FT-IR-
Coupling
Perfect Combination Bridges the Analytical Gap and Facilitates Fingerprint Analysis
Thermal Analysis provides ideal tools for the characterization of all kinds of organic and inorganic
solids and liquids. Thermodynamic transitions, thermal stability, decomposition and chemical
reactions can be detected and quantified with high accuracy over a broad temperature range.
In some cases, however, information about the type of gases evolved is needed in order to gain a
detailed understanding of the chemistry behind the processes. The combination of thermal
analysis with the powerful infrared spectroscopy for gas analysis bridges this analytical gap. It
allows for deeper insight into the material’s behavior and may provide a fingerprint of the analyzed
material.
The Proteus® software for thermal analysis and the OPUS software for the FT-IR measurement are
integrated with one another to support the Thermal Analysis-to-FT-IR coupling. The relationship
to temperature and time of all information produced by the running experiment is meticulously
maintained.
2
Harmonized Instrument Combinations with Fully Integrated Software
from the Specialists in Thermal Analysis and IR Spectroscopy
01 02 03 04
Additional operation for
regular FT-IR experiments
or for hyphenated tests without
modification and disassembling High resolution and
(by use of an external gas cell) high sensitivity
3
Optimized TG/STA/DSC – FT-IR Coupling
Infrared spectroscopy is a
10-13
m 10 m 4 classical technique which depends
Wave length on the interaction of infrared
radiation with the vibrating dipole
moments of molecules. It provides
Gamma X rays UV Visible Infrared Micro- Radio a characteristic spectrum for each
rays waves waves
substance, with the exception of
homonuclear diatomic molecules
or noble gases. Examples of
homonuclear diatomic molecules
are N2 and O2.
Detector
– depending on the frequencies
Intensity
0.05
Transformation. 0.06
Absorbance
0.04
0.05
0.03
Absorbance
0.04
0.02
0.03
0.01
0.02
0.00
0.01 4000 3500 3000 2500 2000 1500 1000
0.00 Wavenumber cm-1
4000 3500 3000 2500 2000 1500 1000
Wavenumber cm-1
4
Sophisticated Gas Cell Design
Windows of the external gas cell for the INVENIO spectrometer with screw thread; dual
window technique for safe cleaning and removal
5
TGA/STA-FT-IR
Coupling for
Maximum
Flexibility
Bruker INVENIO with external gas cell coupled to the NETZSCH TG 209 F1 Libra® with ASC and QMS 403 Aëolos Quadro
Because of this natural gas flow path, top-loading In order to achieve this requirement, all TGA/DSC/
thermobalances are ideally suited for coupling to STA-FT-IR systems are vacuum-tight or gas-tight.
evolved gas analyzers such as FT-IR spectrometers,
mass spectrometers and/or GC-MS systems (gas
chromatograph-mass spectrometers) – and for most
instruments, this can also be combined with an
automatic sample changer (ASC).
∙
with transfer line for
Direct PERSEUS® coupling without transfer line STA-FT-IR coupling
6
FT-IR gas cell Transfer line
(max. 200°C) (max. 230°C) Optional MS
coupling
Coupling
adapter
(max. 300°C)
Outlet
Sample
IR detector
Schematic of the TG 209 F1-FT-IR coupling using an
external gas cell, adapter and heated transfer line
TG cell
An adapter together with a short transfer line should connect the gas outlet of the TGA, DSC or STA furnace to
the gas cell of the FT-IR spectrometer. The evolved gases are transferred using a carrier gas (usually nitrogen).
To prevent cold spots and thus condensation of the gases evolved, the entire gas path should be heated. The
temperatures are up to 300°C for the adapter and max. 230°C for transfer line.
∙
systems. They do not require any additional
cooling and are therefore particularly well- Continuously heated interface and gas cell
suited for TGA/DSC/STA test runs with an
∙ Minimized risk of condensation
∙
automatic sample changer (ASC) or for measure-
ments of longer duration. Mirror-free gas cell
∙
The external gas cells are equipped with MCT
detectors (Mercury-Cadmium-Telluride) by Tests with automatic sample changer
default. They provide a significantly higher for high sample throughput and great
signal-to-noise ratio (compared to DLaTGS) and variability of measurement parametes
require liquid nitrogen cooling.
7
Coupling Via Transfer Line
Low-Volume Gas Path Ensures Fast Response Times
Bruker INVENIO with external gas cell and an ATR TG 209 F1 Libra®
unit in the FT-IR sample compartment
8
STA-FT-IR – Sensitive Evolved Gas Analysis
9
PERSEUS®
Perfect Coupling of Thermal Analysis and FT-IR
10
Space-Saving PERSEUS® STA 449
To gain further information about the evolved gases, the PERSEUS® STA Schematic of the gas path of a
can additionally be coupled to a GC-MS system. PERSEUS® STA system
11
DSC-FT-IR Coupling
12
PulseTA® – A Clever Tool for Calibration,
Quantification and Catalysis Studies
PulseTA® technique is used to injected
Injected gas
gas
inject a defined amount of gas
into the purge gas flow of a
thermobalance (TGA) or simulta-
neous thermal analyzer (STA) and
then monitor the therewith carrier gas
Purge gas
related changes in the sample
mass, enthalpy or changes in the TA
TA MS/FTIR
MS/FT-IR
evolved gases.
pulse device
Pulse device
TGA-DSC Gas composition
0.3
TG curve
Absorbance Units
0.2
0.1
3-D plot for CO2 calibration pulses and CaCO3 decomposition; pulses for
4000 quantification need to be repeatable and temperature-independent
3000
0.0 (a linear relationship to the injected gas concentration)
2000 800
600
Wavenumber cm-1 1000 200
400
Temperature
13
Comprehensive Software
Bruker OPUS and NETZSCH Proteus® –
Unrivaled Combination for Maximum
Ease of Use
The alliance between the NETZSCH Proteus® software and the OPUS FT-IR software is
based on effective data exchange and serves to unify the coupled system functionally.
Measurements are controlled via the NETZSCH Proteus® software. The user only needs
to input the command for data acquisition and for the start of measurement once, and
both the OPUS and Proteus® software will be readied with parameter inputs. Online data
collection is simultaneous and synchronized to ensure precise time and temperature
correlation between all signals from the two coupled instruments during evaluation.
The user operates the two software packages from a single computer and has access to
the full range of possibilities for data evaluation and results display in either package at
any time.
14
∙ Full software integration – online data exchange
between the two instrument software packages
∙ Conjoint presentation of the Gram-Schmidt plot
plus up to four pre-selected traces together with
during the running experiment thermal analysis curves in Proteus® software during
∙
the experiment
∙
Simultaneous instrument control and start-stop
function Online evaluation (SNAP SHOT) of TGA/STA/DSC
∙
measurements already including FT-IR data during
Segmental activation or deactivation of the FT-IR the measurement
∙
coupling with one mouse click
∙
Trace calculations with evaluation of characteristic
Automatic saving of data sets for both temperatures and peak areas together with TGA
measurements (TGA and FT-IR) with identical and DSC curves
∙
file names (but different extensions) in the same
directories Combined analysis graphics of thermal analysis and
∙
FT-IR signals
∙
Measurements with automatic sample changer
allow for individual FT-IR measurement parameter Multi-component search in OPUS
∙
for each position
Identification by various gas phase librarys, e.g.,
TGA-FT-IR library of polymers by NETZSCH
Screenshot of the Proteus® software during evaluation of the same straw experiment:
Temperature-scaled plot of the TGA and DTG curves together with the Gram-Schmidt plot and the calculated
traces of methane, water and carbon monoxide (course of the absorption intensity of a specific band)
15
Advanced
Materials
Characterization
Fourier Transform Infrared Spectroscopy is a well-known technique in
analytics. Since FT-IR spectrometers can be found in so many laboratories,
it is often both convenient and logical to couple these to thermal analysis
systems in order to gain a deeper understanding of the decomposition
or evaporation processes under investigation. Tailored libraries support
evaluation and offer quick and easy spectrum interpretation.
Areas of Application
∙ ∙
Decomposition
Dehydration
∙ ∙
Compositional analysis
Binder burn-out
∙∙
Residual solvent content
Pyrolysis
∙∙
Coal analysis
Polymer content
∙
∙ ∙
Ash content
∙∙
Solid-gas reactions
Combustion Evaporation
∙∙
Oxidation
Corrosion
Outgassing
∙
Catalysis
16
Decomposition Behavior of Ethylene Vinyl Acetate (EVA)
0.20
0.15
TG
Intensity / a.u.
0.10
0.05
0.00
4000 500
400
3000 300
2000 200
1000 100 Temperature
Wavenumber cm-1
TG /% Gram Schmidt AC CH
To identify the decomposition 25
products, 2-D spectra can be 100
0.350
50
extracted from the 3-D cube
0.300 20
shown above and subjected to a 80 40
library search. In the present case, -28.9 % 0.250
analysis of the FT-IR absorption 349.6 °C -71.1 % 15
60 0.200
intensities shows the release of 351.4 °C 30
17
Decomposition Behavior of a Medical Drug Product: Aspirin®
2.0
0.25 TG curve
1.5
0.20
Absorbance Units
Absorbance Units
0.15
1.0
0.10
361 °C
salicylic acid
0.05
0.0
4000 3500 3000 2500 2000 1500 1000
Wavenumber cm-1
18
Differentiation between Diesel Fuel and Biodiesel
-50
There is a clear difference in the
50 100 150 200 250 300
Temperature /°C thermal stability between the
vaporization temperatures of pure
TGA comparison of various diesel fuel/biodiesel mixtures; sample masses: biodiesel (green) and pure diesel
10 to 11.5 mg, Al2O3 crucibles, heating rate: 10 K/min, N2 atmosphere
fuel (red) as discovered by TGA,
but there is no apparent system-
atic relationship with regard to the
Absorbance Units
mixtures. The comparison of the
0.12
corresponding IR spectra, however,
reveals a band at 1759 cm-1 which
1759 cm-1
is specific for biodiesel. It is most
0.10
probably related to the ester
group. The intensity of this band
0.08
even shows a relationship with the
biodiesel proportion: it increases
0.06 with increasing biodiesel content.
Biodiesel
Diesel fuel
0.04
Diesel fuel + 10% biodiesel
Diesel fuel + 30% biodiesel
0.02
0.00
2100 2000 1900 1800 1700 1600
Wavenumber cm-1
19
Simulation of Application and
Urea
100
80 -69 % The use of urea can be divers. It
TG /%
198 °C 310 °C
3.50 Gram Schmidt 244 °C plays an important role.
181 °C 201 °C 246 °C
3.00 NH3
2.50 HNCO 359 °C
311 °C 359
2.00 Trihydoxytriazine °C Investigations by TGA-FT-IR
1.50
1.00
showed a three-step thermal
310 °C
0.50 decomposition. During the first
0
50 100 150 200 250 300 350 400 450
mass-loss step ammonia and
Temperature /°C fulmic acid (HCNO) are formed,
which could be detected by FT-IR.
Measurement parameter: PERSEUS® TG 209 F1, RT – 450°C,
nitrogen atmosphere, heating rate 5 K/min, gas flow 40 ml/min Above 200°C, fulmic acid reacts to
cyanuric acid (trihydroxytriazine)
which decomposes again to fulmic
acid above 230°C. During the third
0.04 mass-loss step, the complete
cyanuric acid was detected in the
Y[Absorbance Units]
100
O
∆
NH3 + O C NH
80
H2N NH2
OH
Transmittance [%]
60
200-230 °C N N
3 O C NH
40 HO N OH
OH
20 N N >230 °C
3 O C NH
HO N OH
0
4000 3500 3000 2500 2000 1500 1000
Wavenumber cm-1
20
Manufacturing Conditions
Firing of Clay
DSC /(mW/mg) TG /%
↓ exo
To save energy in buildings, wall 578.3 °C
1299.1 °C
123.8 °C 720.6 °C
constructions should have low 110.0
0.0
thermal conductivity. One way to 11.49 J/g
37.15 J/g
2.55 J/g
9.74 J/g
achieve this is to use highly porous 929.3 °C
DSC
building bricks. Various organic -0.5 -17.26 J/g
105.0
products, capable of generating -775.08 J/g
a high volume of voids, are mixed 350.0 °C
-1.0
into the clay to form cavities -0.9 %
-0.3 % 100.0
during firing.
-1.5 -3.2 %
In this example, 107.6 mg of a clay
-1.4 % 95.0
green body was heated at a rate of TG
-2.0
10 K/min in Pt/Rh crucibles in -3.0 %
flowing air (50 ml/min). The -0.7 %
-0.6 %
standpoints.
Absorbance Units
0.3
TG curve 0.2
0.1
0.030
0.00
4000
0.025 3000
SO2 1000
1200
800
2000 600
Absorbance Units
400
0.020
Wavenumber cm-1 1000 200 Temperature
0.015
0.010
0.005
Characteristic HF spectrum
0.000
4200 4000 3800 3600 3400
displayed in the enlarged plot
Wavenumber cm-1
21
g
a t h e lp in detectin
n be of g re sing.
T-I R co upling ca g p o ly m er proces
TGA -F rin
failure du
causes of
Process Optimization
TG /%
Manufacturing of Silicones
Sample 1
100 -1.1 %
Sample 2
Here, two control samples of
95 silicones were tested – one
of them (sample 2) revealed
90
manufacturing problems. The
TGA curves exhibit significantly
-25.1 %
different thermal behavior. Sample
85 2 loses approx. 25% of its weight
whereas the mass loss for sample
80 1 is only about 1% up to 250°C.
CH3 CH3
n Cl Sl Cl + m H2O HO Sl O H +
(Me2SiO)y
x
CH3 CH3
Polymerisation mechnism of silicons
22
Technical Specifications
Characteristic Data for All Available FT-IR Coupling Systems
PERSEUS®
DSC 404 F1/F3; TMA 402 F1/F3;
TG 209 F1; STA 449;
Coupling system STA 2500 Regulus; DIL 402 Expedis DSC 204 F1
TG 209 F3 PERSEUS®
STA 449 F1/F3/F5 Select/Supreme
TG 209 F1
RT-1600°C
RT-1000°C (F3)
(Regulus, F5) RT-2000°C (STA) RT-1550°C (TMA)
Temperature range1 10°C(RT)-1100°C RT-700°C
-150°C-2000°C RT-1100°C (TGA) RT-2000°C (DIL)
(F1)
(F1/F3)
Measurements under
No, atmospheric
reduced pressure (for Yes (F1) Yes Yes Yes
pressure only
solvent separation)
Vacuum-tightness of
Yes (F1) Yes Yes Yes No
the complete system
Stand-alone operation
Yes Yes Yes Yes Yes
of the FT-IR
Additional analytical
MS or GC-MS
instruments to be MS or GC-MS MS or GC-MS MS or GC-MS MS or GC-MS
(F1)
coupled
Nickel-coated
Material gas cell Stainless steel Stainless steel Stainless steel Stainless steel
aluminum
1 For most instruments, variable furnaces are required to cover the stated temperature range
2 Larger Dewar recommended for MCT detector
3 For coupling with FT-IR spectrometers from other suppliers, please contact your NETZSCH representative
4 MCT detector requires LN2 cooling
23
The NETZSCH Group is an owner-managed, international technology
company with headquarters in Germany. The Business Units Analyzing &
Testing, Grinding & Dispersing and Pumps & Systems represent customized
solutions at the highest level. More than 3,800 employees in 36 countries and
a worldwide sales and service network ensure customer proximity and
competent service.
NETZSCH-Gerätebau GmbH
Wittelsbacherstraße 42
95100 Selb
Germany
Tel.: +49 9287 881-0
Fax: +49 9287 881 505
at@netzsch.com