Impact of Low Volume Citrate Tubes On Results of First Line Hemostasis Testing

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

| |

Received: 14 February 2019    Revised: 16 March 2019    Accepted: 20 March 2019

DOI: 10.1111/ijlh.13028

ORIGINAL ARTICLE

Impact of low volume citrate tubes on results of first‐line


hemostasis testing

Gian Luca Salvagno1  | Davide Demonte1 | Giovanni Poli1 | Emmanuel J. Favaloro2 |


Giuseppe Lippi1

1
Section of Clinical Biochemistry, University
of Verona, Verona, Italy Abstract
2
Department of Haematology, Sydney Introduction: Pediatric tubes are increasingly used for drawing blood for hemostasis
Centres for Thrombosis and Haemostasis,
testing. This study has investigated the potential impact of low volume citrate tubes
Institute of Clinical Pathology and
Medical Research, NSW Health on results of first‐line hemostasis testing.
Pathology, Westmead Hospital, Westmead,
Methods: The study population comprised 34 patients on warfarin therapy and 17
NSW, Australia
ostensibly healthy volunteers. Blood was collected into five different evacuated
Correspondence
blood tubes from each subject. On right arm, blood was drawn directly into two
Giuseppe Lippi, Section of Clinical
Biochemistry, University Hospital of Verona, standard evacuated blood tubes (3‐mL Vacuette and 2‐mL Vacutest) and one evacu‐
Verona, Italy.
ated low volume blood tube (1‐mL Vacuette) by straight needle venipuncture. On left
Email: giuseppe.lippi@univr.it
arm, blood was drawn using a 5‐mL syringe and then transferred within two none‐
vacuated microtubes (0.5 mL MiniCollect and 0.5 mL Micro Test). Prothrombin time
(PT), activated partial thromboplastin time (APTT), and fibrinogen were assayed on
ACL TOP 700.
Results: Spearman's correlation of PT, APTT, and fibrinogen values obtained using
different tubes was always satisfactory (ie, ≥0.93). A statistically significant bias was
frequently found by comparing values obtained in different tubes. Nevertheless, the
minimum quality specifications for bias were exceeded only by comparing data of
Vacuette 1 mL with those of all other blood tubes for PT, by comparing data of Micro
Test 0.5 mL with those of all other blood tubes for APTT, and by comparing data of
Micro Test 0.5‐mL blood tubes with those of Vacuette 3 mL and Vacuette 1.
Conclusion: First‐line hemostasis testing using low volume citrate tubes may display
differences sometimes exceeding the minimum quality specifications.

KEYWORDS
blood collection, blood tubes, coagulation testing, hemostasis testing, preanalytical variability

1 |  I NTRO D U C TI O N laboratory diagnostics, 3,4 the largest bias being attributable to the
procedure, as well as to the materials, used for drawing venous
Blood sample collection is essentially an unavoidable activity to blood. 5
obtain suitable biological material to be used for laboratory test‐ Previously published data have shown that the type6 and size7
ing.1 Laboratory hemostasis is no exception to this general rule of blood collection needle may impact on hemostasis test results.
and indeed comprises special considerations. 2 Several lines of Additional evidence has been provided that even the type,8,9
evidence now attest that the so‐called preanalytical phase, and brand,10,11 and size or filling volume12 of citrated blood tubes may
especially blood drawing, is the leading source of variability in comprise serious obstacles for harmonization of test results.

Int J Lab Hematol. 2019;1–6. © 2019 John Wiley & Sons Ltd |  1
wileyonlinelibrary.com/journal/ijlh  
|
2       SALVAGNO et al.

The collection of blood samples that are suitable for testing may The procedure consisted of two consecutive venipunctures,
be further challenged in selected categories of patients, additionally performed by the same expert phlebotomist, on both arms of the
magnified for clotting tests because coagulation blood tubes need patient. On the right arm, blood was drawn into the two standard
to be accurately filled, up to their nominal filling volume, to permit evacuated blood tubes (3‐mL Vacuette and 2‐mL Vacutest) and
an appropriate ratio (usually 9:1) between anticoagulant (typically one evacuated low volume blood tube (1‐mL Vacuette) which were
3.2% buffered sodium citrate) and blood. The collection of low blood filled up to the level mark, as for normal local practice (ie, using a
volumes may furthermore be favored to reduce unnecessary diag‐ 19‐gauge straight needle venipuncture, with blood collected directly
nostic‐related blood loss such as in intensive care units (ICUs) (ie, into the evacuated blood collection tubes). The sequence of these
for preventing repeated blood drawing‐related anemia),13 but also three blood tubes was randomized after each following patient. On
in patients with difficult veins (especially pediatric populations and the left arm, blood was drawn using a 5‐mL syringe equipped with
cancer patients). This diagnostic and clinical need has been recently a 19‐gauge needle. After removing the needle from the syringe, the
accomplished by commercialization of low volume tubes, charac‐ blood was then immediately transferred within the two nonevacu‐
terized by draw volumes ≤1 mL. Although these devices have been ated microtubes (0.5 mL MiniCollect and 0.5 mL Micro Test), which
proven effective to decrease phlebotomy‐related blood loss, both in were filled up to the level mark.
pediatric and in adult populations,14 their use has been associated Prothrombin time, APTT, and fibrinogen were assayed
with particular bias in laboratory testing, especially in results of using the same fully automated coagulation analyzer ACL
some immunochemistry15 and hematologic16 parameters. Notably, TOP 700 (Instrumentation Laboratory Bedford, MA, USA),
recent evidence has also shown that the use of partial‐draw citrate with RecombiPlasTin (Instrumentation Laboratory), SynthASil
collection tubes (ie, 2.0 mL) may lead to underestimating unfrac‐ (Instrumentation Laboratory) and Fibrinogen‐CXL (Instrumentation
tionated heparin therapy assessment compared to standard volume Laboratory), respectively. The analytical performance of these tests
tubes (ie, 3.5 mL).17 However, no evidence has been provided, to the has been earlier described elsewhere.18 Immediately after collection,
best of our knowledge, on the impact of pediatric tubes, with draw all blood tubes were centrifuged at 1300 × g for 15 minutes at room
volumes ≤1 mL, on hemostasis testing. Therefore, the present study temperature, in accordance with the current recommendations of
was aimed to investigate the potential impact of low volume citrate the Clinical Laboratory Standards Institute (CLSI).19 The overall time
tubes on results of first‐line hemostasis testing. elapsed between sample collection and testing was always less than
1 hours.
The data obtained in the different blood tubes (median and in‐
2 |  M ATE R I A L S A N D M E TH O DS terquartile range; IQR) were compared among different tubes with
Spearman's correlation. The relative (percent) bias was calculated
The study population comprised 34 consecutive patients on oral with Bland and Altman plots. The relative bias was also compared
anticoagulant therapy (OAT; warfarin; mean age 76 ± 9 years; 13 with the minimum quality specifications for bias, as provided by Ricos
women and 21 men) and 17 ostensibly healthy volunteers recruited et al (ie, ±3.0% for PT, ±3.4% for APTT, and ± 4.8% for fibrinogen). 20
from the laboratory staff (mean age; 45 ± 11 years; 13 women and 4 The statistical analysis was carried out with Analyse‐it (Analyse‐it
men). Blood from each of these subjects was collected into five dif‐ Software Ltd, Leeds, UK). Statistical significance was set at P < 0.05.
ferent evacuated blood tubes, all containing the same sodium citrate The study was based on anonymized routine samples referred for
formulation, as shown in Table 1. routine testing and the study, which was part of a standard process

TA B L E 1   Blood tubes characteristics

Manufacturer Name Anticoagulant Type of tube Draw volume

Greiner Bio‐One, Kremsmünster, Vacuette Buffered sodium citrate Evacuated, 13 × 75 mm 3.0 mL
Austria 3.2% Reference number: 454334
Lot number: A180848F
Greiner Bio‐One, Kremsmünster, Vacuette Buffered sodium citrate Evacuated, 13 × 75 mm 1.0 mL
Austria 3.2% Reference number: 454320
Lot number: A1810395
Greiner Bio‐One, Kremsmünster, MiniCollect Buffered sodium citrate Nonevacuated, 10 × 37 mm 0.5 mL
Austria 3.2% Reference number: 450538
Lot number: A1803QX
Kima, Padova, Italy Vacutest Buffered sodium citrate Evacuated, 13 × 75 mm 2.0 mL
3.2% Reference number: 14074
Lot number: A2908
Kima, Padova, Italy Micro Test Buffered sodium citrate Nonevacuated, 10 × 45 mm 0.5 mL
3.2% Reference number: 814074
Lot number: L2748
|
      3

TA B L E 2   Results of prothrombin time (PT), activated partial thromboplastin time (APTT), and fibrinogen obtained in different citrated
blood tubes

Test Vacuette 3 mL Vacuette 1 mL MiniCollect 0.5 mL Vacutest 2 mL Micro test 0.5 mL

PT (s) 21.9 (15.7) 19.8 (14) 22.2 (15.9) 22.0 (15.3) 21.2 (14.5)
APTT (s) 35.4 (8.6) 33.7 (8.3) 33.8 (8.6) 34.2 (9.2) 30.7 (7.5)
Fibrinogen (g/L) 3.23 (1.26) 3.09 (1.19) 3.28 (1.27) 3.38 (1.09) 3.38 (1.37)

of new blood tubes validation before routine usage in the local hos‐ tubes. Fibrinogen was overall less sensitive to low draw volume be‐
pital, 21 and was cleared by the local Ethical Committee (University cause the minimum quality specifications for bias were only exceeded
Hospital of Verona, n. 35747, July 25, 2016). when comparing 0.5 mL Kima Micro Test with 3.0‐mL or 1.0‐mL blood
tubes produced by the other manufacturer (ie, Grainer Bio‐One). PT
data produced using plasma collected into pediatric tubes with draw
3 | R E S U LT S volumes of 0.5 mL were globally comparable to those generated using
standard volume blood tubes, but were also highly intercomparable
The results of PT, APTT, and fibrinogen obtained in the different between these two pediatric tubes (ie, main bias was 1.6%). The larg‐
blood tubes are shown in Table 2. As regards PT, the Spearman's est variability, always exceeding the minimum quality specifications,
correlation of values obtained in the different tubes was always ex‐ was noted using 1.0‐mL Vacuette, with bias higher between 1.6‐ and
cellent (ie, ≥0.99). Although a statistically significant bias was fre‐ 2.5‐fold that of the quality specifications. Regarding APTT, a variabil‐
quently found by comparing values obtained in the different tubes, ity exceeding the minimum quality specifications was only observed
the minimum quality specifications for bias of PT (ie, ±3.0%) were using 0.5‐mL Micro Test, with bias always >2.8 than the quality spec‐
only exceeded when data obtained in Vacuette 1‐mL blood tubes ifications. This would actually mean that results of PT obtained using
were compared with those obtained in the other blood tubes 1.0‐mL Vacuette and APTT using 0.5 mL Micro Test would not be com‐
(Table 3). The Spearman's correlation of APTT values obtained in the parable to those obtained with standard volume tubes and may hence
different tubes was also satisfactory (ie, always ≥0.93). Even in this disrupt both the diagnostic reasoning and the clinical decision making,
case, a statistically significant bias was frequently found by compar‐ especially when longitudinal patient data are compared using differ‐
ing values obtained in the different tubes (Table 4). Nevertheless, ent tubes. This evidence further highlights the importance for hemo‐
the minimum quality specifications for bias of APTT (ie, ±3.4%) stasis laboratories to prioritize a process for local validation of blood
were only exceeded when data obtained in Micro Test 0.5‐mL collection tubes before introducing new devices. Another important
blood tubes were compared with those obtained in the other blood aspect emerged from this study is that, although hazardous for phle‐
tubes. Importantly, the bias in this tube was on average 3‐fold higher botomists’ safety, blood drawing with a syringe and then transfer of
than the minimum quality specifications. As regards fibrinogen, the that blood into nonevacuated pediatric tubes seems a reliable practice,
Spearman's correlation of values obtained in the different tubes was as attested by the limited bias (always lower than the minimum quality
always excellent (ie, ≥0.96). A statistically significant bias was fre‐ specifications) found for PT, APTT, and fibrinogen values by comparing
quently found by comparing values obtained in the different tubes 0.50 mL MiniCollect with the other two standards tubes (Tables 3-5).
(Table 5), but the minimum quality specifications for bias of fibrino‐ As regards the variability of PT and APTT data observed among
gen (ie, ±4.8%) were only exceeded when data obtained in Micro the tubes we tested, two possible solution can be identified for over‐
Test 0.5‐mL blood tubes were compared with those obtained in two come this drawback in facilities using both standard and pediatric
other blood tubes (ie, Vacuette 3 mL and Vacuette 1 mL). A suba‐ tubes. The first entails the harmonization of test results by calcu‐
nalysis of data in OAT patients and in healthy volunteers revealed lating mean normal PT (MNPT) and mean normal APTT (MNAPTT)
virtually identically results. for both standard and low volume tubes and then defining separate
tests on the analyzer (ie, “standard volume tubes PT or APTT” and
“low volume tubes PT or APTT”). The second approach, which also
4 | D I S CU S S I O N applies to fibrinogen, encompasses the use of separate reference
ranges for standard and low volume tubes.
Low volume blood collection tubes are a valuable resource for both
reducing unnecessary diagnostic‐related blood loss and facilitating
blood drawing in patients with unsuitable/difficult veins. However, 5 | CO N C LU S I O N
local validation of test results is mandatory before these tubes can
be introduced into routine phlebotomy practice, because results Although we acknowledge that this study may have some limitations
21
may otherwise reflect clinically important differences. (ie, drawing blood from different arms, using syringe versus evacu‐
The results of our study show that first‐line hemostasis testing ated blood tubes), our results suggest that results of first‐line hemo‐
may be influenced by the use of some types of low volume, pediatric stasis testing obtained in plasma collected into some pediatric tubes
|

TA B L E 3   Comparison of prothrombin time (PT) test results obtained in different citrated blood tubes
4      

Tubes Vacuette 1 mL MiniCollect 0.5 mL Vacutest 2 mL Micro Test 0.5 mL

Vacuette 3 mL r = 1.00 (95% CI, 0.99‐1.00) r = 0.99 (95% CI, 0.98‐1.00) r = 1.00 (95% CI, 1.00‐1.00) r = 1.00 (95% CI, 0.99‐1.00)
Bias: −7.6% (95% CI, −8.4 to −6.8%) Bias: 0.0% (95% CI, −0.8 to Bias: −1.3% (95% CI, −1.8 to −0.8%) Bias: −2.9% (95% CI, −3.6 to −2.2%)
P < 0.001 0.8%) P < 0.001 P < 0.001
P = 0.916
Vacuette 1 mL ‐ r = 0.99 (95% CI, 0.98‐1.00) r = 1.00 (95% CI, 0.99‐1.00) r = 1.00 (95% CI, 0.99‐1.00)
Bias: 7.6% (95% CI, 6.5to Bias: 6.3% (95% CI, 5.6 to 7.0%) Bias: 4.7% (95% CI, 4.1 to 5.4%)
8.7%) P < 0.001 P < 0.001
P < 0.001
MiniCollect ‐ ‐ r = 0.99 (95% CI, 0.99‐1.00) r = 0.99 (95% CI, 0.98‐1.00)
0.5 mL Bias: −1.3% (95% CI, −2.0 to −0.5%) Bias: −2.9% (95% CI, −3.8 to −1.9%)
P = 0.002 P < 0.001
Vacutest 2 mL ‐ ‐ ‐ r = 0.99 (95% CI, 0.99‐1.00)
Bias: −1.6% (95% CI, −2.3 to −0.9%)
P < 0.001

TA B L E 4   Comparison of activated partial thromboplastin time (APTT) test results obtained in different citrated blood tubes

Tubes Vacuette 1 mL MiniCollect 0.5 mL Vacutest 2 mL Micro Test 0.5 mL

Vacuette 3 mL r = 0.99 (95% CI, 0.98‐0.99) r = 0.97 (95% CI, 0.94‐0.98) r = 0.95 (95% CI, 0.92‐0.97) r = 0.94 (95% CI, 0.90‐0.97)
Bias: −3.4% (95% CI, −4.2 to −2.8%) Bias: −2.9% (95% CI, −4.1 Bias: −1.6% (95% CI, −2.8 to −0.4%) Bias: −13.2% (95% CI, −15.0 to −11.5%)
P < 0.001 to −1.7%) P = 0.009 P < 0.001
P < 0.001
Vacuette 1 mL ‐ r = 0.97 (95% CI, 0.95‐0.98) r = 0.96 (95% CI, 0.93‐0.98) r = 0.95 (95% CI, 0.92‐0.97)
Bias: 0.6% (95% CI, −0.6 to Bias: 1.9% (95% CI, 0.6 to 3.1%) Bias: −9.8% (95% CI, −11.4 to −8.1%)
1.8%) P = 0.005 P < 0.001
P = 0.232
MiniCollect ‐ ‐ r = 0.96 (95% CI, 0.92‐0.97) r = 0.93 (95% CI, 0.88‐0.96)
0.5 mL Bias: 1.3% (95% CI, −0.2 to 2.8%) Bias: −10.4% (95% CI, −12.3 to −8.4%)
P = 0.095 P < 0.001
Vacutest 2 mL ‐ ‐ ‐ r = 0.97 (95% CI, 0.95‐0.98)
Bias: −11.6% (95% CI, −12.8 to −10.4%)
P < 0.001
SALVAGNO et al.
|
      5

TA B L E 5   Comparison of fibrinogen test results obtained in different citrated blood tubes

MiniCollect
Tubes Vacuette 1 mL 0.5 mL Vacutest 2 mL Micro Test 0.5 mL

Vacuette r = 0.98 (95% CI, 0.96‐0.99) r = 0.97 (95% CI, r = 0.97 (95% CI, 0.95‐0.98) r = 0.98 (95% CI, 0.96‐0.99)
3 mL Bias: 0.3% (95% CI, −1.1 to 1.8%) 0.95‐0.98) Bias: 2.7% (95% CI, 1.2 to 4.1%) Bias: 7.1% (95% CI, 5.8 to 8.5%)
P = 0.653 Bias: 2.4% (95% P = 0.001 P < 0.001
CI, 0.7 to 4.1%)
P = 0.007
Vacuette ‐ r = 0.96 (95% CI, r = 0.97 (95% CI, 0.94‐0.98) r = 0.97 (95% CI, 0.94‐0.98)
1 mL 0.94‐0.98) Bias: 2.3% (95% CI, 0.5% to 4.1%) Bias: 6.8% (95% CI, 4.7 to 8.8%)
Bias: 2.0% (95% P = 0.013 P < 0.001
CI, 0.1 to 0.4%)
P = 0.036
MiniCollect ‐ ‐ r = 0.96 (95% CI, 0.93‐0.98) r = 0.97 (95% CI, 0.96‐0.99)
0.5 mL Bias: 0.3% (95% CI, −1.7 to 2.2%) Bias: 4.7% (95% CI, 3.1 to 6.4%)
P = 0.779 P < 0.001
Vacutest ‐ ‐ ‐ r = 0.97 (95% CI, 0.95‐0.98)
2 mL Bias: 4.5% (95% CI, 2.9 to 6.1%)
P < 0.001

may display differences which can sometimes exceed the minimum REFERENCES
quality specification, thus reinforcing the recommendation to per‐ 1. Simundic A‐M, Bölenius K, Cadamuro J, et al. Joint EFLM‐
form local validation studies before introducing new blood tubes COLABIOCLI Recommendation for venous blood sampling. Clin
into routine phlebotomy practice. Chem Lab Med. 2018;56:2015‐2038.
2. Lippi G, Favaloro EJ. Preanalytical Issues in Hemostasis and
Thrombosis Testing. Methods Mol Biol. 2017;1646:29‐42.
AC K N OW L E D G E M E N T S 3. Lippi G, Salvagno GL, Montagnana M, Lima‐Oliveira G, Guidi GC,
Favaloro EJ. Quality standards for sample collection in coagulation
The authors wish to thank Mrs. Cristina Recchi for her technical con‐ testing. Semin Thromb Hemost. 2012;38:565‐575.
tribution to the measurements. All blood tubes used in this experi‐ 4. Favaloro EJ, Lippi G. Pre‐analytical issues that may cause misdi‐
agnosis in haemophilia and von Willebrand disease. Haemophilia.
mental study were directly purchased from the manufacturers with
2018;24:198‐210.
proprietary funding. The companies were not aware of the purposes 5. Lippi G, Franchini M, Montagnana M, Salvagno GL, Poli G, Guidi GC.
of this study and had no role in directing the study or influencing Quality and reliability of routine coagulation testing: can we trust
the outcomes. The views in this report are those of the authors and that sample? Blood Coagul Fibrinolysis. 2006;17:513‐519.
6. Lippi G, Salvagno GL, Guidi GC. No influence of a butterfly device
not necessarily those of the University Hospital of Verona or NSW
on routine coagulation assays and D‐dimer measurement. J Thromb
Health Pathology. Haemost. 2005;3:389‐391.
7. Lippi G, Salvagno GL, Montagnana M, Poli G, Guidi GC. Influence
of the needle bore size on platelet count and routine coagulation
C O N FL I C T O F I N T E R E S T testing. Blood Coagul Fibrinolysis. 2006;17:557‐561.
8. Gosselin RC, Janatpour K, Larkin EC, Lee YP, Owings JT. Comparison
The authors of this paper have no conflicts of interest, including spe‐ of samples obtained from 3.2% sodium citrate glass and two 3.2%
cific financial interests, relationships, and/or affiliations relevant to sodium citrate plastic blood collection tubes used in coagulation
the subject matter or materials included. testing. Am J Clin Pathol. 2004;122:843‐848.
9. Yavaş S, Ayaz S, Köse SK, Ulus F, Ulus AT. Influence of blood collection
systems on coagulation tests. Turk J Haematol. 2012;29:367‐375.
AU T H O R C O N T R I B U T I O N S 10. Lima‐Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth
G, Guidi GC. Sodium citrate vacuum tubes validation: preventing
Giuseppe Lippi and Emmanuel J. Favaloro contributed to the concept, preanalytical variability in routine coagulation testing. Blood Coagul
designed the study, and wrote the manuscript; Gian Luca Salvagno and Fibrinolysis. 2013;24:252‐255.
11. van den Besselaar AM, van Vlodrop IJ, Berendes PB, Cobbaert CM.
Davide Demonte involved in data collection or processing; Giuseppe
A comparative study of conventional versus new, magnesium‐poor
Lippi, Emmanuel J. Favaloro, and Giovanni Poli involved in analysis or Vacutainer® Sodium Citrate blood collection tubes for determina‐
interpretation; and Giuseppe Lippi involved in literature search. tion of prothrombin time and INR. Thromb Res. 2014;134:187‐191.
12. Chuang J, Sadler MA, Witt DM. Impact of evacuated collection tube
fill volume and mixing on routine coagulation testing using 2.5‐ml
ORCID (pediatric) tubes. Chest. 2004;126:1262‐1266.
13. Drews RE. Critical issues in hematology: anemia, thrombocytope‐
Gian Luca Salvagno  https://orcid.org/0000-0003-0487-9051 nia, coagulopathy, and blood product transfusions in critically ill
patients. Clin Chest Med. 2003;24:607‐622.
Giuseppe Lippi  https://orcid.org/0000-0001-9523-9054
|
6       SALVAGNO et al.

14. McPherson RA. Blood sample volumes: emerging trends in clin‐ Coagulation Assays and Molecular Hemostasis Assays: Approved
ical practice and laboratory medicine. Clin Leadersh Manag Rev. Guideline‐Fifth Edition; CLSI document H21–A5. Clinical Laboratory
2001;15:3‐10. Standards Institute: Wayne, PA; 2008.
15. Wang S, Ho V, Roquemore‐Goins A, Smith FA. Effects of blood col‐ 20. Ricos C, Alvarez V, Cava F, et al. Current databases on bio‐
lection tubes, including pediatric devices, on 16 common immuno‐ logic variation: pros, cons and progress. Scand J Clin Lab Invest.
assays. Clin Chem. 2006;52:892‐893. 1999;59:491‐500.
16. Peng Z, Mao J, Li W, Jiang G, Zhou J, Wang S. Comparison of per‐ 21. Lippi G, Cornes MP, Grankvist K, Nybo M, Simundic AM. EFLM
formances of five capillary blood collection tubes. Int J Lab Hematol. WG‐Preanalytical phase opinion paper: local validation of blood
2015;37:56‐62. collection tubes in clinical laboratories. Clin Chem Lab Med.
17. Desconclois C, Eschwège V, Proulle V, Boutekedjiret T, Dreyfus M, 2016;54:755‐760.
Toulon P. Underestimation of unfractionated heparin therapy as‐
sessment due to platelet activation when using partial‐draw (pedi‐
atric) citrate collection tubes. Thromb Res. 2014;134:1117‐1122.
How to cite this article: Salvagno GL, Demonte D, Poli G,
18. Milos M, Herak D, Kuric L, Horvat I, Zadro R. Evaluation and perfor‐
Favaloro EJ, Lippi G. Impact of low volume citrate tubes on
mance characteristics of the coagulation system: ACL TOP analyzer
‐ HemosIL reagents. Int J Lab Hematol. 2009;31:26‐35. results of first‐line hemostasis testing. Int J Lab Hematol.
19. Adcock DM, Hoefner DM, Kottke‐Marchant K, et al. Collection, 2019;00:1–6. https://doi.org/10.1111/ijlh.13028
Transport, and Processing of Blood Specimens for Testing Plasma‐Based

You might also like