20 OT-1 Anesthesia Machine With Ventilator& Monitor

You might also like

Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 12

OT Operation Threater room

20 OT-1 Anesthesia Machine with Ventilator& monitor


1-Anaesthesia Machine
* *Must be suitable for Adult and Pediatrics (Upgradable to neo. )
*Selectatec type allowing for mounting two vaporisers
* *Three inlets for O2 & Air & N2O
* Safety Relief valves in case of high gas inlet pressure
*O2, N2O, Air priority flowmeter block.
*Safety N2O cut-off device
* *O2 pressure failure/Leakage warning device
* *Air pressure failure Leakage warning device
*Jumbo CO2 absorber system with bag/vent valve on swivel bracket
*two vaporizers (Halothane -Isoflorane)
*Ready for mounting 2 gas cylinders O2 & N20
*Three pressure gauges for O2& Air & N2O
Four antistatic castors two of them with brakes
Patient Assembly Unit:
*Patient circuit 3 Reusable tubing complete with connectors, Y-Piece, bag and face mask with difference
adult and Pediatrics
n.1 To and from patient circuit adult & pediatrics
n.1 O2 tube
n.1 N2O tube
n.1 Air tube
* n.1 O2 Transducer and cable for O2 transducer (Calibrate-able)
n.1 Manual ventilation kit
n.1 Cable for O2 transducer
n.1 Silicone adult patient & pediatrics circuit
n.1 O2 tube for cylinder
n.1 N2O tube for cylinder
Multilingual CD & User's Manual
Including: no.1 bag of 2 Lt.
• Inspiratory phase fresh gas decoupling
• Integrated autoclavable exp./insp. flow sensor
• Integrated and controlled emergency air valve
• Manual breathing system with breathing bag 2,3 l (bag size adult and pediatric)
Breathing system group-Real open circuit, semi-closed circuit
All parts that come in contact with patient gas can be autoclaved up to 134°C.
* Backup battery 1 hour or more
* 2-Anaesthesia Ventilator
Ventilation modes:
Volume Controlled Ventilation (VCV),
Pressure Controlled Ventilation(PCV),
Synchronized Volume Control (SIMV),
Pressure Support Manual/Spontaneous modes.
Pressure Support mode.
Wave form displayed : P - T , F - T and V - T .
* Operating modes :
1- Stand-by mode:
Pre-setting the ventilation parameters (VT, I:E, aspiratory pressure limit) & patient info
Able to user control and adjustment
* Functional Parameters:
- Tidal volume (Vt) : 20 to 1500 ml
* - O2 concentration (FiO2) : 21 – 100 %
- Rate : 4 to 60 Bpm
- I:E ratio (variable) : 1:0.5 to 1:6
-Pressure limit : 10 to 70 cm H2O
-Tidal volume adjusting : automatic / Manual
*
Monitoring:
Continues monitoring of inspiratory O2 concentration (can be switched off by Service) breathing frequenc
Tidal volume Exp and Insp.
Minute volume Exp and Insp.
mean or plateau pressure,
Peak airway pressure as well as PEEP.
In addition, all fresh gas flow information should be displayed as virtual flow tubes.
Alarms :
Audible and visual with alarm mute for low air way pressure, high air way pressure,high continuous air wa
The ventilator should be supplied with all accessories and adult and Pediatrics Bellows.
3-Patient Monitor
Better to be from the same manufacturer of anesthesia machine
*
Multi-parameter monitor with 12 inch TFT displaying 6 - 8 waveforms :
HR/PR, ECG/RESP. NIBP-SPO2-2TEMP-2IBP & available Fi CO2,Et CO2 Module, rapid and easy to all
parameter settings, alarm mute on/off freeze on/off recorder
Built-in 2 channel recorder with waveform selector,72 hr trends graphical or more and tabular for HR/PR,
SPO2-2TEMP-2IBP Respiration and Temperature
Alarm recall for last 12 alarms with date and time.
Built-in battery capable for 1 Hour operation or more
Power Parameters: 100 – 240 V / 50 / 60 Hz.
Alarms
Audible and visual alarm for all set parameters with message indicator with Mute and repeat alarm featur
The monitor should be supplied with all accessories & all connections for adult and another one for pedia
wrap) i.e: pulse oximetry (Adult / ped.) size, NIBP cuff (sizes for adult /ped.) CO2 Sensor (Calibrate-able)
Autoclavable.
Supplier with :
O2 cylinder and medical pressure pre-set regulator.
N2O cylinder and medical pressure pre-set regulator.
Operation manual & service manual
Power requirement: 100-240 V/50 HZ.
Shall comply with CE marked and/or FDA Certificate.
Installation & Commissioning :
- 3 Years warranty with Technical Assistance/Maintenance and spare parts from the date of installation a
the CE marked or 2 Years warranty with Technical Assistance/Maintenance spare parts from the date of
commissioning if FDA Certificate.
Guarantee for Availability of Spare parts up to 10 years.
Guarantee for Availability for after sales services by local agency specialist (Biomedical Engineer) beyon
Warranty.
Technical Data Sheet must accompany the offers.
User/Nurses training, by Specialist from the Suppling Agency.
Site setup, Installation, assembly and successful Operation test of Medical machine must be performed b
Biomedical Engineer from supplying agency.
Note:
All equipment needing consumables must allow the possibility to use generic and/or locally made consum
disposables. Compliance to this condition must be declared here by the bidders.
The system offered shall be designed to operate normally under the conditions of the purchaser's country
include Power Supply, Climate, Temperature, Humidity, etc.

20 OT-2 Operation Light (Ceiling) AC/DC Complete/Emergency Power


Two Arms LED Type ( 1 main + 1 back up )
Shadow less Ceiling Suspension Surgical Light
adjustable, dual arm, spring loaded, multi-facets mirror reflector
Working distance range: ~70 - 180 cm or applicable ranges for operation theatres.
minimal heat radiation emission. Absorbing radiators or other heat recovery devices shall be compulsory
Maximum total light intensity : 140.000 - 160.000 Lux at 1.0m ( LED )
Focusable light field: 45 cm, indicative measurements
This equipment shall be supplied with all necessary fixation gears to the building structure
Adjustable lighting focus by a central sterile handle
Lighting control panel integrated in the handle
Automatic change over between the main lamp and the back up one with LED indicator
The control panel should contain:
- on/off push button
- increasing/decreasing lighting push buttons
- main lides failure indicator
Power Supply, 100-220V/50 Hz. Transformer 100-220V/24VAC.
Emergency Power Supply shall be ensured by means of Stand-by UPS.
Spare lamps LED must be provided at least one set for each Arm depending on the type LED. Must be w
according to the lumen/m2
40 OT-3 Operation Light (Mobile)
Surgical Operating Mobile Lamp Led's type. Shall be very stable on 4/5 castors with antistatic rubber ban
Shadow less cold light.
Microprocessor Controlled
Adjustable lighting focus by a central handle (autoclavable)
Controllable intensity
Lamp head diameter shall be 200 to 500mm
Requested luminous intensity: > 90.000 - 110.000 Lux at 1.0m
Shall run on mains and battery . When used on Battery (DC) the decreasing of intensity shall be less than
Lighting autonomy shall not be less than 2 Hrs
Lamp rod shall be adjustable in height. As an indication:
Battery shall be equipped with an automatic charger with uninterruptible connection to mains
20 OT-4 Gyneacolgy (EMO) Operation Table
Gyn. operating table for Caesarean & Gyn. surgeries
foam upholstery pads, electrically conductive. Electrolyte or similar
Elevation of the table top actuated hydraulically with a pedal
Leg plates (Separate) shall be infinitely adjustable
Overall length including head support: at least 2000 mm or 2 meters
* Trendelenburg / reverse Trendelenburg position: at least - 20° /+ 20°
Supplied with:
All Accessories for Caesarean & Gyn. surgeries
20 OT-5 Hydraulic Operating Table For General Surgery
T shaped base,5 sections (divided leg)
together, easily moving wheels with brakes, base and column are constructed from st/st,
top plate are X- ray translucent along the full length of the table, with mattress
antistatic, head rest can be replaced
* Working load: not less than 180 KG
* Adjustments :
- Height : 750 -1000 mm hydrulically using foot pedal or better
- Trendlenburg : +/- 25 degree manually
- Back Section : -15 to + 70 degree manually
- Side Tilting : +/- 15 degree manually
- Leg Section : -90 degree manually
- Head Rest : -40 to + 30 degree manually
Castors: twin antistatic, two with foot brakes
Accessories with necessary clamps:
All ball joints, clamps, rods and fixations should be made from stainless steel
. Flat, square head rest, (table width), infinite slope adjustment, integral foam padding, 1 piece each
. Leg plates, split type, including integral foam padding, 1 piece each
. Patient restraint / body strap, 1 piece
. Arm supports, 2 pieces
. IV rod, 1 set, adjustable height Stainless Steel.
. Shoulder rest / support with pad, 2 pieces
. Narrow arm table with pad, 1 piece
."D" shaped ankle pillow, 1 piece
. Accessory / setting clamps, circular socket, 4 pieces
. Accessory / setting clamps, rotary, 4 pieces
1 Anaesthesia screen with radial support
2 Leg section supports with support
1 Body strap
1 Head support , detachable
1 Set of anti-static ,divided 6 parts
- Paediatric stirrups
-Arm and hand table
- Stainless steel bowl with drain for gynaecologic
40 OT-6 Operating Theater Table "Majo''
40 OT-7Disinfecter Ultravilot Lamp
Bacterial Lamp for Air Sterilization by means of UV Beams, UVC category or far UV (200nm til 280nm) w
strong bactericide effectiveness with a maximum strength situated at 265nm (253.7nm Mercury Spectra )
UVC shall be emitted thru vacuum ionizing tube (Weak permeation). Useful power shall be at least 45% o
nominal power (30W)
Lifetime shall be at least 7500 Hrs. Radiation shall be witnessed by means of blue light emission
Timer shall assert the limit period of the ionized tube running
Encased in steel sheet casing, Epoxy-painted.
mobile trolly option
Shall comply with. CE marked
Power supply: 220V/50 Hz
Supplied with:
Spear lamp 4 pcs
20 OT-8 Electro Surgical Unit,Generator High Frequency general surgery with trolly
Diathermy apparatus, generator high frequency
bipolar and monopolar channels
Up to 350 watts
Cutting and coagulation mode
Operating frequency up to 350 KHz
Microcontroller-based multi-programmable by the user
* Power supply 220-240V 50/60 Hz
Resettable over current breaker shall be fitted for protection
Alarm audiovisual in case of power supply disconnection.
* Power disconnection in case of loose patient plate
Mounted on trolly,stainless steel with swivel castor wheels with brake
* Complete with the following accessories:
Two button handle autoclavable 1
Two button handle cold sterilization 1
cable for plate 1
Reusable patient plate 4
Waterproof double foot switch 1
Short, general surgery blade electrode
Long, general surgery blade electrode
Short, general surgery needle electrode
Short, general surgery ball electrode
Diathermy Quiver Insulated, diathermy Cable Red rubber 5 meter
Supply with :
User manual in English
Service manual in English
Shall comply to CE marked or FDA Certificate.
Installation & Commissioning :
2 Years waranty with spear parts from the date of installation and commissioning .
User /Nurses training, by Specialist from the Suplier.
Note:
All equipment needing consumables must allow the possibility to use generic and/or locally made consum
and/or disposables. Compliance to this condition must be declared here by the bidders.
The system offered shall be designed to operate normally under the conditions of the purchaser's country
conditions include Power Supply, Climate, Temperature, Humidity, etc.
40 OT-9 Suction"Pump, mobile
Suitable for operating rooms, double collecting bottles Glass mounted on the unit made of transparent
autoclavable material with over fill protection. The collecting bottle can be dismantled without turning the
The unit should be adapted with:
* *anti-bacterial filter ( Supplied with extra spares at least 2 additional bacterial filters)
*Stainless Steel Top Tray.
Technical Specifications:
Suction capacity : 50 l/m
Vacuum regulation : from 0 to 650 mmHg
* Containers : 2 x not less than 2 litters each (Autoclavable)
Hose : 3 m silicon rubber (Autoclavable)
Suction Machine (on trolley)
40 OT-10 Suction Pump, Portable
Vacuum range 0- 600mm Hg,
Flow rates – open flow > 20 litres/ minute
Overflow safety protection
* Re-usable collection Jars dual not less than (1500cc) complete with caps and tubing (with disposable lin
Splash proof protection
* Hydrophobic filters (bacteria & exhaust) supplied with extra spares at least 2 additional Filters each.
Adjustable vacuum level control knob (high vacuum)
Suction capacity : 20 l/m
* Containers : 2 x not less than 1.5 litters each (Autoclavable)
Hose : 2 m silicon rubber (Autoclavable)
All accessories – i.e., Catheters, etc..
100 OT-11 Infant Suction Unit
OB/GY Suction Unit.
Suction Pressure = 0 - 750mmHg.
2 Suction Bottles = 1.5L / 3 L.
Supplied With Cups Set L, M & SJ.
Supplied With Complete Accessories.
Suction Machine (on trolley)
nd face mask with difference sizes for

patient info
y Service) breathing frequency,

essure,high continuous air way


ics Bellows.

Module, rapid and easy to all functions and

more and tabular for HR/PR, ECG/RESP-NIBP-

Mute and repeat alarm feature every 3 minutes.


ult and another one for pediatrics (Clamp / finger
CO2 Sensor (Calibrate-able), Adaptor

from the date of installation and commissioning if


spare parts from the date of installation and

(Biomedical Engineer) beyond Expiry of


machine must be performed by Specialized

c and/or locally made consumables and/or

ons of the purchaser's country. The conditions

devices shall be compulsory

lding structure

ED indicator

g on the type LED. Must be with compatible watts

tors with antistatic rubber bands.

g of intensity shall be less than 20%

nnection to mains
ed from st/st,

m padding, 1 piece each

r far UV (200nm til 280nm) with


m (253.7nm Mercury Spectra )
power shall be at least 45% of

of blue light emission

with trolly
c and/or locally made consumables
the bidders.
ons of the purchaser's country. The

e unit made of transparent


smantled without turning the unit off.

and tubing (with disposable liners)