Professional Documents
Culture Documents
Health Technical Memorandum 01-06:: Decontamination of Flexible Endoscopes
Health Technical Memorandum 01-06:: Decontamination of Flexible Endoscopes
01-06: Decontamination of
flexible endoscopes
Part E: Testing methods
March 2016
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
ii
© Crown copyright 2016
You may re-use this information (not including logos) free of charge in any format or me-
dium, under the terms of the Open Government Licence. To view this licence, visit www.
nationalarchives. gov.uk/doc/open-government-licence/ or write to the Information Policy
Team, The National Archives, Kew, London TW9 4DU, or email: psi@nationalarchives.gsi.
gov.uk.
This document is available from our website at www.gov.uk/government/collections/
health-building-notes-core-elements
iii
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
Preface
iv
Best Practice is additional to EQR. Best The HTM 01 suite is listed below.
Practice as defined in this guidance covers
non-mandatory policies and procedures that • HTM 01-01: Management and
aim to further minimise risks to patients; deliver decontamination of surgical instruments
better patient outcomes; promote and (medical devices) used in acute care
encourage innovation and choice; and achieve • HTM 01-04: Decontamination of linen for
cost efficiencies. health and social care
Best Practice should be considered when • HTM 01-05: Decontamination in primary
developing local policies and procedures based care dental practices
on the risk of surgical procedures and available
• HTM 01-06: Decontamination of flexible
evidence. Best Practice encompasses
endoscopes.
guidance on the whole of the decontamination
cycle, including, for example, improved
instrument management, where there is Note
evidence that these procedures will contribute
to improved clinical outcomes. This guidance remains a work in progress
which will be updated as additional evidence
becomes available; each iteration of the
guidance is designed to help to
incrementally reduce the risk of cross-
infection.
v
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
Abbreviations
vi
Executive summary
vii
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
b. use technologies with the best available • New process challenge device efficacy
sensitivity, consistent measurement test introduced.
standards and quantifiable results to • New advice on tests for residual protein
measure effective control of residual detection introduced.
protein levels;
• Previous tests for residual soil using
c. use trend analysis as a tool for self- ninhydrin and alternative methods now
improvement to demonstrate decreasing removed as they are no longer
protein levels over time both on the applicable.
outside of the endoscope and the
lumens using available testing • New test given in Chapter 6 for detection
technologies. of Pseudomonas aeruginosa in final
rinse-water.
• All references updated.
viii
Acknowledgements
ix
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
Contents
Preface������������������������������������������������������������������������������������������������������������������������������������� iv
Abbreviations��������������������������������������������������������������������������������������������������������������������������� vi
Executive summary���������������������������������������������������������������������������������������������������������������� vii
Acknowledgements����������������������������������������������������������������������������������������������������������������� ix
1 Introduction��������������������������������������������������������������������������������������������������������������������������1
2 Terminology��������������������������������������������������������������������������������������������������������������������������2
3 Automatic control test����������������������������������������������������������������������������������������������������������3
4 Drainage��������������������������������������������������������������������������������������������������������������������������������4
5 Venting system���������������������������������������������������������������������������������������������������������������������7
6 Water system������������������������������������������������������������������������������������������������������������������������8
7 Doors and door interlocks�������������������������������������������������������������������������������������������������� 18
8 Chemical dosing�����������������������������������������������������������������������������������������������������������������20
9 Instrumentation fitted to an EWD�������������������������������������������������������������������������������������� 21
10 Load carriers��������������������������������������������������������������������������������������������������������������������� 22
11 Thermometric tests����������������������������������������������������������������������������������������������������������� 23
12 Load dryness�������������������������������������������������������������������������������������������������������������������� 27
13 Chemical additive residual levels�������������������������������������������������������������������������������������28
14 Air quality��������������������������������������������������������������������������������������������������������������������������29
15 Sound pressure test���������������������������������������������������������������������������������������������������������30
16 Cleaning efficacy tests����������������������������������������������������������������������������������������������������� 31
17 Leak and patency testing�������������������������������������������������������������������������������������������������35
18 EWD self-disinfection test������������������������������������������������������������������������������������������������39
19 Disinfectant concentration test���������������������������������������������������������������������������������������� 42
20 Microbiological and temperature tests of disinfection efficacy��������������������������������������43
References������������������������������������������������������������������������������������������������������������������������������45
x
1 Introduction
1 Introduction
1.1 This volume discusses principles and be processed will not withstand the high
methods that are used in the tests described in temperatures required for thermal disinfection. It
this HTM and those detailed in BS EN ISO is necessary to ensure that the disinfectant is
15883 Parts 1 and 4. removed from the endoscope before it is used
on a patient; this is achieved by a post-
1.2 EWDs for flexible thermo-labile endoscopes disinfection rinsing stage. Some EWD process
are machines of Type 1 or 2: cycles are available that add disinfectant to the
final rinse-water. The level of disinfectant at this
• Type 1 EWDs have a single door or lid
stage should be shown to be non¬toxic to
entry.
patients. The microbiological control of this
• Type 2 EWDs have two doors and a stage is of critical importance to the microbial
pass-through facility. status of the reprocessed endoscope. A
number of additional tests are required to
1.3 (Rigid endoscopes able to tolerate terminal ensure that this aspect of the process is
steam sterilization are considered under WDs properly controlled.
for surgical instruments.)
1.5 Periodic storage cabinet tests for flexible
1.4 EWDs are characterised by a chemical endoscopes are given in Chapter 6 of HTM 01-
disinfection stage because the endoscopes to 06 Part D – ‘Validation and verification’.
1
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
2 Terminology
2
3 Automatic control test
3.1 The automatic control test is designed to 3.5 The test should be considered satisfactory
show that the operating cycle functions if the following requirements are met:
correctly as shown by the values of the cycle
variables indicated and recorded by the a. A visual display indicating “cycle
instruments fitted to the EWD. complete” occurs.
b. During the whole of the operational
3.2 In the absence of an independent cycle, the values of the cycle variables
monitoring system (IMS), this test is carried out as indicated by the instruments on the
daily and is the main test for ensuring that the EWD or shown on the batch process
EWD continues to function correctly within the record are within the limits established
validated parameters. The IMS records should as giving satisfactory results either by
be checked in conjunction with the test. the manufacturer or during validation.
c. If elevated temperatures are used during
Test procedure the routine cycle, the time for which the
3.3 Place a test device or a routine endoscope temperature is maintained is not less
in the chamber. Select the routine operating than that established during validation.
cycle and start the cycle.
d. The door cannot be opened until the
3.4 Ensure that a traceability record is made of cycle is complete.
the test. Note chamber temperatures (if e. The person conducting the test does
appropriate), leak test pressure and time at all not observe any mechanical or other
significant points of the operating cycle, for anomaly.
example the beginning and ending of each
stage or sub-stage.
3
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
4 Drainage
4.1 It is important to vent the EWD externally to trap or a large ballvalve is used to obstruct the
prevent any hazardous fumes entering the discharge from the drain.
workspace. This applies both to the drains that
should be sealed within the EWD work area Method for EWDs with sealed doors
and to the EWD cabinet. There is no test to
determine if any gas contents of the EWD 4.6 Introduce the metal sphere (or equivalent
chamber or drain enter the workspace. device) to block the drain, or fit a large ballvalve
Therefore if a potentially hazardous chemical is and close the valve.
used as the disinfectant, such as peracetic
acid, gas measurements should be taken 4.7 Close the door and start the operating
around the EWD during operation to determine cycle. On completion of the operating cycle,
whether there are any leaks in the system. attempt to open the EWD door using the
normal release procedure.
Blocked-drain protection 4.8 If the door opens and the water level is
below the door seal, close the door and start
4.2 If the drain from the EWD chamber another operating cycle.
becomes blocked, continued operation of the
EWD would allow water and suspended soil to 4.9 Repeat the operating cycle as many times
be discharged onto the floor either during the as necessary for either the water level at the
operating cycle or, for cabinet-type EWDs with end of the cycle to be above the level of the
sealed door(s), by sudden discharge when the door seal or a fault to be indicated.
door is opened at the end of the cycle.
4.3 The purpose of blocked drain protection is Method for EWDs without sealed doors
to prevent unacceptable spillage and minimise 4.10 Introduce the metal sphere (or equivalent
the risk of cross-infection. device) to block the drain or ballvalve adjusted
to the closed position.
4.4 In the test, the drain is deliberately blocked,
and successive operating cycles are run until 4.11 Start the operating cycle. On completion of
the water level is above the level of the door the operating cycle, if no water has been spilled
seal. The test is intended for use both as a type from the EWD and no fault has been indicated,
test (and as such is a requirement of BS EN start another operating cycle.
ISO 15883-4) and as an operational or
installation test. 4.12 Repeat the operating cycle as many times
as necessary for either water to be spilled from
Equipment and materials the EWD or a fault to be indicated.
4
4 Drainage
Results for EWDs with sealed doors Efficacy of discharge through the
4.13 A fault should be indicated when, or trap
before, the water level in the chamber reaches 4.17 The test is intended to verify that the
the level of the door seal. The door should not operating cycle is effective in purging the trap of
be capable of being opened by the normal all waste and soil.
release procedure; a tool, key or code should
be required.
Equipment and materials
Results for EWDs without sealed doors 4.18 The following equipment and materials are
necessary:
4.14 A fault should be indicated before water is
spilled from the EWD. The operating cycle • test soil appropriate to the type of EWD
should be stopped preventing further inflow of being tested (see guidance given in
water. A tool, key or code should be required to ‘Performance qualification tests’ in the
restart the EWD. ‘Validation and verification’ volume);
• sampling tube of sufficient length to
Note reach the water trap in the drain of the
This test might not be required if the design EWD and a sampling pump (for example,
of the EWD prevents the water level within a pipette pump or syringe).
the chamber or bowl reaching the level of the
opening. Method
4.19 The test may be carried out as part of the
cleaning efficacy test during operational testing.
Free draining of tanks, chamber,
load carriers and pipework 4.20 On completion of an operating cycle to
test the cleaning efficacy by processing a full
4.15 Residual water that does not drain from
load contaminated with an appropriate test soil,
the internal pipework of the EWD may provide
place the sampling tube into the water trap and
an environment for microbial growth; these
remove a sample for examination.
microorganisms may recontaminate the
disinfected load. 4.21 Examine the water sample from the trap
for residual test soil using the detection method
4.16 The following checks should be carried
appropriate to the test soil.
out during type-testing, works testing and
installation testing to verify that – as designed,
built and installed – the EWD will effectively Results
discharge all the water from the system: 4.22 The water in the trap should be free from
a. free draining of chamber and load residual soil at the same level of detection as
carriers – test by visual observation at that specified for the load items.
end of the cycle;
Residual soil in the trap may present an
b. free draining of tanks – test by visual infection or recontamination hazard.
observation on draining the tanks;
c. pipework flow to discharge point – test
by visual observation including use, Estimation of dead volume of
when necessary, of a spirit level. pipework
4.23 Residual water that does not drain from
the internal pipework of the EWD may provide
5
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
an environment for microbial growth; these reassembly or purging with dry compressed air
microorganisms may then recontaminate the for not less than 30 minutes) is flushed with a
disinfected load. known volume of water (simulating the flow that
would occur in normal use).
4.24 The test is intended primarily as a type
test but may also be of value, if possible, as an 4.28 The volume of water flushed through the
operational test or when investigating microbial system should be twice that determined as the
contamination occurring in an EWD. volume used per operating cycle given in the
manufacturer’s data sheet. The volume of water
4.25 The test should only be carried out after discharged is measured, and the dead volume
completion of the checks for free drainage (see (estimated as the volume retained) is calculated
paragraph 4.15, ‘Free draining of tanks, from the difference between the two values.
chamber, load carriers and pipework’) have
been satisfactorily completed. 4.29 When the EWD has two or more pipework
systems that are entirely separate (for example,
Equipment for flushing water, wash-water, rinse-water, final
rinse-water), each system may be tested
4.26 Volumetric measuring vessels of separately.
appropriate size are necessary.
Results
Method
4.30 The volume of retained water should be
4.27 The pipework of the EWD that is known to less than 1% of the volume of water used.
be dry (either following disassembly and
6
5 Venting system
5 Venting system
Load contamination from ductwork 5.6 If the mirror shows any signs of misting or
the presence of droplets, its position should be
5.1 The evolution of water vapour from the noted.
chamber during the washing stage, disinfection
stage and drying stage may result in
Result
condensation occurring in the ductwork and in
the condenser (if fitted). The ducting is 5.7 There should be no misting of the mirror or
commonly arranged to allow this condensate to droplets present. Misting or droplets on the
drain back into the chamber or drain. This mirror indicates the discharge from around the
condensate is likely to become contaminated door or lid.
and there is a risk that it could come in contact
with the load. The ductwork should be checked
at installation to confirm that any condensate
Chemical vapour emission
from the ductwork cannot drain into the EWD 5.8 EWDs use chemical additives for which
chamber and onto the load. there may be specified exposure limits (usually
disinfectants) under the COSHH Regulations. It
must be ensured that emissions from the EWD
Droplet emissions do not cause personal exposure to exceed the
5.2 Emissions of water droplets present a risk legal limits.
of transmitting infection from contaminated load
items. The test is designed to be carried out as Method
an operational test.
5.9 The method of sampling for airborne
emissions and the method of analysis or
Equipment and materials detection will be specific to the chemical
5.3 The following equipment is necessary: cold additive(s) being used. Advice should be sought
hand mirror. from the EWD manufacturer, the supplier of the
chemical additive(s) and/or the Health & Safety
Method Executive (HSE) in order to determine an
appropriate test method.
5.4 The EWD should be operated with the
chamber empty except for any chamber
Results
furniture (for example, load carrier). Any
chemical additives used should be replaced 5.10 Emissions from the EWD during normal
with water. operation and maintenance, including when
opening the EWD at the end of the cycle or
5.5 Close the door of the EWD. Initiate an when changing or refilling chemical additive
operating cycle. The cold mirror is moved reservoirs, should not expose personnel to
around the edge of the door or lid at a distance concentrations in excess of the legal maxima.
of 50–100 mm from the door or lid.
7
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
6 Water system
8
6 Water system
9
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
10
6 Water system
paper should be examined and a description of in water of high purity may give
the retained material reported. The filtrate unstable or unreliable readings.
should then be visually examined for deposits.
(ii) Only those pH meters specifically
6.32 The appearance should be reported in designed for the measurement of low
terms of both colour and the intensity of any ionic strength solutions should be
colour. If the sample is coloured, it should be used for determining the pH of
examined carefully to see whether evidence of deionised or RO water.
colloidal material is present. b. Colour disc comparator:
6.33 To determine whether there is a problem (i) Colorimetric tests for pH are suitable
with water deposits, draw off approximately for high purity, low conductivity, and
300–500 mL of water into a clean glass beaker water samples of the type to be
or jar. Allow the water to settle and any air tested.
bubbles to disperse. Then examine the sample
against a white sheet with a black line across (ii) Since colorimetric methods are being
the centre under a good light. The water should used for other field tests, this may be
appear crystal clear with a distinct line between the more appropriate method.
the black and white areas. The presence of fine (iii) The accuracy is limited and
deposits will be seen as slight opalescence and discrimination may not be better than
a slightly indistinct edge to the black to white 0.2 pH. This is, however, quite
areas. suitable for field tests.
Results
(iv) Colour, turbidity or strong oxidants in
the sample all interfere with the test.
6.34 All the samples tested should be clear,
bright and colourless. (v) A narrow range indicator (or two for
use on successive samples) should
be chosen to cover the required pH
pH
range from 4 to 8. Manufacturers of
6.35 Two suitable methods for on-site colorimeters usually provide
measurement of pH are available; the colour indicators to cover a range from 2 or
disc comparator and portable pH meter. If 3. Wide range indicators should not
colour comparators are used, the operator be used because of their poor
should undertake a colour blindness test to discrimination.
confirm they are able to read the results.
Method
Equipment
6.37 The test kit should be operated in
6.36 The following equipment is necessary: accordance with the manufacturer’s
instructions. Particular attention should be paid
a. pH meter:
to using accurate volumes of both sample and
(i) There are several commercially reagent, and monitoring both temperature and
available small, portable pH meters. reaction time.
A number of these include built-in
temperature compensation. Although 6.38 The colour of the reacted sample is
they provide suitable accuracy for matched against the calibrated colour disc
most general applications, they are viewed through a blank sample. The value in
not well suited to measurement of pH units is read directly from the disc.
pH in solutions of low ionic strength.
Their use for the determination of pH
11
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
6.39 The calibration should be verified using at 25ºC and traceable to national standard
standard buffer solutions made up in advance reference materials.
and kept in capped bottles until required. The
buffer solutions should be chosen to have a pH 6.47 After calibration, the sample cup, or
in the midpoint of the range of calibrated colour immersion probe, should be thoroughly rinsed
discs to be used in the determination. with pure water.
12
6 Water system
used to measure the potential between the ISE determined using Patton and Reeder’s indicator
and a suitable reference electrode. The as a complexometric indicator.
measured potential is proportional to the
logarithm of the concentration of the ion(s) in 6.60 The commercially available kits often use
solution. novel titration methods instead of burettes. The
test reagents are specific to each kit. The
6.54 The optimum working pH range is 4–9 manufacturer’s instructions should be followed.
and the ionic strength of the sample should be
adjusted for ionic strength. An adjustment Range
buffer consisting of 4 M potassium chloride
solution is often used. Phosphate buffers 6.61 Determinations within the range 5–400
should not be used, since the calcium activity mg/L can be made. The method is not
will be lowered by complexation or precipitation. applicable to purified water or condensate from
clean or pure steam, which should have
6.55 The electrodes are free from any major calcium concentrations well below the range for
interference except zinc ions. They are however accurate determinations.
poisoned by a number of biological fluids.
Results
6.56 The calcium electrode requires a single
junction reference electrode. Calibration is 6.62 The hardness expressed as mg/L CaCO3
made against two or more standard solutions. should be within the following parameters:
These are commercially available. • <50 mg/L for final rinse-water and water
used to dilute chemicals;
Range
• 50–200 mg/L for water used in the
6.57 Calcium-selective electrodes should have rinsing stages;
a Nernstian response for concentrations from 1
M down to about 5 x 10–6 M and a selectivity • >200 mg/L not suitable for use in an
ratio of better than 2000 against magnesium. EWD (if this is the only water grade
This range is suitable for analysis of softened available, pre-softening is essential).
water and purified water (RO or deionised
water). Total viable count
6.63 Endoscopes should be rinsed after the
Method b: titrimetric method disinfection stage to remove any residual
6.58 Commercially available kits for the chemical toxicity. The rinse-water should be
titrimetric determination of both total hardness free from extraneous material, both inorganic
and calcium hardness are available. They are and organic including microorganisms, which
based on the same reaction in which divalent could compromise the patient. Routine total
cations are complexed with the disodium salt of viable counts (TVCs) should be made on the
ethylenediaminetetraacetic acid (EDTA). When final rinse-water. If this water is collected from
the reaction is carried out, at pH 10–11, with the EWD bowl/chamber, it will have passed
Eriochrome Black as the complexometric through the water treatment system and internal
indicator, all the calcium and magnesium ions pipework of the machine. The TVC results will
are chelated by the EDTA, and the absence of give an indication of the water treatment system
free calcium and magnesium ions causes a performance and microbial colonisation of the
colour change in the indicator. EWD pipework. If the TVC is high, additional
samples can be taken to determine the
6.59 At pH values above 12, magnesium ions problem source.
are precipitated as the hydroxide and do not
react with the EDTA. Calcium hardness can be
13
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
6.64 The following test should be carried out as stopped during the final rinse with the bowl full
an installation or operational test and as a of water.
weekly periodic test for EWDs. It should be
carried out by a microbiologist or those trained 6.69 During the final-rinse stage of the EWD
in aseptic techniques. cycle, stop the machine and the open the lid or
door. With gloved hand, remove the sterile
sampling tube from its wrap and place the
Equipment and materials weighted end in the EWD bowl final rinse-water.
6.65 The following equipment and materials are
necessary: 6.70 With the syringe pushed onto the open
end of the sampling tube, draw water down the
a. sterile single-use containers with a tube with the valve open. When the tube is full,
volume of at least 250 mL; close the valve and remove the syringe.
b. sterile filter membranes (47 mm
6.71 Allow water to flow from the sampling tube
diameter, <0.45 µm pore size);
to waste, then open the sampling bottle and
c. suction filtration apparatus; direct the water sample into the bottle. When
the sample bottle is full, cap and label the
d. incubator at 30±2ºC; container with details of the sampling point and
e. R2A agar (see BS EN ISO 15883-1, the time and date the sample was collected.
Annex D) (alternatively yeast extract agar
(2) Using a sample bottle included in the EWD
(YEA) or tryptone soya agar (TSA) may
load
be used;
f. f70% alcohol and non-woven wipes. 6.72 This is an alternative method of sample
collection with two outlets fitted to the lid. One
connection is fitted to an EWD outlet via a
Method: sample collection
length of plastic tube. The other connection is
6.66 Samples of final rinse-water should be fitted with a short length of tube and left open.
taken from the EWD using aseptic techniques
following the manufacturer’s instructions. If 6.73 During the EWD cycle, the bottle is
these are not available, the advice of the AE(D) irrigated by the EWD process liquids, with the
should be sought. One of the following last liquid being the final rinse-water. On
methods may apply. completion of the cycle, or if aborted at the end
of the final rinse stage, the bottle is removed
(1) From the EWD bowl – recommended from the EWD and the connections removed
from the bottle and the openings sealed. The
6.67 The following equipment will be required: full bottle of final rinse-water is labelled with
• A sampling tube made up of a length of details of the sampling point and the time and
heat-resistant tube from the EWD bowl to date the sample was collected and sent to the
a sampling bottle held below the bowl microbiology laboratory.
water level. The tube should be fitted with
6.74 The EWD manufacturer may describe
a normally closed valve and a weight
other methods of final rinse-water collection.
attached to one end. The tube and
fittings should be sterilized before use.
• A water sampling container.
14
6 Water system
Over 100 Risk assessment required to consider taking EWD out of service until
water quality improved Red
Notes:
Microbiological results from weekly tests should be plotted on a graph to give a trend. This will allow the “normal” and “unusual” results
to be distinguished for a particular situation. Investigation of unusual or unsatisfactory results can then be undertaken if results demand
(for example, if routine results are below 10 cfu/100 mL, occasionally some of the results may be above 10 cfu/100 mL).
If a bacterial count above 10 cfu/100 mL is obtained from test water, identification of the species is advised. If a significant proportion of
the microbes appear the same species from their colonial morphology, carry out an oxidase test to presumptively identify Pseudomonas
spp. Then if the test is positive, further investigations are required to determine whether Pseudomonas aeruginosa is present.
Adapted from: Willis, C. (2006). “Bacteria-free endoscopy rinse water – a realistic aim?” Epidemiology and Infection. Vol. 134 No. 2, pp.
279–284.
15
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
6.83 The above levels are suggested values • 70% alcohol and non-woven wipes.
(see the ‘Water quality and water treatment’
section in the ‘Design and installation’ volume); Method: sample collection
if an endoscope requires a higher level of
6.86 The sample is best collected from the
decontamination, a risk assessment will be
EWD chamber or bowl.
needed.
16
6 Water system
17
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
18
7 Doors and door interlocks
19
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
8 Chemical dosing
20
9 Instrumentation fitted to an EWD
21
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
10 Load carriers
10.1 Load carriers come in a variety of forms their connection to endoscopes should be
including cages, carriages and baskets. Their verified by visual observation.
correct functioning is essential to the successful
outcome of an EWD operating cycle. It is 10.3 Load carriers with rotary spray arms
important that they cannot easily be misaligned, should be checked to ensure that the spray
that they function correctly and that, when arms are free to rotate, both when the load
applicable, they make good connection with carrier is empty and when fully loaded.
service supply points in the chamber and with
load items (when necessary). 10.4 Load carriers that have lumen connections
as part of their structure require special
attention. The connection from the body of the
Method
EWD to the load carrier needs to be good else
10.2 The alignment of load carriers, their fluids and air may leak from the connections,
connection to water, air or chemical additive invalidating system checks within the EWD.
supply in the chamber (when applicable) and
22
11 Thermometric tests
11 Thermometric tests
11.1 Thermometric tests are required for both EWD's process control sensor and a probe
thermal disinfection processes and chemical positioned in the centre of the chamber free
disinfection processes if elevated temperature space.
forms part of the process cycle. If during the
self-disinfect or routine cycles elevated 11.6 The temperature attained should be
temperatures form no part in the process, measured throughout three self-disinfect
these tests do not apply. cycles; they should be at least 60 minutes since
the machine was last used (a “cold start”).
11.2 For self-disinfection, the time/temperature
of an A0 of 600 (for example, 80°C for 10 11.7 The EWD should be operated empty
minutes) or higher is suitable (see BS EN ISO except for chamber furniture (for example, load
15883-1, Annex B, for details of the A0 concept carriers).
and time/temperature relationships). The
disinfection temperature is measured at the 11.8 Multi-chamber EWDs may be tested with
coolest surface to be disinfected. each chamber tested consecutively or
concurrently. In the latter case eight sensors will
11.3 The reference test loads to be used for be required for each chamber.
thermometric tests are given in paragraph
16.19, ‘Test loads’. Results
11.9 The results should be the following:
Chamber wall temperature for the a. The temperatures recorded on the
self-disinfection cycle (if applicable) surface of the chamber should be within
0–5°C of the cleaning/disinfection
Equipment and materials temperature throughout the holding
11.4 A temperature recorder complying with the period for the cleaning/disinfection
requirements specified in BS EN ISO 15883-1, stage.
clause 6.2.1, and having no fewer than 10
b. The temperatures recorded on the
sensors is necessary.
surface of the chamber should be within
±5°C of the set temperature for the
Method relevant stage throughout the holding
11.5 Thermocouples should be located one in period for each of the other stages.
each corner of the chamber, one in the centre c. The temperature indicated/recorded by
of the two side walls, one in the centre of the the EWD instruments should be within
roof of the chamber and one adjacent to the ±2°C of that recorded by the test
temperature sensor used as the reference instrument from the sensor adjacent to
sensor for chamber temperature. In addition, a the reference sensor throughout the
probe should be positioned adjacent to the holding period for the disinfection stage.
23
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
d. The temperature profile obtained for the a. The temperatures recorded on the
operating cycle should be consistent surface of the load carrier should be
within ±2°C for three test cycles. within the range 0–5°C of the
disinfection temperature throughout the
holding period for the disinfection stage.
Load carrier temperature during
self-decontamination (if temperature b. The temperatures recorded on the
surface of the load carrier should be
used) within ±5°C of the set temperature for
the relevant stage throughout the
Equipment and materials holding period for each of the other
11.10 A temperature recorder complying with stages.
the requirements specified in BS EN ISO
c. The temperature indicated/recorded by
15883-1, clause 6.2.1, and having no fewer than
the EWD instruments should be within
four sensors is necessary.
±2°C of that recorded by the test
instrument from the sensor adjacent to
Method the reference sensor throughout the
11.11 Temperature sensors should be located holding period for the disinfection stage.
at two diagonally opposite corners of the load d. The temperature profile obtained for the
carrier, in the approximate geometric centre of operating cycle should be consistent
the load carrier and adjacent to the temperature within ±2°C of the reference temperature
sensor used as the reference sensor for on all test points for all three test cycles.
chamber temperature.
11.17 Three independent data-loggers and a
11.12 The temperature attained should be temperature recorder having at least one
measured throughout three operating cycles, sensor may be used as an alternative.
which should be at least 60 minutes since the
machine was last used. The EWD should be
operated empty except for the endoscope Over-temperature cut-out (when
carrier. temperature is used for self-
11.13 The load carrier should be replaced disinfection)
between cycles with a load carrier at ambient 11.18 The EWD is fitted with an over-
temperature. temperature cut-out to ensure that, in the event
of the automatic control failing to control the
11.14 Multi-chamber EWDs may be tested with
temperature in the EWD, the temperature will
each chamber tested consecutively using
not rise to a level that would damage the EWD.
independent data-loggers to record the
temperature of the load carrier. A temperature 11.19 This test can also be used if temperature
recorder with fixed sensors may be used to forms part of either the wash or disinfection
record the temperature adjacent to the stage of an operating cycle. Over-temperature
reference sensor. during these stages is likely to damage
endoscopes being processed.
11.15 This test may be run simultaneously with
the chamber wall temperature test.
Equipment and materials
Results 11.20 A temperature recorder complying with
the requirements specified in BS EN 15883-1,
11.16 The results should be the following.
clause 6.2.1, and having no fewer than four
sensors is necessary.
24
11 Thermometric tests
25
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
26
12 Load dryness
12 Load dryness
12.1 The presence of residual water on cleaned connect to an air supply with a pressure of
and disinfected items may interfere with the 105–120 kPa. Place the distal end of the lumen
correct functioning of the item, promote 50–100 mm above a sheet of coloured crepe
recontamination and microbial growth or paper.
prevent satisfactory subsequent disinfection.
12.4 Pass air down the test lumen and examine
12.2 The ability of the EWD to externally dry an the paper for dampness denoted by dark spots
endoscope may be evaluated visually, but the on the paper.
internal lumens and valves are hidden from view
and require an additional test. If the EWD cycle 12.5 Tests on endoscopes should be made on
claims to include a drying stage, this test all lumens, with the the air supply connected
should be use to determine if the EWD lumens from the control valve and the umbilical end.
have been adequately blown.
Results
Method 12.6 Report any water droplets that are
12.3 Process the test device through a routine discharged from the test device. There should
cycle. On completion remove the load and be no water droplets discharge.
27
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
Method
13.2 The efficacy of the rinse process should
be tested using the normal dose of the
chemical additive on a normal operating cycle
with a surrogate device to simulate an
endoscope (see the ‘Performance qualification
tests’ section in the ‘Validation and verification’
volume). Analysis, by the method
recommended by the manufacturer, should be
carried out on the final rinse-water and on the
surrogate device.
28
14 Air quality
14 Air quality
29
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
30
16 Cleaning efficacy tests
31
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
one month or follow the manufacturer's be aborted, if possible. The test load should be
recommendations. removed and examined for the presence of
residual soil. This test should be run in
Application and use duplicate.
16.7 If the soil has been stored, allow it to 16.12 The manufacturer should establish worst-
equilibrate to room temperature before use. case conditions of temperature, detergent
concentration, surrogate device configuration,
16.8 The test should be carried out using water based on illustrations in BS EN ISO 15883-4
of a quality suitable for the system requirements Annex F, and water pressure/flow rate for use
using a detergent recommended by the EWD during testing.
manufacturer. The EWD manufacturer should
provide guidance on the minimum water quality 16.13 By analysing the fraction of soil removed
suitable for the detergent recommended. All during the cleaning process when operated for
tests should be run in triplicate and for each various time periods shorter than those that will
series of tests each of the three tests should normally be used, a quantitative comparison of
meet or exceed the minimum acceptance cleaning efficacy can be made.
criteria.
16.14 The recommended minimum operating
Method of application conditions given by the manufacturer should be
based on these data, which should be made
16.9 Apply soil to the inner surface of the test available to the User as part of the type-test
pieces by injecting the soil into the tubes of the data sheets.
surrogate device (approximately 5 mL into the 1
mm inner-diameter tube and 20 mL into each of
the 2 mm inner-diameter tubes). Lay the tubes Operational tests
on a horizontal surface and roll them to 16.15 During operational tests of cleaning
distribute the soil over the inner surface. efficacy with test soils, the disinfection stage
and drying stage should be disabled, unless it
16.10 Hold them vertically to allow excess soil
can be demonstrated that inclusion make little
to drain away. Apply a small amount of air to
difference to the results.
each soiled tube from a syringe full of air to
blow out any residual soil. Then apply an even
coat of soil to the outer surface using the Test soils
paintbrush. Allow excess soil to drain from the 16.16 The choice of test soil to be used should
items and allow them to dry at room be that specified in the type tests or that
temperature (15–25°C) for not less than 30 described in paragraph 16.2, which represents
minutes and not more than two hours. To dry a more realistic tissue residue.
the inner surface of the tubes, connect each
tube to a small air pump and pass air down the 16.17 Drying the deposited test soil is
tubes for the drying time. important, as it has an influence on the difficulty
of cleaning. At least two hours plus or minus 15
Method of use minutes’ drying time is required, including
lumens. The soil should be spread on such that
16.11 Connect each surrogate tube to the EWD it appears dry within the specified drying time.
outlets with suitable connectors: one of the 2 After soiling lumen may be dried, by passing
mm inner-diameter tubes connected to the down low pressure dry air for the required time.
biopsy lumen; and one of the 1 mm inner-
diameter tubes connected to the bridge lumen, 16.18 These performance criteria include the
if used. Run a routine EWD cycle. After following:
completion of the wash cycle, the cycle should
32
17 Leak and patency testing
a. The test soil formulation should be there are no standard designs for these test
validated as being of equal or greater pieces, but a design that simulates the internal
difficulty to remove than the naturally structure of a flexible endoscope would be
occurring soils, which it is intended to acceptable.
simulate.
b. The method of application to give a level Procedure for chamber walls and load carriers
of soiling which is reproducible within 16.22 The chamber walls and load carrier
specified limits should be validated and should be contaminated with the test soil in
documented. accordance with the instructions for the test
soil.
c. The method of detection of residual soil
should be quantitative; visual 16.23 The load surfaces are dried in air for 30–
assessment alone should not be 120 minutes, including lumens.
regarded as satisfactory.
d. The limit of sensitivity of the detection 16.24 A normal operating cycle should be run.
method employed should be 30 µg of
16.25 After completion of the wash cycle and
protein or below per test swab.
before the disinfection stage, the cycle should
e. For tests on the chamber walls and load be aborted. The chamber walls and load carrier
carrier and lumen, the detection method should be examined for the presence of
should include a validated method for residual soil by the method prescribed for the
quantitative detection of the residual soil particular test soil.
in situ or a validated method of removal
for example, by swabbing, for 16.26 It is important that the specified
subsequent evaluation. quantities are used.
16.20 A surrogate device for investigation of 16.28 This test should be run only after
cleaning and disinfection may be constructed satisfactory completion of the test for the
from two 1.5 m lengths of polytetrafluorethylene efficacy of soil removal from chamber walls and
(PTFE) tube (2 mm inner diameter) and one 1.5 load carriers. The test load should be
m length of PTFE tube (1 mm inner diameter). contaminated with the test soil in accordance
These should be bound together at intervals of with the instructions for the test soil.
approximately 15 cm. For EWDs with more than
three lumens, two sets of three surrogate 16.29 A normal operating cycle for the load
devices may be required so that each of the type under test should be run.
machine’s lumens can be tested. Advice from
16.30 After completion of the wash cycle and
the EWD manufacture may be required to
before the disinfection stage, the cycle should
select suitable test tube diameters.
be aborted. The test load, lumen, chamber
16.21 As flexible endoscopes with irrigation walls and load carrier should be examined for
lumens include valve systems to control air and the presence of residual soil.
fluids, these need to be simulated in a test
device. Examples of such devices are illustrated
in BS EN ISO 15883-4 Annex F. At present
33
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
34
17 Leak and patency testing
17.1 The leak test is carried out to determine if 17.3 This consists of:
liquid can gain access to the inner cavity of an • Length of flexible tube approximately 1m
endoscope, thereby causing extensive damage. terminated at one end to fit onto the EWD
The leak test is carried out manually before the leak test connection and at the other end
endoscope is cleaned and again by the EWD, terminated with a flow valve.
either at the start of the process cycle or
throughout the cycle. It is important that the air • To construct a test device, the internal
used to pressurise an endoscope does not volume of the largest endoscope in use is
exceed the limit set by the endoscope needed; this may be obtained from the
manufacturer, normally about 250 mbar. manufacturer. The internal volume of the
tube is calculated and additional volume,
17.2 The test described is based on the use of to make up the volume equal to that of an
a test device. This device needs to represent endoscope, can be supplied by an
the inner volume of the largest endoscopes in additional pressure tight container as
use. illustrated in Figure 1.
• A convenient way of adding additional
volume to a tube is to cut the tube in half
and fit a suitable T-piece. To the T-piece
side arm connect a plastic bottle that is
Flow control
Connector to EWD
Endoscope internal
leak test port
volume – TV
CV
Container and tube
internal volume –
CV
Pressure transducer
35
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
approximately the correct internal volume that an EWD fail condition has occurred and a
required. A bottle of internal volume 100– fault has been indicated.
200 mL may be suitable.
17.7 Pass condition – adjust the flow control
• Pressure transducer able to read ±1 mbar valve to give a leak rate 80% less than the fail
(±0.1 kPa) over the range of the EWD value given by the manufacturer. Connect the
operating pressure. test device to the EWD and run a leak test. The
time for the test device pressure to fall to value
Procedure P3 should be greater than time T3, that is, low
or no leak. From the readings confirm that a
17.4 Verify the calibration of the pressure
pass condition has occurred on the EWD and
sensor.
no fault has been indicated.
17.5 Connect the test piece (see Figure 1) to the
17.8 The leak test characteristics are illustrated
EWD leak outlet port with the flow control valve
by Figure 2, where:
fully closed. Disable the pressure regulation
system on the EWD. Start a leak test. Record • Pmax the pressure relief system should
the pressure at which the pressure relief system operate below this pressure;
operates Pmax. Continue the cycle to obtain a
steady reading on the transducer Pb. Check • P1 the leak test pressure;
that the readings of Pmax and Pb do not exceed • T1 the start of the leak test equilibrium
the manufacturer’s recommendations. period;
17.6 Fault condition – connect the test device • P2 pressure after initial equilibrium time;
to the EWD and run a leak test. Adjust the flow
control valve to give a pressure greater than • P3 pressure determined by EWD
that specified by the manufacturer (P1). The manufacturer; the level of maximum
flow valve is opened slowly and the pressure allowable leak within the leak monitoring
transducer reading observed. The aim is to time;
increase the leak so the pressure within the test • T2 start of leak test monitoring time;
device falls below P3 before time T3 is reached,
i.e. a leak is present. Confirm from the readings • T3 end of leak test monitoring time; ∆
Pressure
Pmax
P1
P2
P3
Atmospheric Time
pressure
T0 T1 T2 T3
36
17 Leak and patency testing
• ∆P/T maximum leakage rate permitted using a 1.5 m length of PTFE tubing (1 mm
for the process to continue, i.e. just a inner diameter).
pass.
Method
(P2 – P3 ) 17.11 For each lumen (air/water, biopsy, elevator
P/T
(T3 – T2 ) as relevant), connect a 1.5 m length of tubing of
or the appropriate diameter and run an operating
cycle. On completion of the cycle replace one
(P1 – P3 ) of the tubes with a similar tube that has a Luer-
P/T
(T3 – T1 ) lock male connection. A hypodermic needle of
the appropriate size, with sheath attached and
the end cut off for safety reasons, is attached
to the 1 mm inner-diameter tube. Other small
Note lumen test pieces may be used, attached to the
1.5 m tube, as long as they present a partial
At the start of the test the pressure should be blockage to fluid flow. Another operating cycle
approximately 200 mbar (120 kPa) above should then be run. Repeat the test, changing
atmospheric pressure, unless otherwise the position of the partially obstructed tube on
stated by the manufacturer. The leak test each test.
monitoring period should not be less than 1
minute. The maximum leak rate should be no
greater than 2.5 mbar/min (0.25 kPa/min). Note
Each make of EWD and each lumen may
require a different size needle to obtain a
Lumen patency detection test partial obstruction indication. Some EWDs
may have a very sensitive lumen blocking
17.9 EWDs for endoscopes should be fitted
system. In addition the pressure of fluid down
with means to ensure that each of the lumens
each lumen may be different; this will affect
is patent so that disinfectant and rinse solutions
the partial blockage detection system. Some
will flow through each lumen, even those that
EWDs use air to detect blockage with
have a control wire. In addition a test is required
different sizes of needles from those that use
to demonstrate if an endoscope lumen
fluid flow for operation.
becomes disconnected it will be detected.
needle
Sheath with end cut off glued to needle
37
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
Results Method
17.13 The EWD should indicate a fault for any 17.15 Disconnect each lumen in turn and run a
lumen to which the appropriate partially process cycle. If the EWD can detect several
obstructed surrogate device is fitted. lumen disconnections in one cycle and record
the event, more than one lumen disconnection
can be tested for each process cycle.
Lumen disconnection detection test
17.14 To demonstrate if an EWD can detect Results
disconnected lumens, fit a surrogate device to
each lumen of the EWD and confirm that the 17.16 A satisfactory result is when the EWD fails
EWD can run a routine cycle. due to lumen disconnections and can indicate
which lumens are affected.
38
18 EWD self-disinfection test
18.1 This test is a type test to verify that the have been validated against a test biofilm in a
EWD’s “machine disinfection” mode will similar manner to chemical disinfection (see
disinfect those parts of the EWD that come into below).
contact with fluids that are intended to, or may,
contact the load. The type test will be carried
out by the manufacturer on the chemicals Chemical disinfection
selected as suitable for the EWD cycle. If a 18.5 For chemical disinfection systems a
chemical is chosen that is not listed on the microbiological test will be required. The test is
type-test list, the following test scheme will be designed to ensure that the self-disinfection
required. It is not recommended that live cycle will disinfect contaminated tubing by
organisms are used as test material on EWDs, evaluating the effect of the cycle against a
as there is a risk to patient safety. biofilm containing Pseudomonas aeruginosa.
39
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
40
18 EWD self-disinfection test
41
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
19.1 EWDs employing a chemical disinfection instructions. When this requires the addition of
stage may reuse the chemical disinfectant a water, only distilled or purified water should be
number of times (multi-use). Guidance on the used.
number of cycles for which a particular
disinfectant can be used is provided by the 19.6 A dilution series should be prepared using
chemical manufacturer. Owing to the variable distilled or purified water.
nature of the load, it is difficult to determine
exactly how many cycles the disinfectant will 19.7 If the concentration of fresh disinfectant is
remain active for. If a cycle counter is used to C1 and the minimum concentration of usable
indicate the exhaustion of the disinfectant, the disinfectant is C2,the following dilutions should
number of cycles should be set at a lower be prepared:
number than recommended to allow for a. C2 + [(C1 – C2)/10]
endoscope variation and dilution of the
chemical by residual water. b. C2
42
20 Microbiological and temperature tests of disinfection efficacy
20 M
icrobiological and temperature tests of
disinfection efficacy
20.1 Disinfection efficacy should be verified and then the filter placed onto a R2A agar plate
during type testing using the test described in and incubated at 30±2°C for up to five days.
BS EN ISO 15883-4, clause 6.12.6.2. The test The number of colony forming units (cfu)
should not be carried out until the adequacy of counted will give an indication of the adequacy
the EWD self-disinfection cycle has been of the disinfection process.
established.
20.5 A result of less than 10 cfu per brush
20.2 If temperature forms part of the sample should be obtained if the endoscope
disinfection process, the thermometric test for has been satisfactorily disinfected
assessment of temperature control throughout
the disinfection stage should be completed 20.6 This is the preferred method of lumen
satisfactorily before commencing sampling. An additional less satisfactory test
microbiological tests as described in BS EN ISO that may be useful to measure the
15883-4, clause 6.9. microbiological performance of the EWD cycle,
manual cleaning and possibly the drying
cabinets is to examine an endoscope after it
Disinfection efficacy has been used, decontaminated and dried. The
20.3 Disinfection efficacy should have been test involves passing liquid down the
tested during type-testing with a surrogate endoscope lumens, collecting the effluent and
device (BS EN ISO 15883-4, clause 6.12.6). subjecting it to bacteriological counting.
When necessary the adequacy of the
disinfection process may be verified by using 20.7 The following apparatus is necessary:
the method given below, which is included to • sterile water;
provide additional information should
disinfection problems occur. • sterile 10 mL syringe;
• sterile collection bottles;
Endoscope lumen sampling • lengths of sterile tube to connect the
20.4 The endoscope should be brushed syringe to each endoscope port in turn;
through with a sterile disposable long-stemmed
• sterile gloves.
cleaning brush after processing. The brush end
should be aseptically cut from the stem and 20.8 A suitable procedure is the following (use
aseptically transferred to 10 mL of sterile of sterile gloves and aseptic techniques are
peptone water containing 0.05% polysorbate 80 essential to obtain good results):
and ultrasonicated at between 30 and 50 kHz
for 10 minutes. The eluate should then be
filtered through a 47 mm diameter 0.45 m filter
43
Health Technical Memorandum 01-06: Decontamination of flexible endoscopes: Part E – Testing methods
44
References
References
45