MICROJET RECORDER (180mm)

DATA SHEET PHA

This recorder can record a maximum of 12 channels of


DC voltage, mA, thermocouples and resistance bulbs.
The adoption of an ink jet system makes it possible to
record measured data in analog trace mode or to print in
digital mode at a high speed. This 180mm-wide recorder
performs recording clearly in 6 different colors.

FEATURES
1. Compact size
Compact and lightweight design, 199mm in depth and
about 6kg in mass{weight}.
2. High quality recording
• Ink jet system is used for recording and printing
measured data in 6 different colors at a high speed.
Operating noise is minimized.
• Six and twelve continuous traces without pen offset 6. Easy handling
are possible with this compact size of recorder; a • A cartridge type recording device is used for easy
unique recording system is used for the first time in replacement.
the industry. • Chart paper can be loaded without drawing out the
• Scale of each channel is printed on chart paper, internal unit of the recorder.
elimi- nating the need for scales. • Shortage of ink is detected in early stages and an
3. Easy setting of input signals alarm is given to the operator.
DC voltage input (5mV span, 50V max.), 12 kinds of • The end of chart paper is detected and indicated on
thermocouples (Type B, R, S, K, E, J, T, N, W, L, U, PN) the front panel display.
and resistance bulbs (Pt100) can be set for each chan- • Shortage of ink and the end of chart paper alarm
nel. output is possible.
4. Digital printing 7. Full variety of functions
Beside the analog recording of measured data, digital • Alarm relay output/external control (record start/
printing is also available (periodic printing, list printing, stop, chart speed change, data printing, message
alarm printing, daily report printing, message printing). printing). This unit can easily be connected to the
• Periodic data printing: Channel No., date, time, unit, recorder by user (option).
chart speed, measured value • Chart paper illumination lamp (option): The result of
• List printing: Date, time, unit, recording range, scal- printing can be checked even in lower light.
ing value, alarm set value, chart speed, Tag No. • Burnout function is provided as a standard.
• Alarm printing: Channel No., alarm type, on/off time, • Various recording: Enlarged/reduced recording,
output relay No. auto-range recording, zone recording.
• Daily report and totalized data printing: Printing of • Calculation: Square root extraction, subtraction, en-
maximum, minimum, average and total of data gineering unit conversion, logarithm.
measured during maximum 24 hours • Language: Selectable 3 languages in display and
• Message printing : 10 messages, 16-character user- printing.
entered messages • Passcode security is configurable.
5. Interactive key operation • The message print and alarm print function are op-
Fluorescent indicator is used to clearly indicate alpha- erational, even when the recording mode is off.
numeric characters and symbols. • All parameters of recording format, daily report,
Input mode, recording range, alarm value, chart speed, totalize, message and periodic data printing can be
etc., can be set according to the comments indicated printed cut.
by the display and operating keys. No bothersome op-
eration is required.

EDSX10-60m
Date Nov. 24, 2011
PHA

Chart speeds: Continuous recording type


SPECIFICATIONS 5 to 300 mm/h, continuous recording
301 to 1500 mm/h, intermittent re-
Input system cording
Input points: 6 or 12 continuous recording and inter- Intermitter recording type
mittent recording 5 to 1500 mm/h
Input signal: Thermocouple input…B, R, S, K, E, J, T, Each can be set in 1 mm/h steps.
N, W, L, U, PN Recording cycle: Intermitter recording…30sec/all points
Resistance bulb input…Pt100Ω Continuous recording…Depends on
DC voltage input…50mV, 500mV, 5V, chart speed.
50V range <Calculation formula>
DC current input…4 to 20mA DC, 10 Recording 450
=
to 50mA DC cycle[sec] Chart speed [mm/h]
(Shunt resistor (option) need to be (not faster than 3 seconds.)
connected to the terminal) Measuring cycle: Up to 3 inputs…160ms
Max. input voltage: 6, 12 inputs…320ms
• Thermocouple, resistance bulb and Service life of ink: (Depends on operating conditions)
DC voltage (50mV, 500mV range) About 6 months for 6 points of linear
…±10V DC or less recording at 25 mm/h of chart speed
• DC voltage input (5V, 50V range) Chart handling: Tear off without disturbance of recording.
…±100V DC or less
Input signal setting and change: Indicating system
Setting and change of input signal be-
tween thermocouple, resistance bulb Indication: Fluorescent indication (blue-green),
and DC voltage (50mV, 500mV, 5V, 50V 20 characters × 2 lines
range) is possible for each channel by Characters indicated:
the setting pin in the instrument. 5 × 7 dots, 5.0mm high, 3.5mm wide
Setting of recording range: Contents of indication:
Setting is possible within the reference (1) Measured value:
range by using the keyboard. Temperature…1 digit below decimal
Burnout function:When thermocouple or resistance bulb point;
input is disconnected, the recording is Voltage…6 digits (including sign and
deflected to 100%. decimal point)
Reference range: M easured value of No.1 channel to
No.6 or No.7 channel to No.12 can be
Kind Reference range Reference range indicated simultaneously.
Thermo- B 400 to 1760°C 752 to 3200°F (2) Channel No.: 2 digit (1 to 12)
couple R 0 to 1760°C 32 to 3200°F (3) Engineering unit:
S 0 to 1760°C 32 to 3200°F
K –200 to 1370°C –328 to 2498°F
Max. 7 digits (°C, °F, %, Pa, bar, ppm,
E –200 to 800°C –328 to 1472°F m3/h, etc.)
J –200 to 1100°C –328 to 2012°F (4) Tag No….8 characters
T –200 to 400°C –328 to 752°F
N 0 to 1300°C 32 to 2372°F (5) Time: Year, month, day, hour, minute
W 0 to 1760°C 32 to 3200°F (6) Status indication:
L –200 to 900°C –328 to 1652°F
U –200 to 400°C –328 to 752°F
R
 ecord ON, chart end, battery alarm,
PN 0 to 1300°C 32 to 2372°F alarm, ink shortage alarm, burnout,
Resistance carriage failure

bulb Pt100 –200 to 600°C –328 to 1112°F
Configuration: These can be set according to the
DC voltage –50 to +50mV Scaling is possible comments indicated by operating
–500 to +500mV within the range of
–32767 to +32767 keys as follows,
–5 to +5V
–50 to +50V (decimal point may Passcode
be put as necessary)
Main chart speed
Note: N : NICROSIL-NISIL (IEC584)
Sub chart speed
W : +side 5% Re, –side 26% Re.W (Hoskins Mfg. Co., Alarm setting
U.S.A.) Record mode (trend/logging)
L : +side Fe, –side Cu.Ni alloy (DIN43710)
U : +side Cu, –side Cu.Ni alloy (DIN43710) Recording range
PN : Platinel Input signal
Pt100 : DIN IEC751 List print request
Tag No.
Recording system Daily report setting
Writing system: Ink jet system, 6 colors Totalize function
Chart width: 180mm Communication parameter
Recording color: No. 1,7 channel (orange), No. 2.8 chan- Date and time setting
nel (green), No. 3, 9 channel (purple), Ink monitor clear
No. 4, 10 channel (red), No. 5, 11 chan- Illumination on/off
nel (black), No. 6, 12 channel (blue) Message definition
Recording color can be assigned for Measured value shift
each channel.
Chart length: Z fold 20m
2
Printing system Power supply terminal - ground
…2000V AC, 1 minute
Writing system: Ink jet system, 6 colors Input terminal - ground
Periodic data printing: …500V AC, 1 minute
Measured value, unit, date, time, time Power terminal - input terminal
line, chart speed, channel No. …500V AC, 1 minute
List printing: Between alarm terminals
(1) Measured value list (date, time, channel No., mea- …750V AC, 1 minute
sured value, unit) (Leakage current 5mA or less, how-
(2) Parameter list (date, time, channel No., recording range, ever, if the power supply is 24V DC,
scaling, unit, alarm set value, chart speed, Tag No.) Leakage current of “Power supply
(3) Test pattern (all characters and color patterns) terminal - ground” is 10mA or less)
Message printing: 1 0 m e s s a g e s , 1 6 - c h a r a c t e r u s e r- Reference junction compensation accuracy:
entered messages. K, E, J, T, N, L, U, PN ............. ±0.5°C
Alarm printing: Channel No., alarm type (H, L, RH, (In case of minus input measurement:
RL), output relay No., on/off time ±1.2°C)
Burnout printing: Burnout channel No. and time R, S, B, W .............................. ±1°C
Other: Ink shortage message, automatic (In case of minus input measurement:
range selection mark, recording start ±2.4°C)
mark, chart speed change mark Common mode noise rejection:
Note: Printing is not available for more than 301 mm/h (continuous 120dB at 50, 60Hz ±0.1Hz
recording), or more than 51 mm/h (intermittent recording). Series mode noise rejection:
30dB at 50, 60Hz ±0.1Hz
Performance and characteristics
Physical data
Accuracy and resolution:
Performance under reference condition Mounting: Panel (may be inclined up to 30° back-
(23±2°C, 65±10%RH, power voltage and wards from the vertical.)
frequency variation ±1%, warm-up time
30 minutes or more, vertical mounting,
free from the effect of external noise)

Indication (digital) Recording


Input =90° to 60°
Reso- Reso-
Accuracy Accuracy
lution lution
Thermo- B 0.1°C Material: Case...Steel plate
couple R 0.1°C Front door frame…Polycarbonate with
S
±(0.15% 0.1°C
K digit)
+1 0.1°C glass wool
E
(without 0.1°C Mass{weight}: Approx. 6 kg (without option)
J
reference 0.1°C
T 0.1°C
Approx. 7 kg (with option)
junction
N
compen- 0.1°C Indication Case size: Bezel 288x288mm
W 0.1°C accuracy, 0.1mm
sation mini- Depth 199mm
L
error) 0.1°C ±0.25%
U 0.1°C of record- mum Cutout 281x281mm
PN 0.1°C ing span Finish color: Case…Black, Front door frame…Black
Resist- External terminals: Screw terminal (M4 screw)
±(0.15%
ance Pt100
+1 digit) 0.1°C
bulb
Power requirement
DC –50 to +50mV ±(0.15% 10 µV
voltage –500 to +500mV +1 digit) 100 µV 1) S upply voltage 100 to 240V AC products (9digit of
–5 to 5V 1mV
–50 to 50V 10mV
code symbol = “E”)
Note: Indication accuracy is in % of reference range.
• Power supply voltage:
Indication accuracy of B type TC is ±(0.36%+1digit) between 100V (-15%) to 240V (+10%) AC (Free
400°C to 600°C. power supply)
Indication accuracy of all type TC is ±(0.36%+1digit) between
–200°C to –100°C. • Supply frequency:
Input resistance: Thermocouple:>10MΩ 50/60Hz both employable
50mV range: >10MΩ • Power consumption:
500mV range: >100kΩ 1 00V AC with all options approxi-
5V and 50V range: >1MΩ mately 37VA
Chart speed accuracy: 2 40V AC with all options approxi-
±0.1% (expansion and contraction of mately 56VA
paper is not included) 2) Supply voltage 24V DC products (9digit of code sym-
Clock accuracy: ±50ppm or less (monthly error; about bol = “L”)
2 minutes) • Power supply voltage:
Insulation resistance: 24V (±10%) DC
100MΩ or more (between each termi- • Power consumption:
nal and earth, at 500V DC)
2 6.4V DC with all options 37VA or
Withstand voltage:
less
Input terminal - input terminal
…500V AC, 1 minute
3
PHA

Operating environment Temperature influence:


Change in indication… ±(0.3%+1
(for devices to operate continuously) digit)/10°C, max.
Temperature limits: Change in recording… ±0.5%/10°C,
0 to 50°C max.
Humidity limits: 20 to 80%RH, non condensing is re- Mounting position influence:
quired Inclination within 30°
(temperature x humidity<3200) Change in indication… ±(0.1%+1
Vibration: 10 to 60Hz, 0.2m/s2{0.02G} or less digit) max.
Mounting position: Change in recording… ±0.2% of
Front inclination 0°, rear inclination 30°, recording span, max.
left/right inclination 0° Vibration influence:
Signal source resistance: Linear vibration with 10 to 60Hz of
Thermocouple input…1kΩ or less frequency and 0.2m/s 2 {0.02G} of
Voltage input…Less than 0.1% of in- acceleration is applied to each of 3
put resistance directions for 2 hours.
Resistance bulb input…Less than 10Ω Change in indication… ±(0.1%+1
per line (Resistance of each wire of digit) max.
3-wire system should be balanced Change in recording… ±0.2% of
with others.) recording span, max.
Warm-up time: 30 min or more Effect of external noise:
Shock: No external shock Normal mode noise (50,60Hz±0.1Hz)
Environmental protection: …30dB or more
IEC IP50(Front) / 20(Terminal) C ommon mode noise (50,60Hz±0.1Hz)
Installation category: Ⅱ …120dB or more
Pollution degree: 2 Chart paper influence:
Operating altitude: 2000m max. Standard temperature/humidity: 20°C,
65%RH
Operating environment influence Expansion at 85%RH…0.4% max.
Contraction at 35%RH…0.5% max.
Effect of power supply fluctuation :
(1) Supply voltage 100 to 240V AC products (9digit of
Alarm
code symbol = “E”) Setting method: Setting from keyboard
Voltage variation: 85 to 264V AC Number of alarm levels:
(frequeucy: 50/60Hz) 100VAC base Max. 4 levels for each channel
Change in indication… ±(0.1%+1 Alarm type: High(H), Low(L), High-rate of
digit) max. change(RH), Low-rate of change(RL)
Change in recording… ±0.2% of Alarm action indication:
recording span, max. Kind of alarm and output relay No.
Frequency variation: 47 to 63Hz are indicated for each channel at oc-
(power voltage: 100V AC) 50Hz base currence of alarm.
Change in indication… ±(0.1%+1 Printing: Channel No., kind of alarm, output re-
digit) max. lay No. and on/off time are printed on
Change in recording… ±0.2% of chart paper.
recording span, max. Output: See optional specifications.
(2) S upply voltage 24V DC products (9digit of code Hysteresis: Approx. 0.5% of recording span
symbol = “L”) Alarm timing: Recognition; 1 second (worst case)
With 21.6 to 26.4V DC fluctuation, Action; additional 1 second (worst case)
24V DC base Alarm latch: Hold the alarm display and alarm out-
Change in indication....±(0.1% + 1 put.
digit) max. Others: Shortage of ink and the end of chart
Change in recording....±0.2% of paper alarm output is possible.
recording span, max.
Input signal source resistance or wiring resistance influence: Transportation/storage condition
Thermocouple…10 µV per 100Ω
Vo l t a g e i n p u t … Va r i a t i o n o f 0 . 1 % (Detach a recording head from the main unit before
change of resistance transportation)
Change in indication… ±(0.1%+1 Temperature limits:
digit) max. –10 to +60°C
Change in recording… ±0.2% of Humidity limits: 5 to 90%RH, non condensing is required
recording span, max. Vibration: 10 to 60Hz, 2.45m/s2{0.25G}
Reistance bulb…Variation of resis- Shock: 294m/s2{30G} or less
tance with changes in 10Ω per wire
Change in indication… ±(0.1%+1 Optional specifications
digit) max.
1. Chart illumination:
Change in recording… ±0.2% of
LED
recording span, max.
(3 wires should be balanced.)

4
2. Alarm output/3-point external control:
Periodic data Time, date, chart speed, measured value and unit
This unit can be mounted from the printing function can be printed at fixed intervals.
rear side of the recorder. Printing can be enabled/disabled from keyboard.
(1) Alarm output (DO): Message Maximum 10 messages, 16- character user-
6 or 12 points of relay contact N.O. (1a) printing entered messages can be printed.
output for individual channel opera- Alarm printing Time, channel No., kind of alarm, and output re-
tion or common operation function lay No. can be printed when alarm is on or off.

Maximum contact voltage 240V AC, Unit indication Engineering units such as °C, °F, %, mV, mA,
30V DC Pa,L, etc., are indicated (setting from key board).

Maximum contact current 3A Scaling function Scaling with DC voltage input is possible.
(Setting of decimal point is also possible within
(resistive load) the range of –32767 to +32767).
(2) External control (DI):
Subtract function Difference between any channels is recorded
The following control is possible with (channel is set from keyboard).
external contact signal.
Logarithm Measured value can be displayed and printed by
• Recording start/stop; 10n power
Recording start/stop is effected
Auto-range Recording range is automatically changed for re-
by contact signal. Recording is recording cording in the event of overrange or underrange
started when contact is closed and (setting with keyboard).
This function is not available for combination of
stopped when contact is open. zone recording and enlarged/reduced recording.
• Chart speed change;
Zone recording Recording area is divided into a maximum of 4
Selection between normal and re- zones for recording.
mote chart speeds is effected by This function is not available for combination of
automatic range selection and enlarged/reduced
contact signal. Remote chart speed recording.
is selected when contact is closed
Enlarged/reduced A part of recording area of each channel is ex-
and normal when contact is open. recording panded or contracted for recording.
• Measured value printing; This function is not available for combination of
automatic range selection and zone recording.
Measured value list printing (date,
Square-root ex- Square-root extraction of DC voltage input is pos-
time, channel No., measured value, traction function sible.
unit) is effected by contact signal.
Daily report Measured value of every hour for maximum one
Printing is started when contact is function day (24 data) in each channel is stored for print-
closed. ing. Maximum, minimum and average values are
also printed at the same time.
• Message printing ON-OFF operation, ON-OFF of each channel and
operation start time/stop time can be set from
Note: For external control, use a dry con- keyboard.
tact.
Contact capacity: 12V DC, 0.05A, Totalize function Integrated value of every hour for maximum one
day (24 data) in each channel is stored for print-
N.O. (1a) contact
ing (integration in 1 sec steps). Possible to print
total value only.
3. Transmission function: Total value is also printed at the same time.
ON-OFF operation, ON-OFF of each channel and
T-link interface for transmitting meas- operation start time/stop time can be set from
keyboard.
ured value and receiving the condi-
tion of setting.
Measured value Shift the zero point and inclination of the mea-
shift sured value so that the measured value can be
adjusted according to other instruments.
Transmission
method Half-duplex bit serial
Memory backup Set data and clock function are protected by
Transmission speed 500kbps built-in lithium battery (expected battery life,
about 10 years under normal temperature).
Transmission distance Max. 500m Input filter Response is delayed according to sudden chang-
es in input of each channel (1st order lag filter).
I/O frame 8w or 16w Time constant setting range: 0 to 900 sec (setting
from keyboard).
Available to set/change param-
Message frame Burnout function When thermocouple or resistance bulb input is
eters etc.
disconnected, it is deflected 100%. Also, it is
indicated and printed at the same time.

Passcode 4 digits passcode security is available.

Language English, German, or French is selectable for dis-


play and printing.
Alarm latch The alarm display and alarm output are held even
FUNCTIONS function after the cause of alarming was gone. ON-OFF
operation can be set from keyboard. Cancella-
tion of the held alarm can be made from external
Function Description control (DI).
Range setting Recording range can be set for each channel. Parameter Set parameters on any channel can be copied to
copy any other channels.
Input setting Any input can be set for each channel.

Skip function Used to skip recording, indication and alarm at


any measuring point.

Measured
List printing function

Date, time, and measured value unit can be printed.


value list

Parameter Date, time, recording range, scaling, unit, kind of


list input, alarm set value, chart speed, and Tag No.
can be printed.

Test pattern All characters and color patterns can be printed.

5
PHA

CODE SYMBOLS SCOPE OF DELIVERY


1 2 3 4 5 6 7 8 - 9 10111213
Recorder, panel mounting bracket, accessories (ink car-
PHA 0 0 4 - V Description
tridge (1), fuse (1), chart paper (1), ink absorption cloth (1)).
Recording points
6 6 6 continuous recording Instruction manual (1).
7 6 6 intermittent recording Note: Ink cartridge is not mounted on the recorder at the time of de-
8 8 12 intermittent recording livery.
9 8 12 continuous recording
Power Supply Spare parts
E 100 to 240V AC 50/60Hz
L 24V DC Item Part No. Unit of quantity for sale
Chart paper illumination Ink cartridge PHZH1002 1 pc
A Without
B With Chart paper (0 to 100,
PEX00BL1-1000B 1 box (6 charts)
Alarm output/external control 100 uniform division)
0 Without
1 6-point alarm output/3-point
external control
2 12-point alarm output/3-point Other (optional items)
external control
Transmission function Item Type Specification
Y Without
T With T-Link Shunt resistor PHZT8101 For 10Ω ±0.1%

With 6-point alarm output/3-point


Remarks: Input signal Alarm output/ PHZK8601
external control
Setting prior to delivery is as follows. external control
• Thermocouple K: 0 to 1200 °C With 12-point alarm output/3-point
unit PHZK8201
Note: Contact Fuji Electric for additional features external control
not listed such as Flow integration record
and Calculation of input signals.

The product conforms to the requirements of the


Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TN510405. The applicable standards used to demonstrate
compliance are:-

EN 55011 : 1991 CLASS A


Conducted and Radiated emissions
EN 50082-1 :-1992
Radiated immunity, ESD and FBT

PHA

6
OUTLINE DIAGRAMS (Unit:mm)
Alarm unit(option)
2 T 30
(Alarm 7 to 12)
t
Alarm/external control Mounting
unit(option) bracket Panel
Transmission terminal
(Alarm 1 to 6, DI 1 to 3)
(option)

Input terminal 2
(Input 7 to 12)

280

288
Input terminal 1 Power terminal
(Input 1 to 6)
280 39 160 288

312 199 18

360 min.
Panel cutout
281+20
281+20

320 min.

CONNECTION DIAGRAMS
Input terminal 2 Input terminal 1 Alarm unit Alarm/external control Transmission
(option) unit (option) terminal
(Alarm 7 to 12) (Alarm 1 to 6, DI 1 to 3) (option)
Terminal Terminal
Thermocouple No. No.
Resistance Voltage 71 11
buld + + DI 1 TRX1 TRX2 SHLD
72 Input 7 12 Input 1 11 21
– – + –
73 13 DI 2
12 22
81 21
82 Input 8 22 Input 2 13 23 DI 3 Power terminal
83 23
91 31 31 41 Alarm 7 14 24 Alarm 1 AC AC G

92 Input 9 32 Input 3 32 42 Alarm 8 15 25 Alarm 2


93 33
Power source
101 41 33 43 Alarm 9 16 26 Alarm 3 100 to 240V AC
102 Input 10 42 Input 4 50/60Hz
103 43 34 44 Alarm 10 17 27 Alarm 4
111 51 + - G
35 45 Alarm 11 18 28 Alarm 5
112 Input 11 52 Input 5
113 53 36 46 Alarm 12 19 29 Alarm 6
+ + Power source
121 61 24V DC
– 122 – 62
Input 12 Input 6
123 63 DI 1 Record start or message print 1
DI 2 Chart speed change or message print 2
DI 3 Measured value printing or alarm latch cancel
or message print 3

7
PHA

Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.

International Sales Div


Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/

Information in this catalog is subject to change without notice. Printed in Japan

You might also like