Professional Documents
Culture Documents
Accelerator A3600: Abbott Diagnostics
Accelerator A3600: Abbott Diagnostics
ACCELERATOR a3600
Total Laboratory Automation from an
Automation Leader
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Operating altitude Up to 2000 meters above sea level
Operating temperature range 16°C to 30°C (60°F to 86°F)
Operating humidity Max 80% at 30°C (86°F)
Overvoltage Category 1
Pollution Degree 2
Average heat output in BTU is calculated based upon final
Heat output
system configuration and number of modules.
Dedicated Uninterrupted Power Supply Required. Your local project team will provide requirements for
(UPS) the UPS based upon your track design.
Electrical Main Line Frequency 50/60 Hz
Electrical Main Line Voltage Fluctuation Up to ± 10%
Electrical Main Line Voltage: Single phase
230V ~3680 VA
configuration with UPS-small systems
Electrical Main Line Voltage: Single phase
230V ~9200 VA
configuration without UPS-medium systems
Electrical Main Line Voltage: Three phases
400 V 3N ~ 27600
configuration without UPS-large systems
Required flow rate is calculated based upon final system
External source of compressed air is configuration
required Compressed air must be clean, oil free and dry
Compressed air pressure range 7 to 9 bar
One or more barcode readers emitting laser beams
Bar code readers
Class 2 laser device
Optional system to replace barcode readers on the Input/Output
Module, Bulk Input Module and Rack Input Module
Detects sample tube type (diameter and height)
Lateral Vision System
Detects capped/uncapped status
Detects cap color
Performs barcode label recognition
Page 1 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Code 128
Code 39 with and without check digit
Automation system supported barcode Interleaved 2 0f 5 with and without check digit
symbologies USS Codabar
Note: When check digit is enabled, it is enabled for all
symbologies that allow for check digit configuration (Code 39
and Interleaved 2 of 5)
Code 128: maximum 12 characters
Code 39: maximum 9 characters
Aliquoter Module label printer supported Interleaved 2 0f 5: maximum 20 characters
barcode symbologies USS Codabar: maximum 13 characters
Note: It is not possible to enable check digit for Code 39 and
Interleaved 2 0f 5 symbologies on the Aliquoter Module printer
EN IEC UL 61010 – 1:2001
CAN/CSA-C22.2 No. 61010-1-04
EN IEC 61010-2-081:2001 + A1:2003
Electromechanical Safety:
Automation system meets the requirements EN IEC 61010-2-101:2002
of the following standards EN IEC 61010-2-051:2003
CAN/CSA C22.2 No. 61010 -1-04
CAN/CSA C22.2 No. 61010 -2-081-04
Electromagnetic Compatibility (EMC):
Automation system conforms to relevant Directive 2004/108/EC
provisions of:
1) The device may not cause harmful interference
Radio Frequency Tracking:
2) The device must accept all interference received, including interference that
Automation system complies with Part 15 of may be undesired operation.
the United States Federal Communication
Commission (FCC) rules subject to the
following two conditions: The CANBUS antenna wiring incorporated in this equipment has been modified
to comply with Part 15 (subpart B and C) of the FCC rules. Ferrite Fair Rite
046117645 Bin H4 has been added on transponder cables.
Radio and Telecommunications Terminal
Equipment (R&TTE): The RFID antenna
system mounted in this equipment complies Directive 1999/5/EC
with the essential requirements and other
relevant provisions of:
Page 2 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
• ARCHITECT c16000
• ARCHITECT c8000
Available Abbott analyzer connections • ARCHITECT i2000SR
• CELL-DYN Sapphire
• CELL-DYN SlideMaker Stainer
Page 3 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
Actual Measures of the Allowable Sample Tubes
Actual Measures (with closures)
Name Manufacturer
mm
Greiner Bio-
Vacuette 16.13 x 106.50 17.51 x 106.18 16.30 x 82.7 -
One
Vacutainer BD (Becton,
Hemogard Tube Dickinson and 16.05 x 106.21 17.60 x 105.93 15.92 x 81.66 -
Closure Company)
S-Monovette Sarstedt 15.76 x 106.28 17.81 x 108.33 15.65 x 91.40 17.75 x 91.40
Tube specifications with regular cap are listed in the following tables.
16m m T all T ube
Page 4 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
Page 5 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
Page 6 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
Page 7 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Input/Output Module
Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 64
2.7A 1,768
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 64
Compressed air Power consumption
Weight
(NL/min) (VA)
Page 8 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Page 9 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Page 10 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Page 11 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Centrifuge Module
Nominal Line Frequency
Capacity (Tubes) Throughput Tubes/hour
(Hz)
Up to 80 Up to 300 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 64
12.8A 8,024
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 64
5. Distance 1 m from right side: 63
Compressed air Power consumption
Weight
(NL/min) (VA)
375kg (826.7 lbs.) 9.974 2,950
Wastes Exhaust air Stub length (m)
0 0 1
Dimensions
(LxWxD)
950mm (37.40in) x 1500mm (59.06in) x 1410mm (55.5in)
• Multiple, refrigerated centrifuges with independent settings are possible on each track.
• Accepts capped sample tubes only.
• Configurable speed of the centrifugation spin cycle: 1,000 RPM – 4,500 RPM
Page 12 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Decapper Module
Nominal Line Frequency
Capacity (Caps) Throughput Tubes/hour
(Hz)
Up to 2,000 in internal
Up to 800 n/a
waste bin
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
8A 522.24
3. Distance 1 m from rear side: 62
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 62
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 13.172 192
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The decapper removes both screw and pressure plastic closures from sample tubes.
• Only sample tubes specified in Operations Manual can be decapped by the Decapper Module.
Page 13 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Aliquoter Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
Up to 500 (200µL per tube
Secondary tubes: up to
and 1:4 primary to secondary
500 Tips: up to 1,000 50/60 Hz
tube ratio) includes secondary
Labels: up to 1,700
tube labeling and aliquoting
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
2.2A 1,360
3. Distance 1 m from rear side: 63
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 63
Compressed air Power consumption
Weight
(NL/min) (VA)
300kg (661.4 lbs.) 35.154 500
Wastes Exhaust air Stub length (m)
0 0 1.5
Dimensions
(LxWxD)
701mm (27.59in) x 895mm (35.24in) x 1561mm (61.46in)
• Up to 4 secondary sample tubes can be generated at a time from a presented primary sample tube. If more
than 4 secondary sample tubes are needed from a single primary sample tube, the primary tube is routed to
the Aliquoter Module more than once until all secondary sample tubes are generated.
• Secondary tubes labels are resistant to storing temperature up to -80° C for 4 weeks
• Users should inspect secondary sample tubes for filling and recapping prior to transport.
• The Aliquoter Secondary Tube is made from a medical grade polypropylene (PP). An evaluation of all
potential analytical applications for samples processed with this tube has not been performed. Therefore it is
recommended that the laboratory verifies the suitability of the tube for each intended application.
Page 14 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Secondary Tube Recapper Module
Nominal Line Frequency
Capacity (Caps) Throughput Tubes/hour
(Hz)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 65
2. Distance 1 m from front side: 63
2.5A 163.2
3. Distance 1 m from rear side: 62
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 62
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 16.376 60
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Recapper Module allows automatic positioning of plastic screw caps on secondary sample
tubes produced by the Aliquoter Module.
Page 15 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Sealer Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
Up to 16,000 seals per
Up to 800 50/60 Hz
cartridge
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
Page 16 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Desealer Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a Up to 200 n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
2.5A 163.2 _
Compressed air Power consumption
Weight
(NL/min) (VA)
6kg (13.2 lbs.) 11.89 60
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Desealer Module automates removal of aluminum seals from sample tubes and seals are discarded into
the Desealer waste container.
• Any sample tubes sealed by the Sealer Module can be automatically desealed by the Desealer Module.
Page 17 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
400kg/m2 (81.93
Up to 15,360 Up to 800 50/60 Hz
lbs./ft2)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
Nominal track height = 885mm 1950mm (76.77in) x 2545mm (100.19in) x 1778mm (70 in)
Min. track height = 880mm 1950mm (76.77in) x 2540mm (100 in) x 1778mm (70 in)
• The Storage and Retrieval Module stores sample tubes from the Automation System in a temperature
controlled and protected environment.
Page 18 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
400kg/m2 (81.93
Up to 9,216 Up to 800 50/60 Hz
lbs./ft2)
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 67
2. Distance 1 m from front side: 65
14A 8,840
3. Distance 1 m from rear side: 65
4. Distance 1 m from left side: 63
5. Distance 1 m from right side: 65
Compressed air Power consumption
Weight
(NL/min) (VA)
1,220kg (2,689.64 lbs.) 12.854 3,250
Wastes Exhaust air Stub length (m)
1Liter/hour 0 0.5
Dimensions
(LxWxD)
Max. track height = 910mm 1950mm (76.77in) x 2175mm (85.63in) x 1778mm (70 in)
Nominal track height = 885mm 1950mm (76.77in) x 2150 (84.65in) x 1778 (70 in)
Min. track height = 880mm 1950mm (76.77in) x 2145mm (84.45 in) x 1778mm (70 in)
• The Storage and Retrieval Module stores sample tubes from the Automation System in a temperature
controlled and protected environment.
Page 19 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
U Turn Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
2A 136.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 4.444 50
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
• The Track U-Turn Module provides a short way for sample tubes to reach their Analyzer or Automation
Module destination.
Page 20 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
T Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
4A 272.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 4.444 100
Wastes Exhaust air Stub length (m)
Dimensions
(LxWxD)
1300mm (51.18in) x 1000mm (39.37in) x 510mm (20.08in)
• The Track T-Module is an extension of the Track Module which allows sample tubes to reach their Analyzer
or Automation Module destination.
Page 21 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
L Module
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
8A 544.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
30kg (66.1 lbs.) 0 200
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
500mm (19.7in) x500mm (19.7in) x 100mm (3.9in)
SPECIFICATIONS
Track Head
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
4A 272.0 _
Compressed air Power consumption
Weight
(NL/min) (VA)
17kg (37.5 lbs.) 0 100
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
430mm (16.9in) x 885mm (34.8in) x 196mm (7.7in)
Page 22 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Track Extension
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1. At operator’s normal position
and at bystanders’ positions: 68
2. Distance 1 m from front side: 65
n/a n/a
3. Distance 1 m from rear side: 64
4. Distance 1 m from left side: 64
5. Distance 1 m from right side: 65
Compressed air Power consumption
Weight
(NL/min) (VA)
172kg (379.2 lbs.) n/a n/a
Wastes Exhaust air Stub length (m)
0 0 _
Dimensions
(LxWxD)
2300mm (90.6in) x 1020mm (40.2in) x 430mm (16.9in)
SPECIFICATIONS
Standard Slot
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
1A 68 _
Compressed air Power consumption
Weight
(NL/min) (VA)
2kg (4.4 lbs.) 4.444 25
Wastes Exhaust air Stub length (m)
0 0 0.5
Dimensions
(LxWxD)
Included in the track
Page 23 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Dimensions
(LxWxD)
465mm (18.3in) x 200mm (7.9in) x 1500mm (59.1in)
• The connection to the ARCHITECT i2000SR is a point in space connection which means that the
samples will be aspirated automatically while on the spur and then route back to the main track.
• Samples may be front loaded manually by the operator onto the track spur in either Priority or
Routine locations if needed. Samples manually loaded are removed manually by the operator.
Page 24 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
0 0 0.5
Dimensions
(LxWxD)
200mm (7.9in) x 200mm (7.9in) x 1442mm (56.8in)
• The connection to the ARCHITECT c8000 and c16000 is a point in space connection which means
that the samples will be aspirated automatically while on the spur and then route back to the main
track.
• Samples may be front loaded manually by the operator directly into the sample carousel within the
analyzer if needed. Samples manually loaded are removed manually by the operator.
Page 25 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Dimensions
(LxWxD)
443mm (17.4in) x 1313mm (51.7in) x 1456mm (57.3in)
• The interface module for CELL-DYN systems is a pick and place design which uses a robotic arm
to remove samples from the track and place them into analyzer racks. After analysis, the robotic
arm will place the sample tubes back into a carrier on the main track for further routing as needed.
• Up to 4 CELL-DYN Sapphire analyzers and 1 CELL-DYN SlideMaker Stainer may be connected to
a single Hematology Island Module. Instrument specific connections are required.
• Multiple Hematology Island Modules may be connected to an ACCELERATOR a3600 as needed.
Page 26 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Connection- CELL-DYN Sapphire
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
8.2 5,168 _
Compressed air Power consumption
Weight
(NL/min) (VA)
90kg (198.4 lbs.) 0.1 1,900
Wastes Exhaust air Stub length (m)
0 0 0
Dimensions
(LxWxD)
1220mm (48.03in) x 960mm (37.8in) x 1456mm (57.3in)
SPECIFICATIONS
Connection- CELL-DYN SlideMaker Stainer
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a 50/60 Hz
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
5.4 3,400 _
Compressed air Power consumption
Weight
(NL/min) (VA)
90kg (198.4 lbs.) 0.144 1,250
Wastes Exhaust air Stub length (m)
0 0 0
Dimensions
(LxWxD)
855mm (33.7in) x 960mm (37.8in) x 1456mm (57.3in)
Page 27 of 28
Refer to the Operations Manual for operational precautions, limitations and hazards. Manuals
can be found on www.abbottdiagnostics.com. Specifications are subject to modification, revision
and improvement.
SPECIFICATIONS
Fake Node
Nominal Line Frequency
Capacity Throughput Tubes/hour
(Hz)
n/a n/a n/a
Maximum
Continuous Heat (BTU/h) Noise (dBA)
Current AMPS
n/a n/a _
Compressed air Power consumption
Weight
(NL/min) (VA)
0 2 n/a
Wastes Exhaust air Stub length (m)
0 0 0
Dimensions
(LxWxD)
Included in the track