PT.

Draeger Medical Indonesia Phone


+62 21 8066 9030
Technical Data E-mail
BHP@draeger.com
Lembar Data Spesifikasi

Spesifikasi
Seattle PAP Bubble CPAP System With Compressor

Helping infants with respiratory distress breathe easier. The Seattle-Positive Airway Pressure (PAP) system is an innovation which uses
the proven advantages of Bubble CPAP therapy, such as oscillatory effects similar to high frequency ventilation
combined with a unique design. Affordable and easy healthcare for those who need it most.

Oxymixer
Main flow outlet 0 – 15 l/min
Secondary flow outlet with bleed 0 – 3 l/min
Bleed flow ≤ 3 l/min at 3.4 bar (49.3 psi)
Maximum combined flow (all outlets) ≥ 30 l/min
Emergency flow (in the event of the air or oxygen supply failing) >15 l/min
Alarm signal in the event of loss of supply pressure Alarm on when the pressure difference of both gases is
0.9 – 1.8 bar (13 – 26.1 psi).
Alarm volume ≥80 dB at a distance of 30 cm
Adjustment range of oxygen concentration 21 – 100%
Gas supply pressure 3.2 – 6.5 bar (46.4 – 94.3 psi) air and oxygen with max. 0.7 bar
(10.1 psi) pressure difference between the two
Accuracy of the gas mixture (FIO2) ± 3 % oxygen
Connection types DISS outlets for gas mixture and NIST inlets for air and oxygen
Dimensions LxWxH 13 x 16.5 x 12.2 cm (5.1 x 6.5 x 4.8 inch)
Weight approx 1,600 g
Operating temperature +5 °C to +50 °C (41 to 122 °F)

Seattle PAP
Length of breathing hoses 1.8 m (70.86 in), including 40 cm
(15.74 in) hose extension
Material
Breathing hoses PP, EVA, ABS
Humidifier chamber PC, SBC, silicone, aluminum
Connector PP
Pressure relief valve PC, PE, PP, stainless steel
Bubble CPAP chamber PP, POM
O2 hose PVC (DEHP free)

Performance characteristics
Insp. resistance at 15 L/min <2 mbar* (or hPa or cmH2O)
Compliance at 60 mbar <1 mL/hPa (or mL/mbar)
Leakage at 60 mbar <50 mL/min
Electrical connection data 22 Ohm

Humidifier chamber (N)


Humidifier output at 4 to 12 L/min >10 mg/L
Volume (air) of humidifier chamber with water/without water 190/300 mL
Warm-up time 30 min
Ambient conditions
During storage
Temperature –20 to 60 °C (–4 to 140 °F)
Relative humidity 5 to 95 % (non-condensing)
Ambient pressure 500 to 1200 hPa (7.3 to 17.4 psi)
During operation
Temperature 18 to 26 °C (64 to 79 °F)
Hose extention for use in the incubator 18 to 40 °C (64.4 to 104 °F)
Relative humidity 10 % to 90 % (non-condensing)
Inside the incubator 10 % to 95 % (non-condensing)
Ambient pressure 700 to 1060 hPa (10.2 to 15.4 psi)
Classification according to Directive 93/42/EEC Annex IX
Class IIa
UMDNS code Universal Medical 11001
Device Nomenclature System –
Nomenclature for medical devices
Protection class Type BF

Include Set :
Oxymixer Low Flow 1 Unit
Seattle PAP System 1 Box
Trolley set 1 Set
O2 Hose 1 Pcs
Air Hose 1 Pcs
Humidifier Set 1 Set
Medical Compressor 1 Set
Interface Babyflow Demo Package 1 Set

PT. Draeger Medical Indonesia Alamanda Tower – 32nd floor Jl. TB Simatupang 23-24 Jakarta, 12430, Indonesia

Tel: +62 21 8066 9030

Fax : +62 21 8066 9099

www.draeger.com