Product Specifications

M-30PCFL lyse breaks down the red cells and achieve WBC 3-part differential.
Probe Cleanser

Probe Cleanser is used for the regular cleaning of the analyzer

2.11.2 Reagent Consumption Volume

Table 2-10 Reagent Consumption Volume

Sample Presentation Mode Diluent (ml) Lyse(ml) Probe Cleanser (ml)


Whole blood mode 19.96 0.31 0
Predilute mode 19.93 0.31 0
Dispensing diluent 1.4 0 0
Shutdown 36.1 0 1
Startup (exclude the consumption 0
44.04 0
during background check)
Exit standby1 4.29 0 0
Exit standby2 33.28 0 0
Exit standby3 44.04 0 0
Exit standby4 44.35 0.31 0

2.11.3 Controls and Calibrators


The controls and calibrators are used for the analysis quality control and calibration of the
analyzer.
The controls are suspension of stimulated human blood, specifically manufactured to
monitor and evaluate the analysis precision of the analyzer. The controls are prepared with
three levels, namely low, normal and high. The calibrators are also suspension of stimulated
human blood, specifically manufactured for the calibration of the analyzer, so as to build the
metrological traceability of analysis results. For the use and storage of controls and calibrators,
please refer to the Instruction for Use of each product.
All references related to the controls and calibrators in this manual refer to the "controls"
and "calibrators" Mindray specifically formulated for BC20s/30s by Mindray.

2.1. Information Storage Capacity


Table 2-11 Data storage requirements

Data BC-20s: no less than 200,000 samples


storage BC-30s: no less than 500,000 samples
capacity
Information The information stored should at least include the following:
(histogram), sample information, patient information, flags as
well as any special information of the analyzer.

2-11