This document contains:
1) A test record sheet for an e360 ventilator that includes test parameters, settings, and tolerances to verify proper operation.
2) Sections to test things like LED displays, pneumatic leaks, alarms, flow rates, pressures, volumes, and more.
3) Columns to record the set value, measured value, and whether it is within the specified tolerance range for each test.
This document contains:
1) A test record sheet for an e360 ventilator that includes test parameters, settings, and tolerances to verify proper operation.
2) Sections to test things like LED displays, pneumatic leaks, alarms, flow rates, pressures, volumes, and more.
3) Columns to record the set value, measured value, and whether it is within the specified tolerance range for each test.
This document contains:
1) A test record sheet for an e360 ventilator that includes test parameters, settings, and tolerances to verify proper operation.
2) Sections to test things like LED displays, pneumatic leaks, alarms, flow rates, pressures, volumes, and more.
3) Columns to record the set value, measured value, and whether it is within the specified tolerance range for each test.
This document contains:
1) A test record sheet for an e360 ventilator that includes test parameters, settings, and tolerances to verify proper operation.
2) Sections to test things like LED displays, pneumatic leaks, alarms, flow rates, pressures, volumes, and more.
3) Columns to record the set value, measured value, and whether it is within the specified tolerance range for each test.
Hospital or Organization ______________________________________________________________
e360 Serial Number ___________________ Hour Meter ________ Software Rev. __________
Service Technician ________________________________________
Date __________________ Released by ____________________________________________
Test Equipment: ID No. Electrical Safety Tol. Value
Test Calibration Analyzer Ground resistance <0.1? Oxygen Analyzer Max leakage current <100 µA Electrical Safety Analyzer FIO2 Test Pass Set Displayed Measured Tol. Range FI02 FI02
Front Panel Visual LED Check 21 + 3 Variance
Pneumatic Leak Test 40 + 3 Variance Loss of Gas Alarms (Air/O2 ) 60 + 3 Variance FIO2 80 + 3 Variance Main Flow 100 + 3 Variance Bias Flow Main Flow (L/min) FIO2 = 21% Set V? peak I Measured Tol. Range Flow/89% Inspiratory time
Respiratory Rate 5 + 1 Variance
Pressure Control 10 + 1 Variance Pressure Support 50 + 4 Variance Ptrig sensitivity 80 + 5 Variance Main Flow (L/min) FIO2 = 100% V? peak I PEEP/CAP Set Measured Tol. Range Flow/89% Manual Inflation Insp. Tidal Volume Exp. Tidal volume (VTE) 5 + 1 Variance Alarm Silence 10 + 1 Variance Loss of power/battery operation 50 + 4 Variance 80 + 5 Variance Bias Flow (L/min) Set Bias Flow Tol. Range 3 2.5 - 3.5
SER360 A1106 6-13
OPERATIONAL VERIFICATION PROCEDURE
5e360 OVP TEST RECORD SHEET (CONTINUED)
Inspiratory Time (sec) Ptrig Sensitivity (cmH2O) Set tI Measured Tol. Set Actual Tol. Range Insp. Time Range 0.25 0.2 - 0.3 1 +0-2 0.5 0.4 - 0.6 3 +2-4 Respiratory Rate (b/min) PEEP/CPAP (cmH2O) Set ƒtot Measured Tol. Set PEEP Measured Tol. Range Resp. Rate Range Pressure 5 +1 3 + 1 Variance Variance 10 + 1 Variance 20 +1 20 + 1 Variance Variance Pressure Control (cmH2O) Set Ppeak Measured Tol. Inspiratory Tidal Volume (mL) Range + 10% Set VTI Measured Tol. Range Volume 10 Variance + 10% 200 + 10% 20 Variance Variance + 10% 500 + 10% 50 Variance Variance Pressure Support (cmH2O) Expiratory Tidal Volume (mL) Set Ppeak Measured Tol. Set VTE Tol. Measured Tol. Pressure Range Range Volume Range + 10% 250 212-259 + 10% 10 Variance Variance