Customer Technical Bulletin: CTB Information
Customer Technical Bulletin: CTB Information
Customer Technical Bulletin: CTB Information
Technical Bulletin
CTB Information
Tools Information
Revision Information
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no representation to the qualifications of any service technicians other than its own employees. Improper or
uninformed maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including but not limited to one or more of the following: mechanical collision, electrical shock, and radiation.
Any of these hazards could cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
SHEET
1 OF 11
P1027348 B
DWG NO. REVISION
GENERAL INFORMATION
PERFORMANCE CHECK
Complete a thorough performance verification before disassembling and modifying any system. Make a
note of any operational discrepancies.
SAFETY PROCEDURES
Become familiar with Customer Support Service’s safety policies before working on Oncology Systems
Products. Ensure compliance with CSS’s safety practices and procedures while performing this
modification. Refer to the Product Users’ Manual for safety instructions.
CONTACT INFORMATION
If there are any questions, issues, or concerns contact Varian at 1.888.VARIAN.5 (1.888.827.4265) or go
to http://my.varian.com and click on Contact Us.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 2 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
Note
This document is not all-inclusive and beam quality tests may also be required for other
part replacements, alignments, adjustments and upgrades that are not listed.
It is the responsibility of all Halcyon users to understand and practice these guidelines.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 3 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
General Rules:
1. If using a beam array device and the system is operational, always collect beam reference
data with the beam array device before performing the maintenance.
2. If a beam array device is not available, all profiles must be verified after the maintenance
using a water tank scanning system.
NOTE 3 If there is an unexpected change in the dose calibration of ≥ 2%, beam energy must be verified.
NOTE 4 Any repairs or adjustments/alignments that change SSMM output level, gun current, or magnetron
current (MAG I), can result in a change to the output beam energy. Unlike the high energy Clinac,
Halcyon system design does not have an energy filter (example BMAG), so any change in
magnetron current or gun current will result in a corresponding change in output energy. Therefore,
these types of repairs require a beam energy check after maintenance.
Note 5 Compare the MPC result to the baseline or the recent successful MPC session before any repair.
Unexpected or significant deviation of comparison results may indicate requirement of additional
customer QA or service alignments.
Note 6 Imager must be removed to perform service task, so requirements are the same as imager
removal/replacement. FRUs listed for RTI4343L kV Imager are the same as DMI MV Imager but
only list “DMI” in their descriptions.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 4 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 5 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 6 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
Table 4: Beam Quality Guidelines for HAL Printed Circuit Board (PCB) Replacements
"Completed necessary beam quality tests and alignments per TT-AL-01682, and informed
customer to perform required beam QA tests per CTB-AL-1037 before starting treatments."
Varian employees are not authorized to adjust the dose calibration without customer
approval and guidance.
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 7 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
"Completed necessary beam quality tests and alignments per TT-AL-01682, and
informed customer to perform required beam QA tests per CTB-AL-1037 before
starting treatments."
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 8 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
MV Imaging Calibration
"Completed necessary beam quality tests and alignments per TT-AL-01682, and informed
customer to perform required beam QA tests per CTB-AL-1037 before starting
treatments."
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 9 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
This table provides the minimum test requirements, in sequential order, that are
required after completing a specific maintenance task.
kV Imaging Calibration
If a Varian Customer Support Representative (CSSR) was involved with these repairs,
CBCT Calibrations
then that CSR is required to enter the following comment in the work order / FSR to
define the customer's responsibilities.
"Completed necessary beam quality tests and alignments per TT-AL-01682, and
informed customer to perform required beam QA tests per CTB-AL-1037 before
starting treatments."
7.1 kV Panel (Imager) Replacement and/or Removal & Reinstallation
7.2 PCB DMI Readout Module Replacement or Module Position Change NOTE 6
7.3 PCB DMI Control Board Replacement NOTE 6
7.4 PCB DMI Gate Drive Module Replacement or Module Position Change NOTE 6
7.5 DMI PSU PCB Spare including Pad Replacement NOTE 6
7.6 X-Ray Generator Assembly (EMD) Replacement
7.7 EMD EPS-LP Controller Panel Assembly Replacement
7.8 EMD EPS-LP Controller & PS Assembly (Basket) Replacement
7.9 Norm Chamber Replacement or Adjustment
7.10 HT Cables Replacement
7.11 X-Ray Tube Assembly Replacement
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 10 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.
Bill Of Materials
ATTENTION: This information is provided for use by qualified service technicians. Varian Medical Systems, Inc. makes no
SHEET
representation to the qualifications of any service technicians other than its own employees. Improper or uninformed P1027348
maintenance or repair of the Product may expose a maintenance technician, operator and/ or patient to safety hazards including
but not limited to one or more of the following: mechanical collision, electrical shock, and radiation. Any of these hazards could DWG NO.
AL-1037 Rev B 11 of 11
cause serious injury and even death. Copyright © 2019, Varian Medical Systems, Inc. All rights reserved.