SPECIFICATIONS
Operating altitude Up to 2000 meters above sea level
Operating temperature range 16°C to 30°C (60°F to 86°F)
Operating humidity Max 80% at 30°C (86°F)
Overvoltage Category 1
Pollution Degree 2
Average heat output in BTU is calculated based upon final
Heat output
system configuration and number of modules.
Dedicated Uninterrupted Power Supply Required. Your local project team will provide requirements for
(UPS) the UPS based upon your track design.
Electrical Main Line Frequency 50/60 Hz
Electrical Main Line Voltage Fluctuation Up to ± 10%
Electrical Main Line Voltage: Single phase
230V ~3680 VA
configuration with UPS-small systems
Electrical Main Line Voltage: Single phase
230V ~9200 VA
configuration without UPS-medium systems
Electrical Main Line Voltage: Three phases
400 V 3N ~ 27600
configuration without UPS-large systems
Required flow rate is calculated based upon final system
External source of compressed air is configuration
required Compressed air must be clean, oil free and dry
Compressed air pressure range 7 to 9 bar
One or more barcode readers emitting laser beams
Bar code readers
Class 2 laser device
Optional system to replace barcode readers on the Input/Output
Module, Bulk Input Module and Rack Input Module
Detects sample tube type (diameter and height)
Lateral Vision System
Detects capped/uncapped status
Detects cap color
Performs barcode label recognition

Page 1 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS
Code 128
Code 39 with and without check digit

Automation system supported barcode Interleaved 2 0f 5 with and without check digit
symbologies USS Codabar
Note: When check digit is enabled, it is enabled for all
symbologies that allow for check digit configuration (Code 39
and Interleaved 2 of 5)
Code 128: maximum 12 characters
Code 39: maximum 9 characters
Aliquoter Module label printer supported Interleaved 2 0f 5: maximum 20 characters
barcode symbologies USS Codabar: maximum 13 characters
Note: It is not possible to enable check digit for Code 39 and
Interleaved 2 0f 5 symbologies on the Aliquoter Module printer
EN IEC UL 61010 – 1:2001
CAN/CSA-C22.2 No. 61010-1-04
EN IEC 61010-2-081:2001 + A1:2003
Electromechanical Safety:
Automation system meets the requirements EN IEC 61010-2-101:2002
of the following standards EN IEC 61010-2-051:2003
CAN/CSA C22.2 No. 61010 -1-04
CAN/CSA C22.2 No. 61010 -2-081-04
Electromagnetic Compatibility (EMC):
Automation system conforms to relevant Directive 2004/108/EC
provisions of:
1) The device may not cause harmful interference
Radio Frequency Tracking:
2) The device must accept all interference received, including interference that
Automation system complies with Part 15 of may be undesired operation.
the United States Federal Communication
Commission (FCC) rules subject to the
following two conditions: The CANBUS antenna wiring incorporated in this equipment has been modified
to comply with Part 15 (subpart B and C) of the FCC rules. Ferrite Fair Rite
046117645 Bin H4 has been added on transponder cables.
Radio and Telecommunications Terminal
Equipment (R&TTE): The RFID antenna
system mounted in this equipment complies Directive 1999/5/EC
with the essential requirements and other
relevant provisions of:

Page 2 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS
• ARCHITECT c16000
• ARCHITECT c8000
Available Abbott analyzer connections • ARCHITECT i2000SR
• CELL-DYN Sapphire
• CELL-DYN SlideMaker Stainer

Connections to the following testing modalities are available:


Coagulation
Open Connectivity: Third party analyzer
connections are available in some countries Immunoassay
and all require agreement by the respective Clinical Chemistry
vendor prior to Abbott offering the
Hematology
connection.
HPLC
Specific information regarding 3rd party connections is
available from your Abbott Sales Representative.
PET (Polyethylene terephthalate
Acceptable types of plastic tubes that can be
sealed using the Sealer Module include: PS (Polystyrene)
PP (Polypropylene)

Nominal Measures of the Allowable Sample Tubes


Nominal Measures (with closures)
Name Manufacturer
mm
Greiner Bio-
Vacuette 13 x 100 16 x 100 13 x 75 -
One
Vacutainer BD (Becton,
Hemogard Tube Dickinson and 13 x 100 16 x 100 13 x 75 -
Closure Company)
S-Monovette Sarstedt 13 x 100 16 x 100 13 x 75 16 x 75
Venosafe Terumo 13 x 100 16 x 100 13 x 75 -
Vacutest Kima 13 x 100 16 x 100 13 x 75 -

Page 3 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
Actual Measures of the Allowable Sample Tubes
Actual Measures (with closures)
Name Manufacturer
mm
Greiner Bio-
Vacuette 16.13 x 106.50 17.51 x 106.18 16.30 x 82.7 -
One
Vacutainer BD (Becton,
Hemogard Tube Dickinson and 16.05 x 106.21 17.60 x 105.93 15.92 x 81.66 -
Closure Company)

S-Monovette Sarstedt 15.76 x 106.28 17.81 x 108.33 15.65 x 91.40 17.75 x 91.40

Venosafe Terumo 14.60 x 105.31 17.24 x 105.32 14.66 x 80.83 -

Vacutest Kima 15.22 x 106.53 17.36 x 104.75 15.17 x 81.49 -

Tube specifications with regular cap are listed in the following tables.
16m m T all T ube

Tube Nominal Height = 100 mm


Tube Nominal Diameter = 16 mm
Heights (mm)
A- 14 to 20 Cap height
B- 82 - 102 Uncapped height
C- 110 Total tube height
Widths (mm)
D- 12.95 Interior diameter of tube
E- 17.24 to 17.81 Cap width

Page 4 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

13m m T all T ube

Tube Nominal Height = 100 mm


Tube Nominal Diameter = 13 mm
Heights (mm)
A- 13 to 19 Cap height
B- 82 - 102 Uncapped height
C- 110 Total tube height
Widths (mm)
D- 10.25 Interior diameter of tube
E- 14.60 to 16.30 Cap width

13mm Short Tube

Tube Nominal Height = 75 mm


Tube Nominal Diameter = 13 mm
Heights (mm)
A- 13 to 19 Cap height
B- 70 - 80 Uncapped height
C- 86 Total tube height
Widths (mm)
D- 10.25 Interior diameter of tube
E- 14.60 to 16.30 Cap width

Page 5 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

16m m T all T ube

Tube Nominal Height = 100 mm


Tube Nominal Diameter = 16 mm
Heights (mm)
A- 24.4 Cap height
B- 82 to 102 Uncapped height
C- 108.0 Total tube height
Widths (mm)
D- 12.95 Interior diameter of tube
E- 17.8 Cap width
F- 8.3 Cap width

13m m T all T ube

Tube Nominal Height = 100 mm


Tube Nominal Diameter = 16 mm
Heights (mm)
A- 24.4 Cap height
B- 82 to 102 Uncapped height
C- 108.0 Total tube height
Widths (mm)
D- 10.25 Interior diameter of tube
E- 15 Cap width
F- 8.3 Cap width

Page 6 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

16m m Short T ube

Tube Nominal Height = 75 mm


Tube Nominal Diameter = 16 mm
Heights (mm)
A- 24.4 Cap height
B- 74.7 Uncapped height
C- 91.3 Total tube height
Widths (mm)
D- 12.95 Interior diameter of tube
E- 17.8 Cap width
F- 8.3 Cap width

13m m Short T ube

Tube Nominal Height = 75 mm


Tube Nominal Diameter = 13 mm
Heights (mm)
A- 24.4 Cap height
B- 74.7 Uncapped height
C- 91.3 Total tube height
Widths (mm)
D- 11.0 Interior diameter of tube
E- 15.0 Cap width
F- 8.3 Cap width

Page 7 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Input/Output Module
Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)

Up to 780 Up to 750 50/60 Hz

Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 64
2.7A 1,768
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 64
Compressed air Power consumption
Weight
(NL/min) (VA)

400kg (881.8 lbs.) 13.457 650


Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
1970mm (77.56in) x 1380mm (54.33in) x 1085mm (42.71in)
• The Input/Output Module provides a single point of sample tube input and output, barcode tube
identification and tube type identification.
• Configurable lanes for sample input or output using capped, uncapped, un-spun or pre-spun sample tubes.

Page 8 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Rack Input Module

Nominal Line Frequency


Capacity (Tubes) Throughput Tubes/hour
(Hz)

Up to 288 Up to 800 50/60 Hz


Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
4A 2,584
3. Distance 1 m from rear side: 63
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 63
Compressed air Power consumption
Weight
(NL/min) (VA)
200kg (440.9 lbs.) 13.113 950
Wastes Exhaust air Stub length (m)
0 0 1
Dimensions
(LxWxD)

1224mm (48.19in) x 1204mm (47.40in) x 702mm (27.64in)


• The Rack Input Module provides a point of sample tube input, barcode tube identification

Page 9 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Bulk Input Module


Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)
Up to 800 Up to 1,000 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 72
2. distance 1 m from front side: 68
0.7A 435.2
3. distance 1 m from rear side: 66
4. distance 1 m from left side: 66
5. distance 1 m from right side: 66
Compressed air Power consumption
Weight
(NL/min) (VA)
200kg (440.9 lbs.) 23.95 160
Wastes Exhaust air Stub length (m)
0 0 1
Dimensions
(LxWxD)
981mm (38.62in) x 1250mm (49.21in) x 920mm (36.22in)
• The Bulk Input Module (BIM) provides a point of sample tube input, barcode tube identification and tube
type identification (if equipped with Lateral Vision System). It allows direct loading onto the track of capped
sample tubes without using racks.
• Accepts unspun, capped samples. The use of pre-spun sample tubes in the BIM is not recommended.
• If pre-spun sample tubes are used in the BIM the following applies:
o The User must verify the suitability and performance of each tube for the intended application.
o Further, if pre-spun sample tubes are loaded into the BIM, it is recommended that only serum-
separating tubes with gel separator are used. Refer to sample tube package inserts from respective
tube manufacturer for specific guidance on sample handling and processing.
o One of the following conditions must be met:
 A centrifuge module is not present on the Automation System
 Pre-spun tube barcodes have a different prefix on order to distinguish them from unspun
tubes.

Page 10 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Rack Ouput Module


Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)
Up to 288 Up to 800 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
4A 2,584 _
Compressed air Power consumption
Weight
(NL/min) (VA)
200kg (440.9 lbs.) 15.732 950
Wastes Exhaust air Stub length (m)
0 0 1
Dimensions
(LxWxD)
1224mm (48.19in) x 1204mm (47.40in) x 702mm (27.64in)
• The Rack Output Module provides a point of sample tube output.

Page 11 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Centrifuge Module
Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)

Up to 80 Up to 300 50/60 Hz

Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 64
12.8A 8,024
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 64
5. Distance 1 m from right side: 63
Compressed air Power consumption
Weight
(NL/min) (VA)
375kg (826.7 lbs.) 9.974 2,950
Wastes Exhaust air Stub length (m)
0 0 1
Dimensions
(LxWxD)
950mm (37.40in) x 1500mm (59.06in) x 1410mm (55.5in)
• Multiple, refrigerated centrifuges with independent settings are possible on each track.
• Accepts capped sample tubes only.
• Configurable speed of the centrifugation spin cycle: 1,000 RPM – 4,500 RPM

Page 12 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Decapper Module
Nominal Line Frequency
Capacity (Caps) Throughput Tubes/hour
(Hz)
Up to 2,000 in internal
Up to 800 n/a
waste bin
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
8A 522.24
3. Distance 1 m from rear side: 62
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 62
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 13.172 192
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The decapper removes both screw and pressure plastic closures from sample tubes.
• Only sample tubes specified in Operations Manual can be decapped by the Decapper Module.

Page 13 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Aliquoter Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
Up to 500 (200µL per tube
Secondary tubes: up to
and 1:4 primary to secondary
500 Tips: up to 1,000 50/60 Hz
tube ratio) includes secondary
Labels: up to 1,700
tube labeling and aliquoting
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
2.2A 1,360
3. Distance 1 m from rear side: 63
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 63
Compressed air Power consumption
Weight
(NL/min) (VA)
300kg (661.4 lbs.) 35.154 500
Wastes Exhaust air Stub length (m)
0 0 1.5
Dimensions
(LxWxD)
701mm (27.59in) x 895mm (35.24in) x 1561mm (61.46in)
• Up to 4 secondary sample tubes can be generated at a time from a presented primary sample tube. If more
than 4 secondary sample tubes are needed from a single primary sample tube, the primary tube is routed to
the Aliquoter Module more than once until all secondary sample tubes are generated.
• Secondary tubes labels are resistant to storing temperature up to -80° C for 4 weeks
• Users should inspect secondary sample tubes for filling and recapping prior to transport.
• The Aliquoter Secondary Tube is made from a medical grade polypropylene (PP). An evaluation of all
potential analytical applications for samples processed with this tube has not been performed. Therefore it is
recommended that the laboratory verifies the suitability of the tube for each intended application.

Page 14 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS
Secondary Tube Recapper Module
Nominal Line Frequency
Capacity (Caps) Throughput Tubes/hour
(Hz)

Up to 1,000 Up to 400 n/a

Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
2.5A 163.2
3. Distance 1 m from rear side: 62
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 62
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 16.376 60
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Recapper Module allows automatic positioning of plastic screw caps on secondary sample
tubes produced by the Aliquoter Module.

Page 15 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Sealer Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
Up to 16,000 seals per
Up to 800 50/60 Hz
cartridge
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS

1. At operator’s normal position


and at bystanders’ positions: 68
2. Distance 1 m from front side: 64
0.8A 489.6
3. Distance 1 m from rear side: 63
4. Distance 1 m from left side: 65
5. Distance 1 m from right side: 63

Compressed air Power consumption


Weight
(NL/min) (VA)
15kg (33.1 lbs.) 70.578 180
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Sealer Module allows automatic sealing of plastic sample tubes after analytical processing.

Page 16 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Desealer Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a Up to 200 n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
2.5A 163.2 _
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 11.89 60
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Desealer Module automates removal of aluminum seals from sample tubes and seals are discarded into
the Desealer waste container.
• Any sample tubes sealed by the Sealer Module can be automatically desealed by the Desealer Module.

Page 17 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS

Storage and Retrieval Module (15,360 tubes)


Nominal Line Frequency Required Floor
Capacity (Tubes) Throughput Tubes/hour
(Hz) Capacity

400kg/m2 (81.93
Up to 15,360 Up to 800 50/60 Hz
lbs./ft2)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS

1. At operator’s normal position


and at bystanders’ positions: 67
2. Distance 1 m from front side: 65
14A 8,840
3. Distance 1 m from rear side: 65
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 65

Compressed air Power consumption


Weight
(NL/min) (VA)

1,390kg (3,064.43 lbs.) 13.413 3,250


Wastes Exhaust air Stub length (m)
1Liter/hour 0 0.5
Dimensions
(LxWxD)
Max. track height = 910mm 1950mm (76.77in) x 2570mm (101.18in) x 1778mm (70 in)

Nominal track height = 885mm 1950mm (76.77in) x 2545mm (100.19in) x 1778mm (70 in)

Min. track height = 880mm 1950mm (76.77in) x 2540mm (100 in) x 1778mm (70 in)
• The Storage and Retrieval Module stores sample tubes from the Automation System in a temperature
controlled and protected environment.

Page 18 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

Storage and Retrieval Module (9,216 tubes)

Nominal Line Frequency Required Floor


Capacity (Tubes) Throughput Tubes/hour
(Hz) Capacity

400kg/m2 (81.93
Up to 9,216 Up to 800 50/60 Hz
lbs./ft2)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 65
14A 8,840
3. Distance 1 m from rear side: 65
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 65
Compressed air Power consumption
Weight
(NL/min) (VA)
1,220kg (2,689.64 lbs.) 12.854 3,250
Wastes Exhaust air Stub length (m)
1Liter/hour 0 0.5
Dimensions
(LxWxD)

Max. track height = 910mm 1950mm (76.77in) x 2175mm (85.63in) x 1778mm (70 in)

Nominal track height = 885mm 1950mm (76.77in) x 2150 (84.65in) x 1778 (70 in)
Min. track height = 880mm 1950mm (76.77in) x 2145mm (84.45 in) x 1778mm (70 in)
• The Storage and Retrieval Module stores sample tubes from the Automation System in a temperature
controlled and protected environment.

Page 19 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

U Turn Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
2A 136.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 4.444 50
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Track U-Turn Module provides a short way for sample tubes to reach their Analyzer or Automation
Module destination.

Page 20 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS

T Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
4A 272.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 4.444 100
Wastes Exhaust air Stub length (m)

0 0 1 Stub length = stub module + stub sampling

Dimensions
(LxWxD)
1300mm (51.18in) x 1000mm (39.37in) x 510mm (20.08in)
• The Track T-Module is an extension of the Track Module which allows sample tubes to reach their Analyzer
or Automation Module destination.

Page 21 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.

SPECIFICATIONS
L Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
8A 544.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 0 200
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
500mm (19.7in) x500mm (19.7in) x 100mm (3.9in)

SPECIFICATIONS
Track Head
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
4A 272.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
17kg (37.5 lbs.) 0 100
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
430mm (16.9in) x 885mm (34.8in) x 196mm (7.7in)

Page 22 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Track Extension
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 68
2. Distance 1 m from front side: 65
n/a n/a
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 64
5. Distance 1 m from right side: 65
Compressed air Power consumption
Weight
(NL/min) (VA)
172kg (379.2 lbs.) n/a n/a
Wastes Exhaust air Stub length (m)
0 0 _

Dimensions
(LxWxD)
2300mm (90.6in) x 1020mm (40.2in) x 430mm (16.9in)

SPECIFICATIONS
Standard Slot
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1A 68 _
Compressed air Power consumption
Weight
(NL/min) (VA)
2kg (4.4 lbs.) 4.444 25
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track

Page 23 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS

Connection- ARCHITECT i2000SR


Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a 220Vac = 50/60Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
0.8A 516.8 _
Compressed air Power consumption
Weight
(NL/min) (VA)
40kg (88.2 lbs.) 10.57 190
Wastes Exhaust air Stub length (m)
1.5 Stub Length = external stub + internal stub
0 0

Dimensions
(LxWxD)
465mm (18.3in) x 200mm (7.9in) x 1500mm (59.1in)
• The connection to the ARCHITECT i2000SR is a point in space connection which means that the
samples will be aspirated automatically while on the spur and then route back to the main track.
• Samples may be front loaded manually by the operator onto the track spur in either Priority or
Routine locations if needed. Samples manually loaded are removed manually by the operator.

Page 24 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS

Connection- ARCHITECT c8000/c16000


Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1.45 95.2 _
Compressed air Power consumption
Weight
(NL/min) (VA)
21kg (46.3 lbs.) 4.444 35
Wastes Exhaust air Stub length (m)

0 0 0.5

Dimensions
(LxWxD)
200mm (7.9in) x 200mm (7.9in) x 1442mm (56.8in)
• The connection to the ARCHITECT c8000 and c16000 is a point in space connection which means
that the samples will be aspirated automatically while on the spur and then route back to the main
track.
• Samples may be front loaded manually by the operator directly into the sample carousel within the
analyzer if needed. Samples manually loaded are removed manually by the operator.

Page 25 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS

Interface Module- CELL-DYN


Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a - 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
16A 10,064 _
Compressed air Power consumption
Weight
(NL/min) (VA)
65kg (143.3 lbs.) 11.118 3,700
Wastes Exhaust air Stub length (m)
0 0 0.5

Dimensions
(LxWxD)
443mm (17.4in) x 1313mm (51.7in) x 1456mm (57.3in)
• The interface module for CELL-DYN systems is a pick and place design which uses a robotic arm
to remove samples from the track and place them into analyzer racks. After analysis, the robotic
arm will place the sample tubes back into a carrier on the main track for further routing as needed.
• Up to 4 CELL-DYN Sapphire analyzers and 1 CELL-DYN SlideMaker Stainer may be connected to
a single Hematology Island Module. Instrument specific connections are required.
• Multiple Hematology Island Modules may be connected to an ACCELERATOR a3600 as needed.

Page 26 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Connection- CELL-DYN Sapphire
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
8.2 5,168 _
Compressed air Power consumption
Weight
(NL/min) (VA)
90kg (198.4 lbs.) 0.1 1,900
Wastes Exhaust air Stub length (m)
0 0 0

Dimensions
(LxWxD)
1220mm (48.03in) x 960mm (37.8in) x 1456mm (57.3in)

SPECIFICATIONS
Connection- CELL-DYN SlideMaker Stainer
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
5.4 3,400 _
Compressed air Power consumption
Weight
(NL/min) (VA)
90kg (198.4 lbs.) 0.144 1,250
Wastes Exhaust air Stub length (m)
0 0 0

Dimensions
(LxWxD)
855mm (33.7in) x 960mm (37.8in) x 1456mm (57.3in)

Page 27 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Fake Node
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
n/a n/a _
Compressed air Power consumption
Weight
(NL/min) (VA)
0 2 n/a
Wastes Exhaust air Stub length (m)
0 0 0

Dimensions
(LxWxD)
Included in the track

The ACCELERATOR a3600 is manufactured by Inpeco S.A. and distributed by Abbott


Laboratories.

ACCELERATOR a3600, ACCELERATOR, ARCHITECT, i2000SR, c8000, c16000 and Choose


Transformation are trademarks of Abbott Laboratories in various jurisdictions.

© 2015 Abbott Laboratories ADD-00056777, 10/2015

Page 28 of 28

You might also like