CALIBRATION CERTIFICATE & INSPECTION CERTIFICATE
OF
MASTERBALL CALIBRATION UNIT
CARL ZEISS INDIA (BANGALORE) PVT. LTD.,Page _1__of _1_pages
N Certificate No, : 19-999335EP
4 < INSPECTION Control Number : AQ02679700
ELH! CERTIFICATE Date : September 17,2019
ACRE
roby Set
Customer ; CARL ZEISS INDIA (BANGALORE) PVT. LTD...
Equipment Name ; MASTERBALL CALIBRATION UNIT.
‘Type :
Quantity: 1
Serial Number ; 999335EP
Calibration Date: September 17,2019
Note) Accuracy is guranteed under the 20°C temperature.
Certified by Measurement made by
TOKYO SEIMITSU CO., LTD.
Metrology Company, Quality Assurance Division
4-2, Higashinakanuki-machi, Tsuchiura-city, Ibaraki, Japan
Tel: 81(Japan)-29-831-1234 Fax: $1(Japan)-29-831-8051
DT78034-T001-00 1/1 TOKYO SEIMITSU CO.,LTD.Page | __of _2 pages
\ Certificate No. : 19-999335EP_
A CERTIFICATE OF Control Number: A002679700
HUES CALIBRATION Date : September 17,2019
eee
a
Customer : CARL ZEISS INDIA (BANGALORE) PVT. LTD.
Equipment Name : MASTERBALL CALIBRATION UNIT
Type
MC-S65B,
Quantity : 1
Serial Number : 999335EP
Manufacture ; TOKYO SEIMITSU CO.,LTD
Environmental
artimay : Temperature : 20°C £0.5°C Humiditty : 50% 10%
Calibration Date: September 17,2019
Calibration standard used
Vieasaring Speciicaions Conver Calibration Calibration Laboratory
Instruments (Type) No. Validity (Manufacturer)
Surfeom s1400D 79.2201 | —Mar-2020 [Tokyo Seimitsu
Optical Flat 0130 792065 | Mar-2022 |JQA
Rondcom ROSA 79-2235 | Dee-2020 [Tokyo Seimitsu
Gauge Blocks Class TQ-2026 | Oct-2019 [Tokyo Seimitsu
Electronies Micrometer] E-MD-MSI 79-0223 | _Feb-2020 [Tokyo Seimitsu
‘The measuring instruments used are traceable to the national measurement standards controlled by
National Institute of Advanced Industrial Science and Technology of Japan.
‘The above product was calibrated in compliance with a quality system registered to ISO 9001. using
applicable Tokyo Seimitsu's procedure based on ISO.
We certify that the result of the calibration is in accordance with the following pages.
This certificate shall not be reproduced except in full, without the written approval of Tokyo
Seimitsu Co., Ltd.
Higashinakanuki-machi, Tsuchiura-city, Ibaraki, Japan
TOKYO SEIMITSU CO., LTD.
Metrology Company, Quality Assurance Division
Quality Control Manager
_—a beste
TSQ3833 TOKYO SEIMITSU CO., LTD.Page 2 of 2 pages
Type
Results of measurement :
CERTIFICATE OF CALIBRATION
RESULTS OF MEASUREMENT
E-MC-S65B
Certificate No. : 1
199335EP,
Control Number : A002679700
Serial Number : 999335EP
Calibration Procedure : TOKYO SEIMITSU engineering standard DT78034-K0O1
according to JIS B07506 Electric micrometer
No tem Result Remarks
1 [Diameter of ball 12.7003 mm
2 |Roundness (P-P) 0.03 um
5 |Ftamess of al reference bases 0.3 um
TghraTTaTCe
* Toetwieen center base and right base passdb
Measuring Condit
No tem Instrument Measuring condition
Block Gauge
1 |Diameter of ball
Etccronies Micromet
Stylus: 0.8 mm radius Sphere
2 |Roundness (P-P) RONDCOM 654 } Filter: SO (up). Speed: 4 /min
MZC : Minimum zone reference cirles
3 |Ftamess of al reference bases Optical Flat
eighr arene
a [between center base and right bas ne
TSQ3835
TOKYO SEIMITSU CO., LTD.a) Jess Page 1 of 2 pages
aes Certificate Number TQ19676
Joss 0047
Tois67
~~ J6880047
Certificate of Calibration
Aalepan _
2019-09
Customer : CARL ZEISS INDIA (BANGALORE) PVT. LTD.,
“Ades PLOT NO33, JIGANI LINK ROAD, BOMMASANDRA INDSUTRIAL
AREA, BANGALORE - 560099, INDIA
Object : Gauge blocks
Quantity : 3
Serial Number 2 88740,86654,87021
Manufacturer : KOLB & BAUMANN GMBH & CO. KG
Calibration em; Length
Comparison Measurement with Calibration Procedure(YQ5181 1)
by TOKYO SEIMITSU CO., LTD.
Reference Standard(s) : —TQ2392,7Q2226 oe -
Calibration Method
4-2, Higashinakanuki-machi, Tsuchiura-shi, Ibaraki, Japan
TOKYO SEIMITSU CO., LTD. Tsuchiura Plant Calibration Laboratory
Calibration Site
Environmental Condition 20°C 0.5 °C ( Room temperature )
of Calibration Room = * 50% + 10% ( Relative humidity )
Date of Calibration; September 13, 2019
We certify that the result of calibration is in accordance with the next page.
Date of Issue : September 17, 2019
4-2, Higashinakanuki-machi, Tsuchiura-shi, Ibaraki, Japan
TOKYO SEIMITSU CO., LTD.
Metrology Company, Tsuchiura Plant, Quality Assurance Department
Calibration Laboratory
‘This certificate is based on article 144 of the Measurement Law and indicates the result of calibration in
accordance with measurement standards traceable to Primary Measurement Standards (National Standards)
which realizes the physical units of measurement according to the International System of Units (SI). The
accreditation symbol is attestation of which the result of calibration is treceable to Primary Measurement
Standards (National Standards).
‘The certificate shall not be reproduced except in full, without the written approval ofthe issuing laboratory,
‘The ealibration laboratory who issued this calibration certificate conforms to ISO/IEC 17025.
This calibration certificate was issued by the calibration laboratory accredited by IAJapan who is a signatory to
the Mutual Recognition Arrangement (MRA) of International Laboratory Accreditation Cooperation (ILAC)
and Asia Pacifie Laboratory Accreditation Cooperation (APLAC). This (These) calibration result(s) may be
accepted internationally through ILAC/APLAC MRA. Calibration label is to identify calibration status easily by
attaching same information of Calibration Certificate to the calibrated item,
TSQaz10P TOKYO SEIMITSU CO.,LTD.ew, ule) css Page 2 of 2 pages
seme Certificate Number TQ19676
“eS JOSS 0047
Results of Calibration
Nominal Seri) PDevation of each Tengih fom nominal length Gam] Variation
length Number | *i€ eal pont [tthe highest point] atthe owes point | in length Remark
Zn (mm) ALe ALmax 4Lmin V (um)
4 88740 -0.03 -0.02 -0.08 0.06
16 86654 “0.11 0.11 0.16 0.05
40 87021 0.00 0.00 -0.03 0.03
1) Expanded uncertainty of the calibration
Expanded uncertainty of the length deviations from the no
from 0.5 mm up to 100 mm U = 0.08 um
‘over 100 mm up to 250 mm. U = (0.05 + £ £2500) um
(L Nominal length of calibrated gauge block ( mm ))
‘The above reported expanded uncertainty of the calibration is stated as the standard uncertainty of calibration
multiplied by the coverage factor & = 2 such that, for an approximately normal distribution, the coverage
probability corresponds to approximately 95 %.
2) The calibrated lengths are values on 20 °C.
3) Material and coefficient of thermal expansion
“Material of object is steel with rich carbon and crome.
Coefficient of thermal expansion is (11.9 #0.5 )*10°/K
(The coefficient is not the actual number but the number by the gage suppliet:)
4) Symbols used inthe table (see the figure below.)
Abmax ‘Nominal length at the central point Ln
lengths are as follows.
Length at the central point Le
Length at the highest point Emax
erin Deviation from nominal length at the central point ALe=Le-Ln
al t Deviation from nominal length at the highest point AL max~£max-Ln
L ! Deaton om nominal engh she loves point AL n=Lninen
Variation in length VoL max-Lmin=AL max-AL- min
‘Note: Variation in length is only a reference calculated with ALmax and ALmin, nd
TSQ32175 TOKYO SEIMITSU CO., LTD.What is JCSS ?
Confident Evidence of
‘Traceability and Competence!
Japan libration Service System
Jucss is...
the Calibration Laboratory Accreditation System based on the Japanese Measurement
Law in order to ensure the metrological traceability in Japan.
IAJapan, National Institute of Technology and Evaluation, Accreditation Body of JCSS
assesses and approves the technical competence and conformity to the requirements of
ISO/EC17025U1SQ17025) for calibration laboratories.
[CSS Calibration Certificate means...
JCSS Calibration Certificates issued by JCSS Laboratories, accredited by [AJapan under
the ILAC and APLAC MRA, certify formally the metrological traceability to National
Measurement Standards of Japan provided by NMIJ.
There are no needs to trace back to upper traceability hierarchy if you are with a JCSS
Calibration Certificate.
A JCSS Calibration Certificate is more confident “Evidence of Traceability" with
Technical Competence and an appropriate Measurement Uncertainty than a Certificate
with Legal metrology, QMS Certification, and Manufacturers’ self declaration.
As lAJapan is one of signatories for ILAC and APLAC MRA, JCSS calibration
certificates are equivalent to those of laboratories accredited by other ILAC and
APLAC MRA signatories, such as NVLAP, A2LA, UKAS, DKD, NATA, etc.
Also for calibration laboratories with overseas metrological traceability, [AJapan assesses
‘them and grants accreditation under “ASNITE program.”
|S LAC & APLAC MRA
NATA
: (Grange 085 Curatn Canta)
Sas
1AJapan, National Institute of Technology and Evaluation
2-49-10 Naina Seba ku, Tokyo 151