seca inscale 748

Technical
Documentation/
Service Manual
Service Manual
seca 748 1 Contents

Valid from: October 1996

1 Contents 17-05.01.243A

2 Operation
2.1 Operating instructions seca 746 17-10-07-3218

3 Function
3.1 Functional description 30-34-00-4658
3.2 Functional diagram 746 25-Ol-02-392A

4 Maintenance and repairs


4.1 Service instructions 30-34-00-574
4.2 Fault Oescription 3&3POO-437G
4.3 Replacement instructions 30-34-OO-489A
4.4 Adjustment instructions 3C-34-OC-467E
4.5 Spare Parts list seca 746 30-34-0043%

5 Electronics
5.1 Principle circuit diagram 08-09-01-276

6 Appendix

Manual testing and calibration instructions 30-34-00-446F


Wiring diagram, microprocessor 0%Ol-21.322D
Wiring diagram, display / buzzer 06-01-21-3230
Wiring diagram, ND converter 06.01.21-321G
Wiring diagram, power supply 06-01-21-3548
Component mounting diagram main circuit board 0%03-Ol-022c
Component mounting diagram AKI converter board 06-03-Ol-006D

Gable plan 746 06-04-04-0640

12.03.97-Hart 17-05-01-243 Index A


Service Manual
seca 748 2 Operation

Operation
Service Manual
seca 748 2.1 Operating instructionr

17-10-07-321 Index B
Service Manual
seca 748 3 Function

Function
Service Manual
seca 717/727/728/737/748/757 3.1 Functional Description

Display
Functional Description The S-digit display consists of individual LED%
combined into a plastic frame to form a 7-Segment
The forte is measured by a platform weighing cell display. The digit height is 13.2 mm.
which is designed as a double torsion spring with
strain gauges. The strain gauges arc interconnected Power supply
to form a full bridge. The weighing cell delivers a The power supply to the electrical components is
weight-proportional voltage which undergoes analog realized with a discrete switching controller. This is
/ digital conversion and is then further processed by characterised by high efficiency and a wide input
a microcontroller. voltage range. lt tan therefore be supplied from a
rechargeable battery a banery or a power supply
The electronics is divided up into the following unit.
components:

AD converter
CPU
Display
- Power suppty

AD oonvortrr
The AD converter operates on the Dual-Slope
principle, i.e. the rising edge of an integrator is
determined by 8 reference resistance. while the
falling edge is determined by the differential voltage
at the strakt gauge measuring bridge. The AD
converter is of a discrete design and is controlled by
a microcontroller. The AD converter is temperature-
compensated and provides a resolution of 16 bit.

In thamodel 746 the AD converter is housed on a


separate board close to the weighing cell. It is
encapsulated by a special cast, which must be
undamaged for use in the incubator.

CPU
The core of the CPU section is an 6-bit
microcontroller. This fulfills the following tasks:
- Conversion into a digital value of the forte
introduced into the weighing cell.
- Calculation. automatic control and compensation
of the zero Point. If the measured value falls
below the current zero Point, ‘-----* appears in
the display.
Conversion of the digital measured value into a
5digit decimal display.
Control of the display in the multiplex
Self-test of the program memory and data
memory and also of the ALU,
Checking the maximum weighing range. ‘Stop’
appears in the display if the load is greater than
Fmax
Supply voltage and rechargeable battery
monrtoring; ‘bAn’ appears in the display if the
supply voltage has reached a lower threshold.
Operation of a serral interface (not available
to the User).

27.05~97 -Hart BD-34-00-485 Index B


Elektronik mit Anzeige
Electronics with display

\
Wfigezelle

I I

ARI - Wandler Platine


A/D transducer board

SeCtl HAMBURG

Funktionsschema 25-01-02-392
Service Manual
seca 748 4.1 Service Instructions

The electronie incubator sc& seca 740 is a medical measurement and monitoring unit, with the aid of which it is
possibla to monitor the vital life functions of new-born babies. This places a special responsibility on the Service
technician who is entrusted with repairing or maintaining the sc&. All werk carried out on and in the sc& should
therefore only be performed by sufficiently trained. specialist personnel. The fixtures and units (load table, forte
introduction. weights etc.) necessary for Service work must be designed in accordance with the unit’s requirements.
The seca customer service should always be consulted in cases of doubt in order to provide full clarification of any
questions.
Depending on the workshop’s technical equipment. the scale electronics tan be repaired with tha aid of the enclosed
Plans and descriptions. However. it is essential to observe the following points:

. All manipulations in the analog part of the Signal processing of the electronics (see shaded area. adjustmant
instructions) merke adfustment of the scale necessary. Gradient and cornerstona calibration should be
performed in all casas (see adjustment instructions).

. The entire temperature coefficient of the sc& (elactronics + sensor) is compensated for at the factory for
the range +lo’C to +4O’C. Spezial care must therefore be taken when electronie components are replaced.

a) Resistors influencing gain should always have a temparature coefficient of c = TK 15 ppm.

b) Only the stated types may be used as operation amplifiers.

C) Only the stated types may be used as FET switches (low ON resistance).

. The design of this sc& has been confirmed as conforming to the guidelines and legal regulations. Repairs
should therefore only be made with original spare Parts. Installation, assembly and calibration must not result
in a Change to the ur?t’s design. lf in doubt. please contact seca customer service who will then help you
further.

27.05.97.Hart Page 1 of 1 3&-34-CG-574


Service Manual
seca 717/727/728/737/748/757 4.2 Fault Description

Fault search Measure the voltage at pin 1 of IC 131. lt should be


approx. 5V. If not, check and if nacessary raplace
A visual inspection must always be performed with diode 132. Otherwise. replace IC 131.
respect to solder Points. corrosion, contamination
and assembly. Check the reset circuit:
Check pin 13 of IC 126 with an oscilloscope. When
1. Fault domxlptlon: rochargoablr the Start key is pressed. it must be possible to
battwy chargr not corroct measure a brief positive pulse. Otherwise, replace
Connect scale to the plug-in power supply unit. IC 126.

Refer to list 30-34-00-561 for the permissible Check pin 9 of IC 520 with an oscilloscope. If it is
rechargeable battery current. not possible to maasure a brief positive pulse when
the Start key is pressed, replace transistor 128.
Check the rechargeable battery current Otherwise, replace microprocessor IC 520.

Remove the rechargeable battery and connect a Check the oscillator of the microprocessor:
current measuring unit to the connections for the A frequency of 12 MHz must be measured at pins
rechargeable battery 18 and 19 of the microprocessor IC 520.

If the charging current is not in the defined range. If this is not the case. replace the quartz
check the polarity of diode 106 and correct it if
necessary (anode to input voltage. LED lights up If the scale still does not function. replace the
when rechargeable battery current flows). microprocessor.

If the polarity is in Order. measura the charging


current with resistor 103.

If no charging current is flowing. check transistor


104 and replace it if necessary

2 Fault dowrlptlom Th. scalr tannot b.


stmtod. Check durlng oporatlon only wlth
rochargoabl. battory:
Measure the rechargeable battery voltage and
Charge it up if necessary

Check during operation with plug-in power supply


unit:
The voltage at Z diode 102 must correspond to the
power supply unit voltage.

If the voltage collapses, there is a short circuit (e.g


in diode 102) or the plug-in power supply unit is
faulty

Check the 5 V supply voltage:


The voltage at capacitor 516 must be in the range
4.7 v 5.3.

Measure the voltage at pin 6 of IC 131. If it is less


than 6V, replace the diode 110 (in the case of
rechargeable battery Operation), or diode 111 (in the
case of power supply unit Operation).

Measure the voltage at pin 6 of IC 131. If it is less


than 6V after pressing the Start key. check pin 6 of
IC 126 with an oscilloscope. When the Start key is
pressed. it must be possible to sec a brief positive
pulse. If it is missing, check the Start key and the
diodes 115 and 116. Otherwise replace IC 126~

23.04,97 Jen Page 1 3&34-CU437 Index G


Service Manual
seca 7 17/727n20/737i74arr57 4.2 Fault Dercription

3. Foult doocrlptlon: Scol. dlsplayo Check the output of the microprocessor, IC 520, or
‘bAtt’ l ftor thr Stut the input of IC 419:
‘bAtt’ is initiated at an input voltage of below approx. Using an oscilloscope, rneasure at the relevant pin
6 v. for controlling the Segment.

Check during Operation with the plug-in power


supply unit: UN InputoflC419

The voltage at 2 diode 102 must correspond with


the power supply unit voltage.

If the voltage collapses. there is short circuit (e.g. in


diode 102) or the plug-in power supply unit is faulty

Check for oparation only with rechargeable battery.


without plug-in power supply unit: - -
Tha voltage at capacitor 120 must be approx. 9V in
the case of Seca 746. and must be approx. 7.2 V in
the case of the Seca 727 and 757.

If this is not the case, Charge or replaca the


rechargeable battery.

4. Foult doocrlptlonr Dlglt leulty


Check the driver transistor (201-205):
On the basis of the associated transistor. a square-
wave Signal with an amplitude of approx. 0.7 V and
-
L
4
-
,
-
4 Ihm

an offset of approx. + 4 V must be measured. If the output does not switch, replace the
microprocessor.
A square-wave Signal with an amplitude of approx. 5
V must be measured at the collector of the 6. Fault domcrlptlon: Melno oporotlon
associated transistor. rocognltlon functlon dooo not werk
corr.otly.
If this is not the case. replace the transistor. The unloaded scale should switch off after 10
minutes in the case of Operation with a plug-in
5. Feult doocrlptlom Sogmont teulty power supply unit.
Check control of the segments with an oscilloscope. The following applies to Operation with a
rechargeable battery:
Check the output of IC 419: seca 717. 727, 726. 737, 757: 1 Minute
Using an oscilloscope, measure at the relevant pin seca 748: 2 Minutes
for controlling the Segment.
Checktransistor 109:
In the case of mains operation, there must be a
UN voltage of approx. 0.2V at the collector.

If this is not the case. replace transistor 109.


5 -
If there is a voltage of approx. 0.2 V at the collector
4 in the case of rechargeable battery operation. an
incorrect transistor was used, PNP instead of NPN
transistor,
3
On the basis of transistor 109. there is approx. 0.65
i V in the case of mains Operation
2+
If this IS not the case, check LED 106 and resistors
1 + - 107. 108 for correct fitting

i 4 6 8 10 12
I 14 tlms
If the output does not swtch. replace IC 419

23.04.97 -Jen Page 2 303~C0437 Index G


Service Manual
seca 717/727/728/737/748/757 4.2 Fault Description

7. Fault descrlptlon: Senslng l lommt Model 746


mod. 748 doem not tunotlon. Check the lines from the main board to the auxiliary
Check the output of the microprocessor, IC 520: board:
- Start the scale - check the solder points
- there must be a voltage of 5 V at pin 34 Measure Signals 5V, GND. REF. TEST on the
load the scale with 5 kg and restart auxiliary board
’ UP ’ flashes - Measure the signal COM on the main board
- relieve the load on the scale by at least 1 kg If a Signal is not transmitted, replace the line
0 V must be present at pin 34 for approx. 1 sec.
if this is not the case, replace the resistor network Refer to the oscillograms below for checking the
519 or the Signals. The pulse widths of the Signals REF and
mlcroprocessor, IC 520 COM Change with the load.

Checktransistor 324:
approx. 0.2 V must be present at the collector of
transistor 324 UN
load the scale with 5 kg and restart
’ UP ’ flashes
5
- relieve the load on the scale by at least 1 kg
- 5 V must be present at the collector of transistor
4
324 for approx. 1 sec.
- if this is not the case. replace transistor 324
3
Checktransistor 325:
- there must be approx. 12 V (or 9 V in the case of 2
rechargeable battery operation) st the collector of
transistor 325
- load the scale with 5 kg and restart
’ UP ’ flashes
- relieve the load on the scale by at least 1 kg
0.2 V must be present at the collector of transistot
325 for approx. 1 sec.
if this is not the case, replace transistor 325
1

l 4

If the sensing element still does not function,


replace it

8. Feult domxlptlonr Soel. do.8 not


dlspley corroctly
Check the gradient and adjust if necessary, see
manual calibration instructions:
seca 727: No. 30-34-00-446
seca 7371757: No. 30-34-00-464
S=C~748: No. 30-34-00-446

0. Feult ckscrlptlonr Scelo dlsplays


‘EEEE’ l fter l tertlng
Check the fuse 450 and the solder bridge 450 and
replace if necessary
+
3 7 tlms
Check the connections of the weighing cell on the
rna~n board,

Check the zero Point and adjust if necessary see


manual calibratlon instructions:
seca 7271726, No, 30-34-00-446
seca 71717371757: No, 30-34-00-464
seca 740 No 30-34-00-446

23,04.97 -Jen Page 3 30-34-00437 Index G


Service Manual
seca 717i727/72a~737/74a~757 4.2 Fault Description

Check the connection of the forte transducer:


Check the solder Points (for seca 748 s8e drawing
06-06-04-541).

If all lines are in Order, there is a fault in the AD


converter or the microprocessor is not operating
correctly.

When replacing the AD converter auxiliary board in


model 748, the sc& must be calibrated in
accordance with the calibration instructions 30-34-
00-446 and the adjustment instructions 30-34-00.
467.

23.04.97 - Jen Page 4 30-34-rXS437 Index G


Service Manual
seca 748 4.3 Replacement Instructions

1. Open thr scalr 3. Roplaco tho oallbrstod l lmctronlcm wlth


1.1 Unplug the low-voltage connector from the display thr wmighlng toll, basr plsto and brldgo
housing. holdw.

1.2Press the 4 cover taps (1) out of the housing (2) 3.1 Turn the scale and undo the 5 screws (4) on the
and undo the screws (3) Remove the housing (2). cover plate (5). Remove the cover plate (5).
Turn the scale back again.
1.3Undo the 3 fastening screws (6) and rernove the
rear panel (7). 3.2 Undo the two fastening screws (11) with the nuts
(12) on the housing carrier (13) and the base plate
1.4Undo the 4 fastening screws (6) in the display (14). Remove the housing carrier (13) and the
housing (9) in Order to rernove the display board connecting cable (15).
(10).

4. Seal. assombly.
2 Fault soaroh and roplacrmrnt of
rlrctronlo oomponwnts 4.1 Fit the new calibrated electronics (16) with housing
carrier anr connecting cable (15) in reverse order to
‘LIDepending on the fault type. perform the fault that described in points 3.3-3.2.
search in accordance with the fault description The housing carrier and the base plate must be
30-4c-0&437. parallel to each other.

4.2Align the display board (10) in the display housing


(9) and assemble the scale in reverse order that
described in Points 1.2 - 1.4.

4.3Adjust and inspect the scale in accordance with the


calibration instructions 30-34-00-446

4.4Seal the AD converter board (17). apart from the 3


adjustment potentiometers. with the Silicone casting
compound Sylgard, components A BE.

4.5Recalibrate the scale.

4.6 Seal the 3 adjustment potentiometers on the AD


converter board (17) with the silicone casting
compound Sylgard A 8 B.

4.7Close the cover (18).

12.03.97.Hart Page 1 (3) 30-34-00-489 Index A


L
Service Manual
seca 748 4.4 Adjustment Instructions

Asoembly and final inspection


Completely assemble the scale.
Insert the rechargeable battery into the holder provided for it.
Connect the scale to the plug-in supply unit and switch on.
‘0.000’ appears in the display.

Lateral Error

Start the unloaded scale. ’ 0.0 0 f? appears in the display. Next. load the Center of the scale and all
four corners with 3 kg in turn. The displayed value must not differ by more than 5 g.

Maximum Weighing Range


Start the scale with a preload of 3.5 kg and an additional load of 9.5 kg. The preload is intended to
simulate the bed in the incubator.
‘- u P-’ appears in the display.
Relieve 1 kg load from the scale. ’ 0.0 0 0’ appears in the display shortly afterwards.
Load the scale with 3 kg. 3.OCxlf 5g appears in the display.
If the scale is loaded with more than 3.300 kg fl00 g. ’ 5 T 0 P’ must appear in the display.
If this does not happen, the zero Point is incorrectly set.

Gradient

Load the scale with the preload and a load of 0.5 kg and Start it. ‘-IJ P-’ appears in the display.
Relieve 0.5 kg of load from the scale. ’ 0.0 0 0’ appears in the display shortly afterwards.
Load the scale with 9.99 kg and set the gradient with Potentiometer SO4until
F = 9.99 kg, display = 9.990 kg.
Relieve the scale and restart with a preload and 0.5 kg. ‘-UP-’ appears in the display.
Relieve the scale of 0.5 kg load. ’ 0 0 0 0’ appears in the display shortly afterwards.
Load the scale with 9.99 kg and check the gradient.
F = 9.99 kg, display = 9.990 kg.
Repeat this procedure until the gradient check is successful.

Linearity
Load with a preload and 0.5 kg and Start it. ‘- u P-’ appears in the display.
Relieve 0.5 kg of load from the scale. ’ 0 0 0 0' appears in the display shortly aftetiards.
Load the scale in the following Steps and check the following tolerantes:

F= Okg, display = O.OCKlkg f 5 g


F= 5kg. display = 5.OKl kg + 5 g
F= 9kg, display = Q.OOflkg f 5 g

23.04.96 - Jen Page l(2) 30-34-00-487 Index: E


Service Manual
seca 748 4.4 Adjustment Instructions

Checking the gradient and the Hold function

Load the scale with the preload and the load of 0.5 kg and Start it. ‘-u P-’ appears in the display.
Relieve 0.5 kg of load from the scale. * 0.0 0 0’ appears in the display shortly afterwards.
Load the scale with 9.99 kg and check the gradient.
F = 9.99 kg, display = 9.990 kg.
If the gradient is not correct, it must be Set again.
Press Hold and relieve the scale. The displayed value must remain and the Hold LED must flash.
Press Hold again: The scale must now display O.ooO kg again.

Rechargeable battery charging current


Remove the rechargeable battery insert and Start the scale.
Connect the plug-in power supply unit of the test circuit to the mains --> the green LED lights up.
Insert the test circuit into the rechargeable battery compartment, press gently until the small red LED
lights up.
Error criterion The battery charging current is correct. when the green and small red LED light
up.
The battery charging current is too high when both red LEDs lights UP.
If the small red LED does not light up, there is still no contact to the rechargeable
battery connection.
Remove the test circuit and replace the rechargeable battery ccmpartment.

Charge the rechargeable batteries


Connect the scale to the mains for at least 20 hours with the plug.-in power supply unit or Charge the
rechargeable batteries wfth an external charger for 10 hours with 11 mA.

23.04.96 Jen Page 2(2) 3c-34-CO487 Index: E


Service Manual
seca 748 4.5 Spare Parts list

l .ca 748 Valid from: October 1996

Item Ani&-No. Designation Price category

01 50-00-06-467 Fillister-head screw DIN 64 - M6 x 16 14

02 01-13-05-270 Dummy plug 02

03 02-12-04-246 Housing 41

04 060406317c Front film 38

05 01-10-04-005 Levelling unit 19


01-10-04209 Foot screw M6
50-90-00-550 Knurled button M6
50-00-93-460 Hexagon nut M6
50-90-00-210 Star leck

06 02-03-01-246 Bridge support 30

07 06-04-05-2658 Display housing 36

08 06-0616-017 Gable harness, battery holder 27

09 08-0407-244 Aear panel 20

10 04-02-05-203 Housing support 46

11 01-16-02-251 Cover plate 22

12 50-00-06-419 Fillister-head screw DIN 64 - M3 x 6 A2 03

13 0600 14.202 Cover for battery compartment 17

14 50-00-92-110 Fiat-head SCreW DIN 921 M4 X 10 5.6 VZ 11

15 01-13-04-216 Captive washer 08

16 66-21-06-090 Rechargeable battery 9V Block Ni-Cd 29

17 66-32-10-256 Plug-in power supply unit 36

16 06-0404-114 Electronics. calibrated 70

19 56-01-01-060 Cover compound. Sylgard (0.1 kg) 31

28.01.97.Hart Page 1 30-34-00436 Index C


Service Manual
seca 748 4.5 Spare Parts list

Item Article-No. Designation Price categoiy

20 08-06-16-069 Gable harness, power supply unit 26

21 65-90-12-020 Pieze sensing element 25

22 61-70-67-511 IC microcontroller 32

23 1605-01-639 Information plate 10

24 14-05-01-640 Information plate 10

25 14-05-01-669 Information plate 10

26 01-13-05353 Sliding washer 14

Page 2 30-34-00436 Index C


r-
N

x-

t
Service Manual
seca 748 5 Electronics

Electronics
Service Manual
seca 748 5.1 Principle circuit diagram

display

microcomputer
-

A/D-converter

24.04.97 Hart 0809-0 1-278


Service Manual
seca 748 6 Appendix

Appendix
seca 748 Manual calibration instructions

l Remove the positive supply voltage (still under 6V) from the cathode of diode 102 and
connect to the anode of diode 110 (+RECHARGEABLE BATTERY).

l Start the scale.

- “bAtt” must now appear again afier the scale is started. If necessary, the supply voltage
must still be regulated down slightly. The scale switches itself off afler 2 seconds, otherwise
the power supply recognition function is faulty.

l Reconnect the positive supply voltage to the cathode of diode 102 and increase to 12V.

l Remove the voltage measurement unit fiom capacitor 5 18.

Checking the rechargeable battery charging current4 and Segment test

l Connect the ammeter to the anode of 110 (+RECHARGEABLE BATTERY ) and the
ground to the anode of 102 and Start the module.

l After starting, check whether all Segments and the Hold LED light up during the Segment
test.

l Check the rechargeable battery charging current.

- The rechargeable battery charging current may be in the range

7.0mA < Iaccu < 9.0mA

Otherwise, calibrate resistor R103 accordingly,

- The indicator lamp diode for the rechargeable battev charging must light up

Ageing

l Set the module to continuous Operation with bridge 618. Put assembled boards with
weighing cells into the bum-in chamber, switch on and nm for at least 36 hours in lO-
second cycles.

Calibrating the A/D converter

l Move all trimmers into the Center Position.

l The intemal resolution is displayed with 5 digits. Therefore, plug in the 7-Segment display at
Position 30 1.

4 NOI necessaryifinsircuit testwascarried out first.


23.04.96 -JEN Page 2(6) 30-34-00-446 Index: F
seca 748 Manual calibration instructions

l DO not set any program bridge (intemal resolution with zero Point display).

l Insert a zero Point decade for resistor 509 and gradient decade for resistor 5 14. The
decades should be set to values obtained in the previous calibrations J.

l Place a preload of 850 g onto the scale.

Zero Point presetting

l Start the module

l A weight value or “EEEE” (zero Point outside the limits) appears in the display.

l Actuate the zero Point decade until the display Shows a value of between 3000 and 6000

l If “EEEE” is still in the display, the Zero Point decade for the resistor 508 must be inserted
and then adjusted.

Gradient calibration6

l DO not set any solder bridge l Set solder bridge 609


l Start the module l Start the module
l The intemal zero Point is in the display.
Note this value, or enter in the
Computer.
l Add another 8 kg of load to the scale in . Add another 8 kg of load to the scale in
addition to the preload. Also note this addition to the preload.
value or enter it in the computer.
l Calculate the value to be set as follows l Set the display to precisely 8kg ? log
(with pocket calculator program) with the gradient decade.
16000
V = value =
zero Point value
l-
8kg value
l Precisely set the calculated value to +20 l Relieve the load from the scale apart
Counts with the gradient decade. With from the preload, restart the module,
good linearity, the gradient is now set check the gradient and if necessaty
suficiently accurately. readjust the decade.

5 Ifno values arc known, the zwo point decadeshould be setto lh4.Q and the gradient decadeto 5Okn
6 Calibration tan be in both countsand kg. Refer to the lefi column for calibration in count.sand to the tight
column for calibration in kg.

23.04.96 -JEN Page 3(6) 30-34-00-446 Index: F


seca 748 Manual calibration instructions

Zero Point calibration

l Relieve the load from the scale apart from the preload of 850 kg

l DO not set any solder bridge.

l Correct the zero Point with the zero Point decade until:

Display at 3600 (tolerante: -100 / +50)

l If it is not possible to adjust the value with the gradient decade at 509, the zero Point decade
must be connected at 508 and the calibration must be repeated as from the zero Point
presetting.

l Solder in values of the two decades as metal-film resistors (TK _ 50 ppm) from series E96
(next due).

Important: allow the resistorsto cool down for approx. 1


minute.

Checking the zero Point and gradient

l DO not set any solder bridge

l Restart the module if necessaty

- Check the zero Point:

Display at 3600 (tolerante: -100 / +50)

l Set program bridge 609.

l If “0.000” is not in the display, restart the module.

l Load the scale with 8 kg,

- Check the gradient:

F=Skg Display = 8.000 kg f log

Important: In the case of values greater than 9.995 kg, “STOP” appears in the display.

l Relieve the scale of 8 kg load

23.04.96 -JEN Page 4(6) 30-34-00-446 Index: F


seca 748 Manual calibration instructions

F=Okg T = 40°C FOT40K = (Steps)

. Set program bridge 609. ’

. Load the scale with 8 kg in addition to the preload.

. Note the following values:

F=Skg T = 40°C FST40K = .<,._(kg)

. Put the module into the climate chamber 1OT

. Remove the program bridge 609. (Intemal resolution with zero Point display).

. Apply a preload of 85Og.

. Start the scale.

. Note the following values:

F=Okg T = 10°C FOTlOK = .___,(Steps)

. Set the program bridge 609.8

. Load the scale with 8 kg in addition to the preload.

. Note the following vahres:

F=Skg T = 10°C FSTlOK = ._.<.(kg)

The following tolerantes arc permissible:

Zero Point drift:

1 FOT4OK - FOTIOK 1 I 100 counts

Gradient drift:

1 FST40K - FSTlOK 1 I5g

. If the module is outside the permissible error, the drift must be compensated for. Then
repeat the temperature test.

. Reset bridge 609 and finish the module.

7 Altematively,another program modetan be selectedwhich offers the Sameresolution.


s Alternatively,another program modetan be selectedwhich offen the sameresolution.

23.04.96 -JEN Page 6(6) 30-34-00-446 Index: F


Alle Rechte an und aus diaren Unterlagen verbleiben Vogel 8 Halke
Oatul,~ame : 06.06.91 Oeh 07.04.95 Jen 12.11.96 Jen 8.1.97 Jen
Acndcrung : 1967 2719 2909 2918
Index : A E C d

r seca Eenennung : Nr.:


wjring diagram
mlcroprocessor 08-01-21-322
HAMBURG 717//727/728/737/748/757
- W-A
- REF
TEST

Alle Flechte an und aus diesen Unterlagen verbleiben Vogel P Halke


oatu. : 09.11.1990 Oatum/Name : 16.10.93 Oeh 07.04.94 Och 10.4.95 Jen 8.1.97 Jen 23.4.97 Jen
Gezeichnet : Zcssin Aenderung : 2492 2596 2719 2910 2938
Ceprueft : Index : C 0 E F G
Stueckliste:

Nr.:
I
08-01-21-321
HAMBURG 717/727/728/737/748/757
DC-ADAP

VIN+

VVVY power suppiy 08-01-2-354


HAMBURG 717/727/728/737/748/757
Alle Rechte an und aus diesen Unterlagen verbleiben Vogel L Halke

08-03-01-022
cf-bleiben Vogel P Halke

Nr.: 08-03-01-006
component mounting diagram
HAMBURG analog-digital-converter