My Terragene’
November Tith, 2021
To whom it may concern
We, TERRAGENE S.A, located in Alvear, Santa Fe, Argentina, hereby declare that the
following specifications apply for IG/OG/PQ Process validation schemes for
Terragene Bionova” IC10/20FR Auto-readers.
Installation qualification (IQ) specifications
1Q Spe
Terragene does no specify any special requirements for the installation of Bionova*
IC10/20FR Auto-readers. The installation and start-up of the Auto-reader is carried
out by the user by following the steps detailed on the user manual.
ations applicable for the device: NO.
Operational qualification (OQ) specifications
04 Specifications applicable for the device: NO
Bionova® IC10/20FR Auto-reader performs automated internal tests to verify that the
device is working according specifications. If the tests reveal that the device cannot
perform its intended use correctly, the device will not allow the user to proceed with
an incubation, either by canceling a single incubation position or by canceling the
possibility to perform incubations on the device altogether. Such findings are
informed to the user by an audible & visual alarm and a printed ticket.
Performance qualification (PQ) specifications
PQ Specifications applicable for the device: NO
Bionova* ICIO/20FR Auto-reader performs automated internal tests to verify that the
intended use of the device can be correctly carried cut at any moment. The
automated tests verify that both the readout system and the incubation temperature
are according to specifications. Also, the use of positive control indicators is
recommended on the Auto-reader’s User manual. A positive control is anon-
sterilized SCBI used as a reference during the incubation process. The use of a
positive control is a recommended practice as it helps to ensure: + The correct
incubation temperature is reached. + Viability of spores has not been altered due to
improper storage temperature, humidity, or proximity to chemicals. + The aptitude of
the media to promote rapid growth and fiuorescence generation. + Proper
functioning of the incubator. Please refer to the Auto-reader User manual and to the
Biological Indicator instructions for use for more information,