Professional Documents
Culture Documents
7890 Chemstation Workshop - 101-200
7890 Chemstation Workshop - 101-200
7890 Chemstation Workshop - 101-200
101
Split Flow Diagram: Liner Overload
(SEPTUM)
PURGE VENT
TOTAL FLOW
COLUMN
102
Injection Processes
For Methanol:
A = .79 / 32
C = ( 275 + 273)/273 = 2
At 15 psi Head Pressure:
Expansion = 555
At 60 psi Head Pressure:
Expansion = 222
103
Pressure In GC
104
Various Capillary Liners
105
Split/Delayed Split Liners
Syringe In
Column Out
106
Delayed Split Injection
107
Delayed Split Injection
108
Delayed Split Injection
109
Delayed Split Flow Diagram
• Purge valve closed during injection (No split vent flow).
• At a specified time after injection, the purge valve opens to sweep remaining
vapors out of inlet .
110
When To Use Delayed Split Injections
The most important benefit of delayed split injection is that a majority of the
injected sample is introduced into the column; this results in much higher
sensitivity than can be normally achieved using split injection. Delayed split
injection is routinely used in areas such as environmental analysis, pesticide
monitoring of foods, and drug screening where low levels (low ppm levels) of
target analytes need to be detected. In certain applications, sample clean up
may be necessary since more “raw” sample is introduced into the column. So
column protection becomes as important as sensitivity.
111
Delayed Split Mode Problems
Several of the problems with split inlets have parallels for
splitless inlets. These include peak shape distortions due 1
to poor column installation techniques and liner/gold seal
1
contaminations. Liner volume overload can occur more 91 2
4
easily due to more sample being present longer in the 1
5 67 1
inlet. Leaks will also plague the inlets when in splitless 8
0
mode.
Unique to delayed split injections are problems
associated with improper purge times (when split flows
are initiated) and incorrect initial oven temperatures that
are at or not below the boiling point of the solvent. The
optimum purge times and purge (split) flows need to
always be experimentally verified for the sample that is
being analyzed. The chromatograms on the right show
the results for no purging at all (above) and the purge flow
being turned on at 0.75 minutes (below).
Since more of the sample is introduced into the inlet for
longer periods of time during delayed split injections,
overall inlet contamination effects are going to be more
severe. This also can cause column overloading. Column
overload peak shapes are very conspicuous, exhibiting
front tails (also called shark fin shapes). This is shown on
the next chromatogram. Such peak shapes seriously
compromise integration as well as reproducibility.
112
Flow versus Temperature
• One physical property of gases is that their viscosity
increases with and increase in temperature. This has a
number of effects on gas chromatography.
• For isothermal analyses all inlet flows should be set with the
column at the analysis temperature. If the flows are set at a
lower temperature and then the temperature is raised to the
analysis temperature the column flow will be decreased,
which can alter resolution, run time, and ultimately
precision. For split injections reducing column flow will
also increase split vent flow, altering the split ratio.
• For temperature programmed analyses the variation in
column flow with temperature means linear velocity cannot
be optimized through the entire run if the inlet is set in the
constant pressure mode. The column flow will decrease
with elevated oven temperature.
• The new generation of GC’s allows a constant flow setting,
which compensates for this.
113
Changes in Column Flow versus Temperature
0 0
0 50 100 150 200 250 300 350 400
Temperature degrees C
114
Operating Modes For EPC Systems
• Newer GC’s like the 6890 and 7890 can now operate with
constant flow as well.
• The user can now also utilize pressure programming. This may
or may not establish a constant flow mode. This is typically
used with isothermal column oven conditions.
115
Improvements to the Capillary Split/Delayed Split
Inlet for the 7890
116
6890’s Insert Weldment Assembly
117
Now Easier Access to the Inlet
118
7890A Insert Weldment
119
Benefit of Septum Purge
120
6890 Septum Purge
Vent to atmosphere
Septum purge
gas
mechanical forward fixed resistor
pressure regulator (fixed)
121
New 7890A Version
Split Carrier Supply
Septum Purge
EPC Module
Frit Frit
Valve Valve Valve
PS
FS
PS
Septum
PS = Pressure Sensor
FS = Flow Sensor
– a fixed restrictor
122
Septum Purge
Design includes:
EPC Module
Vent Frit
Valve
Proportional valve
PS
Septum
Pressure sensor
123
Control the Septum Purge Flow
124
Program the Septum Purge Flow
125
The Heart of a GC – The Column
126
Things to Know: Typical Capillary Column
127
Separation is a Partitioning Process
128
Capillary Columns
129
Separation Mechanisms
Generally:
Like dissolves like
or
Polar interacts with polar
130
Cause – Analyte Polarity
131
Packed GC Columns
132
Column Separation Characteristics
GOOD GOOD
POOR POOR
133
Capillary Column Efficiencies
134
Calculating Efficiency
t ' 2
t ' = t - t n = 5.545 R
Start R R m
Wh
tm t 'R Time
tR n = effective theoretical
plates.
Let's relate “n” to the length of the column.
n
Plates per meter (N) = or
L
L
Height equivalent to a theoretical plate (HETP) =
n
135
Measuring Linear Velocity and Flow Rate
136
Efficiency & Carrier Gas Linear Velocity
μ ( opt μ)
137
Optimum Flow Rate
1/
8” 20-30 ml/min. 4.2-6.3 cm/sec.
Capillary Column
HP Series 530 (megabore) 3-5 ml/min. 22-38 cm/sec.
139
For Capillary Columns:
140
Comparison of Columns of Different
Dimensions
141
Isothermal vs. Temperature Programming
142
Column Temperature Operation
• Isothermal:
– Oven is maintained at a constant temperature throughout the analysis.
– Stop time is set with the initial time.
– Rate is set to zero.
– Excessive broadening of later eluting peaks.
• Temperature Programmed:
– Used when components have a wide range of boiling points (>100°).
– Reduces analysis time and produces sharper peaks.
– Increased column bleed causing baseline drift.
– Can have Multiple Ramps.
– 6890 can "Fast Ramp" up to 120°/min. at temperatures between 50°C
and 70°C. Even at temperatures above 300C, ramp rates are 35C/min.
143
Peak Efficiency
144
Resolution
RT = 4.59
RT = 5.10
6.0E5 peaks.
• Resolution is measured in terms of
5.0E5
two adjacent peaks which we want
4.0E5
to separate. Generally, the most
Abundance
145
Decreasing Run Time
146
Using Fast GC
toluene
C8 C9 plus aromatics
aromatics
0.32 mm i.d., helium
benzene
0.5탅, 200/1
naphthalenes
147
Using Fast GC: Constant Pressure
70C at 17C-min to 170C, 15psi
C14,15,16 in hexane
FID2 B, (JML\SIG20151.D)
1.059
1.072
1.120
6.065
pA
6.890
Conventional GC
800
700
Parameters (i.e. slow
temperature program, long
column, low column
7.891
600
pressure/flow)
500
6.770
1.177
5.815
0.998
100
1 2 3 4 5 6 7 8 min
FID2 B, (PITTCON\PITTCO12.D)
pA
C14,C15,C16 in hexane
9
21
0.
700
Less than 17 seconds
43
2
0.
600
8
27
0.
500
R=4.3
400
149
Using Fast GC: DB-5 (15 meter x0.25 mm ID, 0.25 um)
210C, 87 psi, 20.6 ml/min, 243 cm/sec
FID2 B, (PITTCON\PITTCO14.D)
pA
72
1000
1
0.
C14,C15,C16 in hexane
0
19
0.
Less than 13 seconds
800
7
21
0.
At higher pressures 600
(and consequently
higher column flows
and linear velocities),
R=5.8
400
good resolution is
maintained and in
fact, improved.
200
8
15
0.
0.2 150
Type of Carrier Gas Effect on Efficiency and
Resolution
OV-101 0.4µ
1.0
25 m x 0.25 mm
He 15 m x 0.25 mm
.8
Glass WCOT
.6 SE - 52
H2
.4 Isothermal, 150º C
.2
10 20 30 40 50 60 70 80 90
R = 1.17 R = 1.37
Average Linear Velocity (cm/sec)
151
Reformate Gasoline
0 5 10 15 min
0 1 2 3 4 5 min
0 1 2 3 4 min
152
Column Selectivity
153
Considerations In Column Selection
• Phase coating
• Column diameter
• Film thickness
• Length
154
Column Polarity
Dispersion Interaction
155
Film Thickness
156
What is a GC Detector?
157
Detector Types
158
Detector Sensitivity Comparison
ELCD (S or N)
ELCD (X)
AED
TCD
FID
µECD
NPD (N)
NPD (P)
FPD (S)
PID
IRD
(SIM) (SCAN)
MSD
160
7890 Flame Ionization Detector
Theory of Operation
• H2 – Air Flame
• Sample is burned in
flame.
• Charged Ions
produced.
• Ions attracted to
collector.
• Collector current
converted to output
via Electrometer.
161
FID Detector Cutaway
Igniter
Insulators Collector
Air
Jet
Column nut
162
FID Response/Selectivity CHO+
CO2 He H2O
N2 CHO+ N2
CHO+
CHO+ He CO2
H2O CHO+
N2
Air N2 He Air
He
N2 He H2
Air FID air supply CH
H2 He 4
CH4 Methane in sample N2 N2
He
CHO+ Methyl cations H2 H2
CH4
H2 FID hydrogen N2 N2
He
He Carrier gas H2 H2
CH4
H2O Water N2 N2
He
N2 Makeup gas H2 H2
CH4
Silicon Halides CO O2
H2O CO2 N2
Perhalogenated HCOH HCOOH
compounds
164
7890 Flame Ionization Detector
• Little or no Response H2O, CO, CO2, N2, O2, CS2, COS, NH3, and all
inert gases. Formaldehyde and heavily
halogenated compounds give little response.
165
7890 Flame Ionization Detector Jets
Adaptable FID Jets
Jet Type Part# Jet Tip ID
0.29 mm
Capillary 19244-80560
0.011 in.
0.47 mm
Packed 18710-20119
0.018 in.
0.79 mm
Paced Wide Bore 18789-80070
0.030 in.
0.47 mm
High Temp G1531-80620
0.018 in.
0.47 mm
High Temp G1531-80620
0.018 in.
166
FID Parameters
167
FID Flow Rates
30 ml/min
Packed columns 10 – 60 ml/min
Carrier (1/8” column)
Gases
(H2, He, N) 1.5 ml/min
Capillary columns 0.5 – 5 ml/min
(0.32 mm column)
Detector
Air 200 – 600 ml/min 450 ml/min
Gases
168
FID Pressure/Flow Relationships
800 80
600 60
400 40
200 20
169
Exploded Parts View of the FID
170
7890 Flame Ionization Detector
171
7890 Flame Ionization Detector EPC Flow
Module
172
FID EPC Module
173
FID EPC Module Replaceable Parts
Valves
“O” Rings
Fittings
Restrictors
Valves
174
Thermal Conductivity Detector (TCD)
175
7890 Thermal Conductivity Detector
Benzene 0.11
Hexane 0.12
Argon 0.12
Methanol 0.13
Nitrogen 0.17
Helium 1.00
Hydrogen 1.28
Flow
Flow
178
7890 TCD - Theory of Operation
The filament
current is set to
keep the filament
at a fixed Constant
Current
temperature above Supply
the block
temperature.
Filament
Column Reference
179
7890 TCD - Theory of Operation
Filament
Column Reference
180
TCD Normal Flow Ratio
20 ml/min
Col + MUG
+ 30 ml/min
Reference Sample
Filament
Filament
10 ml/min Ref Reference
Reading Reading
20 ml/min
Reference 20 ml/min
Col + MUG
30 ml/min 30 ml/min
Reference 20 ml/min 20 ml/min Reference
Column + MUG Column + MUG
10 ml/min
Col + MUG
+ 24 ml/min
Reference Sample
Filament
Filament
14 ml/min Ref Reference
Reading Reading
16 ml/min 10 ml/min
Reference Col + MUG
+
6 ml/min Ref
30 ml/min 30 ml/min
Reference 10 ml/min 10 ml/min Reference
Column + MUG Column + MUG
The low column flow, relative to reference flow, causes the carrier to be diluted with reference
flow. This decrease in concentration reduces the size of the TCD signal.
182
Reference Flow Too Low
8 ml/min
Reference
+
18 ml/min 10 ml/min
Col + Col +
Reference Sample
Filament
Filament
MUG MUG
Reading Reading
10 ml/min 12 ml/min
Reference Col + MUG
+
2 ml/min
Col + MUG
The loss of column flow to the bypass channel does not alter the concentration of sample
in the filament channel. Therefore, no significant change in the sample reading, but the
reference flow is contaminated with sample increasing the size of the reference reading.
183
TCD Parameters
Set polarity to + or -.
184
TCD Flow Rates
Reference 15 – 60 ml/min
Detector
Gases Packed MUG 2 – 3 ml/min
(H2,He)
Capillary MUG 5 – 15ml/min
185
Choosing Reference Flow Rate
Ratio of Ref flow to Column + MUG.
3.5
2.5
1.5
0.5
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
186
7890 TCD EPC Flow Diagram
187
7890 TCD EPC Module
Reference Gas Line.
Switching Valve.
188
7890 TCD Filament Drive - Delta T Sensor
500 ΔT 50C
400
189
TCD Board
TCD Board Connections
Filament
Connections
PRT sensor
Connections
190
TCD Parts Explosion
Detector Cover
Insulation
PRT Sensor
To EPC Module.
Heater/Sensor Assembly.
TCD Cell.
191
Thermal Conductivity Detector Cell
Filament Wires to
Detector Board.
Heater Slot.
192
Electron Capture Detector
193
ECD Theory of Operation
Detector Anode -
Pulse frequency increases
to maintain constant current.
Secondary Emission
of Electrons. Gas Exit.
-e
-e -e
X
N2 -e -e As electronegative
-e
X
-e -e sample enters cell -
-e
X -e free electron population
is depleted.
Ni63
Source.
Makeup Gas -
N2 or ArMe. Column - He, N2 or H2
Carrier.
194
7890 uECD Cell
Vent
Cell
Ni63
Hydrocarbons 1
Ethers,esters 10
196
7890 µECD
197
7890 Electron Capture Detector
µECD Gas Flows
198
7890 µECD EPC Module
Vent
Ni63
199
ChemStation Configuration
200