constant while
NOTE 20-The pore pressure parameter B- advancing the axial load piston
NOTE 19-For saturated clays. percolation used to measure the degree of saturation-is a downward against the specimen cap
may not be necessary and water can be added function of soil stiffness. If the degree of using controlled axial strain as the
simultaneously at both top and bottom. saturation of the sample is 100 percent, the B-
loading criterion. Specimen drainage is
value measurement will increase with
decreasing soil stiffness. Therefore, when not permitted during shear.
8.2.2.2 Starting with Initially Saturated testing soft soil samples, a Bvalue of 95 8.2.5.1 Prior to Axial Loading-Before
Drainage System-After filling the burette percent may indicate a degree of saturation initiating shear, perform the following
connected to the top of the specimen with less than 100 percent. steps:
de-aired water, apply a chamber pressure NOTE 21-The back pressure required to 8.2.5.1.1 By opening or closing the
of 5 psi (35 kPa) or less and open the saturate a compacted specimen may be higher
for the wet mounting method than for the dry
appropriate valves, isolate the specimen
specimen drainage valves. When the pore so that during shear the specimen
pressure at the bottom of the specimen mounting method and may be as high as 200
psi (1400 kPa). porewater pressure will be measured by
stabilizes, according to the method the pore-pressure measurement device
described in Section 8.2.2.1 or when the NOTE 22-Many laboratories use differential
pressure regulators and transducers to achieve and no drainage will occur.
burette reading stabilizes, back
the requirements for small differences between 8.2.5.1.2 Place the chamber in position
pressuring of the specimen porewater chamber and back pressure.
may be initiated. in the axial loading device. Be careful to
align the axial loading device, the axial
8.2.3 Applying Back Pressure- load-measuring device, and the triaxial
Simultaneously increase the chamber and 8.2.4 Measurement of the Pore
Pressure Parameter B-Determine the chamber to prevent the application of a
back pressure in steps with specimen lateral force to the piston during shear.
drainage valves opened so that de-aired value of the pore pressure parameter B in
water from the burette connected to the accordance with Sections 8.2.4.1 through 8.2.5.1.3 Bring the axial load piston
top and bottom of the specimen may flow 8.2.4.3. The pore pressure parameter B is into contact with the specimen cap to
into the specimen. To avoid undesirable defined by the following equation: permit proper seating and realignment of
prestressing of the specimen while B = ∆u/∆σ3 the piston with the cap. Remember that
applying back pressure, the pressures during this procedure, care should be
where: taken not to apply an axial load to the
must be applied incrementally with
∆u = the change in the specimen pore specimen exceeding 0.5 percent of the
adequate time between increments to
pressure that occurs as a result of a estimated axial load at failure. If the axial
permit equalization of pore-water
change in the chamber pressure load-measuring device is located outside
pressure throughout the specimen. The
when the specimen drainage valves of the triaxial chamber, the chamber
size of each increment might be 5 psi (35
are closed, and pressure will produce an upward force on
kPa), 10 psi (70 kPa), or even 20 psi (140
kPa), depending on the compressibility of ∆σ3 = the change in the chamber pressure. the piston that will react against the axial
the soil specimen, the magnitude of the loading device. In this case, start shear
8.2.4.1 Close the specimen drainage
desired effective confining stress, and the with the piston slightly above the
valves and increase the chamber pressure
degree of saturation of the specimen just specimen cap, and before the piston
by 10 psi (70 kPa).
prior to the addition of the increment. comes into contact with the specimen
8.2.4.2 After approximately 2 minutes cap, either (1) measure and record the
The difference between the chamber determine and record the maximum value
pressure and the back pressure during initial piston friction and upward thrust of
of the induced pore pressure. For many the piston produced by the chamber
back pressuring should not exceed 5 psi specimens, the pore pressure may
(35 kPa) unless it is deemed necessary to pressure and later correct the measured
decrease after the immediate response axial load, or (2) adjust the axial load-
control swelling of the specimen during and then increase slightly with time. If
the procedure. The difference between measuring device to compensate for the
this occurs, values of ∆u should be friction and thrust. The variation in the
the chamber and the back pressure must plotted with time and the asymptotic pore
also remain within ± 5 percent when the axial load-measuring device reading
pressure used as the change in pore should not exceed 0.1 percent of the
pressures are raised and within ± 2 pressure. A large increase in ∆u with time estimated failure load when the piston is
715
T 296 METHODS OF SAMPLING AND TESTING

moving downward prior to contacting the content.) In cases where there is NOTE 25-In the event that the application of
specimen cap. If the axial load-measuring insufficient material from trimmings for the chamber pressure results in a change in the
device is located inside the chamber, it index property tests, that is, where specimen length, Ao should be corrected to
reflect this change in volume. Frequently, this
will not be necessary to correct or specimens have the same diameter as the
is done by assuming that lateral strains are
compensate for the uplift force acting on sampling tube, the specimen should be equal to vertical strains. The diameter after
the axial loading device or for piston weighed prior to removing material for volume change would be given by D = Dσ (1 -
friction. However, if an internal load- index property tests and a representative ∆L/L).
measuring device of significant flexibility portion of the specimen used to
is used in combination with an external determine its final water content. Prior to
deformation indicator, correction of the placing the specimen (or portion thereof) 10.4 For determination of total stress
deformation readings may be necessary. in the oven to dry, sketch a picture or (no pore pressure measurement):
In both cases, record the initial reading take a photograph of the specimen 10.4.1 Calculate the principal stress
on the pore-water pressure measurement showing the mode of failure (shear plane, difference (deviator stress), a, - a3, for a
device immediately prior to when the bulging, etc.). given applied axial load as follows:
piston contacts the specimen cap and the σ1 - σ3 = P/A
reading on the deformation indicator
10. CALCULATIONS where:
when the piston contacts the specimen
cap. P = measured applied axial load
10.1 Initial Specimen Properties- (corrected for uplift and piston
8.2.6 Axial Loading-Apply axial load
Using the dry mass of the total specimen, friction, if required see Section
to the specimen using a rate of axial
calculate and record on the appropriate 8.2.5.1.3), and
strain that will produce approximate
equalization of pore pressures throughout data sheet the initial water content, A = corresponding average cross-
the specimen at failure. Strain rates of volume of solids, initial void ratio, initial sectional area.
about 1 percent per minute (for plastic degree of saturation, and initial dry unit 10.4.2 Stress-Strain Curve-Prepare a
soil) to about 0.3 percent per minute (for weight. Calculate the specimen volume graph showing the relationship between
brittle soils) have been used successfully. from values measured in Section 6.2 or principal stress difference (deviator
At a minimum, record load, deformation, 6.3. Calculate the volume of solids by stress) and axial strain, plotting deviator
and pore-water pressure values at dividing the dry mass of the specimen by stress as ordinate and axial strain (in
increments of 0.1 percent to 1-percent the specific gravity of the solids (Note percent) as abscissa. Select the
strain and, thereafter, at every 1 percent. 23) and dividing by the density of water. compressive strength and axial strain at
Take sufficient readings to define the Calculate the void ratio by dividing the failure in accordance with the definitions
stressstrain curve; hence, more frequent volume of voids by the volume of solids in Sections 3.2.3 and 3.2.6.
readings may be required in the early where the volume of voids is assumed to
10.4.3 Correction of Strength Due to
stages of the test and as failure is be the difference between the specimen
Stiffiiess of Rubber Membrane-Assuming
approached. Continue the loading to 15- volume and the volume of the solids.
units are consistent, the following
percent strain, except loading may be Calculate dry density by dividing the dry
equation, or other acceptable equations,
stopped when the principal stress mass of the specimen by the specimen
shall be used to correct the principal
difference (deviator stress) has dropped volume.
stress difference or deviator stress for the
20 percent or when 5-percent additional effect of the rubber membrane if the error
axial strain occurs after a peak in NOTE 24-The specific gravity of solids can in principal stress difference due to the
principal stress difference (deviator be determined in accordance with Test Method stiffness of the membrane exceeds 5
stress). T 100 or it may be assumed based on previous percent:
test results.
4 Em t ε1
NOTE 23-The use of a manually adjusted ∆ (σ1 - σ3 ) =
D
null-indicating device will require nearly 10.2 Calculate the axial strain, a
continuous attention to ensure the criterion for (expressed as a decimal), for a given where:
undrained shear.
applied axial load, as follows: ∆ (σ1 - σ3 ) = correction to be
ε = ∆L/Lo subtracted from the
9. REMOVING SPECIMEN
where: measured principal stress
difference,
∆L = change in length of specimen as
9.1 When shear is completed, perform read from deformation indicator, and 4A
the following steps: D = diameter of
Lo = initial length of test specimen π
9.1.1 Remove the axial load and reduce minus any change in length prior to specimen,
the chamber and back pressures to zero. loading. Em = Young's modulus for the
9.1.2 With the specimen drainage 10.3 Calculate the average cross- membrane material,
valves remaining closed, quickly remove sectional area, A, for a given applied t = thickness of the
the specimen from the apparatus so that axial load as follows: membrane, and
the specimen will not have time to absorb
A = Ao/(l - ε) ε1 = axial strain.
water from the porous discs.
where: 10.4.3.1 The Young's modulus of the
9.1.3 Remove the rubber membrane
and determine the water content of the A = initial average cross-sectional membrane material may be determined
total specimen in accordance with the area of the specimen, and by hanging a 0.5-in. (12.5-mm) wide
procedure outlined in ASTM D 2216. ε = axial strain for the given axial strip of membrane over a thin rod,
(Free water remaining on the specimen load (expressed as a decimal). placing an other rod along the bottom of
after removal of the membrane should be the hanging membrane, and measuring
blotted away before obtaining the water the force per unit strain obtained by

716
T 296 METHODS OF SAMPLING AND TESTING

stretching the membrane. The modulus A = corresponding cross-sectional 10.5.4.1 The Young's modulus of the
value may be computed using the area, in.2 (mm2). membrane material may be determined
following equation assuming units are 10.5.1.1 Calculate the effective minor by hanging a 0.5-in. (12.5-mm)
consistent: principal stress, σ´3 for a given applied circumferential strip of membrane using
FL axial load as follows: a thin rod, placing another rod through
Em = the bottom of the hanging membrane, and
Am ∆L σ´3 = σ3 - ∆u measuring the force per unit strain
where: where: obtained by stretching the membrane.
Em = Young's modulus of the membrane σ´3 = effective confining stress, psi The modulus value may be computed
material, (kPa) and u = induced pore-water using the following equation:
F = force applied to stretch the pressure at the given axial load
membrane, (total pore-water pressure minus where:
the total back pressure), psi (kPa).
Am = twice the initial thickness of the Em = Young's modulus of the
membrane multiplied by the width 10.5.2 Principal Stress Difference membrane material, psi (kPa),
of the membrane strip, (Deviator Stress) and Induced Pore-
Water Pressure versus Strain Curves- F = force applied to stretch the
L = unstretched length of the Prepare graphs showing relationships membrane, lbf (kN),
membrane, and between principal stress difference L = unstretched length of the
∆L = change in length of the membrane (deviator stress) and induced pore-water membrane, in. (mm),
due to application of F. pressure with axial strain, plotting ∆L = change in length of the membrane
A typical value of E for latex membrane deviator stress and, induced pore-water due to application of the force, F,
is 200 psi (1400 kPa). pressure as ordinates and axial strain as in. (mm), and
abscissa. Select the principal stress
difference (deviator stress) and axial Am = area of the membrane = 2 tm Ws,
NOTE 26-The effect of the stiffness of the strain at failure in accordance with a in.2 (m2 = mm2/1000000)
membrane on the lateral stress is usually definition in Section 3.2.3. where:
assumed to be negligible.
10.5.3 p' - q Diagram-Prepare a graph tm = thickness of the membrane, and
NOTE 27-The correction for rubber
membranes is based on simplified assumptions showing the relationship between p', (σ´1 Ws = width of circumferential strip
concerning their behavior during shear. Their + σ´3)/2 and q, (σ1 - σ3)/2, plotting q as (0.5-in. or 15-mm).
actual behavior is complex and there is not a ordinate and p' as abscissa using the same
consensus on more exact corrections. scale. The value of p' for a given axial
load may be computed as follows: NOTE 29-A typical value of E,. for latex
membranes is 200 psi (1400 kPa).
10.4.4 Calculate the major and minor p' = ((σ1 - σ3) + 2 σ´3)/2 NOTE 30-The effect of the membrane on the
principal total stresses at failure as where: lateral stress is usually assumed to be
follows: σ1 - σ3 = principal stress difference negligible (see Note 27).
σ3 = minor principal total stress = (deviator stress), psi (kPa),
chamber pressure, and and 10.5.4.2 Determine the major and
σ1 = major principal total stress = σ´3 = effective minor principal minor principal stresses at failure based
deviator stress at failure plus chamber stress, psi (kPa). on total stresses, σ1f and σ3f respectively,
pressure. 10.5.4 Correction for Rubber and on effective stresses σ´1f and σ´3f
10.4.4.1 Calculate the initial degree of Membrane-Use the following equation to respectively, as follows:
saturation of the test specimen using the correct the principal stress difference σ3f = effective consolidation stress, psi
initial mass and dimensions. (deviator stress) for the effect of the (kPa),
rubber membrane if the error in principal σ1f = (σ1 - σ3) at failure + σ3f, psi (kPa),
stress difference (deviator stress) due to
NOTE 28-The specific gravity determined in σ3f = σ3f - ∆uf, psi (kPa), and
accordance with Test Method T 100 is the strength of the membrane exceeds 5
percent: σ´1f = (σ1 - σ3) at failure + σ´3f, psi
required for calculation of the degree of
saturation, or an assumed value may be used (kPa), where ∆u is the induced
∆( σ1 - σ3) = (4Emtmε1)/D4
provided it is noted in the test report that an pore-water pressure at failure.
assumed value was used. where:
10.6 Mohr Stress Circles-Construct
∆(σ1 - σ3) = the correction to be Mohr stress circles at failure based on
subtracted from the total and effective stresses on an
10.4.5 Proceed to Section 10.6. measured principal stress arithmetic plot with shear stress as
10.5 For determination of effective difference (deviator ordinate aid normal stress as abscissa
stress (with pore pressure measurement): stress), psi (kPa), using the same scales. As shown in
10.5.1 Calculate the principal stress D4 = 4Aσ / π diameter of Figure 1 the circle based on total stresses
difference (deviator stress), σ1 - σ3, for a is drawn with a radius of one-half the
specimen, in. (mm),
given applied axial load as follows: principal stress difference (deviator
Em = Young's modulus for the stress) at failure with its center at a value
σ1 - σ3 = P/A
membrane material, psi equal to one-half the sum of the major
where: (kPa), and minor total principal stresses. The
P = given applied axial load tm = thickness of the Mohr stress circle based on effective
(corrected for uplift and piston membrane, in. (mm), and stresses is drawn in a similar manner
friction if required as obtained in except that its center is at a value equal to
ε1 = axial strain (decimal
Section 8.4.1.3), lb (kN), and one-half the sum of the major and
form).
minoreffective principal stresses.

717
T 296 METHODS OF SAMPLING AND TESTING

11. REPORT the values of the effective minor and 11.1.16 Mohr stress circles based on total
major principal stresses at failure. and effective stresses.
11.1 The report shall include the 11.1.17 Failure sketch or photograph of
following (as applicable): the specimen.
11.1.1 Identification data and visual 11.1.18 Remarks and notations regarding
description of specimen, including soil any unusual conditions or other
classification and whether the specimen information necessary to properly
is undisturbed, compacted, or otherwise interpret the results obtained, including
prepared. any departures from the procedure
outlined.
11.1.2 Values of plastic limit and liquid
limit, if determined in accordance with
Test Methods T 89 and T 90. 12. PRECISION AND BIAS
11.1.3 Value of specific gravity of solids
and notation if the value was determined 12.1 The variability of soil and
in accordance with Test Method 100 or resultant inability to determine a true
assumed. reference value prevent development of a
11.1.4 Particle-size analysis, if meaningful statement of bias. Data are
determined in accordance with Method D being evaluated to determine the
422. precision of this test method. In addition,
11.1.5 Initial specimen dry unit weight, the subcommittee is seeking pertinent
void ratio, water content, and percent data from users of this test method.
saturation.
11.1.6 Initial height and diameter of 13. KEYWORDS
specimen.
11.1.7 Method followed for specimen 13.1 Back pressure saturation;
saturation (that is, dry or wet method). 11.1.12 Axial strain at failure, percent. unconsolidated-undrained triaxial
11.1.8 Total back pressure. compression test; effective stress; pore
11.1.13 Rate of strain, percent/min.
11.1.9 The pore pressure parameter B at pressure; shear strength; strength
11.1.14 Principal stress difference envelope; total stress.
the end of saturation.
(deviator stress) and induced pore-water
11.1.10 Failure criterion used. pressure versus axial strain curves as
11.1.11 The value of the principal stress described in Section 10.4.
difference (deviator stress) at failure and 11.1.15 The p' - q diagram for the test
results.

718

You might also like