Professional Documents
Culture Documents
Attachment 0
Attachment 0
Ares(2020)592229 - 30/01/2020
D2.2 Risk Analysis
REVISION HISTORY
Version Date Nature of modifications Writer
A 20012020 Creation Zo RAOLISON
Table of contents
1 Purpose............................................................................................................................................ 2
2 List of documents ............................................................................................................................ 2
3 TF-510 RISK MANAGEMENT PLAN................................................................................................... 3
3.1 PURPOSE ................................................................................................................................... 4
3.2 INTRODUCTION ......................................................................................................................... 4
3.3 SCOPE OF THE RISK MANAGEMENT PLAN ................................................................................ 4
3.4 RESPONSIBILITIES AND AUTHORITIES ....................................................................................... 4
3.5 PRODUCT DEFINITION............................................................................................................... 5
3.5.1 Medical device description ................................................................................................ 5
3.5.2 Intended use ...................................................................................................................... 6
3.5.3 Indications for use ............................................................................................................. 7
3.5.4 Expected performance ...................................................................................................... 7
3.5.5 Preliminary hazard analysis ............................................................................................... 7
3.5.6 Medical device life cycle steps......................................................................................... 13
3.6 RISK ANALYSIS ......................................................................................................................... 14
3.6.1 Method used to analyze risk ........................................................................................... 14
3.6.2 Risk control options ......................................................................................................... 14
3.6.3 Risk quotation criteria ..................................................................................................... 14
3.6.4 The severity scale ............................................................................................................ 15
3.6.5 The probability of occurrence scale ................................................................................ 15
3.6.6 Risk acceptability Criteria ................................................................................................ 16
3.6.7 Risk-benefit evaluation criteria ....................................................................................... 17
3.7 Verification and evaluation of overall residual risk ................................................................ 17
3.7.1 Review of all residual risks ............................................................................................... 17
3.7.2 Method used to analyze overall residual risk .................................................................. 18
3.7.3 Review for conflicting requirements ............................................................................... 18
3.7.4 Review of warnings in the manuals/ instructions for use ............................................... 18
3.7.5 Residual risk interactions ................................................................................................. 18
3.7.6 Global residual risk .......................................................................................................... 18
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 1 sur 26
D2.2 Risk Analysis
1 Purpose
This document is a report on the Risk analysis work carried out on the MRI PADS 7TLF. It is the
deliverable D2.2 of the work package CE documents. It describes the risk management plan for the
device, the completed risk analysis matrix and the risk management report.
2 List of documents
The risk analysis files are organized into several parts:
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 2 sur 26
D2.2 Risk Analysis
Only the files TF-510, TF-520 and TF-530 are provided in this current document. In each file, additional
documents are cited and can be provided upon request.
MRI PADS
RISK MANAGEMENT PLAN
EDITION
Function Name Position
Tryfon
Approver Director of Multiwave Imaging
ANTONAKAKIS
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 3 sur 26
D2.2 Risk Analysis
3.1 PURPOSE
The purpose of this risk management plan is to provide a description of the risk management file and
the method used to identify and reduce risks related to the MRI Pads product family throughout the
products life-cycle. It defines the scope, responsibilities, requirements and review of risks according to
the EN ISO 14971:2012 standard.
3.2 INTRODUCTION
The risk management process is conducted according to the principles outlined in EN ISO 14971:2012.
When subcontractors have already competed a risk analysis pertaining to their manufacturing
activities, it becomes part of our risk analysis after evaluation and approval.
- knowledge of and experience with the medical device (or similar medical devices), its use and the
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 4 sur 26
D2.2 Risk Analysis
technologies involved;
- knowledge of and experience with risk management techniques;
- knowledge of the state of the art concerning the use of the medical device.
NOTE Risk management tasks can be performed by representatives of several functions, each
contributing their specialist knowledge.
Head of
Multiwave
Elodie Multiwave Expert and
Imaging and 1-2-3
GEORGET-PARIS Imaging reviewer
Medical
Technologies
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 5 sur 26
D2.2 Risk Analysis
During an MRI scan, the patient is in supine position on a bed which is introduced in the scanner.
The PADS 7TNS, PADS 7TSC and PADS 7TLF are lateral pads to be used with 7T MRI scanners. They are
positioned on each side of the patient head close to the temporal lobes (Cf. Figure 2).
The pads may come into contact with the skin, or the hair of the patient for brain imaging pads, and
must remain in position throughout the MRI scan.
Figure 2. Patient in supine position with two lateral 7T Pads on each side of the head.
The PADS 7TSC are used to improve local magnetic B1+ field homogeneity in 7T MRI scanners for
imaging the temporal lobes of the brain in an adult population.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 6 sur 26
D2.2 Risk Analysis
The PADS 7TLF are used to improve local magnetic B1+ field homogeneity in 7T MRI scanners for
imaging the temporal lobes of the brain in an adult population.
C.2.1 What is the intended use and how is the medical device to
be used?
what is the medical device's role relative to: The medical device is an MRI accessory used
- diagnosis, prevention, monitoring, treatment or alleviation for MRI body imaging in a generic MRI
of disease, scanner at 7 Tesla. It can be used to image
- compensation for injury or handicap or adults of both genders.
- Replacement or modification of anatomy, or control of
conception?
What are the indications for use (e.g. patient population)?
Does the medical device sustain or support life?
-is special intervention necessary in the case of failure of the
medical device?
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 7 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 8 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 9 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 10 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 11 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 12 sur 26
D2.2 Risk Analysis
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 13 sur 26
D2.2 Risk Analysis
1. Identification of safety features, hazards associated with the device, hazardous situations, or
possible harm to the patient or users or environment (§4.3). The list of hazards provided in
annex E of the standard may be used for hazard identification.
2. Evaluation of the probability of occurrence of each hazardous situation, and the severity of the
resulting harm (§4.4 and 4.5)
3. Description of the control options (by design, protective measures, information for safety)
(§6.2 and 6.3)
4. Residual risk evaluation and risk/benefit analysis (§6.4, 6.5 and 6.6)
The 4 steps are documented in an Excel file TF-520 Risk analysis matrix, allowing the risk project team
to follow the entire risk assessment process.
Note: one hazardous situation may cause more than one type of harm. The most serious harm is always
considered when performing risk evaluation.
According to the EU Directive 93/42/EEC, control measures are put in place in order to reduce the risks
as far as possible (i.e. the AFAP method: As Far As Possible).
According to the EU Directive 93/42/EEC and Annex ZA 7.c) of the EN ISO 14971:2012 standard, the
information given to users does not solely allow for additional risk reduction.
If needed, an action plan may be written regarding the application of risk control methods.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 14 sur 26
D2.2 Risk Analysis
Even if the MRI Pad is a reusable medical device, the probability of harm is inferred from the number
of devices put on the market and not from the total number of uses. Indeed, most risks will occur only
once, making the MRI Pad useless and requiring its return to the manufacturer.
In fact, the probability of occurrence of harm is computed as the product of two different probabilities:
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 15 sur 26
D2.2 Risk Analysis
P1 and P2 are estimated separately and the overall probability of occurrence P1xP2 is then determined
following the table below.
P2
P1 x P2 Low Moderate High
0-1% 1-10% 10-100%
Low
1: Improbable 1: Improbable 2: Rare
0-1%
Moderate
P1 1: Improbable 2: Rare 3: Probable
1-10%
High
2: Rare 3: Probable 4: Frequent
10-100%
Since we have no historical, production or PMS data, probabilities are initially estimated based on
expert knowledge and common sense, always considering a worst-case scenario (highest possible
probability of occurrence). Experimental data, analytical techniques or simulation may be used in this
assessment.
Probabilities will be re-evaluated using historical, production and PMS data.
According to the EU Directive 93/42/EEC, all risks should be considered and reduced as far as
possible (even if the risk is in a broadly acceptable region).
The matrix is used to evaluate risks before and after risk control measures are implemented.
Severity
Probability 1: Negligible 2: Moderate 3: Significant
1: Improbable 1 5 7
2: Rare 2 6 10
3: Probable 3 8 11
4: Frequent 4 9 12
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 16 sur 26
D2.2 Risk Analysis
The risk-benefit ratio is evaluated by the risk management team who establish the acceptability of the
risk as a function of the expected performance of the device and the actual experience. The benefits
determined in the clinical evaluation report are used as the basis of the evaluation.
- Performance 1: homogenization and enhancement of the transmit B1+ field in the temporal
lobes in the vicinity of the pads.
A risk-benefit analysis is used to justify a risk once all practicable measures to reduce the risk have
been applied. If, after applying these measures, the risk is still below acceptability level, a risk/benefit
analysis is needed to establish whether the medical device is likely to provide more benefit than harm.
Generally, if all practicable risk control measures are insufficient to satisfy the risk acceptability criteria
in the risk management plan, the design or the manufacturing steps shall be modified.
At least once per year, a risk management review is performed; it reviews the risk management plan,
the risk analyses, and all clinically obtained data.
A team is put together to perform the risk review. Note that it is possible to have an external person
on the team.
At the end of the review, a new risk management report is written and archived.
If there is a lack of evidence, an action plan shall be established and a provisional deadline to obtain
the required evidence will be indicated in the corresponding risk analysis Excel file. A summary, in the
form of two risk acceptability tables, is then made based upon the risk analysis files. These two tables
provide an overview of risk reduction through risk control methods. The types of risks are described
and it is ensured that these risks are considered in the clinical evaluation. These tables are part of the
risk management report.
It should be noted that a decrease in the risk score is only effective if the associated evidence is
available; in the absence of this evidence, the initial risks core applies.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 17 sur 26
D2.2 Risk Analysis
During the risk review, special attention should be focused on evaluating any conflicting requirements
between the control options chosen.
If one conflicting requirement is found during the review, the risk controls should be reviewed and a
new risk analysis must be performed until no to eliminate conflicting requirements are found.
During the risk analysis review, the team in charge should determine if the warnings are adequate, and
easy to understand, in order to have an acceptable overall residual risk. The team must also ensure
that all risk control measures related to information for safety specified in the risk analysis are included
in the instructions for use.
A combined risk is a new risk resulting from the simultaneous or concomitant appearance of two or
more dangerous situations.
During the risk management review, the team will separate residual risks that are « combinable » from
those that are « impossible to combine ».
The team must then evaluate if the residual risks can be combined, by asking two main questions:
1. Does a new risk emerge from the combination of the residual risks?
2. Does the combination of the residual risks lead to a higher risk level than each individual residual
risk?
If the answer to one of the two questions is YES, a new risk-benefit analysis of the combined residual
risks must be conducted.
If the answer is NO for both questions, there is no need for a further risk-benefit analysis.
At the time of the review, the working group will ensure that:
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 18 sur 26
D2.2 Risk Analysis
Average risk: Ar = ∑ Rn / N with N total number of risks and Rn the individual risk value.
If a new risk or a higher risk is identified, new risk control options must be applied in order to reduce
the risk to an acceptable level, and a new risk analysis performed.
If conflicting requirements are found, the team in charge of the review must find a way to reduce them
until the associated risk becomes acceptable
If inadequate warnings are found in the manual, or the instructions are difficult to understand or
incoherent, the team in charge of the review must modify the manual such that no inadequate
warnings remain.
If no solutions can be found to reduce risks to acceptable levels, the design and/or the production
should be modified or stopped.
N.B.: In accordance with Article 7 of the standard EN ISO 14971:2012, in the case where the overall
residual risk is deemed to be unacceptable according to the criteria set out in the Article, the team
must establish whether the medical advantages of the intended usage of the device outweigh the
overall residual risk.
The risk-benefit ratio for the overall residual risk is evaluated with the help of the clinical evaluation.
This risk-benefit ratio is qualitative and is based on the literature about the product itself and similar
products.
If the medical benefit outweighs the overall residual risks, then the latter are deemed to be acceptable.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 19 sur 26
D2.2 Risk Analysis
Any corrective and preventative actions implemented are evaluated using the product review (PMS)
to verify that the agreed actions are effective.
Additional risk management reviews may take place as required upon the occurrence of particular
events: customer complaints, adverse events, ANSM website alerts, literature reviews, etc.
A team must be established for this additional review; afterwards, a new risk management report is
written and archived.
3. Appropriate methods are in place to obtain relevant production and post-production information
Once the product is on the market, a review of the risk management file is carried out on a yearly basis
or whether a change occurs that could impact patient safety (field feedback, adverse events, changes
in the design, etc.).
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 20 sur 26
D2.2 Risk Analysis
Material science
Writer Zo RAOLISON
engineer
Head of Multiwave
Auditor/ Elodie Innovation and
Approver Georget-Paris Medical
Technologies
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 21 sur 26
D2.2 Risk Analysis
5.1 Purpose
This document forms an integral part of the Technical File of the medical device MRI pads 7TLF. The
purpose of this document is to present the risk management report of the medical device "MRI PADS
7TLF" according to EN ISO 14971 : 2013.
5.3 Participants
The present report also presents the final risk management review.
• Preliminary hazard analysis was performed using Annex C of EN ISO 14971: 2012
3. Description of the control options (by design, protective measures, information for
safety)
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 22 sur 26
D2.2 Risk Analysis
2) The risk control options were defined according to the risk management plan (inherent safety by
design, protective measures, information for safety).
3) The team determined whether or not the risk control options could generate new risks.
4) The effectiveness of the risk control options was verified by the team and the associated
documentation (reports, validation files, technical file sections, instructions for use, etc.) was noted.
5) The team evaluated the evidence in order to decide whether the evidences are sufficient to
decrease the risk
6) Then, the “individual” risk was re-evaluated using the same criteria
7) The risk-benefit ratio was determined by the team using the intended performance as part of the
benefits.
The tables below present the risk assessments and levels of acceptability as determined from the risk
analysis file and the usability risk analysis file, thus providing an overview of the effectiveness of risk
reduction through the implementation of risk control methods.
2 2
32 1 14 84 27 27
Occasional Occasional
3 3
14 12 27 16 5 4
Probable Probable
4 4
88 21 20 2 0 0
Frequent Frequent
Comparing the two tables above, the group found that, in general, the risks were reduced by reducing
the likelihood of appearance. In order to do so, the following risk management measures have been
implemented:
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 23 sur 26
D2.2 Risk Analysis
During the review, the effectiveness of risk reduction measures was verified and was determined to
be appropriate. Each residual risk was deemed acceptable in view of patient benefit.
To the best of our knowledge all use errors and hazards associated have been identified at this time.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 24 sur 26
D2.2 Risk Analysis
By combining a set of scenarios for the use of a non-compliant device by untrained practitioners, the
group arrived at the same risks as the risks already identified. No new or higher-level risks were
identified by this combination analysis.
Furthermore, the conclusions of the clinical evaluation also do not show any new risks from clinical
use. Based on the clinical evaluation, reflecting all known clinical situations, the benefit-risk ratio for
the overall residual risk remains acceptable.
5.6 Global residual risk
During the review, the global residual risk was investigated according to the Risk Management Plan,
the §5.1 of the present document and the risk acceptability criteria table.
The benefit of using the MRI pads for 7T head scans is an enhancement of the B1+ field in the vicinity
of the pads leading to images that may help for improved diagnosis and treatment. Used with due
caution by the intended user, the benefits of the pads outweigh the risks of its use.
Furthermore, the usability file demonstrates that the MRI pads for 7T Head are easy to use. The patient
benefits outweigh the risks identified.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 25 sur 26
D2.2 Risk Analysis
· Customer complaints
· Customer listening
· Changes in the literature
· Adverse events of similar products
· Production information directly related to the product
6 CONCLUSION
At the date of the risk management review, the state of the risk management file is as follows:
1. The Risk Management Plan is deemed to be appropriate: the responsibilities and the
method used by all members of the risk analysis team are well defined
2. Risk Analysis files: risk control measures are appropriate
3. The Instructions for use were updated
4. Overall Residual Risk was reviewed and deemed acceptable
5. Appropriate Methods have been put in place to collect relevant production and post-
production information
Considering all the data obtained and the expected benefits for the patient, the working group
concluded that the overall residual risk is acceptable and that the benefit / risk ratio is in the patient’s
favor for both product risk analysis and usability risk analysis.
Proprietary and confidential. Not to be redistributed without prior permission of Multiwave Innovation.
Page 26 sur 26
Accept
TECHNICAL FILE
Ref : Risk analysis
Overall Probability
Overall Probability
Severity
Severity
Step of the medical Device Harm to the user / patient / Initial Information for safety
Activity Risk number Hazard Foreseeable sequence of events Hazardous situations P1 P2 Inherent safety by design Protective measures Efficacy of the risk control measures P1 P2
device Life cycle concerned environment risk (IFU)
Inappropriate storage case for the pads that fold sure the material in your pad is
Insufficient energy is delivered to the Dissatisfaction, loss of time
C8.02 Functional ABC during storage/shipment causing poor distribution H H 4 1 4 evenly distributed throughout L H 2 1
body. (user/patient)
of the composite material. its plastic container, manually
level it if necessary)
An unhomogeneous/stiff/uneven pad is
C8.03 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
The patient's and/or user's airways are Respiratory tract irritation, other
A H H 4 3 12 L H 2 3
exposed to particles and/or solvent limited health risks (user/patient)
C8.05
material during product use or Respiratory tract irritation, other
BC M H 3 3 11 L H 2 3
manipulation. limited health risks (user/patient)
specifications (wrong particles / solvent). Insufficient energy is delivered to the Dissatisfaction, loss of time
S1.02 Functional ABC H H 4 1 4 L H 2 1
The resulting pad is inhomogeneous (appearance body. (user/patient)
of "stones" and/or "lumps") or too stiff, with varying
permittivity. An unhomogeneous/stiff/uneven pad is
S1.03 Operational ABC Discomfort (patient) H H 2 1 2 L H 2 1
positioned next to the patient's head. Visual check before use
(do not use if filling is stiff or not
permittivity.
An unhomogeneous/stiff/uneven pad is
S1.06 Operational ABC Discomfort (patient) H H 4 1 4 Product design validation [1] L H 2 1
positioned next to the patient's head.
Supplier control Stability report [3]
S - Supply-chain of
Discomfort, dissatisfaction, loss of Control of materials at reception Technical specifications [5,6,7,8,9,10]
raw materials and S1 - Selection of ABC H H 4 1 4 L H 2 1
time (user/patient) Traceability Supplier selection procedure [11]
intermediate suppliers
QC inspection before releasing Receipt files [12,13,14,15]
products Minor local skin irritation
B H H 4 2 9 finished pads IFUs [26,27,28] L H 2 2
(user/patient)
The patient's and/or user's skin is PMS report [29]
Severe eye/skin irritation, allergic
B exposed to particles and/or solvent H M 3 3 11 Visual check before use (use L M 1 3
S1.07 Choosing an unreliable supplier. skin reaction (user/patient)
material during product use or only if no leaks and pouch
Reception of pouches that do not comply with
manipulation. Minor local skin irritation integrity intact and correctly
Biocompatibility AC specifications. H L 2 2 6 L L 1 2
(user/patient) sealed)
The resulting pad breaks resulting in a leakage of
Safety warning concerning
solvent or particles. Severe eye/skin irritation, allergic
AC H L 2 3 10 contact with composite material L L 1 3
skin reaction (user/patient)
Inapropriate quantity of composite inserted in the Insufficient energy is delivered to the Dissatisfaction, loss of time
P3.01 Functional ABC H H 4 1 4 L H 2 1
pouch, resulting in a thinner pad. body. (user/patient)
Minor local skin irritation sealing process Sealed filled pouch visual inspection [22]
P3 - Composite B M H 3 2 8 L H 2 2
(user/patient) QC inspection before releasing Production sheet [23]
P - Production insertion in the
The patient's and/or user's skin is finished pads Operators training records [30]
pouch Severe eye/skin irritation, allergic
B exposed to particles and/or solvent M M 2 3 10 Training sessions for Visual check before use (use IFUs [26,27,28] L M 1 3
P3.04 skin reaction (user/patient)
Inapropriate quantity of composite inserted in the material during product use or manufacturing operators only if no leaks and pouch PMS report [29]
pouch, resulting in a thicker pad. manipulation. Minor local skin irritation integrity intact and correctly
Biocompatibility AC M L 1 2 5 L L 1 2
Pad breaks during use, resulting in a leakage of (user/patient) sealed)
solvent or particles. Safety warning concerning
Severe eye/skin irritation, allergic contact with composite material
AC M L 1 3 7 L L 1 3
skin reaction (user/patient)
An unhomogeneous/stiff/uneven pad is
P5.09 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
Sealing process validation [4]
Discomfort, dissatisfaction, loss of
ABC H H 4 1 4 Seal resistance tests every 3 Stability report [3] L H 2 1
The vacuum process is too long, resulting in time (user/patient)
months Manufacturing procedure [18]
powder residues on the seals.
Minor local skin irritation QC inspection before releasing Sealed filled pouch visual inspection [22]
B The seals are not resistant/are incomplete/have H H 4 2 9 Sealing process validation L H 2 2
(user/patient) finished pads Production sheet [23]
holes. The patient's and/or user's skin is
Training sessions for Operators training records [30]
A leaking pad is used on a patient. Severe eye/skin irritation, allergic
B exposed to particles and/or solvent H M 3 3 11 Visual check before use (use L M 1 3
P5.10 manufacturing operators IFUs [26,27,28]
skin reaction (user/patient)
material during product use or only if no leaks and pouch
PMS report [29]
manipulation. Minor local skin irritation integrity intact and correctly
Biocompatibility AC H L 2 2 6 L L 1 2
(user/patient) sealed)
Safety warning concerning
Severe eye/skin irritation, allergic contact with composite material
AC H L 2 3 10 L L 1 3
skin reaction (user/patient)
An unhomogeneous/stiff/uneven pad is
P6.03 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
Sealing process validation [4]
Discomfort, dissatisfaction, loss of Annual calibration verification
ABC H H 4 1 4 Stability report [3] L H 2 1
time (user/patient) Seal resistance tests every 3
The sealing machine is not calibrated. Manufacturing procedure [18]
months
The seals are not resistant/are incomplete/have Minor local skin irritation Sealed filled pouch visual inspection [22]
B H H 4 2 9 Sealing Process Validation QC inspection before releasing L H 2 2
holes. (user/patient) Production sheet [23]
The patient's and/or user's skin is finished pads
A leaking pad is used on a patient. Operators training records [30]
Severe eye/skin irritation, allergic Training sessions for
B exposed to particles and/or solvent H M 3 3 11 Visual check before use (use L M 1 3
P6.04 IFUs [26,27,28]
skin reaction (user/patient) manufacturing operators
material during product use or only if no leaks and pouch
PMS report [29]
manipulation. Minor local skin irritation integrity intact and correctly
Biocompatibility AC H L 2 2 6 L L 1 2
(user/patient) sealed)
Safety warning concerning
Severe eye/skin irritation, allergic contact with composite material
AC H L 2 3 10 L L 1 3
skin reaction (user/patient)
Insufficient energy is delivered to the Dissatisfaction, loss of time only if no leaks and pouch
P6.07 Functional ABC H H 4 1 4 L H 2 1
body. (user/patient) integrity intact and correctly
sealed)
An unhomogeneous/stiff/uneven pad is
P6.08 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
Insufficient energy is delivered to the Dissatisfaction, loss of time only if no leaks and pouch
P6.12 Functional ABC H H 4 1 4 L H 2 1
body. (user/patient) integrity intact and correctly
sealed)
An unhomogeneous/stiff/uneven pad is
P6.13 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
Sealing process validation [4]
Discomfort, dissatisfaction, loss of Seal resistance tests every 3 Stability report [3]
ABC H H 4 1 4 L H 2 1
The sealing time is too short or too long. time (user/patient) months Manufacturing procedure [18]
The seals are not resistant/are incomplete/have Minor local skin irritation QC inspection before releasing Sealed filled pouch visual inspection [22]
B H H 4 2 9 Sealing Process Validation L H 2 2
holes. (user/patient) finished pads Production sheet [23]
A leaking pad is used on a patient. The patient's and/or user's skin is Training sessions for Operators training records [30]
Severe eye/skin irritation, allergic
B exposed to particles and/or solvent H M 3 3 11 manufacturing operators Visual check before use (use IFUs [26,27,28] L M 1 3
P6.14 skin reaction (user/patient)
material during product use or only if no leaks and pouch PMS report [29]
Presence of harmless material (dirt/water) on the An dirty pad is positioned next to the Discomfort, dissatisfaction, loss of
P7.03 Operational ABC H H 4 1 4 L H 2 1
outer face of the pad. patient's head. time (user/patient)
P8.03 Operational ABC The pad is wrongly positioned. Discomfort (patient) H H 4 1 4 manufactured one by one L H 2 1
without ever mixing (line
Discomfort, dissatisfaction, loss of clearance) Label specifications [10]
ABC H H 4 1 4 L H 2 1
time (user/patient) QC inspection before releasing Manufacturing procedure [18]
Minor local skin irritation finished pads Sealed filled pouch visual inspection [22]
P8 - Labeling B H H 4 2 9 L H 2 2
P - Production (user/patient) Training sessions for Production sheet [23]
process
The patient's and/or user's skin is manufacturing operators Operators training records [30]
Severe eye/skin irritation, allergic
B exposed to particles and/or solvent H M 3 3 11 Label on pad Visual check before use (use IFUs [26,27,28] L M 1 3
P8.04 skin reaction (user/patient)
Mislabeling or missing label. material during product use or Additional label added in only if no leaks and pouch PMS report [29]
The pad is used/kept on storage long after expiry manipulation. Minor local skin irritation storage case (brand name, integrity intact and correctly
Biocompatibility AC H L 2 2 6 L L 1 2
date. The pad gets fragile and break, leading to (user/patient) reference, expiry date) sealed)
composite leakage. Only used by skilled Safety warning concerning
Severe eye/skin irritation, allergic contact with composite material
AC H L 2 3 10 professionals L L 1 3
skin reaction (user/patient)
Insufficient energy is delivered to the Dissatisfaction, loss of time its plastic container, manually
P9.02 Functional ABC H H 4 1 4 M H 3 1
body. (user/patient) level it if necessary / do not use
P9.03 Operational ABC pad is positioned next to the patient's Discomfort (patient) H H 4 1 4 homogeneous) M H 3 1
head.
Pads are incorrectly placed in storage case and sure the material in your pad is
Insufficient energy is delivered to the Dissatisfaction, loss of time
P10.02 Functional ABC folded during storage/shipment causing poor H H 4 1 4 evenly distributed throughout L H 2 1
body. (user/patient)
distribution of the composite material. its plastic container, manually
level it if necessary)
An unhomogeneous/stiff/uneven pad is
P10.03 Operational ABC Discomfort (patient) H H 4 1 4 L H 2 1
positioned next to the patient's head.
The patient's and/or user's airways are Respiratory tract irritation, other
A H H 4 3 12 L H 2 3
exposed to particles and/or solvent limited health risks (user/patient)
P10.05
material during product use or Respiratory tract irritation, other
BC M H 3 3 11 L H 2 3
manipulation. limited health risks (user/patient)
P10.08 Operational ABC The pad is wrongly positioned. Discomfort (patient) H H 4 1 4 manufactured one by one L H 2 1
without ever mixing (line
Discomfort, dissatisfaction, loss of clearance) Label specifications [10]
ABC M H 3 1 3 L H 2 1
time (user/patient) Training sessions for Manufacturing procedure [18]
Minor local skin irritation manufacturing operators Production sheet [23]
B M H 3 2 8 L H 2 2
(user/patient) Label on pad Case and foam preparation protocol [24]
The patient's and/or user's skin is Additional label added in Operators training records [30]
Severe eye/skin irritation, allergic
B exposed to particles and/or solvent M M 2 3 10 storage case (brand name, Visual check before use (use IFUs [26,27,28] L M 1 3
P10.09 Mixing of products. skin reaction (user/patient)
material during product use or reference, expiry date) only if no leaks and pouch PMS report [29]
The wrong product is placed on the storage case.
manipulation. Minor local skin irritation Labels verification and matching integrity intact and correctly
Biocompatibility AC The pad is used/kept on storage long after expiry M L 1 2 5 L L 1 2
(user/patient) during conditioning sealed)
date. The pad gets fragile and break, leading to
Only used by skilled Safety warning concerning
composite leakage. Severe eye/skin irritation, allergic
AC M L 1 3 7 professionals contact with composite material L L 1 3
skin reaction (user/patient)
F2.09 Operational ABC The pad is wrongly positioned. Discomfort (patient) H H 4 1 4 Only used by skilled L H 2 1
professionals
The pad is repeatedly folded for better positioning, Insufficient energy is delivered to the Dissatisfaction, loss of time sure the material in your pad is
U1.02 Functional ABC H H 4 1 4 M H 3 1
causing poor distribution of the composite material. body. (user/patient) evenly distributed throughout
its plastic container, manually
An unhomogeneous/stiff/uneven pad is
U1.03 Operational ABC Discomfort (patient) H H 4 1 4 level it if necessary) M H 3 1
positioned next to the patient's head.
The patient's and/or user's airways are Respiratory tract irritation, other
A H H 4 3 12 M H 3 3
exposed to particles and/or solvent limited health risks (user/patient)
U1.05
material during product use or Respiratory tract irritation, other
BC M H 3 3 11 L H 2 3
manipulation. limited health risks (user/patient)
Discomfort, dissatisfaction, loss of transport and storage of the pad Stability report [3]
ABC H H 4 1 4 M H 3 1
time (user/patient) when not in use Usability report [31]
Only used by skilled
U - Clinical use U2 - Storage Minor local skin irritation The pad is held tight against the Transport resistance Report [25]
B H H 4 2 9 professionals M H 3 2
(user/patient) foam IFUs [26,27,28]
Storage instructions
The patient's and/or user's skin is Pouch cannot be bend for more PMS report [29]
The case containing the pads is stored in poor Severe eye/skin irritation, allergic Visual check before use (use
B exposed to particles and/or solvent H M 3 3 11 than 20° M M 2 3
U2.04 conditions at customer's site (excessive skin reaction (user/patient) only if no leaks and pouch
material during product use or
Biocompatibility temperature, wrong position or excessive integrity intact and correctly
manipulation. Minor local skin irritation
AC manipulations of the case) causing premature wear H L 2 2 6 sealed) M L 1 2
(user/patient)
and tear. Safety warning concerning
Severe eye/skin irritation, allergic contact with composite material
AC H L 2 3 10 M L 1 3
skin reaction (user/patient)
The patient's and/or user's airways are
Respiratory tract irritation, other
U2.05 ABC exposed to particles and/or solvent H H 4 3 12 M H 3 3
limited health risks (user/patient)
material during product use or
Discomfort, dissatisfaction
ABC M H 3 1 3 L H 2 1
(user/patient)
The patient's and/or user's airways Minor local skin irritation
ABC Use of inappropriate cleaning products causing M M 2 2 6 Usability report [31] L M 1 2
and/or skin are exposed to harmful (user/patient) Only used by skilled
U3.01 Biocompatibility presence of harmful product on the outer face of Cleaning instructions IFUs [26,27,28]
products during product use or Severe eye/skin irritation, allergic professionals
the pad. PMS report [29]
manipulation. skin reaction, respiratory tract
ABC M M 2 3 10 L M 1 3
irritation, other limited health risks
(user/patient)
#N/A
Device: A Initial risk cotation Severity Device: A Residual risk cotation Severity
1 2 3 1 2 3
1 27 4 7 1 59 29 37
2 32 24 30 2 81 4 23
Occurrence Occurrence
3 7 5 5 3 19 1 5
4 96 1 23 4 3 0 0
Device: B Initial risk cotation Severity Device: B Residual risk cotation Severity
1 2 3 1 2 3
1 27 0 3 1 59 2 33
2 32 1 14 2 84 27 27
Occurrence Occurrence
3 14 12 27 3 16 5 4
4 88 21 20 4 2 0 0
Device: C Initial risk cotation Severity Device: C Residual risk cotation Severity
1 2 3 1 2 3
1 27 7 10 1 59 29 37
2 32 21 27 2 84 4 24
Occurrence Occurrence
3 14 5 8 3 16 1 4
4 89 1 20 4 3 0 0
B/R
ment according
Benefit/risk ratio
agement plan
1 N YES
3 N Ditto Favorable
3 N Ditto Favorable
2 N Ditto Favorable
7 N Ditto Favorable
4 N Ditto Favorable
4 N Ditto Favorable
4 N Ditto Favorable