When calibration, take the calibrator out from 2-8℃, mix well, aspirate
the required amount into the sample cup, and immediately put the
AFP、CEA、CA125、CA15-3、CA19-9、FPSA、t-PSA、FERR
I、PG II、CA50,CA242、HE4、SCCA、ProGRP;
Total β HCG、FSH、LH、TESTO、PROG、E2、E3;
HBsAg、Anti-HBs、HBeAg、Anti-HBe、Anti-HBc、HIV;
Insulin、C-Peptide、DHEA-S、Cortisol、Tg、Anti-Tg、Anti-TPO;
After the calibrator is opened, use the EP tube to aliquot it (one aliquot is
the volume required for one calibration + dead volume), and then store it
at -20°C. Take it out before use and equilibrate to room temperature.
TSH、PRL
frozen:
MB、ACTH、NSE、Renin、AMH
1. Take out the calibrator and equilibrate for 15-30 minutes to the room
2. Unscrew the outer cover and carefully remove the rubber stopper. Place
EP tube (one aliquot is the volume required for one calibration + dead
volume), then store at -20°C. Take it out before use and equilibrate to
room temperature.
QC storage:
bottles in the dispensing container evenly, use EP tubes for aliquot (one
aliquot is the volume required for one QC + dead volume), and then store
and freeze them. Take it out from 2-8℃ during use, mix well and aspirate
the required amount into the sample cup, immediately put the remaining
QC products back to 2-8℃; or insert the original tube into the sample
rack directly on the machine, close the bottle cap immediately after
Involved product: ACTH QC, NSE QC, cardiac marker QC, hypertension
2. Unscrew the outer cover and carefully remove the rubber stopper. Place
4. Cover the rubber stopper and balance for 20-30 minutes. During this
tubes (one aliquot is the volume required for one QC + dead volume), and
then store at -20°C. Take it out before use and equilibrate to room
temperature.
Tumor Marker
Opened
Target Value Unopened Stability (2- Bottle Capacity/ Aspiration
Calibrator Frequency Stability (2- Frozen Storage
C0/C1/C2 8 ℃) Volume
8 ℃)
t-PSA lot of the ≈0/2/60 ng/ml Within the validity 3 months — 2 ml/ 20 μl
reagent. period
ProGRP ≈0/40/800 pg/ml Within the validity 2 months — C0: 1.2 ml; C1、C2: 1.0mL /25 μl
period
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Fertility
Bottle Capacity/
Unopened Stability Opened Stability
Calibrator Frequency Target Value C0/C1/C2 Frozen Storage Aspiration
(2-8℃) (2-8℃)
Volume
Within the validity
Total β HCG ≈0/50/4500 mIU/ml 3 months — 2 ml/ 15 μl
period
Within the validity
FSH ≈0/10/100 mIU/ml 3 months — 2 ml/ 30 μl
period
Within the validity
LH ≈0/10/100 mIU/ml 3 months — 2 ml/ 20 μl
period
1. When necessary, Within the validity 1 time: -20℃ 3
PRL ≈0/10/100 ng/ml 2 months 2 ml/ 15 μl
such as the QC is period months
out of range; Within the validity
TESTO 2. When changing ≈0/1.5/16 ng/ml 3 months — 2 ml/ 25 μl
period
the reagent lot; Within the validity
PROG 3. When the ≈0/4/40 ng/ml 3 months — 2 ml/ 30 μl
period
calibration is Within the validity
E2 expired for the ≈0/500/4800 pg/ml 3 months — 2 ml/ 35 μl
period
same lot of the Within the validity
E3 reagent. ≈0/2/30 ng/ml 3 months — 2 ml/ 25 μl
period
Within the validity
Cortisol ≈0/5/55 μg/dl 2 months — 2 ml/10 μl
period
Within the validity
DHEA-S ≈0/20/500 μg/dl 2 months — 2 ml/10 μl
period
ACTH ≈0/100/1000 pg/ml Within the validity 7 days 1 time: -20℃ 2 2 ml/50 μl
period months
Within the validity C0: 0.6 ml;
AMH ≈0/0.2/8 ng/ml 8h 1 time: -20℃ 28 days C1、C2: 0.5mL
period /50 μl
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Thyroid
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Infectious Disease
Unopened Stability Opened Stability Bottle Capacity/
Calibrator Frequency Target Value C0/C1/C2 Frozen Storage
(2-8℃) (2-8℃) Aspiration Volume
Within the validity
HBsAg ≈0/5/100 IU/ml 1 month — 2 ml/ 55 μl
period
Within the validity
Anti-HBs ≈0/50/500 mIU/ml 1 month — 2 ml/ 55 μl
period
Within the validity
HBeAg Negative/Positive 1 month — 2 ml/ 55 μl
period
1. When
Within the validity
Anti-HBe necessary, such as Negative/Positive 1 month — 2 ml/ 55 μl
period
the QC is out of
Within the validity
Anti-HBc range; Negative/Positive 1 month — 2 ml/ 55 μl
period
2. When changing
Within the validity
HIV the reagent lot; Negative/Positive 1 month — 2 ml/ 105 μl
period
3. When the
Within the validity
Anti-TP calibration is Negative/Positive 2 months — 2 ml/ 45 μl
period
expired for the
Within the validity
Anti-HCV same lot of the Negative/Positive 2 months — 2 ml/ 55 μl
period
reagent.
Within the validity —
nCoV NTAb ≈0/20/350 AU/ml 30 days 0.5 ml/ 10 μl
period
Within the validity — C0: 0.6 ml; C1、C2:
nCoV S-RBD ≈0/20/400 AU/ml 30 days
period 0.5mL /10 μl
Within the validity — C0: 0.6 ml; C1、C2:
nCoV TAb ≈0/20/350 AU/ml 30 days
period 0.5mL /10 μl
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Cardiac Marker
Bottle Capacity/
Unopened Opened Stability
Calibrator Frequency Target Value C0/C1/C2 Frozen Storage Aspiration
Stability (2-8℃) (2-8℃)
Volume
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Diabetes
Bottle Capacity/
Target Value Unopened Stability Opened Stability
Calibrator Frequency Frozen Storage Aspiration
C0/C1/C2 (2-8℃) (2-8℃)
Volume
1. When necessary, such as
the QC is out of range;
Within the validity
Insulin 2. When changing the ≈0/25/500 μIU/ml 2 months — 2 ml/10 μl
period
reagent lot;
3. When the calibration is
expired for the same lot of Within the validity
C-Peptide ≈0/0.5/10 ng/ml 2 months — 2 ml/ 10 μl
the reagent. period
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Inflammation
Bottle Capacity/
Target Value Unopened Stability Opened Stability
Calibrator Frequency Frozen Storage Aspiration
C0/C1/C2 (2-8℃) (2-8℃)
Volume
1. When necessary, such as
the QC is out of range;
2. When changing the C0: 1.2 ml;
Within the validity
PCT reagent lot; ≈0/0.5/20 ng/ml 30 days — C1、C2: 1.0mL
period
3. When the calibration is /25 μl
expired for the same lot of
the reagent.
Note:
Liver Fibrosis
Bottle Capacity/
Target Value Unopened Stability Opened Stability
Calibrator Frequency Frozen Storage Aspiration
C0/C1/C2 (2-8℃) (2-8℃)
Volume
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Gastritis
Bottle Capacity/
Target Value Unopened Stability Opened Stability
Calibrator Frequency Frozen Storage Aspiration
C0/C1/C2 (2-8℃) (2-8℃)
Volume
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Anemia & bone metabolism
Bottle Capacity/
Unopened Stability Opened Stability (2- Frozen
Calibrator Frequency Target Value C0/C1/C2 Aspiration
(2-8℃) 8 ℃) Storage
Volume
Within the validity
VD-T ≈0/20/140 ng/ml 1 month — 2 ml/40 μl
period
1. When necessary, such as
Within the validity
Folate the QC is out of range; ≈0/2.5/20 ng/ml 1 month — 2 ml/80 μl
period
2. When changing the
Within the validity
VB12 reagent lot; ≈0/200/2000 pg/ml 1 month — 2 ml/50 μl
period
3. When the calibration is
Within the validity
PTH expired for the same lot of ≈0/150/2400 pg/ml 1 month — 2 ml/30 μl
period
the reagent.
Within the validity
CT ≈0/100/1000 pg/ml 1 month — 2 ml/30 μl
period
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Hypertension
Bottle Capacity/
Target Value Unopened Stability
Calibrator Frequency Opened Stability Frozen Storage Aspiration
C0/C1/C2 (2-8℃)
Volume
Note:
Suggested allocation volume for one-time consumption=3* Aspiration volume + Dead volume
Control
2 Levels,
Thyroid Function Multi Within the validity
6 or 12 bottles for one 5.0 ml /bottle 30 days 1 time: -20℃ 3 months
Control period
level
2 Levels,
Within the validity
Tumor Marker Multi Control 6 or 12 bottles for one 5.0 ml /bottle 30 days 1 time: -20℃ 3 months
period
level
2 Levels,
Within the validity
Reproductive Multi Control 6 or 12 bottles for one 5.0 ml /bottle 30 days 1 time: -20℃ 3 months
period
level
3 Levels
Negative:3 or 6 bottles
for one level Within the validity
HIV Ag&Ab Control 2.0 ml /bottle 30 days —
Antigen positive and period
antibody positive: 3 or 6
bottles for each level
Cardiac Marker Multi Control 2 Levels, 3 or 6 bottles 2.0 ml /bottle Within the validity Troponin I: 24 hours 1 time: -20℃ 2 months
for one level period BNP: 24 hours
CK-MB: 30 days
Myo: 2 days
Note: