Professional Documents
Culture Documents
Laycock 2001 PERFECTscheme
Laycock 2001 PERFECTscheme
by J Laycock
Pelvic Floor Muscle D Jerwood
Assessment:
The PERFECT Scheme
Summary
Introduction
Aims of study
The pelvic floor muscles (PFM) consist of
1. To develop a digital technique to assess pelvic floor
approximately 70% slow-twitch (type 1)
muscles (PFM). and 30% fast-twitch (type 2) muscle fibres
2. To validate the technique and test for validity and (Gilpin et al, 1989). Although the PFM are
reliability. thought to work tonically and reflexly
3. To translate the assessment into an exercise-based during routine daily activities, voluntary
regimen. contractions are required for training.
PFM weakness is recognised as one of
Method and Results PERFECT is an acronym with the problems encountered in patients
P representing power (or pressure, a measure of strength with urinary and faecal incontinence, and
using a manometric perineometer), E = endurance, re-education should address the perceived
R = repetitions, F = fast contractions, and finally ECT = every deficit, be it in the slow- or/and fast-twitch
muscle fibres. There are several ways by
contraction timed. The scheme was developed to simplify
which to assess the PFM, including digital
and clarify PFM assessment. The pressure (referred to palpation (Brink et al, 1989), using a
hereafter as power, although actually a surrogate measure pressure perineometer (Laycock and
of muscular strength) of a contraction was validated by Jerwood, 1994), electromyography (EMG)
examining perineal lift and perineometric pressure during (Haslam, 2002) ultrasound scanning
a maximum voluntary contraction (MVC). Data analysis (Vierhout and Jansen, 1989) and
magnetic resonance imaging (MRI)
demonstrated significant positive correlations between power
(Khullar, 2002). Evaluation of muscle
and both lift (r = 0.864; p = 0.031) and perineometric strength and endurance provides
pressure (r = 0.786; p = 0.001); digitally assessed endurance information on the severity of muscle
correlated with the area under the pressure curve of a weakness and forms the basis of patient-
sophisticated perineometer (r = 0.549; p = 0.001). A study specific exercise programmes.
of inter-examiner reliability demonstrated highly significant Research involving PFM assessment
probably started with the work of Kegel
positive correlations between two examiners for power
(1948), who advocated clinical assessment
(r = 0.947; p < 0.001), endurance (r = 0.946; p < 0.001), of the pubococcygeus as part of a routine
repetitions (r = 0.730; p < 0.005) and number of fast gynaecological examination, using one
contractions (r = 0.909; p < 0.001). Scatter diagrams finger to palpate the pubococcygeus per
confirmed a lack of systematic bias between examiners. vaginam. He maintained that two fingers
Test-retest reliability produced highly significant correlations placed the vaginal tissues under tension
and distorted the anatomic relationships
(p < 0.001) between power (r = 0.929) and endurance values
(Kegel, 1956). However, the premise that
(r = 0.988) recorded on two different occasions, with no stretching the vaginal tissues during ass-
convincing evidence of significant discrepancies between the essment may adversely affect outcomes
pair of assessments. was challenged by Chiarelli (1989) who
maintained that stretching the muscle
Conclusions The PERFECT scheme has demonstrated may produce an enhanced response. In
reliability and validity as an assessment tool. Furthermore, support of this hypothesis, Jahnke et al
it is proposed that this scheme provides guidelines for the (1989) showed that the initial phase of
planning of patient-specific exercise programmes which passive muscle stretching is associated
satisfy the principles of muscle training. with a rise in reflex tone.
The one-finger assessment technique
was again described by Hendrickson can easily slip out, a score of 1 is given.
(1981), who defined a mild contraction If the finger can be gripped somewhat
as slight unsustained pressure on the but remains in the same position, a
examiner's finger; a moderate contraction score of 2 is given, and if the finger is
as firm pressure held for one to three forcibly gripped, expelled, or pulled
seconds; and a strong contraction as anteriorly, a score of 3 is given.
forceful pressure on the examiner's finger Test-retest reliability of this scoring
sustained for four seconds or more. system showed a statistically significant
A different method was devised by relationship of observation between ten
Graber et al (1981) based on four com- patients examined and then re-examined
ponents: control, sustained strength, 10 days later.
atrophy and tone. Control is measured Another scoring system, this time linked
using a perineometer and indicates the with incontinence, was described by Brink
patient's ability to contract and relax the et al (1989). This measure uses concepts
PFM to command. Sustained strength is of pressure (rated 1 to 4), time and
an indication of the time (up to 10 displacement. These were evaluated by
seconds) that a strong contraction could the index and middle fingers in the
be held. Atrophy is used to describe the antero-posterior position (index finger
development and wastage of the muscle. resting on the middle finger) introduced
This involved the technique of mapping, 4 cm to 6 cm into the vagina. Test-retest
which necessitated palpating the PFM and for this scoring method was r = 0.65,
charting areas of reduced muscle bulk. p < 0.01, with inter-rater reliability
Tone was determined by the degree of r = 0.91, p < 0.01. A negative correlation
resistance of the muscle against moderate between muscle strength and both urine
pressure from the examining finger. loss and age was demonstrated.
A further report (Worth et al, 1986) The authors of all the aforementioned
described another one-finger digital studies did not attempt to use any
vaginal assessment with four components: standard international muscle-grading
pressure, duration, ribbing and position scheme or to differentiate between
(of the examining finger). slow- and fast-twitch muscle activity.
■ They defined pressure as the strength Furthermore, they did not relate the
of contraction. If no pressure is felt, a assessment findings to an individual
score of 1 is given. If moderate pressure exercise programme. Moreover, co-
is felt, a score of 2 is given, and firm contraction of the abdominal muscles
pressure is recorded as 3. was discouraged and this is now
acknowledged to be inappropriate
■ Duration is based on the length of time (Sapsford et al, 2001).
a contraction can be sustained. If no
contraction is felt or the contraction Aims of Study
lasts no longer than 1 second, a score 1. To develop a digital technique for
of 1 is assigned. A contraction held for quantitative assessment of the
2 to 3 seconds is scored 2, and a score voluntary contractility of the PFM.
of 3 is assigned to a contraction held
for 4 or more seconds. 2. To validate the above technique and
■ An assessment of ribbing refers to the test for observer reliability.
tone and texture of the PFM during
a contraction. If the muscle feels soft 3. To translate the measurements from
and flabby, it is assigned a score of 1; the digital assessment into the
a score of 2 describes a muscle that planning of a patient-specific exercise
feels different from the surrounding programme.
tissues but not ribbed, and a score of
3 is assigned if the muscle feels distinct, Methods and Materials
Laycock, J and PERFECT is an acronym to remind all
like rings of ribbing or ribbed muscle
Jerwood, D (2001).
tissue. health professionals of the need to
‘Pelvic floor muscle
assessment: The ■ Assessment of position refers to the assess the main components of PFM
PERFECT Scheme’, contractility. This assessment scheme was
plane the examining finger is in,
Physiotherapy, 87, 12, developed to provide a simple, reliable
in relation to the vaginal introitus.
631-642. method of PFM evaluation and involves
If no force is exerted and the finger
four components, as shown in table 1.
Table 1: The PERFECT assessment scheme stronger can be observed as an in-drawing Authors
of the perineum and anus. J Laycock PhD FCSP
P Power (pressure)
Grade 4 Increased tension and a good is a specialist
E Endurance
contraction are present which are capable continence
R Repetitions
of elevating the posterior vaginal wall physiotherapist in
F Fast private practice who
E Every against resistance (digital pressure
carrried out the
C Contraction
applied to the posterior vaginal wall). research for this
T Timed Grade 5 Strong resistance can be applied article.
to the elevation of the posterior vaginal D Jerwood BSc PhD
Although the study was carried out on wall; the examining finger is squeezed FSS is head of
women by palpating the perivaginal and drawn into the vagina (like a hungry mathematics, School
muscles per vaginam, the PERFECT baby sucking a finger). of Computing and
Mathematics,
assessment can also be used for PFM
University of
assessment per rectum in men and women. Table 2: Proposed modified Oxford grading
scheme Bradford, and was
To ensure reproducibility, the following responsible for the
factors were adopted throughout the Grading Muscle response statistics.
study:
0 Nil
■ The location and action of the pelvic This article was
1 Flicker
floor muscles were described to the received on January
2 Weak 17, 2000, and
subjects in enough detail for adequate
3 Moderate accepted on August
understanding of this muscle group.
4 Good 22, 2001.
■ Whenever possible, subjects were
5 Strong
positioned in supine with their head on
two pillows. The hips were flexed and Address for
abducted, and the knees bent. Correspondence
Consequently, in a specific case, the
■ The PFM were examined using the power could simply be recorded as Jo Laycock,
index finger placed approximately grade 3 for a moderate contraction. The Culgaith Clinic,
Pea Top Grange,
4 cm to 6 cm inside the vagina and However, the registered grade is
Culgaith, Penrith
positioned at 4 o'clock and 8 o'clock permitted to be augmented with a symbol
CA10 1QW.
to monitor muscle activity. Moderate + or –, when the need arises. Thus 3+
pressure was applied over the muscle could be translated as there being more
bulk to assist in the initiation of the than a moderate contraction but less than
appropriate muscle contraction. a good contraction (grade 4). Similarly, a
■ Verbal informed consent was obtained 3-- is recorded when the contraction is less
than a grade 3, but more than a grade 2.
from all subjects.
This augmentation is to allow for an
Power element of doubt to be introduced and
Power is measured on a modified Oxford thereby to soften this (partly subjective)
scale (table 2). The authors acknowledge six-point ordinal scale. There will
that digital palpation during a maximal admittedly be a learning curve involved
voluntary contraction (MVC) evaluates in assessing the strength of a PFM
muscle strength, not power. However, with contraction.
this caveat, the misnomer ‘power’ will be
used throughout the text. Endurance
The following definitions are proposed: Endurance is expressed as the length of
time, up to 10 seconds, that an MVC can
Grade 0 No discernible muscle
be sustained before the strength is
contraction.
reduced by 35% or more. In other
Grade 1 A flicker or pulsation is felt words, the contraction is timed until
under the examiner's finger. the muscle starts to fatigue. A further
possible indication of PFM fatigue may
Grade 2 An increase in tension is
be the simultaneous contraction of hip
detected, without any discernible lift.
adductors and glutei, and the stronger co-
Grade 3 Muscle tension is further contraction of transversus abdominis.
enhanced and characterised by lifting of Breath-holding should be discouraged; if
the muscle belly and also elevation of the detected, the subject should be instructed
posterior vaginal wall. A grade 3 and to contract the pelvic floor on expiration.
Inter-examiner Reliability Study Table 5: Analysis of results of digital assessment of ten patients
by two examiners (SHD and JL) on the same visit
Reliability between different examin-
ers involved two physiotherapists (JL Patient P E R F
and SHD) each performing complete SHD JL SHD JL SHD JL SHD JL
PERFECT assessments on ten subjects
1 3.0 2.5 7 6 6 5 7 6
on the same visit, and the results are
shown in table 5. Analysis shows highly 2 4.0 4.0 9 10 3 6 8 7
significant positive correlation coefficients 3 5.0 5.0 10 10 8 6 10 10
between the two examiners for P 4 4.5 4.0 10 10 8 7 10 10
(r = 0.947; p < 0.001), E (r = 0.946; 5 2.0 2.0 4 3 3 2 6 5
p < 0.001), R (r = 0.730; p < 0.005) and F 6 2.0 3.0 3 2 4 5 5 6
(r = 0.909; p < 0.001). 7 3.0 3.0 3 3 6 5 5 6
When testing for significance of 8 1.0 1.0 2 2 1 2 3 2
correlation coefficients, the P values 9 1.5 1.0 4 5 3 3 4 5
relate to rejection of a null hypothesis 10 4.0 4.0 3 3 4 5 7 7
claiming that the true correlation is zero Spearman's r = 0.947 r = 0.946 r = 0.730 r = 0.909
which is not a useful concept within correlation
validity studies. Subjects (such as patients Coefficient
3 and 4 in table 5) who are performing at significance p < 0.001 p < 0.001 p < 0.005 p < 0.001
level
the boundary of data censorship (that is,
holding an MVC for at least 10 seconds, 5
and being able to produce at least 10
fast contractions) will enhance any
correlation coefficients. Removing pat- 4
ients 3 and 4 from these correlation
studies will reduce every one of the
Repetitions (SHD)
assessments of endurance (E) and the two Table 6: Test-retest results of digital assessment
of power (P) and endurance (E)
counts of fast contractions (F) (not
reproduced here), however the weakest Patient Power Endurance
inter-assessor agreement was found with Test 1 Test 2 Test 1 Test 2
repetitions (R) and this is illustrated in
1 3.0 4.0 2 2
figure 2. Now, there is only one (perfect)
2 0.0 0.0 0 0
agreement, with four over-assessments
(totalling 5 units) and three under- 3 3.0 3.0 3 3
assessments (also totalling 5 units); the 4 4.0 3.5 10 10
greatest discrepancies occurr with two 5 3.0 3.5 5 6
subjects who recorded 6 repetitions for 6 3.5 4.0 10 10
JL. The mean nett difference is now zero 7 2.5 2.0 5 6
but (with a correlation coefficient of only 8 2.0 2.0 2 3
0.730), the impression is given of a lack of 9 0.0 0.0 0 0
systematic bias between examiners with 10 3.0 3.5 7 7
less convincing consistency. Notice, 11 2.5 2.5 4 4
however, when recording repetitions the 12 2.5 3.0 2 3
censoring limit of 10 which had been 13 4.0 4.5 8 9
imposed by the physiotherapist was never 14 4.0 4.0 10 10
actually attained for any of these ten 15 4.5 4.0 4 6
subjects, although other imposed maxima 16 2.5 2.5 8 8
were regularly attained by subjects 3 and 4. 17 2.0 2.0 10 10
This could imply that the R component 18 3.0 3.5 10 10
was found to be a particularly fatiguing 19 5.0 5.0 5 5
exercise by those subjects, and its 20 3.5 4.0 10 10
implementation introduced additional Spearman's r = 0.929 r = 0.988
sources of intra-patient variation to be correlation
confounded with inter-assessor variation. Coefficient p < 0.001 p < 0.001
The inter-examiner assessments of significance
number of repetitions (R) apart, taken level
over the remaining three digital scores,
this study demonstrated 46.7% exact minimum unit of graduation (of 0.5).
agreement of the digital scores. However, Regarding endurance, 70% exact
disagreement in all but two cases (both in agreement was demonstrated, with a
the repetition test) did not exceed one variation of only one second in all except
unit of measurement. These encouraging one, which was two seconds. In all cases,
results provide global evidence in support endurance remained the same or
of the inter-examiner reliability of these increased, despite the fact that these
components of the digital assessment of women were not instructed to practise
the PFM. pelvic floor exercises. The complete
absence of patients with shorter
Test-retest Reliability Study endurance periods is surprising, but the
The results of re-testing 20 women two to mean nett shift is only 0.4 seconds per
five weeks after the initial assessment patient and (again) well within the
are shown in table 6. These results show minimum unit of evaluation (of one
that for muscle strength (P), 45% exact second).
agreement was demonstrated. Further- Evaluating Spearman's rank correlation
more, of the 11 measures that differed, coefficients for these test-retest results,
ten of these differed by only 0.5 grade and the coefficient for power (P) is found to
the remainder by one grade. In addition, be 0.929 (p < 0.001) and for endurance
it is shown that 8/11 demonstrated an (E) stands even higher at 0.988. These
increase in strength whereas 3/11 extremely high levels of correlation
demonstrated a decrease. together with mean discrepancies which
Scatter graphs (similar to figures 1 and 2, cannot be reduced by systematic adjust-
but not reproduced here) illustrate a ments provide convincing evidence that,
slight tendency for some patients to for an experienced physiotherapist, the
improve over the period between the digital evaluation of power and
tests. However, the mean nett shift of only endurance of PFM contractions following
+0.125 per patient is well within the a short delay is a reliable technique.
before the muscles fatigued, was incorp- using transrectal ultrasound on 17 women
orated into the scheme to provide (15 with stress incontinence and two
information on the number of con- with other bladder problems). They rep-
tractions an individual should per- orted a mean lift of 5.5 mm with a PFM
form at each exercise session to predict contraction which compares well with
‘overload’. Previous studies have sugg- the 5.0 mm lift observed in incontinent
ested a wide range of daily contractions; women in this study (6.0 mm for cont-
Benvenuti et al (1987) proposed 120 daily inent subjects).
PFM contractions whereas Bø et al (1989)
used 8 to 12 daily contractions, and these Validity Study 2
recommendations appeared to be made This study aimed to validate the subjective
at random. The present study proposes digital assessment of P (pressure/power)
patient-specific exercise programmes, as and E (endurance) and to compare
practised in other physiotherapy regimens the sensitivity between digital and
for muscle dysfunction. The number of perineometric techniques in a large
repetitions gives further information on sample of women (n = 233), most of
fatigue; Edwards (1978) states that fatigue whom were incontinent.
is failure to maintain the expected force Accuracy of the digital method was
with continued or repeated contractions. demonstrated by the highly significant
It is postulated that the last component, correlation of P with maximum pressure
F -- number of fast contractions, provides (r = 0.786; p < 0.001), and E with the area
a measure of fast-fibre activity. Millard under the pressure curve (r = 0.549;
(1987) was possibly the first to recognise p < 0.001) as recorded by a perineo-
the importance of practising fast and slow meter. This evidence lends support to the
PFM contractions. validity of the digital method. These
Compared with other studies (Hend- findings are reinforced by the study by
rickson, 1981; Graber et al, 1981; Worth et Brink et al (1989) of 388 older women
al, 1986; Brink et al, 1989), the PERFECT (mean age 67.5), who showed significant
method gives the examiner more flex- positive correlation between digital
ibility and is less ambiguous. pressure scores and electromyography
(EMG) scores (r = 0.60; p < 0.01).
Validity Study 1 Further support for the digital tech-
PFM digital assessment of strength is nique is given in a study (N = 263)
represented by P (power/pressure), comparing digital scores (modified
which is thought to manifest itself Oxford scale) and pressure (using the
with both squeeze pressure and lift. pelvic floor exerciser/perineometer –
Consequently, it was decided to examine PFX), where good agreement between
the ‘lift’ component of an MVC in this the two techniques was demonstrated
study. The sample number is small (n = 8) (Isherwood and Rane, 2000).
and so the results should be interpreted To assess endurance digitally the
with caution, but analysis of the data examiner must be able to detect a
showed a positive and significant reduction in pressure during a 10-second
correlation between perineal lift and MVC, at which point the time (in
digital assessment scores (r = 0.864). This seconds) is recorded. Digital measures
theory would endorse the postural are admittedly less sensitive than
function of the PFM in supporting the perineometric measures, due largely to
proximal urethra, with reduced support, the subjective nature of the test and the
causing bladder neck descent on difficulty in gauging small changes in
coughing, precipitating incontinence. It is pressure. In addition, the area under the
postulated that continuous activity pressure curve is a record of total pressure
(resting tone) of the slow-twitch fibres recorded during the ten-second MVC
should maintain the advantageous (even though it may fluctuate), whereas
position of the bladder neck and E is an estimation of the time an MVC
proximal urethra, with the fast-twitch can be held at the maximum level. This
fibres reflexly recruited to provide a disparity of measurement is reflected in
quick, strong re-enforcement of urethral the reduced (but still highly significant)
lift and squeeze, during, for example, correlation (r = 0.549; p < 0.001). The
coughing. Bladder neck lift was also highly significant differences in both
reported by Vierhout and Jansen (1989) P and E values between continent
Key Messages
■ Digital assessment of the contractility ■ The PERFECT assessment is easy to
of PFM can be carried out during perform, reliable and reproducible.
vaginal examination.
■ The assessment scheme described
■ Assessment should include evaluation provides information for a patient-
of PFM strength and endurance, and specific exercise programme.
reflex activity (during cough).