DR-ID 800CL Service Manual: IN: Installation
DR-ID 800CL Service Manual: IN: Installation
DR-ID 800CL Service Manual: IN: Installation
IN: Installation
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-I
Control Sheet
Issue date Revision number Reason Pages affected
08/30/2013 03 Revision for V7.3 (FM6189) 4, 6, 10, 13–15, 19, 21, 22, 24–30, 32, 36–38
11/30/2014 06 Revision for V8.0 (FM6261) 4, 6, 10, 11, 19–24, 31, 34, 36, 40
11/30/2014 06 Changes in pagination (FM6261) 25–30, 32, 33, 35, 37–39, 41–46
05/31/2016 08 Revision for V9.0 (FM9412) 1, 4–8, 11–15, 21–24, 26–29, 40, 41, 47
05/31/2016 08 Changes in pagination (FM9412) 9, 10, 16–20, 25, 30–39, 42–46, 48–50
11/22/2017 09 Revision for V11.0 (FM9492) 1-2, 4-6, 13, 15, 18-19, 25-27, 30-31, 39-40, 43
11/22/2017 09 Changes in pagination (FM9492) 7-12, 14, 16-17, 20-24, 28-29, 32-38, 41-42, 44-54
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-II
1. DR-ID 800CL 1.3 Places of Installation
Installation Conditions ■ Environmental conditions
Avoid installing the machine at the following places:
This section states the PC and monitor installation conditions.
Places where the temperature changes drastically
NOTE Places near heaters or other heat sources
Listed below are the abbreviations adopted in this chapter. Places where the machine is likely to be wet or submerged
Places where the machine is likely to come into contact with any
Application : AP corrosive gas
DR-ID 800CL configuration data : Config Dusty or dirty places
Places where the machine is subjected to constant vibration or shock
For the dimensions and weights of the peripherals, see the respective Places where the machine is subjected to excessive vibration or
manuals that are supplied with them. shock
Places exposed to direct sunlight
1.1 Dimensions and Weight Also, be sure that the monitor is not installed at the following places:
➥ REFERENCE
The above conditions are to fulfill requirements for both PC and monitor.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-1
2. Installation Procedure Overview 1. Check the supplied component items, install
the DR-ID 800CL accessories, and make cable
connections to the FPD.
➤ IMPORTANT
Be sure that the following equipment is completely installed and set up.
2. Set up the software related to the DR-ID 800CL
- FPD
and FPD. The major procedures are indicated
- Other equipments
below:
$ See the installation instructions set forth in the respective
Make preparations for DR-ID 800CL-AP
service manuals.
startup (e.g., driver, option key, and AP key
Be sure to prepare the following before installation. installation).
- DR-ID 800CL-AP key (Disk/License)*1 Set up the software for FPD operations
- USB-connected disk drive (JIG) (e.g., FPD-AP installation).
- USB-connected mouse (JIG)*2 Set up the DR-ID 800CL system information
- USB-connected keyboard (JIG)*2 (e.g., hospital name).
*1 In case of the license, follow the procedures to acquire. Perform Config setup that is needed for the
*2 As necessary, remove the USB adapter cover at the front and the connection of the other equipment.
access point before connecting directly to the PC.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-2
3. DR-ID 800CL Hardware Setup
010-307-01E
12.14.2012 FM6115
DR-ID 800CL Service Manual IN-3
3.1 Checking the Component Items Product code Description Qty. Remarks*1
DX CL GENERAL RIS General RIS Browser key
Check that the following component items are fully supplied. 1 O
BROWSER (Disk/License)
AP (Disk) Virtual Grid function key
DX CL VIRTUAL GRID 1 O
DR-ID 800CL Standard Kit Software (Disk) (Disk/License)
Operation Manual (Disk) Dynamic Visualization II function
DX CL DYN VISUAL II UPG SWL 1 O
HDD recovery media (Disk) key (License)
DR-ID 800CL Reference Guide (Disk) Image Combining Processing
DX CL IM COMPOSITION LC E function key (License) 1 O
■ Others (separately packed items, options, etc.) (V12.0 or later for US)
Bluetooth barcode reader
Product code Description Qty. Remarks*1 DR-XD 1000 BT BCR KIT 1 O
(OPTOELECTRONICS OPN-3002i)
DX CL DICOM PRINT DICOM print key (Disk/License)*2 1 O *1
O : Option
*2
DX CL DICOM ORDER MWM
DICOM worklist (MWM) key
1 O Configuration option of DX CL 800CL OP KIT
(Disk/License)*2
DICOM worklist (MWM) key
DX CL DICOM ID MWM (Disk/License) 1 O
(patient information only)
DX CL DICOM MPPS DICOM PPS key (Disk/License) 1 O
DICOM Commitment Push key
DX CL DICOM COMMITMENT 1 O
(Disk/License)
Shuttering processing key
DX CL SHUTTER-PROC 1 O
(Disk/License)*2
Grid pattern removal processing
DX CL GPR 1 O
key (Disk/License)*2
Retake Analysis key
DX CL RETAKE ANALYSIS 1 O
(Disk/License)*2
Image Reader QC key
DX CL RUQC 1 O
(Disk/License)
Simple ordering key (Disk/
DX CL SIMPLE ORDERING 1 O
License)
General exposure motion
DX CL MOTION DETECTION detection function key (Disk/ 1 O
License)
DX CL REMOTE IMAGE Remote Image Display function
1 O
DISPLAY key (Disk/License)
010-307-10E
11.30.2018 FM9540
DR-ID 800CL Service Manual IN-4
3.2 Installing the Available Items 3.3 Making a Cable Connection to the FPD
Mount the available items on the PC. Use a SE cable to make connection between the IU and FPD.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-5
3.4 Connecting to the Other Equipment
Connect the following equipment to the DR-ID 800CL.
● DICOM server
$ See “1. Hardware Setup” under “Appendix 10 DICOM CR
Storage” of “Appx IN”.
● DICOM DX Storage
$ See “1. Hardware Setup” under “Appendix 11 DICOM DX
Storage” of “Appx IN”.
● DICOM printer
$ See “1. Hardware Setup” under “Appendix 12 DICOM Print” of
“Appx IN”.
● RIS
$ See “1. Hardware Setup” under “Appendix 13 RIS: DICOM
MWM/DICOM PPS” of “Appx IN”.
● X-CON
$ See “1. Hardware Setup [Setting Done Prior to Shipment]”
under “Appendix 19 X-Ray Controller” of “Appx IN”.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-6
4. DR-ID 800CL Software Setup 4.10 Setting the DX Parameter
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-7
4.12 Changing the Password of the Service Utility 4.1 DR-ID 800CL-AP Startup
(V8.0 or Later)
The following procedures must be performed for DR-ID 800CL-AP
startup:
4.13 Batch setting of VG parameter (V9.0 or Later) OS startup
OS setup
Touch panel monitor setups/operational verification
Option key installation
AP key reinstallation
AP startup verification
➥ REFERENCE
The following OS setup and software installation steps are completed at
factory prior to shipment:
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-8
4.1.1 OS Startup
NOTE
When the DR-ID 800CL is started for the first time, the following window
may open. Place a checkmark to the “Public Network” and click
.
DXL03070009.ai
CLP74010.EPS
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-9
3. Click [Login]. 4.1.2 OS Setup
➔ The DR-ID 800CL-AP starts running.
Perform the following OS-related setup.
Setting the date, time and time zone.
DCL41002.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-10
4.1.3 Touch Panel Monitor Setups/ ■ Checking the contrast and brightness settings
Operational Verification
Check the contrast and brightness of the display.
Set the touch panel monitor and verify its operation.
NOTE
The shape and name of the icon may vary with the versions of Intel(R)
■ Verifying touch panel Graphics and Media. When a different version is installed, perform the
Verify the the touch panel. work by replacing the names in the following procedures.
1. Touch at the upper right corner of the window. 1. Select “Control Panel” from the menu.
➔ The luncher window opens. ➔ The “All Control Panel Items” window opens.
➔ Soft keyboard opens. ➔ The “Intel(R) Graphics and Media Control Panel” window opens.
3. Touch on soft keyboard. 3. Select “Color Correction” tab and check the set up data is
as follows.
➔ The Windows Start menu opens.
I.
5. Touch [OK].
II.
➔ The system returns to the desktop screen.
III.
IV.
D80C010704001.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-11
4. Close the “Intel(R) Graphics and Media Control Panel” window. 4.1.4 Setup for Addition of Descriptive Characters to
➔ The system returns to the “All Control Panel Items” window. Icons [Applicable only in Japan]
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-12
010-307-10E
11.30.2018 FM9540
DR-ID 800CL Service Manual IN-13
4.1.5 Option Key Installation Function Product Abbreviation
DX CL DICOM ORDER MWM DICOM worklist function*1
The following option keys are available. Install necessary option keys in
compliance with the DR-ID 800CL installation specifications. DX CL DICOM ID MWM DICOM worklist function (ac
Acquisition of patient information
DICOM PPS function
DX CL DICOM MPPS
➥ REFERENCE To be installed basically in co
DX CL RETAKE ANALYSIS A function that stores/calcula
• For V8.1 or earlier, or licence unsupported version in V9.0 or later
DX CL RUQC Image Reader QC function
- After option key installation, be sure to install the AP key. The reason
is that installing the option keys disables the AP key that has been Other DX CL MOTION DETECTION General exposure motion de
enabled. DX CL REMOTE IMAGE DISPLAY Remote Image Display func
(V7.4 or later)
- Each option key is supplied on one dedicated CD for each product.
DX CL GENERAL RIS BROWSER General RIS Browser (V7.4
• For licence supported version in V9.0 or later *1
Configuration option of DX CL OP KIT
- Apply the licence key instead of installing the option key in this step.
- If the option key has been installed by mistake, it is required to
reapply the option key together with the licence key.
- For details on acquisition and application of licence, see following
links.
$ See “Appendix 54 Software Licence” of “Appx IN: Installation–
Appendix”. (Other than US)
$ See “Appendix 66 Software Licence (For US)” of “Appx IN:
Installation–Appendix”.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-14
The installation procedure is described below.
The same installation procedure applies to all the option keys.
$ For the option key uninstallation procedure, see
“Appendix 7 Uninstalling the AP” under “Appx IN”.
1. Connect the USB connection Disk drive (jig) to the USB port
on the Mobile X-ray Unit.
DCL70410.eps
3. Click .
D80C010704011.ai
4. Click .
6. To install other option key, change the Disk and repeat steps
3 through 5.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-15
4.1.6 AP Key Installation
2. Click .
NOTE
• For V8.1 or earlier, or licence unsupported version in V9.0 or later ➔ A window opens to indicate the completion of installation.
Before AP key installation, be sure to install a standard key (standard
kit) and option keys.
• For licence supported version in V9.0 or later
- Apply the licence key instead of installing the AP key in this step.
(Option and AP will be enabled while applying the licence key. For DCL44010.ai
NOTE
In case the installation start window does not open automatically,
double-click the execution file (Filename extension: exe) in the Disk.
DXL309001.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-16
4.1.7 DR-ID 800CL Startup Verification 3. Click [Login].
Verify the normal startup of the DR-ID 800CL. ➔ The DR-ID 800CL-AP starts running.
NOTE
When the DR-ID 800CL is started for the first time, the following window
may open. Place a checkmark to the “Public Network” and click
.
DXCL140419005.ai
DXL03070009.ai
CLP74010.EPS
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-17
4.2 Settings on the DR-ID 800CL for ➥ REFERENCE
Connection with the FPD If you cannot complete step 2. from the initial window within a
3-second period and allow the AP to start up, perform the following
Use the Service Utility to set up the following items: steps to start the Service Utility.
DR-ID 800CL system information
1. Click “Shut Down” in the menu.
Optimizing the monitor display
Menu database replacement (required only for equipment used 2. Click while holding down the <Shift> key.
abroad)
3. From the menu, sequentially choose “All Programs”,
4.2.1 Starting the Service Utility “Fuji Film”, and “Console Advance.
4. When the initial window opens, sequentially click its upper left
1. Click and then sequentially choose “All Programs”, and upper right corners within a period of 3 seconds.
“Fuji Film”, and “Console Advance”.
➔ The Service Utility starts. The “IIP Service Utility” window opens.
➔ The DR-ID 800CL-AP starts up.
DXCL100810002.ai
DXL03100001.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-18
4.2.2 DR-ID 800CL System Information Setup 4. Enter setup data as directed below.
Complete setup concerning DR-ID 800CL system information as
❍ Make a date format selection by double-clicking
directed in this section.
for .
■ Setting the system information
1. From the “IIP Service Utility” window, click [Setup 0: Japanese date (S62.10.07) [Default]
Configuration Item]. 1: ANSI long date (1987.OCT.07)
2: ANSI short date (1987.10.07)
3: American long date (OCT.07.1987)
4: American short date (10.07.1987)
5: European long date (07.OCT.1987)
6: European short date (07.10.1987)
NOTE
Use “System Common Date Format” to specify the date format.
If you change the date format, the following items of information
are changed to match the newly selected format.
DXCL100803001.ai
Date output on film
➔ The “Setup Configuration Item” window opens. Format for birth date input
Date displayed by DR-ID 800CL-AP
2. Click the sign within in the “Setup
Other date indications
Configuration Item” window.
D80C010704010.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-19
❍ Use to specify the language to 5. Change the other items of system information as needed.
be used. Item
Description
number
0 : Japanese 11 : ChineseSimplified Setup of a process to be performed after an abnormality
1 : English 12 : ChineseTraditional 3
occurrence.
2 : German 13 : Portuguese
3 : French 14 : Polish 13 Specifying the series generation logic.
4 : Spanish 15 : Hungarian Setting the system ID. Assign a unique ID individually for each
14
5 : Italian 16 : Czech equipment unit.
6 : Swedish 17 : Russian Specifying the font sizes of the ID number and other items for
7 : Finnish 19 : Turkish 22
CRT display and film annotation character printing.
8 : Danish 21 : Greek 34 Specifying whether or not to output an exposure result data file.
9 : Norwegian 22 : Romanian
35 When “0:DX” is specified for item no. 54, set this item to “0:MG”.
10 : Korean
37 Specifying whether or not to use the user authentication function.
❍ Double-click for Specifying whether the radiographer name field should display the
38
name of a user who logs in with the user authentication function.
to enter a hospital name
Setup of a study modality handled on the DR-ID 800CL.
(consisting of two-byte characters). 54
When “0:DX” is selected for the “Modality setup”, use the User
Utility or Menu Information Import Export Tool to set up the DX
NOTE parameter.
Use “Institute/Site Name in Two-Byte Characters” to enter a Specifying whether to enable or disable the Service Utility
67
hospital name string of two-byte characters for use in the password function.
DR-ID 800CL system. 301
Specifying whether or not to enables the free size trimming
The default hospital name is “ ”. Ch ange this function.
entry to a hospital name string of two-byte characters as Password for the guest account of User Utility. (The user name is
defined by a specific hospital. fixed as “Guest”.)
305
When “Yes” is specified for item no. 37, the guest account
You can enter a string of up to 15 two-byte characters.
becomes invalid.
Since this setting value is accompanied by the stored image Specifying whether or not to display the Selector Setting window
and will be applied during image display or external output, 315
in the User Utility.
make sure to confirm that the facility name is correct with the Specifying the trimming size used when determining the trimming
user and the connected equipment manufacturer. Also it is 318 size with the exposure field auto recognition.
recommended that the confirmed contents be written in the To connect DR-ID 613SE, set to “24CMX30CM”.
report. Setting the password for the admin account of the User Utility.
(The user name is fixed as “VIP”.)
340
When “Yes” is specified for item no. 37, the admin account
becomes invalid.
351 Setting allowable trimming sizes for the manual trimming.
$ For details on individual setup items, see “3.4 Configuration
Details — 1. IMAGE MODALITY” under “MU: Maintenance
Utility”.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-20
■ Setting the film annotation character format
Use the following procedures to change the image information output
on film, film annotation character format and presence/absence of
image frame.
4. Click .
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-21
■ Setting the CSL/IDT FUNCTION No. Name Selection Meaning
To change the way how to display various items manipulated on the 16 Examination 0 : Pad with space to Specifying the Accession
DR-ID 800CL or the number of characters to be displayed, use the No. Padding the head ( 1234567) No. padding method.
Method Right-padded with
procedure described below. spaces to the head.
1 : Pad with zero to the
1. Click on the “Setup Configuration head (0001234567)
Item” window. Right-padded with zeros
➔ Setup items relating to the IDT function will be displayed on the to the head.
right-hand side of the window. 2 : Pad with space to
the tail (1234567 )
IDT/Console function-related setup items Left-padded with spaces
to the tail.
3 : Pad with zero to the
tail (1234567000)
Left-padded with zeros
to the tail.
17 Patient ID 1 to 64 ( 10) Setting the patient ID
Length length.
* Be sure to enter one-byte
characters on the
keyboard.
* To specify the number of
00000160.EPS characters that exceeds
11, set “Yes” for No. 22
2. Set up the CSL/IDT function, as required. Long Type Character
Format of IMAGE
Table CSL/IDT Function Setup Item List MODALITY.
No. Name Selection Meaning 18 Patient ID 0 : Pad with space to Specifying the patient ID
Padding the head ( 1234567) padding method.
15 Examination 1 to 16 ( 10) Setting the Accession No.
Method Right-padded with
No. Length length (1 to 16). * The setup value becomes
spaces to the head.
invalid if the mass
* Set 6 or larger when 1 is 1 : Pad with zero to the
head (0001234567) screening order issuance
set for “29. Study Number
Right-padded with zeros function is activated.
Issuance Logic Setup”
and 14 or larger when 2 to the head.
is set 2 : Pad with space to
* Be sure to enter one-byte the tail (1234567 )
characters on the Left-padded with spaces
keyboard. to the tail.
3 : Pad with zero to the
: Default setting, *: Note
tail (1234567000)
Left-padded with zeros
to the tail.
: Default setting, *: Note
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-22
No. Name Selection Meaning
22 Requesting 0 : Code Setting the content of the
Dept. Name Displays the requesting menu to be displayed in the
Field Setup department code. requesting department
1 : NameSbcs name field.
Displays the requesting
department name in
alphanumerics.
2 : NameDbcs
Displays the requesting
department name in
kanji.
23 Technician’s 0 : Code Setting the content of the
Name Field Displays the technician’s menu to be displayed in the
Setup code. technician’s name field.
1: NameSbcs
Displays the technician’s
name in alphanumerics.
2 : NameDbcs
Displays the technician’s
name in kanji.
348 Display No Setting whether or not to
Calibration Not displayed. display a warning message
Dialog Yes at the time of calibration.
Displayed.
: Default setting, *: Note
4. Click .
010-307-10E
11.30.2018 FM9540
DR-ID 800CL Service Manual IN-23
■ Setting the FPD 6. Click .
Perform the HostName and IP Address settings for the FPD registered
to the DR-ID 800CL. ➔ The setup data will be saved accordingly.
DXCL140704004.ai
5. For any setup data changed, select “Save” from the “Config”
menu.
➔ A window opens prompting you to confirm the setup data.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-24
■ Setting for DR standard size image creation (V8.0 or later) ● Setting the system information
To use the DR standard size image creation function, perform the 1. From the “IIP Service Utility” window, click [Setup
following settings. Configuration Item].
D80C010704010.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-25
● Setting the CSL / IDT FUNCTION ● Setting the User Utility
1. Click on the “Setup Configuration 1. Click and then sequentially choose “All Programs”,
Item” window.
“Fuji Film”, and “Console Advance”.
➔ Setup items relating to the IDT function will be displayed on the
➔ The Log-in window for DR-ID 800CL-AP opens.
right-hand side of the window.
IDT/Console function-related setup items
00000160.EPS
4. Click .
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-26
4. Enter the following. 8. Open “Exposure parameter 3” tab and set up “Various Fill
User name : **** Density [%]”.
Password : ****
5. Click [OK].
➔ The “Main Menu” window opens.
DXCL120704006.ai
DXCL120850007.ai
9. Click [OK].
➔ The system returns to the “Exposure Menu Setting” window.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-27
4.2.3 Registering the Image Detector I. Set the name of the Image Detector with 1 to 5 characters
without using a same name as other Image Detectors.
II. Set the ID of the Image Detector with 1 to 16 one-byte
1. Start the DR-ID 800CL Service Utility. alphanumeric characters without using a same ID as other Image
➔ The “IIP Service Utility” window opens. Detectors.
III. Select the device type from the following:
2. Click [Device Setting]. BENEO
FDR AcSelerate
3. Click [Add]. CALNEO U/MT/MB
➔ The “Data Type” window opens. CALNEO-C/FDR D-EVO (DR-ID 601SE/DR-ID 602SE)*1
1
CALNEO-C SQ/FDR D-EVO plus (DR-ID 611SE/DR-ID 612SE)*
4. Perform the following settings. FDR AcSelerate_CSI
CALNEO HC SQ (only in Japan)
1
CALNEO-C mini/D-EVO plus C24 (DR-ID 613SE)*
CALNEO Smart G/D-EVO II G
I (DR-ID 1201SE/DR-ID 1202SE)*1
CALNEO Smart C/D-EVO II C
II (DR-ID 1211SE/DR-ID 1212SE/DR-ID 1213SE)*1
CALNEO GL1 (DR-ID 1305SE only in Japan)
III FDR ES G (DR-ID 1271SE/DR-ID 1272SE)
FDR ES C (DR-ID 1273SE/DR-ID 1274SE/DR-ID 1275SE)
IV
*1 : Device which can be connected to the DR-ID 800CL
V IV. Set the color to identify the device (as required).
Pink Vivid Pink
Blue Vivid Blue
Yellow Lime Yellow
Purple Vivid Purple
DXCL040415001.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-28
5. Click .
6. Click [Cancel].
➔ The system returns to the main menu.
010-307-10E
11.30.2018 FM9540
DR-ID 800CL Service Manual IN-29
4.2.4 Setting the Selectors
This section describes the procedure used for setting the selectors.
II
III
IV
VI
VII
VIII
IX
X
XI
D8CS070704003.ai
010-307-10E
11.30.2018 FM9540
DR-ID 800CL Service Manual IN-30
I. Choose a selector tab. VI. Choose a icon to be used.
II. Set the Technique Code paying attention not to set the same NOTE
code with other selectors.
Select a file with “_S” at the end of the name, which is a small
NOTE icon for 0.7M monitor.
When connecting with the X-CON, set up in accordance with
the X-CON. Set up the exposure menu also in accordance
➥ REFERENCE
with the X-CON. If the X-ray tube/technique code which has
been set to the exposure menu does not exist in the Since the cassette designs of D-EVO and D-EVO2 are different,
selectors, the selector which has been set as the set a dedicated icon for each of them.
InitialActiveButton will be set.
D-EVO2 D-EVO
The default value of the Technique Code is “2”.
When multiple readers have been registered, also register
multiple selectors. However, set the Technical Code to a
number other than “2”.
When the Technical Code has been set as a number other
than “2”, edit “TechinquieCodeConvertTbl.ini” and specify the
Technique Code as “2”. VII. Set the name of the Image Detector.
$ For details of setting the Technique Code, see VIII. Select the Tube Number.
“Appendix 9 Setting the Technique Code” under
“MU: Maintenance Utility”.
➥ REFERENCE
The tube number can be selected but it is invalid.
III. Select a device to be added to the selector (use the host name
for selection). IX. Command for enabling or disabling the Virtual Grid function.
IV. Choose a Reader Type. (V8.1 or later)
CR-Cassette : CR Cassette For grid (oscillation type) device, set to OFF.
Velocity U : Velocity upright position For grid (steady type) device or exposing without grid, set to
Velocity T : Velocity supine position ON.
FPD-U : DR built-in upright position
FPD-T : DR built-in supine position ➥ REFERENCE
1
FPD-C* : DR Cassette
It is possible to set this item only when the “Technique Code”
*1 : ReaderType which can be set to the DR-ID 800CL is not “-1”.
For the study reread from the media, the Virtual Grid function
V. Enter the character string (consisting of up to five characters) to
be displayed on the selector button.
does not operate. Set the Virtual Grid function to OFF on the
QA screen if necessary.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-31
X. Sound setting for “Ready” notices (V8.0 or later) 4.2.5 Optimizing the Monitor Display
Makes sound from DR-ID 800CL once, when the status switches [Setting Done Prior to Shipment]
from “Not Ready” to “Ready”.
NOTE Set the LUT file as follows to optimize the image display of the monitor
that is being used.
This function can not be used in DR-ID 800CL.
NOTE
XI. To restore the currently edited selector to the status that
prevailed at DR-ID 800CL startup, turn this radio button ON. When using the TECNART-made 15" LCD with touch panel for V7.4 or
earlier, be sure to install LUT file in advance.
NOTE
$ For details of procedures, see “4.2 Installing LUT for the
When multiple readers have been configured, repeat I through TECNART-made 15" LCD equipped with touch panel” under “MU:
XI to perform the configuration of selectors. Maintenance Utility”.
NOTE
If the host name is changed for the equipment whose “ReaderType” is
selected as other than “Cassette”, you must select the “hostName”
again since the “Shot Ready” icon will be hidden when the host name
has been changed.
If the same technique code is to be set to different selectors, the Technique DXCL100704001.ai
Code needs to be set by editing the “TechniqueCodeConvertTbl.ini” file. ➔ The “Monitor List” window opens.
$ For details on the setup procedures, see “Appendix 9 Setting
the Technique Code” under “MU: Maintenance Utility”.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-32
2. Select the connected display and then click . 4.2.6 Changing the Menu Database
To change an exposure menu or image processing parameter, the
menu database must be replaced following individual procedures.
■ For V7.0
NOTE
Pay attention to the fact that exposure menu and selector settings will
revert to the default if a menu database is replaced.
D8CS080316001.ai
➔ Settings are thus completed, and the system returns to the “IIP
Service Utility” window.
DXL03040002.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-33
2. Enter “3” and press the <Enter> key. 4. Enter the item number of an image processing parameter (for
➔ A window like that shown below opens. Monitor/for Film) you wish to use, and then press the
<Enter> key.
➔ A window like that shown below opens.
DXCL070305001.ai
➔ A window like that shown below opens. 8. Insert the Standard Kit Disk into the disk drive.
AWS35001.ai
DXL03070010.ai
➔ An explorer opens.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-34
10. Copy the following files and overwrite files with the same ■ For V7.3 or later
names in “C:\Program Files\FujiFilm\IIP\Param”.
“D:\Config\Selector.mdb” NOTE
“D:\Param\XconParam.mdb” Pay attention to the fact that exposure menu and selector settings will
revert to the default if a menu database is replaced.
11. Remove the Disk from the disk drive.
For overseas, the default settings at shipment are intended for the U.S.
Select proper locale at the steps 3 and 12 when installing overseas
other than the U.S.
➥ REFERENCE
If the language of the OS is Japanese, the “Japanese” exposure menu
for filming has be installed. If it is English, the “English(U.S.A without
mammo menu)” exposure menu for filming has be installed.
NOTE
In case the installation start window does not open automatically,
double-click the execution file (Setup.bat) in the Disk.
The items displayed in the “DR Console Setup Tool” vary with the
versions of DR-ID 800CL.
DXL03040002.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-35
2. Enter “3” and press the <Enter> key. 4. Enter the item number of an image processing parameter (for
➔ A window like that shown below opens. Monitor/for Film) you wish to use, and then press the
<Enter> key.
➔ A window like that shown below opens.
DXCL070305001.ai
DXL03070010.ai
➔ An explorer opens.
AWS35001.ai
10. Click “Setup.bat”.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-36
11. Enter “4” and press the <Enter> key.
➔ A window like that shown below opens.
DXCL080305001.ai
12. Enter the item number of a language you wish to install and
press the <Enter> key.
D80C020309004.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-37
4.2.7 Exiting the Service Utility 4.3 Registration of Connection Settings of
1. From the “Setup Configuration Item” window, choose the In-Hospital Network (V7.3 or Later)
“Close” on the “Config” menu.
➔ The “IIP Service Utility” window opens. When using the network setting switching function, register connection
settings of the in-hospital network as necessary.
2. Click . NOTE
The network setting switching function cannot be used on the in-
➔ The system returns to the desktop screen. hospital network in the following cases.
- If it is connected with the in-hospital network only by a wired
connection
- If DHCP is used for connection with the in-hospital network
Do not use the network setting switching function on the following
networks.
- The image local network
- The network which shares the in-hospital network with the image
local network
If the same SSID is registered to the connection settings of multiple
networks inside the hospital, set “No” in “368 Enable radio map in
hospital network dialog” of “5. CSL/IDT FUNCTION”.
If the in-hospital network is being connected by using a wireless network
adapter and the function for switching network setting is used, do not
start the utility application which is attached to the wireless network
adapter. In addition, delete the utility application from menu
– “All Programs” – “Startup” so that the application does not start
automatically when the DR-ID 800CL is started.
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-38
4.4 Setting Network Connection in ■ Wired connection to both the in-hospital and
PC with 2 NIC image local networks
(1) (2)
An example of the setting network connection for PC with 2 NIC is
shown below. DR-ID 800CL
Change the IP address of the DR-ID 800CL as required.
HUB/ TCP/IP
Access SE Access
NOTE Point (DR Cassette) Point
Different IP addresses must be set to each NIC. They also must not
belong to the same subnet.
PACS RIS
The LAN cables of the in-hospital network and the image local network
must not be switched.
In-hospital Network Image Local Network
IP addresses of wired connection and wireless connection of DR-ID D80C010704008.ai
010-307-09E
11.22.2017 FM9492
DR-ID 800CL Service Manual IN-39
■ Wired and wireless connection to the in-hospital network and
wired connection to the image local network
(1) (3)
DR-ID 800CL
(2)
HUB/ TCP/IP
Access SE Access
Point (DR Cassette) Point
PACS RIS
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-40
4.5 Setting the Priority Order of 5. Set the connections in the “Connection” field and the “Bindings
the Network Binding for xxx” field (xxx is the connection name) so that they are
displayed in the following priority order.
1. Wired connection to the in-hospital network
Perform the priority order of the network binding settings.
2. Wireless connection to the in-hospital network
4.5.1 For Windows 7 3. Wired connection to the image local network
➥ REFERENCE
1. Select “Control Panel” from the menu. By setting the priority order for the network binding, “This Host”
becomes the connection with the highest priority order of the
➔ The “All Control Panel Items” window opens.
connected network connections.
➥ REFERENCE
If the menu bar is not displayed, press the <Alt> key to display it.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-41
4.5.2 For Windows 10 2. Select “Network and Sharing Center”.
For Windows 10 laptop PC / desktop PC that uses two or more network
adapters, the address of “This Host” in Service Utility becomes the link
local address (169.254.xxx.xx).
To avoid this phenomenon, perform the following settings.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-42
4. Right-click “Eth1” and select “Properties”. 6. Click [Advanced].
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-43
7. Set the following settings, and click [OK].
Automatic metric : Remove the checkmark.
Interface metric : Input “1”.
8. Click [OK].
➔ The “Ethernet Properties” window opens.
9. Click [OK].
➔ Returns to the “Network Connections” window.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-44
4.6 Setting the Automatic Execution 4. Save and close the Notepad.
Batch of Ping (V7.4 or Later)
5. Right-click “ping.bat” and select “Create Shortcut”.
➔ The shortcut file of “ping.bat” is created.
NOTE
Perform the procedures only when it is applicable to all the following 6. Right-click the shortcut file of “ping.bat” and select “Properties”.
conditions. ➔ The “ping - Shortcut Properties” window opens.
The MWM option and the MWM-PI option are enabled.
Both wired and wireless connection to the in-hospital network are used. 7. Select “Minimized” from the pull-down menu of “Run” and then
click .
When switching the multiple NIC (Network Interface Card) for the in-
hospital network to operate communications, MMM communications
with RIS may irregularly fail.
In order to avoid this situation, perform the following procedures.
2.
D80C040704004.ai
Right-click “ping.bat” and then select “Edit”.
➔ The “ping - Shortcut Properties” window closes.
➔ Notepad opens.
8. Select “All Programs” from the menu.
3. Change the IP address of Ping notification destination to the IP
address of the RIS terminal.
9. Right-click “Startup” and select “Explore”.
➔ The “Startup” window opens.
D80C040704003.ai
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-45
4.6.2 Operational Verification
Verify that it can be connected to RIS from each NIC.
In case of both wired and wireless connection to the in-hospital network,
that which one is used for communication depends on the network
settings. Therefore, the operational verification will be performed in the
situation that the NIC except the target NIC from the network have been
cut off.
$ For details of network setting, see “4.5 Setting the Priority Order
of the Network Binding”.
1. Cut off the NIC except the target NIC from the network.
NOTE
Cut off the NIC from the network using the following method. It is
not required to cut off the NIC for image local network.
NIC for wired connection: Remove the LAN cable.
NIC for wireless connection: Move the DR-ID 800CL if there is no
communication with the AP.
3. Make sure that the order information and patient information can
be received.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-46
4.7 Procedures for Changing 4.9 Barcode Reader Connection
the DR-ID 800CL Host Name
When using the barcode reader, perform the following connection and
Procedures observed when the DR-ID 800CL host name is changed setup.
follow.
● Barcode reader connection and setup
$ See “Appendix 5 Changing the DR-ID 800CL Host Name” under $ For details on the connection, see “Appendix 1 Connecting
“Appx IN”. the Barcode Reader [Option]” under “Appx IN”.
4.8 DR-ID 800CL Setup for Connection to 4.10 Setting the DX Parameter
Other Equipment When “0:DX” is set up for the Modality setup (No. 54 of IMAGE
MODALITY), DX parameter needs to be set up for the exposure menus.
This section explains about the software setup procedures to be
Provide the user with appropriate explanations with this regard.
performed when connecting the following equipment to the DR-ID
As requested by the user, use the User Utility or Menu Information
800CL.
Import Export Tool to set up the DX parameter for the DR-ID 800CL
● DICOM server exposure menus.
$ See “2. Software Setup” under “Appendix 10 DICOM CR $ For details on how to operate the Menu Information Import Export
Storage” of “Appx IN”. Tool, see “18. Menu Import Export Tool” under
“MU: Maintenance Utility”.
● DICOM DX Storage
$ See “2. Software Setup” under “Appendix 11 DICOM DX
Storage” of “Appx IN”. 4.11 Setting the Connection Device
Status Icon (V7.3 or Later)
● DICOM Printer
$ See “2. Software Setup” under “Appendix 12 DICOM Print” of Change the threshold of the status icon displayed on the DR-ID 800CL
“Appx IN”. as necessary.
● RIS $ For details, see “Appendix 17 Setting the Connection Device
Status Icons (V7.3 or Later)” under “MU: Maintenance Utility”.
$ See “2. Software Setup” under “Appendix 13 RIS: DICOM
MWM/DICOM PPS” of “Appx IN”.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-47
4.13 Batch setting of VG parameter 4.14 Setting when using the grid pattern
(V9.0 or Later) removal processing function (GPR)
To use a Virtual Grid function, set the VG parameter for each exposure $ See “17.2 The Grid Pattern Removal Processing” of “Maintenance
menu by batch. Utility (MU)”.
$ For the VG parameter batch setting procedure, refer to “17.4 Batch
Setting of VG Parameters (V9.0 or later)” of “Maintenance Utility (MU)”.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-48
5. Checkout Procedures 5.1 Checking the Connection between
the DR-ID 800CL and FPD
5.1 Checking the Connection between 1. Turn ON the FPD (DR system) power. If it is already ON, you
the DR-ID 800CL and FPD do not have to do anything here.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-49
5.2 Checking the Connection between 5.3 Checking the Output Image
the DR-ID 800CL and Other Equipment
Run the following checks on the output image.
Check the connection to the following equipment. Format check
Irregularity, sensitivity and density checks
● DICOM server Film annotation character format
$ See “3. Connection Checkout” under “Appendix 10 DICOM
CR Storage” of “Appx IN”. ■ Exiting the DR-ID 800CL-AP
● DICOM DX Storage 1. Shut down the DR-ID 800CL-AP.
$ See “3. Connection Checkout” under “Appendix 11 DICOM DX
Storage” of “Appx IN”. ➔ The DR-ID 800CL exits returning then to the desktop.
● DICOM Printer
$ See “3. Verifying the Connection” under “Appendix 12 DICOM
Print” of “Appx IN”.
● RIS
$ See “3. Verifying the Connection” under “Appendix 13 RIS:
DICOM MWM/DICOM PPS” of “Appx IN”.
● X-CON
$ See “3. Verifying the Connection/Data Exchange” under
“Appendix 19 X-Ray Controller” of “Appx IN”.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-50
5.4 Operation Check of ■ If “×” is always displayed with the icon which indicates
the Network Switching Function the electric wave intensity
1. Move close to the access point, and check if there is any
If the network switching function is used, check the following items. change after several minutes.
Receipt of orders
Exposure 2. Check if there is any mistake with the SSID registered in
Distribution and printing of images “Service Utility” – “Network Adapter Configuration”.
Transfer of the area and switching of the network connection
If any problem occurs, check the settings according to the detail of the
3. Check if the network of the concerned SSID is registered in
problem. “Start” menu – “Control Panel” – “Network and Sharing
Center” – “Manage Wireless Networks”.
■ If it is impossible to communicate with a machine in the ■ If the electric wave intensity does not improve/ become
in-hospital network stable after executing network repair
1. Check if it is possible to communicate with the concerned 1. Check if it changes when repair is executed at a different place.
machine by a wired connection.
2. Check if a wrong adapter name of the in-hospital network is
2. Check if there is any mistake with the settings of “Service registered.
Utility” – “Setup Configuration Item” – “NETWORKCONFIG”.
3. Check the possibility of radio frequency interference with
3. Check if a ping command is executed successfully with the another machine.
concerned machine in the state it is connected wirelessly.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-51
6. Setup File Backup, Image
Database Initialization and
Error Log Verification/Deletion
This section explains about the procedures for backing up the
DR-ID 800CL setup files, initializing various image databases and
verifying/deleting the error log.
NOTE
Never turn OFF the DR-ID 800CL and FPD power during a backup
process.
If the power turns OFF, the machine may fail to start up.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-52
6.1 Backing Up the DR-ID 800CL Setup Files 6. Click .
8. Click .
5. Click .
➔ You are then returned to the “Configuration Restore/Backup”
window.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-53
6.1.3 Making backup of files 6.1.4 Making image backup after installation
To make backup of several files that are not covered with the backup To make backup of hard drive image after installation, refer to the
function of ServiceUtility, refer to the following. following.
$ For details on the backup procedure, see “■ Backing up each type $ See “Making Recovery Media” under “How to Perform Restoration
of the files” in “Appendix 8 Upgrading the AP” under “Appx IN: from Recovery DVD” in “MOBILE X-RAY SYSTEM FDR Go
Installation – Appendix”. Installation Manual”.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-54
6.2 Initializing Various Image Databases 4. Click .
Initialize the image database, output queue and patient database as ➔ A confirmation window opens.
follows.
2. Click . 5. Click .
➔ A confirmation window opens.
NOTE
When the “Press any key to continue...” message appears, press
the <Enter> key.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-55
6.2.2 Clearing the Image Output Queue 6.2.3 Initializing the Patient Database
1. Click . 1. Click .
MU000020.ai
MU000019.ai
2. Click . 2. Click .
NOTE NOTE
When the “Press any key to continue...” message appears, press When the “Press any key to continue...” message appears, press
the <Enter> key. the <Enter> key.
➔ The images are then deleted from the output queue and the ➔ The patient database is then initialized and the system returns to
system returns to the “DataBase Utility” window. the “DataBase Utility” window.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-56
6.3 Checking and Deleting 6.4 Turning OFF the DR-ID 800CL
the DR-ID 800CL Error Log
NOTE
Check the DR-ID 800CL error log and then delete it as appropriate.
Do not turn OFF the FPD power.
1. Click .
1. Shut down the DR-ID 800CL system.
➔ The “Event Viewer” main window opens.
2. Turn OFF the power switch of the DR system on the Mobile
2. Confirm the error log. X-ray Unit.
$ For details on how to check the error log, see “2.4 Details of
Events” under “MT: Machine Troubleshooting”.
5. Click .
➥ REFERENCE
You can save the backup of HDD image at this point for just in case.
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-57
BLANK PAGE
010-307-11E
07.22.2019 FM9582
DR-ID 800CL Service Manual IN-58