Godrej & Boyce Mfg. Co. Ltd.

,
CALIBRATION CERTIFICATE Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

FR-QT-21
Certificate No : C-220519-5-1 Date of Issue : 21-May-22
Date of Calibration : 19-May-22 Recom. Due Date : 19-May-23
Customer Details : Receipt No. : C-220519-5
IAI MALAT DIVISION., Challan No : Lab
C/O Elcom Systems Pvt Ltd,SP 86-88, RICO Industrial Area, Date of Receipt : 19-May-22
Bhiwadi,Alwar, Cond. On Receipt : Satisfactory
Rajasthan-301019.
Details of Machine Under Calibration:
Description : Wireless Weighing machine With Indicator Sr. No. : 0311PK2001
Max. Capacity : 700 kg Model : 100406-RFX
Make : Intercomp Mode : Compression
Load cell Sl.No : 0210LB202274 Channel : 1

Details of Force Proving Instrument (Master Load cell )Used For Calibration:
Name Load Cell with Indicator
ID No QSC/F/06
Make/Model/capacity Sensortron / SEN-10KN-C / 10KN
Certificate Number TP/FCM/2021/004
Due Date 3-Nov-22
Uncertainty in % at 95% C.L @k=2 0.11
Traceability Techno Products
Calibrated at Temp 26 °C

Work Instruction No: WI-FC-01,02 IS Std: 1828(Part1):2015 /ISO 7500-1


Environmental Details: Temperature : 26.9 °C Relative Humidity: 57% RH

AUTHORIZED SIGNATORY : P. Muthukumar


DESIGNATION : S M
CALIBRATED BY : Lab Engineer : T. Krishnan
ULR No.: CC255922000012166 N

% DSUnknown
q
1G
1g
0.1 0 0 0.1 9 0 cm
0 J 0 j 4 M []0 d
1i
0g
313 292 m
313 404 325 453 432 529 c
478 561 504 597 504 645 c
504 736 440 760 391 760 c
286 760 271 681 265 626 c
265 625 l
100 625 l
100 828 253 898 381 898 c
451 898 679 878 679 650 c
679 555 628 499 538 435 c
488 399 467 376 467 292 c
313 292 l
h
308 214 170 -164 re
f
0.44 G
1.2 w
1 1 0.4 rg
287 318 m
287 430 299 479 406 555 c
451 587 478 623 478 671 c
478 762 414 786 365 786 c
260 786 245 707 239 652 c
239 651 l
74 651 l
74 854 227 924 355 924 c
425 924 653 904 653 676 c
653 581 602 525 512 461 c
462 425 441 402 441 318 c
287 318 l
h
282 240 170 -164 re
B
Q

Digitally signed by DS M/S GODREJ AND


BOYCE MFG CO LTD 1
Date: 2022.05.21 16:36:31 +05:30
Reason: Calibration Certificate
Location: Chennai

Godrej & Boyce Mfg. Co. Ltd.


Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature) Page 1 of 2
Godrej & Boyce Mfg. Co. Ltd.,
CALIBRATION CERTIFICATE Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

FR-QT-21 ULR No.: CC255922000012166 N


Certificate No : C-220519-5-1 Date of Issue : 21-May-22
Date of Calibration : 19-May-22 Recom. Due Date : 19-May-23
Calibration Results (Mechanical - Force)
Load cell (0210LB202274)
Mode Load Cell Range : 700 kg Least Count Temperature in °C: 26.9
Compession Certified range : 700 kg 0.5 kg Humidity in % RH: 57
Force Expanded
Observed Observed Observed Avg. Test
Proving Force Applied Std In Equvalent Force Uncert-
Test Reading Test Reading in Test Reading Reading Error in kg
Instrum- N in Kg ainty %
in kg(1) kg (2) in kg(3) in -kg
ent Used (±)
686.5 70.0 70.0 70.0 70.0 70.0 0.00 0.30
10 kN Compression Load cell

1372.9 140.0 140.0 140.0 140.0 140.0 0.00 0.30


2059.4 210.0 210.0 210.0 210.0 210.0 0.00 0.30
2745.9 280.0 280.0 280.0 280.0 280.0 0.00 0.30
3432.3 350.0 350.0 350.0 350.0 350.0 0.00 0.30
4118.8 420.0 420.0 420.0 420.0 420.0 0.00 0.30
4805.3 490.0 490.0 490.0 490.0 490.0 0.00 0.30
5491.7 560.0 560.0 560.0 560.5 560.2 0.17 0.31
6178.2 630.0 630.5 630.0 630.5 630.3 0.33 0.40
6864.7 700.0 700.5 700.5 700.5 700.5 0.50 0.41

Load Test Nos 1 2 3 Relative zero error in %


Residual Indication on Force Indicator in kg 0.0 0.0 0.0 0.00
Relative Resolution error in %
Resolution of the force Indicator in kg 0.5
0.01

Remarks:
1. Equipment used for calibration were calibrated & traceable to National & International Standards.
2. This report refers only to particular item (s) submitted for calibration.
3. Godrej Calibration Services is not liable for any change in calibration data & performance specifications on account of malfunctioning of Standards /
Instruments / Equipment covered by this report due to damage caused to it after issuance of this report.
4. The calibration results reported are valid at the time of and under the stated conditions of the measurements.
5. This report shall not be reproduced in full / part without prior permission of Godrej Calibration Services.
6. Satisfactory calibration report in no ways implies that the equipment calibrated is approved by NABL.
7. All precautions have been taken for any error or omission while calibrating the instruments and issuing this certificate. However Godrej Calibration
Services shall not be liable for any loss or liability that may be arise to any party in this regards.
8. Readings given are average of 3 Measurements.
9. The reported Expanded Uncertainty is calculated at 95.45 % Confidence level with coverage factor k=2.
10.This report does not cover under the scope of Accreditation.

Godrej & Boyce Mfg. Co. Ltd.


Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature) Page 2 of 2